K233276 · Suzhou Minhua Medical Apparatus Supplies Co., Ltd. · DXQ · Dec 21, 2023 · Cardiovascular
Device Facts
Record ID
K233276
Device Name
Reusable Blood Pressure Cuff
Applicant
Suzhou Minhua Medical Apparatus Supplies Co., Ltd.
Product Code
DXQ · Cardiovascular
Decision Date
Dec 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1120
Device Class
Class 2
Indications for Use
The Reusable Blood Pressure Cuff is an accessory used in conjunction with non-invasive blood pressure measurement systems. The cuff is non-sterile and reusable. It is available in adult size.
Device Story
Reusable blood pressure cuff; accessory for non-invasive blood pressure measurement systems. Device wraps around patient limb; secured via hook and loop closure. Air hose connects cuff to external blood pressure monitor. Used in clinical or home settings; operated by clinicians or patients. Cuff inflates to occlude blood flow; monitor detects oscillations to calculate blood pressure. Output displayed on monitor; informs clinical decisions regarding hypertension management or cardiovascular status. Benefits include non-invasive, repeatable blood pressure assessment.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing, including compliance with IEC 80601-2-30:2018, ISO 81060-1:2007, and biocompatibility testing (ISO 10993-5:2009, ISO 10993-10:2010).
Technological Characteristics
Polyester cuff with internal bladder; hook and loop closure. Pressure range 0-300 mmHg. Single-tube configuration. Biocompatibility per ISO 10993-5 and ISO 10993-10. Non-sterile, reusable. Mechanical device; no energy source or software.
Indications for Use
Indicated for adult patients requiring non-invasive blood pressure monitoring when used with compatible measurement systems.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 21, 2023
Suzhou Minhua Medical Apparatus Supplies Co., LTD % Alex Wang Regulatory Affairs Specialist Suzhou Minhua Medical Apparatus Supplies Co., Ltd. No.789, Wu Fang Road, Friendship Industrial Zone, Song Ling Town, Wu Jiang Suzhou, Jiangsu 215222 China
Re: K233276
Trade/Device Name: Reusable Blood Pressure Cuff Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXO Dated: August 28, 2023 Received: September 29, 2023
Dear Alex Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233276
Device Name Reusable Blood Pressure Cuff
Indications for Use (Describe)
The Reusable Blood Pressure Cuff is an accessory used in conjunction with non-invasive blood pressurement systems. The cuff is non-sterile and reusable. It is available in adult size.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Tab #6 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number:
- Date of Preparation: 08/28/2023 1.
- Sponsor Identification 2.
### Suzhou Minhua Medical Apparatus Supplies Co., LTD
You Yi Industrial Park, Songlin Town, Wujiang, 215222, CHINA Establishment Registration Number: 3011529836
Contact Person: Michael Ying Position: Project Manager Tel: +86-0512-63091066 Fax: +86-0512-63091066 Email: yingjiaqing@mhmedical.com.cn
3. Designated Submission Correspondent
Mr. Alex Wang Position: Regulatory Affairs Specialist Tel: +86-0512-63091066 Fax: +86-0512-63091066 Email: alex.wang@pedesign.com.cn
{5}------------------------------------------------
- Identification of Proposed Device 4.
Trade Name: Reusable Blood Pressure Cuff Common Name: Blood Pressure Cuff
Regulatory Information Classification Name: Blood Pressure Cuff Classification: 2 Product Code: DXQ Regulation Number:870.1120 Review Panel: Cardiovascular
Intended Use Statement:
The Reusable Blood Pressure Cuff is an accessory used in conjunction with non-invasive blood pressure measurement systems. The cuff is non-sterile and reusable. It is available in adult size.
Device Description
The proposed device is an accessory used in conjunction with non-invasive blood pressure measurement systems, and wrap around the patient's limb. The cuff is non-sterile and reusable.
The model difference of the proposed device includes bladder type, tube, color, joint model and size.
#### న్. Identification of Predicate Device(s)
Predicate Device 1 510(k) Number: K133117 Product Name: Andon Blood Pressure Cuff
Predicate Device 2 510(k) Number: K171889 Product Name: Reusable Blood Pressure Cuff
6. Non-Clinical Test Conclusion
6.1 Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
{6}------------------------------------------------
- > IEC 80601-2-30: 2018. Medical Electrical Equipment - Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Non-Invasive Sphygmomanometers.
- > ISO 81060-1:2007 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type
- > ISO 10993-5: 2009, Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity.
- > ISO 10993-10: 2010, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
- 6.2 Human Factors / Usability Testing
Usability testing is not required per the risk profile of this product.
- Clinical Test Conclusion 7.
No clinical study is included in this submission.
- 8. Substantially Equivalent (SE) Comparison
| Item | Proposed Device<br>Cuff with Internal Bladder | Predicate Device 1<br>K133117 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device class | 2 | 2 |
| Product code | DXQ | DXQ |
| Regulation<br>Number | 21 CFR 870.1120 | 21 CFR 870.1120 |
| Intended use | The Reusable Blood Pressure<br>Cuff is an accessory used in<br>conjunction with non-invasive<br>blood pressure measurement<br>systems. The cuff is non-sterile<br>and reusable. It is available in<br>adult size. | Blood Pressure Cuffs are<br>intended to be wrapped on the<br>upper arm and used with a non-<br>invasive blood pressure<br>monitor system to complete the<br>measurement of blood<br>parameters on adults. |
| Patient<br>populations | adult | adult |
| Tube<br>configuration | One tube | One tube |
| Bladder | Internal Bladder | Internal Bladder |
Tak1a 1 Campariaan for Creff mith Internal D
{7}------------------------------------------------
| Principles of Operation | Bladder is wrapped around the patient's limb and secured by hook and loop closure. Air hose is connected to the noninvasive blood pressure measurement systems. | Bladder is wrapped around the patient's limb and secured by hook and loop closure. Air hose is connected to the noninvasive blood pressure measurement systems. |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Size | Small adult: 17-22cm Small adult long / Adult: 22-30 cm<br>Large Adult: 30-42 cm<br>Large Adult long: 42-48 cm<br>Universal: 22-42 cm | KD-525-P34: 15-24cm<br>KD-525-P31: 20-34 cm<br>KD-525-P32: 30-44 cm<br>KD-525-P33: 40-48 cm |
| Reusable | Yes | Yes |
| Sterile | No | No |
| Pressure range | 0-300mmHg | 0-300mmHg |
| Material | Polyester | Not available |
| Biocompatibility | Comply with ISO 10993-5 and ISO 10993-10 | Comply with ISO 10993-5 and ISO 10993-10 |
There are two differences for the proposed cuff and predicate device 1.
For size:
The cuff size range of proposed devices is included in the range of the predicate device 1. And the size is indicated in user manual clearly for selection. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.
For material:
The material of proposed device and predicate device 1 may be different, but both of them comply with ISO 10993-5 and ISO 10993-10. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.
| Item | Proposed Device<br>Cuff with Internal Bladder | Predicate Device 2<br>K171889 |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device class | 2 | 2 |
| Product code | DXQ | DXQ |
| Regulation<br>Number | 21 CFR 870.1120 | 21 CFR 870.1120 |
| | | |
| Intended use | The Reusable Blood Pressure<br>Cuff is an accessory used in<br>conjunction with non-invasive<br>blood pressure measurement<br>systems. The cuff is non-sterile<br>and reusable. It is available in<br>adult size. | The Reusable Blood Pressure<br>Cuff is an accessory used in<br>conjunction with noninvasive<br>blood pressure measurement<br>systems. The cuff is non-sterile<br>and reusable. It is available in<br>infant, pediatric but not<br>neonatal, and adult sizes. |
| Patient<br>populations | adult | infant, pediatric but not<br>neonatal, and adult sizes |
| Tube<br>configuration | One tube | One or Two tube |
| Bladder | Internal Bladder | Internal Bladder or Integrated<br>Bladder |
| Principles of<br>Operation | Bladder is wrapped around the<br>patient's limb and secured by<br>hook and loop closure. Air hose<br>is connected to the noninvasive<br>blood pressure measurement<br>systems. | Bladder is wrapped around the<br>patient's limb and secured by<br>hook and loop closure. Air hose<br>is connected to the noninvasive<br>blood pressure measurement<br>systems. |
| Size | Small adult: 17-22cm Small<br>adult long / Adult: 22-30 cm<br>Large Adult: 30-42 cm<br>Large Adult long: 42-48 cm<br>Universal: 22-42 cm | Small adult: 17-22cm Small<br>adult long / Adult: 22-30 cm<br>Large Adult: 30-42 cm<br>Large Adult long: 42-48 cm<br>Infant / infant long: 9-14.8cm<br>Child /child long: 13.8-21.5cm<br>Small adult / small adult long:<br>20.5-28.5cm<br>Small adult long / Adult:<br>27.5-36.5cm<br>Big adult/big adult long:<br>35.5-46cm<br>Thigh: 45-56.5cm |
| Reusable | Yes | Yes |
| Sterile | No | No |
| Pressure range | 0-300mmHg | 0-300mmHg |
| Material | Polyester | Polyester or TPU leather with<br>nylon |
| Biocompatibility | Comply with ISO 10993-5 and<br>ISO 10993-10. | Comply with ISO 10993-5 and<br>ISO 10993-10. |
## Table 2 Comparison for Cuff with Internal Bladder
{8}------------------------------------------------
There are six differences for the proposed cuff and predicate device 2.
For intended use:
{9}------------------------------------------------
The difference of intended use between proposed device and predicate device 2 is the patient population. and the patient populations of proposed devices is included in the range of the predicate device 2. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.
### For patient populations:
The patient populations of proposed devices is included in the range of the predicate device 2. And the patient populations are indicated in user manual clearly for identification. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.
## For tube configuration:
The tube configuration range of proposed devices is included in the range of the predicate device 2. And the configuration is indicated in user manual clearly for selection. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.
## For bladder:
The bladder type of proposed devices is included in the types of the predicate device 2. And the type is indicated in user manual clearly for selection. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.
### For size:
The cuff size range of proposed devices is included in the range of the predicate device 2. And the size is indicated in user manual clearly for selection. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.
### For material
The materials of proposed devices is identical to the materials of predicate device 2 with internal bladder type. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.
#### 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.