K254251 · Shanghai AnQing Medical Instrument Co., Ltd. · HIH · Feb 26, 2026 · Obstetrics/Gynecology
Device Facts
Record ID
K254251
Device Name
InnovexView (GC146-17, GC150-20, GC155-23)
Applicant
Shanghai AnQing Medical Instrument Co., Ltd.
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Feb 26, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The InnovexView is intended to be used for direct viewing of the cervical canal and the uterine cavity, or female urinary tract, including the bladder for the purpose of performing diagnostic and therapeutic surgical procedures. The types of procedures where the InnovexView could offer visualization include: • Assessment of abnormal bleeding, pelvic pain, amenorrhea and abnormal findings from hysterosalpingogram; • Assessment of infertility and pregnancy wastage; • Confirmation of the presence of intrauterine foreign body; • Assist in locating submucosal fibroids and polyps targeted for removal; • Provide visual guidance during directed biopsy, polypectomy, submucosal myomectomy, transection of intrauterine adhesions and septa. The InnovexView can also be used to permit viewing of the female adult urinary bladder through a minimally invasive approach by utilizing natural orifices for the purpose of performing diagnostic and therapeutic procedures.
Device Story
Single-use endoscope system; includes handle with deflection lever, insertion section, and cable. Captures anatomical images via distal CMOS sensor; illuminates via distal LEDs. Connects to external video processor (K243497) for signal processing and display on monitor. Used in healthcare facilities by clinicians for minimally invasive diagnostic/therapeutic procedures in uterine cavity or urinary bladder. Provides real-time visual guidance for biopsies, polypectomies, and myomectomies; assists in locating pathology. Benefits include improved visualization for minimally invasive gynecological and urological interventions.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601 series), photobiological safety (IEC 62471), mechanical/optical performance (ISO 8600), and biocompatibility (ISO 10993).
Technological Characteristics
Single-use endoscope; CMOS image sensor; LED illumination; deflection lever for distal tip control. Patient-contacting materials compliant with ISO 10993-1. Sterilized via Ethylene Oxide (SAL 10^-6). Connectivity: wired connection to external video processor. Dimensions: 4.6mm-5.5mm outer diameter; 240mm working length.
Indications for Use
Indicated for adult female patients requiring diagnostic or therapeutic visualization of the cervical canal, uterine cavity, or urinary bladder, including assessment of abnormal bleeding, pelvic pain, amenorrhea, infertility, pregnancy wastage, foreign bodies, fibroids, and polyps, or for visual guidance during surgical procedures like biopsy, polypectomy, myomectomy, and adhesion/septa transection.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 26, 2026
Shanghai AnQing Medical Instrument Co., Ltd.
Shuwen Fan
RA Manager
3 & 4 Floor, #2 Bldg., 366 Huiqing Rd.
East Zhangjiang High-Tech Park
Shanghai, 201201
CHINA
Re: K254251
Trade/Device Name: InnovexView (GC146-17, GC150-20, GC155-23)
Regulation Number: 21 CFR 884.1690
Regulation Name: Hysteroscope and accessories
Regulatory Class: II
Product Code: HIH, FAJ
Dated: December 29, 2025
Received: December 29, 2025
Dear Shuwen Fan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254251 - Shuwen Fan
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K254251 - Shuwen Fan
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JASON ROBERTS -S
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PBC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K254251 | |
| Device Name InnovexView (GC146-17, GC150-20, GC155-23) | |
| Indications for Use (Describe) The InnovexView is intended to be used for direct viewing of the cervical canal and the uterine cavity, or female urinary tract, including the bladder for the purpose of performing diagnostic and therapeutic surgical procedures. The types of procedures where the InnovexView could offer visualization include: • Assessment of abnormal bleeding, pelvic pain, amenorrhea and abnormal findings from hysterosalpingogram; • Assessment of infertility and pregnancy wastage; • Confirmation of the presence of intrauterine foreign body; • Assist in locating submucosal fibroids and polyps targeted for removal; • Provide visual guidance during directed biopsy, polypectomy, submucosal myomectomy, transection of intrauterine adhesions and septa. The InnovexView can also be used to permit viewing of the female adult urinary bladder through a minimally invasive approach by utilizing natural orifices for the purpose of performing diagnostic and therapeutic procedures. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K254251
Page 1 of 6
510(k) Premarket Notification Submission
InnovexView
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
Date Prepared: 2026-02-25
## 510(k) submitter
Company Name: Shanghai AnQing Medical Instrument Co., Ltd.
Company Address: 3 & 4 Floor, No.2 Building, 366 Huiqing Rd,
East Zhangjiang High-Tech Park, 201201, Shanghai, China
## Contact
Name: Shuwen Fan
Telephone: +86-21-61117375
Email: ra_dept@anqing-sh.com
## Device Identification
Device trade name: InnovexView (GC146-17, GC150-20, GC155-23)
Common Name: Hysteroscope and accessories
Classification Name: Hysteroscope (And Accessories)
Regulatory Class: Class II
Regulation Number: 21 CFR 884.1690
Product Code(s): HIH, FAJ
## Legally Marketed Predicate Device
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K200038 | Endosee System | HIH |
The predicate device has not been subject to a design-related recall.
## Device Description Summary
InnovexView (Model: GC146-17, GC150-20, GC155-23) is intended to be used with the video processor (model: EA101 cleared via K243497). When used with the compatible video processor and monitor, the endoscope system can be operated as intended and indicated. InnovexView is a single-use endoscope, which consists of a Handle, an Insertion Section, and an Endoscope Cable. The handle includes a deflection lever, an inflow port and an outflow port, to which luer connectors with flow control valve can be connected, a working channel port to which, a biopsy valve can be connected. The insertion section contains a working channel for access of endoscopic accessories and distention media
{5}
K254251
Page 2 of 6
510(k) Premarket Notification Submission
InnovexView
outflow, space for inflow, and wiring to transmit the image signals to the video processor. The bending section at the distal portion of the insertion section is controlled by the user via the deflection lever on the handle. The distal tip of the insertion section contains a CMOS image sensor for capturing the anatomical image and transmitting video signals to the video processor, LEDs for illumination, and the distal openings of working channel and inflow. The endoscope cable connects the endoscope handle to the video processor for transmitting video signals and providing power to the LEDs.
## Mechanism of action:
The light emitted from the distal tip of the InnovexView is irradiated into the body cavity, and the light reflected from the cavity enters the optical system and is captured by the CMOS image sensor. The CMOS acquisition image is controlled by the CMOS drive circuit, and the RGB video signal is output to the video processor. The video processor receives video signals from the endoscope, processes the video signals, and outputs the processed video signal to the build in monitor or attached monitor. The video processor also controls the brightness of the LED on the endoscope.
InnovexView has the following physical and performance characteristics:
- Maneuverable tip controlled by the user
- Rigid insertion cord
- LED and CMOS image sensor at the distal tip
- Sterilized by Ethylene Oxide
- For single use
## Intended Use/Indications for Use
The InnovexView is intended to be used for direct viewing of the cervical canal and the uterine cavity, or female urinary tract, including the bladder for the purpose of performing diagnostic and therapeutic surgical procedures.
The types of procedures where the InnovexView could offer visualization include:
- Assessment of abnormal bleeding, pelvic pain, amenorrhea and abnormal findings from hysterosalpingogram;
- Assessment of infertility and pregnancy wastage;
- Confirmation of the presence of intrauterine foreign body;
- Assist in locating submucosal fibroids and polyps targeted for removal;
- Provide visual guidance during directed biopsy, polypectomy, submucosal myomectomy, transection of intrauterine adhesions and septa.
The InnovexView can also be used to permit viewing of the female adult urinary bladder through a minimally invasive approach by utilizing natural orifices for the purpose of performing diagnostic and therapeutic procedures.
{6}
K254251
Page 3 of 6
510(k) Premarket Notification Submission
InnovexView
# Indications for Use Comparison
The InnovexView device has the same intended use as the predicate Endosee System (K200038). Both devices are intended to visualize the uterine cavity for diagnostic and therapeutic procedures, including the removal of fibroids and polyps. The indications of the subject device are within that of the subject device. Slight difference in wording does not result in new intended use of the subject device.
## Comparison of Technological Characteristics
| | Subject Device,
InnovexView (K254251) | Predicate Device,
Endosee System (K200038) |
| --- | --- | --- |
| | Physical Characteristics | |
| Outer diameter | 4.6mm(13.8Fr) to
5.5mm(16.5Fr) depending on
model | 4.5mm |
| Working Length | 240mm | 276mm |
| Working Channel | Yes | Yes |
| Deflection range | 35°left/35°right | N/A |
| | Optical Characteristics | |
| Type of Imager | CMOS | CMOS |
| Field of View | 110° | 100° |
| Direction of View | Forward viewing | 20° |
| Depth of Field | 5mm~100mm | 5mm~90mm |
| Light Source | LED at distal tip | LED at tip of cannula |
| | Patient Contacting Materials | |
| General material
type of main
patient-contact part | Compliance with ISO10993-1 | Cannula Compliance with
ISO10993-1 |
| Duration and type
of contact | “Surface –Mucosal Membrane”
with a contact duration of
“Limited (< 24 hours)” | Cannula “Surface –Mucosal
Membrane” with a contact
duration of “Limited (< 24
hours)” |
| | Sterilization Methods | |
| Number of Users | Single-Use | Single-Use Cannula |
| Sterilization | EO Sterilized,
SAL 10^{-6} | EO Sterilized Cannula, SAL
10^{-6} |
| | Technological Characteristics | |
| Environment of use | Healthcare facility/hospital | Healthcare facility/hospital |
| Energy source | Electricity | Electricity |
3 / 6
{7}
K254251
Page 4 of 6
510(k) Premarket Notification Submission
InnovexView
| | Subject Device,
InnovexView (K254251) | Predicate Device,
Endosee System (K200038) |
| --- | --- | --- |
| Video Processor | External video processor
previously cleared, not included
in current 510k. | Handheld, battery-operated a
reusable, detachable Display
Module as an integral part of
the Endosee System. |
| Note:
The subject and predicate devices have the same fundamental technology, type of
imager, number of uses and sterilization method.
The subject device differs from the predicate in working length, distal end outer
diameter, deflection, field of view, direction of view, depth of field and patient-
contacting materials. These differences do not raise different questions of safety and
effectiveness as compared to the predicate and have been evaluated through
performance testing. | | |
# Non-Clinical Performance
1. Electrical Safety and Electromagnetic Compatibility Summary
The electrical safety and EMC data included in the submission is in compliance with the following FDA recognized standards:
- ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021]
- ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021]
- IEC 60601-2-18 Edition 3.0 2009-08
- IEC/TS 60601-4-2 Edition 1.0 2024-03
2. Photobiological safety
The proposed device was tested according to the following FDA recognized standards:
- IEC 62471:2006 Medical electrical equipment, Photobiological safety of lamps and lamp systems
3. Mechanical and Optical Performance
The proposed device was designed to comply with applicable parts of ISO 8600. Optical measurements were performed according to applicable part of ISO 8600 standard.
Nonclinical mechanical, functional, and performance testing included:
- Appearance (visual inspection)
- Dimensional characteristics
- Deflection range
- Water resistance (leakage tightness)
- Angle control system performance
- Stiffness
{8}
K254251
Page 5 of 6
510(k) Premarket Notification Submission
InnovexView
- Torque transmission
- Inflow and outflow performance
- Connection compatibility
- Instrument channel compatibility
- Tensile strength (insertion tube–handle joint)
- Connection force (endoscope connector–video processor)
- Connection force (handle–cable interface)
- Bending fatigue
- Coaxiality
- Optical performance (field of view, direction of view, resolution, depth of field, image quality, illumination)
- Electrical safety testing
Comparative optical performance testing, including evaluation of image quality parameters such as modulation transfer function (MTF), color performance, signal-to-noise ratio(SNR)/dynamic range, distortion, and image intensity uniformity, was performed to support substantial equivalence.
## 4. Biocompatibility
The biocompatibility evaluation for the patient-contacting components of the proposed device was performed in accordance with ISO 10993-1:2018, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, and FDA guidance. The following testing was conducted using EO-sterilized, finished devices:
- **Cytotoxicity (MEM Elution):**
Conducted in accordance with ISO 10993-5:2009, Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity, and USP <87> Biological Reactivity Tests, In Vitro. Extraction conditions were prepared consistent with ISO 10993-12:2012, Sample preparation and reference materials. Results demonstrated slight reactivity (Grade 1), within acceptable limits.
- **Skin Sensitization (Guinea Pig Maximization Test):**
Performed in accordance with ISO 10993-10:2021, Tests for irritation and skin sensitization. Polar and non-polar extracts were evaluated under exaggerated conditions. No evidence of sensitization was observed.
- **Intracutaneous Reactivity (Irritation):**
Conducted in accordance with ISO 10993-23:2021, with extraction conditions prepared per ISO 10993-12:2021. Mean irritation scores were below the threshold for classification as an irritant.
- **Acute Systemic Toxicity:**
Performed in accordance with ISO 10993-11:2017, Tests for systemic toxicity. No treatment-related mortality or clinical signs of toxicity were observed.
5 / 6
{9}
K254251
Page 6 of 6
510(k) Premarket Notification Submission
InnovexView
- **Material-Mediated Pyrogenicity:**
Evaluated in accordance with ISO 10993-11:2017 and USP <151> Pyrogen Test. No pyrogenic response was observed.
All testing met applicable acceptance criteria. No adverse biological responses were identified under the conditions of the studies. Collectively, these results demonstrate that the materials and manufacturing processes used in the device do not present unacceptable biological risks for its intended short-term mucosal contact.
5. Sterilization and shelf life testing
The sterilization method has been validated to ISO 11135:2014, which has thereby determined the routine control and monitoring parameters.
EO/ECH residual test was performed according to ISO 10993-7:2008.
The shelf life of the proposed device is determined based on stability study which includes aging test according to ASTM F1980-21, Standard Guide for Accelerated Aging of Sterile Barrier.
6. Package Validation
Package validation was conducted according to ISO 11607-1:2019 [including AMD1:2023] and ISO 11607-2:2019 [including AMD1:2023], and ASTM F88/F88M-23, ASTM F1929-23.
Transport and shipping testing as per ASTM D4169-22.
**Summary of Clinical Tests**
Clinical Tests Not Applicable.
**Conclusion**
The conclusions drawn from the nonclinical tests demonstrate that the subject device, the InnovexView is as safe, as effective, and performs as well as the legally marketed device identified above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.