Endosee System

K200038 · CooperSurgical, Inc. · HIH · Feb 5, 2020 · Obstetrics/Gynecology

Device Facts

Record IDK200038
Device NameEndosee System
ApplicantCooperSurgical, Inc.
Product CodeHIH · Obstetrics/Gynecology
Decision DateFeb 5, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 884.1690
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Endosee System is used to permit viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and therapeutic procedures. The types of procedures where the Endosee System could offer visualization include: Abnormal bleeding Infertility and pregnancy wastage Evaluation of abnormal hysterosalpingogram Intrauterine foreign body Amenorrhea Pelvic pain Directed biopsy Removal of fibroids and polyps Transection of intrauterine adhesions Transection of intrauterine septa The Endosee System can also be used to permit viewing of the adult urinary bladder through a minimally invasive approach by utilizing natural orifices for the purpose of performing diagnostic and therapeutic procedures.

Device Story

Handheld, battery-operated endoscope; consists of reusable LCD display module and sterile, single-use cannula. Cannula features distal LED light source and camera; captures video/images of cervical canal, uterine cavity, or bladder. Signals transmitted via electrical connector to display module for real-time visualization. Used in OR, outpatient, or office settings by clinicians. Docking station recharges battery and facilitates image/video download to external computers. Enables diagnostic and therapeutic procedures; assists clinicians in identifying pathologies and performing interventions; improves patient outcomes via minimally invasive visualization.

Clinical Evidence

Bench testing only. Verification testing included biocompatibility (ISO 10993), sterilization (ISO 11135), shelf-life (ASTM F1980), electrical safety (ANSI/AAMI ES 60601-1), EMC (IEC 60601-1-2), and performance testing (optical distortion, flow rate, bend tests, drop tests, IPX2 ingress). Software and cybersecurity verification performed per IEC 62304 and FDA guidance.

Technological Characteristics

Handheld endoscope; reusable LCD display module; sterile, single-use flexible cannula (276mm length, 4.5mm OD). Materials: PEEK, glass, silicone, polycarbonate, ABS, TPE, copper. LED light source. Connectivity: Wi-Fi, electrical cable to display. Sterilization: Ethylene oxide (SAL-6). Power: Battery-operated with docking station. Software: Embedded firmware.

Indications for Use

Indicated for adult patients requiring visualization of the cervical canal, uterine cavity, or urinary bladder for diagnostic and therapeutic procedures, including evaluation of abnormal bleeding, infertility, pregnancy wastage, abnormal hysterosalpingograms, intrauterine foreign bodies, amenorrhea, pelvic pain, and performance of biopsies or surgical interventions like fibroid/polyp removal and adhesion/septa transection.

Regulatory Classification

Identification

A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. February 5, 2020 CooperSurgical, Inc. Christine Kupchick Regulatory Affairs Associate 95 Corporate Drive Trumbull, CT 06611 Re: K200038 Trade/Device Name: Endosee System Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: Class II Product Code: HIH, FAJ Dated: January 7, 2020 Received: January 8, 2020 ## Dear Christine Kupchick: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jason Roberts, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200038 Device Name Endosee System Indications for Use (Describe) The Endosee System is used to permit viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and therapeutic procedures The types of procedures where the Endosee System could offer visualization include: - Abnormal bleeding - Infertility and pregnancy wastage - · Evaluation of abnormal hysterosalpingogram - · Intrauterine foreign body - · Amenorrhea - · Pelvic pain - · Directed biopsy - Removal of fibroids and polyps - · Transection of intrauterine adhesions - · Transection of intrauterine septa The Endosee System can also be used to permit viewing of the adult urinary bladder through a minimally invasive approach by utilizing natural orifices for the purpose of performing diagnostic and therapeutic procedures. Type of Use (Select one or both, as applicable) | <div> <span> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |---------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K200038 Image /page/3/Picture/2 description: The image shows the CooperSurgical logo. The logo is in blue and features the company name in a stylized font. Below the company name is the tagline "Healthy women, babies, and families™". # 510(k) SUMMARY ### 510(k) Submitter | Company Name: | CooperSurgical Inc. | |------------------|-----------------------------------------| | Company Address: | 95 Corporate Drive<br>Trumbull CT 06611 | ### Contact | Name: | Christine Kupchick | |------------|---------------------------------------| | Telephone: | 203-601-5200 Ext. 3370 | | Fax: | 203-601-9870 | | Email: | christine.kupchick@coopersurgical.com | Date Prepared: January 07, 2020 ### Device Identification | Trade Name: | Endosee® System | |--------------------|------------------------------| | Common Name: | Hysteroscope; Cystoscope | | Regulation Number: | 21 CFR 884.1690 | | Regulation Name: | Hysteroscope and Accessories | | Product Code: | HIH; FAJ | | Regulatory Class: | Class II | ### Predicate Device Information The subject Endosee System is substantially equivalent to the predicate Endosee System (K190639). The predicate device has not been subject to a design-related recall. ### Device Description The Endosee System is a handheld, battery-operated, portable endoscope that consists of a reusable Display Module with an LCD touchscreen monitor and a sterile, single-use cannula. It is intended for use in viewing the adult cervical canal, uterine cavity, or female urinary tract, including the bladder, during diagnostic or therapeutic hysteroscopy or cystoscopy procedures in an operating room, outpatient, or office setting. The Cannula has a light source and camera at the distal end, which are used for visualization and to capture image and video of the diagnostic area. The image and video signals are electronically transferred from the Cannula to the Display module via an electrical connector and cable, and the LCD monitor is used for viewing. The Endosee is provided with a Docking Station that recharges the battery in the Display Module and allows the user to download images/video from the internal device memory to an external computer. {4}------------------------------------------------ ### Indications for Use The Endosee System is used to permit viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and therapeutic procedures. The types of procedures where the Endosee System could offer visualization include: - Abnormal bleeding - Infertility and pregnancy wastage - Evaluation of abnormal hysterosalpingogram ● - . Intrauterine foreign body - . Amenorrhea - Pelvic pain - Directed biopsy - Removal of fibroids and polyps - . Transection of intrauterine adhesions - Transection of intrauterine septa The Endosee System can also be used to permit viewing of the adult urinary bladder through a minimally invasive approach by utilizing natural orifices for the purpose of performing diagnostic and therapeutic procedures. ### Substantial Equivalence Discussion | Attribute | Subject Endosee System | Predicate Endosee System | |---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | CooperSurgical, Inc. | CooperSurgical, Inc. | | Indications for Use | The Endosee System is used to permit viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and therapeutic procedures. The types of procedures where the Endosee System could offer visualization include: abnormal bleeding, infertility and pregnancy wastage, evaluation of abnormal hysterosalpingogram, intrauterine foreign body, amenorrhea, pelvic pain, directed biopsy, removal of fibroids and polyps, transection of intrauterine adhesions, and transection of intrauterine septa. The Endosee System can also be used to permit viewing of the adult urinary bladder through a minimally invasive approach by utilizing natural orifices for the purpose of | The Endosee System is used to permit viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic and therapeutic procedures. The types of procedures where the Endosee System could offer visualization include: abnormal bleeding, infertility and pregnancy wastage, evaluation of abnormal hysterosalpingogram, intrauterine foreign body, amenorrhea, pelvic pain, directed biopsy, removal of fibroids and polyps, transection of intrauterine adhesions, and transection of intrauterine septa. The Endosee System can also be used to permit viewing of the adult urinary bladder through a minimally invasive approach by utilizing natural orifices for the purpose of | ### Table 1: Substantial Equivalence Comparison {5}------------------------------------------------ | | performing diagnostic and<br>therapeutic procedures. | performing diagnostic and<br>therapeutic procedures. | |------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Handheld, battery-operated<br>endoscope that consists of a<br>reusable, detachable Display<br>Module and sterile, disposable<br>Cannula with a handle; Wi-Fi<br>capability. | Handheld, battery-operated<br>endoscope that consists of a<br>reusable, detachable Display<br>Module and sterile, disposable<br>Cannula with a handle; Wi-Fi<br>capability. | | Fundamental<br>Technology | | | | Cannula Type<br>and Dimensions | Type: Flexible<br>Working Length: 276mm<br>Largest Outer Diameter: 4.5mm | Type: Flexible<br>Working Length: 276mm<br>Largest Outer Diameter: 4.5mm | | Cannula<br>Working<br>Channel | Working channel for fluid infusion<br>and instrument access | Working channel for fluid infusion<br>and instrument access | | View | Field of View: $100° ± 5°$<br>Direction of View: $20° ± 3°$ | Field of View: $100° ± 5°$<br>Direction of View: $20° ± 3°$ | | Light Source | LED | LED | | Image Display | Handheld LCD display module | Handheld LCD display module | | Cannula<br>Patient-<br>Contacting<br>Materials | Direct: PEEK, ink, glass, adhesive,<br>lens coating, surface black coating,<br><br>Indirect: ABS, TPE, polycarbonate,<br>adhesive, ink, silicone, PVC (non-<br>phthalate), copper | Direct: PEEK, ink, glass, adhesive,<br>lens coating, surface black coating,<br><br>Indirect: ABS, TPE, polycarbonate,<br>adhesive, ink, silicone, PVC, copper | | Number of Uses | Cannula: single-use, disposable<br>Display Module: reusable | Cannula: single-use, disposable<br>Display Module: reusable | | Cannula<br>Sterilization | Ethylene oxide, SAL-6 | Ethylene oxide, SAL-6 | The subject and predicate devices have the same intended use. The subject and predicate devices have the same fundamental technology, cannula type and dimensions, cannula working channel, field and direction of view, light source, image display, number of uses and sterilization. The subject Cannula differs from the predicate in indirect patient-contacting materials and an additional layer of shielding on the Cannula board. These differences do not raise different questions of safety and effectiveness as compared to the predicate. ### Non-Clinical Performance As part of demonstrating substantial equivalence to the predicate, a risk analysis was completed to identify the risks associated with the Endosee Cannula material changes and additional shielding. Verification testing were conducted to evaluate the modifications. The following tests associated with the device modifications were performed on the subject device according to methods and acceptance criteria outlined in the predicate device (K190639). The subject device passed all the testing in accordance with internal requirements and applicable standards to support substantial equivalence of the subject device. {6}------------------------------------------------ #### ● Sterilization Evaluation per ISO 11135:2014 - o Bioburden per ISO 11737-1:2018 #### ● Biocompatibility Testing per ISO 10993-1:2009 - Cytotoxicity per ISO 10993-5:2009 о - Irritation per ISO 10993-10:2010 o - Sensitization per ISO 10993-10:2010 o #### . Stability and Shelf Life of Cannula The following tests were performed on samples accelerated aged to 6-months per ASTM F1980-16: - o Leak Test #### Performance Testing . - o Fixed Pattern Noise Test The following testing was leveraged from the predicate device (K190639). Test results from the predicate were used to support the subject device because the conditions were identical or the subject device modifications did not introduce a new worst-case configuration or scenario for testing. #### ● Sterility Testing per ISO 11135-1:2014 - o EO Residuals per ISO 10993-7:2008 #### Stability and Shelf Life of Cannula . The following tests were performed samples accelerated aged to 6-months per ASTM F1980-16: - o Seal Tensile Strength - Seal Peel o - Performance (Image Quality, Flow Rate, Leak, Bend, Tip Retention, Instrument o Access) #### ● Ship Testing per ISTA 3A:2008 - Bubble Leak per ASTM 2096-11 #### ● EMC and Electrical Safety - UL Electrical Safety per ANSI/AAMI ES 60601-1:2005/(R)2012 and IEC 60601-2 18:2009 with Essential Performance per IEC 60601-2-18 - o UL EMC per IEC 60601-1-2:2014 and IEC 60601-2-18:2009 with Essential Performance per IEC 60601-2-18 #### ● Performance Testing - Optical Distortion Test o - Saline Flow Rate Test o - O 90° Bend Test - 3-Point Bend Test O - LED Temperature Test o {7}------------------------------------------------ - Drop Test per IEC 60601-1-15 o - o IPX2 Ingress Test per IEC 60529:2004 - System Level Verification - . Software Verification per IEC 62304:2006 and as recommended in the FDA guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" - . Cybersecurity/Interoperability Information provided as recommended in FDA guidance, "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" - . Wireless information provided as recommended in FDA guidance, "Radio Frequency Wireless Technology in Medical Devices Guidance" - Design Validation ### Conclusion The results of the testing described above demonstrate that the subject Endosee System is as safe and effective as the predicate and supports a determination of substantial equivalence.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...