LUSTRE GoGlow Advanced LED Patches (PR8001, PR9001)
K253878 · Shenzhen Kaiyan Medical Equipment Co., Ltd. · OLP · Mar 4, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253878
Device Name
LUSTRE GoGlow Advanced LED Patches (PR8001, PR9001)
Applicant
Shenzhen Kaiyan Medical Equipment Co., Ltd.
Product Code
OLP · General, Plastic Surgery
Decision Date
Mar 4, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
LUSTRE GoGlow Advanced LED Patches (Model: PR8001, PR9001) is an Over-the-Counter (OTC) device that emits energy in the blue and red spectrum for the treatment of mild to moderate inflammatory acne.
Device Story
LUSTRE GoGlow Advanced LED Patches (PR8001, PR9001) are OTC light-based devices for treating mild to moderate inflammatory acne on the face. Each patch contains 24 LEDs (12 blue at 415nm, 12 red at 630nm) delivering approximately 30 mW/cm² intensity. The system includes two LED patches, two rechargeable lithium-ion battery controllers, a charging case, and adhesive 'Mini Dots' for skin attachment. Operation involves connecting the controller to the patch via magnets, applying to the skin, and pressing the power button. The device delivers a 3-minute treatment session and automatically shuts off. The output is light energy intended to reduce acne inflammation. The device is designed for patient self-use at home.
Clinical Evidence
No clinical data was provided. Substantial equivalence is supported by non-clinical bench testing, including electrical safety (IEC 60601-1, IEC 60601-1-11, IEC 60601-2-83), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), and battery safety (IEC 62133-2). Biocompatibility was established by demonstrating identical materials and processing to the predicate device.
Technological Characteristics
LED-based light therapy device; 24 LEDs per patch (12 blue 415nm, 12 red 630nm); Irradiance ~30 mW/cm²; Powered by 3.7Vdc 200mAh Li-ion battery; Dimensions vary by model (PR8001 bean-shaped, PR9001 triangle); Adheres to skin via acrylate Mini Dots; Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-83, IEC 62471, and IEC 62133-2; Biocompatibility per ISO 10993-1, -5, -10, -11, -23.
Indications for Use
Indicated for the treatment of mild to moderate inflammatory acne in patients using the device as an Over-the-Counter (OTC) product.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 4, 2026
Shenzhen Kaiyan Medical Equipment Co., Ltd.
Dijkstra Alain
CEO
Bldg.#3 And Bldg.#5, 40th Of Fuxin St.
Huaide Community Fuyong Town
Shenzhen, Guangdong 518103
China
Re: K253878
Trade/Device Name: LUSTRE GoGlow Advanced LED Patches (PR8001, PR9001)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OLP
Dated: November 21, 2025
Received: December 4, 2025
Dear Dijkstra Alain:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253878 - Dijkstra Alain
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253878 - Dijkstra Alain
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2026.03.04
16:02:16 -05'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253878 | ? |
| Please provide the device trade name(s). | | ? |
| LUSTRE GoGlow Advanced LED Patches (PR8001, PR9001) | | |
| Please provide your Indications for Use below. | | ? |
| LUSTRE GoGlow Advanced LED Patches (Model: PR8001, PR9001) is an Over-the-Counter (OTC) device that emits energy in the blue and red spectrum for the treatment of mild to moderate inflammatory acne. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary of K253878
This summary of 510(k) is being submitted in accordance with the requirement of 21 CFR 807.92.
1. Submitter's Information
Company Name: Shenzhen Kaiyan Medical Equipment Co., Ltd
Establishment Registration Number: 3011644607
Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China
Contact Person (including title): Alain Dijkstra(CEO)
E-mail: registrar01@kaiyanmedical.com
Manufacturer:
Company Name: Lustre Skin Ltd
Establishment Registration Number: 3018188770
Address: Alba Innovation Centre, Alba Campus, Livingston Scotland, UK, EH547GA
Phone: +44 (0)1506 592239
Contact Person (including title): Blessing Nwabude, Head of Regulatory
Email: Blessing.Nwabude@Lustreskin.com
Application Correspondent:
Contact Person: Alain Dijkstra
Company: Shenzhen Kaiyan Medical Equipment Co., Ltd
Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China
Tel: +86 755 82129361
Fax: +86 755 25024651
Email: registrar01@kaiyanmedical.com
2. Subject Device Information:
Device/Trade Name: LUSTRE GoGlow Advanced LED Patches(Model: PR8001, PR9001)
Classification Name: Over-The-Counter Powered Light Based Laser For Acne
Review Panel: General & Plastic Surgery
Product Code: OLP
Regulation Number:21 CFR 878.4810
Regulation Class: II
3. Predicate Device Information
Predicate Device (K230720)
Sponsor: Light Tree Ventures Europe B.V.
Trade Name: LUSTRE 3XPRESS Light Beauty Therapy patches, model: PR6001, PR5001
Classification Name: Over-The-Counter Powered Light Based Laser For Acne.
Review Panel: General & Plastic Surgery
Product Code: OLP
Regulation Number: 21 CFR 878.4810
1 / 5
{5}
Regulation Class: II
# Reference Device(K160691)
Sponsor: Zuko, Inc.
Trade Name: Acne Light Therapy Wand
Classification Name: Over-The-Counter Powered Light Based Laser For Acne.
Review Panel: General & Plastic Surgery
Product Code: OLP
Regulation Number: 21 CFR 878.4810
Regulation Class: II
# 4. Device Description
The LUSTRE GoGlow Advanced LED Patches (Model: PR8001, PR9001) is an over-the-counter light emitting diode (LED) device indicated for the treatment of mild to moderate acne. It works by emitting energy in the blue (415nm) and red (630nm) spectrum. Each patch incorporates 24 LEDs with a power intensity of approximately 30 mW/cm².
The two models differ only in shape: the PR9001 is triangle, while the PR8001 is bean-shaped.
The device package includes two LED patches, two rechargeable controllers, a charging case, a USB cable, 56 Mini Dots (for holding the patch safely to the skin), a pair of goggles and a user manual. To use, simply connect a controller to each patch via the built-in magnets, apply the patches to the targeted skin area using the Mini Dots, then press and hold the power button on the controller to activate the device and start the treatment session. The patches will automatically shut off after the 3-minute session is complete.
# 5. Intended Use / Indications for Use
LUSTRE GoGlow Advanced LED Patches (Model: PR8001, PR9001) is an Over-the-Counter (OTC) device that emits energy in the blue and red spectrum for the treatment of mild to moderate inflammatory acne.
# 6. Comparison to Predicate Devices
Compare with the predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards.
| Elements of Comparison | Subject device | Predicate device (K230720) | Reference device(K160691) | Remark |
| --- | --- | --- | --- | --- |
| Company | Shenzhen Kaiyan Medical Equipment Co., Ltd | Light Tree Ventures Europe B.V. | Zuko, Inc. | -- |
| Trade Name | LUSTRE GoGlow Advanced LED Patches | LUSTRE 3XPRESS Light Beauty Therapy patches | Acne Light Therapy Wand | -- |
| Model | PR8001, PR9001 | PR5001, PR6001 | unknown | -- |
| 510 (K) Number | Applying | K230720 | K160691 | -- |
| Product Code | OLP | OLP | OLP | Same |
| Regulation Class | Class II | Class II | Class II | Same |
| Indications for | LUSTRE GoGlow | The LUSTRE 3XPRESS | The Acne Light | Same |
{6}
| Elements of Comparison | Subject device | Predicate device (K230720) | Reference device(K160691) | Remark |
| --- | --- | --- | --- | --- |
| Use / Intended use | Advanced LED Patches (Model: PR8001, PR9001) is an Over-the-Counter (OTC) device that emits energy in the blue and red spectrum for the treatment of mild to moderate inflammatory acne. | Light Beauty Therapy patches (Model: PR6001, PR5001) is an Over-the-Counter (OTC) device intended for treatment of mild to moderate inflammatory acne. | Therapy Wand is indicated to treat mild to moderate inflammatory acne. | |
| Power Source | Controller Lithium-ion Battery: 3.7Vdc, 200mAh, 0.74Wh | Input: 100-240Va.c., 50Hz/60Hz, output: DC 5V, 2A | 1 AAA alkaline batteries | Different Note 1 |
| Wavelengths | 630 ± 10nm
415 ± 10nm | 630 ± 10nm
415 ± 10nm | 442 ± 4nm
633 ± 4nm | Same |
| Irradiance Source | LED | LED | LED | Same |
| LED Number | For each model:
Blue: 12
Red: 12
Total: 24 | For each model:
Blue: 12
Red: 12
Total :24 | unknown | Same |
| Intended Location Use | Face | Face | Face | Same |
| Irradiance (mW/cm²) | Red: 5-8mW/cm²
Blue: 23-27mW/cm²
Total: 28-35mW/cm² | Red: 5
Blue: 25
Total:30 | Red:7.5
Blue:26.5
Total:34.0 | Similar Note 2 |
| Dose / per time(J/cm²) | 5.04-6.3 | 5.4 | 12.2 | Similar Note 2 |
| Treatment Time | 3 minutes per treatment | 3 minutes per treatment | unknown | Same |
| EMC | IEC60601-1-2 | IEC60601-1-2 | IEC60601-1-2 | Same |
| Safety | IEC 60601-1
IEC 60601-2-83
IEC 60601-1-11
IEC 62471
IEC 60601-1-6 | IEC60601-1
IEC60601-2-57
IEC60601-1-11
IEC62471 | IEC60601-1
IEC62471 | Same |
| Biocompatibility | ISO 10993-1
ISO 10993-5
ISO 10993-10
ISO 10993-23
ISO 10993-11 | ISO 10993-1
ISO10993-5
ISO10993-10
ISO 10993-23
ISO 10993-11 | unknown | Same |
3 / 5
{7}
Comparison in Detail(s):
Note 1: Although the power supply of the subject device differs slightly from the predicate device, they all comply with the safety requirements of IEC 60601-1, IEC 60601-1-2 and IEC 62133-2. Therefore, this minor difference does not raise any safety or effectiveness issues.
Note 2: The subject device's irradiance and dose/per time are slightly different from those of the predicate device and reference device, but they are very close. This minor difference does not raise any safety or effectiveness issues.
7. Test Summary
7.1 Non-Clinical Tests Performed
1) Electrical safety, and electromagnetic compatibility Test
Non-clinical testing was performed on the subject device to validate the design and ensure compliance with the following standards related to medical device electrical safety, electromagnetic compatibility, photobiological safety and battery safety:
- IEC 60601-1 2020-08 Edition 3.2 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-11:2015+AMD1:2020 Edition 2.1 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance --Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment.
- IEC 60601-2-83:2019+AMD1:2022 Edition 1.1 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
- IEC 60601-1-2 Edition 4.1 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
- IEC TS 60601-4-2:2024 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems.
- IEC 62133-2:2017+AMD1:2021 Edition1.1 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems.
2) Biocompatibility Test
The Acrylate Mini Dots of the LUSTRE GoGlow Advanced LED Patches (Model: PR8001, PR9001) in its final finished form is identical to Mini Dots of the predicate device, the “LUSTRE 3XPRESS Light Beauty Therapy patches(model: PR6001, PR5001)” in formulation, processing, sterilization, and geometry, and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents). The predicate device is manufactured by Shenzhen Kaiyan Medical Equipment Co., Ltd and was cleared under K230720, dated July 3, 2023. There is no change in biocompatibility since the previously cleared devices. Therefore, based on this information, no biocompatibility testing was needed to support substantial equivalence of this device.
3) Software Verification and Validation
{8}
Software verification and validation testing consistent with basic documentation were conducted in accordance with the recommendations provided in the FDA guidance document "Content of Premarket Submissions for Device Software Functions." The corresponding documentation has been provided.
## 4) Usability Validation
Usability testing was conducted on the LUSTRE GoGlow Advanced LED Patches (Models: PR8001, PR9001), which complies with IEC 62366-1 and IEC 60601-1-6.
## 7.2 Summary of Clinical Performance
Clinical testing was not needed for this 510(k). The non-clinical performance testing provides sufficient evidence to support that the device can be used safely and effectively.
## 8. Date Prepared: February 11, 2026
## 9. Final Conclusion
The subject device is as safe, as effective, and performs as well as the legally marketed predicate device K230720 and K160691.
5 / 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.