K253666 · Shenzhen Mlay Intelligent Technology Co., Ltd. · OHT · Feb 23, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253666
Device Name
IPL Hair Removal Device (Models: T31A, T32A)
Applicant
Shenzhen Mlay Intelligent Technology Co., Ltd.
Product Code
OHT · General, Plastic Surgery
Decision Date
Feb 23, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IPL Hair Removal Device is an over-the-counter device indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long term, stable reduction in the amount of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Device Story
Personal, light-based hair reduction device; intended for OTC home use. Inputs: skin contact and skin tone via integrated sensors. Operation: Quartz glass Xenon lamp emits intense pulsed light (IPL) to reduce hair growth; semiconductor cooling chip (Peltier effect) with sapphire cooling panel provides skin cooling. Features: auto-recognition skin color mode adjusts intensity; safety interlock prevents light emission unless full skin contact is detected; automatic mode for continuous flashing. Output: light pulses. Benefit: hair reduction with minimal pain; no cutting or pulling. User: patient self-use. Healthcare provider role: none; device is OTC.
Clinical Evidence
Bench testing only. No clinical data provided. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, irritation, sensitization). Electrical and EMC safety verified per IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-11, 60601-2-57, and 60601-2-83. Photobiological safety verified per IEC 62471.
Technological Characteristics
Xenon Arc Flashlamp (510-1200nm); max fluence 6.48J/cm2; pulse duration 0.4-0.92ms (multipulse). Cooling: semiconductor chip (Peltier effect) with sapphire cooling panel. Sensors: skin contact and skin tone. Power: external adapter. Operation: finger switch or automatic mode. Biocompatibility per ISO 10993. Standards: IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-11, 60601-2-57, 60601-2-83, IEC 62471.
Indications for Use
Indicated for removal of unwanted hair and permanent reduction in hair regrowth in adults. Suitable for use on small areas (e.g., underarms, facial hair below chin line) and large areas (e.g., legs).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 23, 2026
Shenzhen Mlay Intelligent Technology Co., Ltd.
Libin Liu
General Manager
201, 301, Bldg. 28, Cuigang Industrial Zone 2
Huaide Community, Fuyong St., Baoan District
Shenzhen, Guangdong 518103
China
Re: K253666
Trade/Device Name: IPL Hair Removal Device (Models: T31A, T32A)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHT
Dated: November 20, 2025
Received: November 21, 2025
Dear Libin Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253666 - Libin Liu
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253666 - Libin Liu
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YAN FU-S
Digitally signed by YAN FU-S
Date: 2026.02.23 16:41:49
-05'00'
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253666 | |
| Device Name IPL Hair Removal Device (Models: T31A, T32A) | |
| Indications for Use (Describe) The IPL Hair Removal Device is an over-the-counter device indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long term, stable reduction in the amount of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K253666 510(k) Summary Prepared on: 2026-02-22
Contact Details 21 CFR 807.92(a)(1)
| Applicant Name | Shenzhen Mlay Intelligent Technology Co.,Ltd. |
| --- | --- |
| Applicant Address | 201, 301, Building 28, Cuigang Industrial Zone 2, Huaide Community, Fuyong Street, Baoan District, Shenzhen City, Guangdong Province, China. Shenzhen 518103 China |
| Applicant Contact Telephone | +86 18344147065 |
| Applicant Contact | Mr. LIBIN LIU |
| Applicant Contact Email | wuxiaotian@mlays.com |
Device Name 21 CFR 807.92(a)(2)
| Device Trade Name | IPL Hair Removal Device (Models: T31A, T32A) |
| --- | --- |
| Common Name | Laser surgical instrument for use in general and plastic surgery and in dermatology |
| Classification Name | Light Based Over-The-Counter Hair Removal |
| Regulation Number | 878.4810 |
| Product Code(s) | OHT |
Legally Marketed Predicate Devices 21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K250942 | Ice Cooling IPL Hair Removal Device (UI20S DB, UI20S RE, UI20S PW, UI20S GP, UI20S GK, UI20S BK, UI20 DB, UI20 RE, UI20 PW, UI20 GP, UI20 GR, UI20 BK) | OHT |
| K241998 | Ice Cooling IPL Hair Removal Device (UI20 DB, UI20 RE, UI20 GP, UI20 PW, UI20S DB, UI20S RE, UI20S PW, UI20S GP,UI20 WH, UI20 GR, UI20 BS, UI20 MP, UI20 BL, UI20 PN, UI20 BR, UI20WG, UI20S WH, UI20S GR, UI20S BS, UI20S MP) | OHT |
| K251176 | IPL Hair Removal Device (Model(s): T14B,T16B,T19B, T15B,T17C,T18B, T21A,T21B,T21C, T21D,T22A,T22B, T25B,T25C) | OHT |
Device Description Summary 21 CFR 807.92(a)(4)
The IPL Hair Removal Device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user.
The device provides hair reduction using Intense Pulsed Light technology and cooling technology (suitable for model T31A, T32A) The Intense Pulsed Light technology works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Quartz glass Xenon lamp and a skin sensor to detect appropriate skin contact. If the device is not properly and fully applied to the skin of the treatment area, the device will not emit light pulses; If the device is properly and fully applied to the skin of the treatment area, the device can emit light pulses in as quickly as 0.5 seconds. In automatic mode, it supports continuous flashing and
{5}
automatic light emission. It is designed with dual lamps that work together and can emit mutipulses per shot.
In auto-recognition skin color mode, the skin tone sensor can detect and identify the color of skin, and determine the required intensity based on the recognized skin color. Make sure the skin tone sensor is in full contact with the skin. If a valid skin color is detected, the corresponding energy level is displayed. If it is not in full contact with the skin, the energy level is 0 and no light pulses are emitted.
The cooling technology based on the temperature difference electrical phenomenon through the semiconductor cooling chip inside the IPL main device and uses the principle of the Peltier effect to achieve the purpose of cooling function. The cooling panel is constructed with sapphire, and does not affect the irradiated area (spot size) of the light outlet.
Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The IPL Hair Removal Device is an over-the-counter device indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long term, stable reduction in the amount of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Indications for Use Comparison
21 CFR 807.92(a)(5)
The indications for use of the subject device is the same as the predicate device's, the contents as below:
The Home Use Hair Removal Device is an over-the-counter device intended for removal of unwanted hair such as but not limited to small areas such as underarm and facial hair below the chin line and large areas such as legs.
Technological Comparison
21 CFR 807.92(a)(6)
The IPL Hair Removal Device has the same intended use, energy medium (Xenon Arc Flashlamp), and similar operational characteristics as the predicate devices.
The wavelength range (510-1200nm) is the same as K251176.
The maximum fluence 6.48J/cm2 is within range of those of the predicate devices (K251176, 5.90J/cm2; K250942, 6.92J/cm2).
The spot size 5.4 cm2 is larger than the that of the predicate devices (e.g., K250942, 3.9cm2) but does not raise different questions in safety or effectiveness.
The pulse duration (0.4ms - 0.92ms multipulse) is within the range of those of the predicate devices (K250942, 0.86-6.32ms, double pulse, quadruple pulse; K251176, 0.4-5.5ms).
Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or effectiveness. Performance data supports that the device is as safe and as effective as the predicate device for its intended use. Therefore, the IPL Hair Removal Device (Models: T31A, T32A) is substantially equivalent to its predicate devices.
Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
1) Biocompatibility Testing
The biocompatibility evaluation for the body-contacting components of the subject device was conducted in accordance with the "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing Within a Risk Management Process", as recommended by FDA. The battery of testing was performed to, and passed, including:
ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation
The biocompatibility evaluation described above was conducted on a device model with the same body-contacting materials, material sources (from the same manufacturer), and contact nature as the subject device for this registration. The production process is only different in terms of the molds, which does not involve changes to material composition. Although there are minor differences in production processes, no pollutants or residues are generated or introduced during production. Therefore, the biocompatibility data from the tested model can be directly referenced to demonstrate that the subject device also possesses excellent biocompatibility and will not pose biological hazards to the human body.
2) Electrical Safety and EMC Safety
Electrical safety and Eye safety testing was performed to, and passed, the following standards:
IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
IEC 60601-1-11 Edition 2.1 2020-07 CONSOLIDATED VERSION Medical Electrical Equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral Standard: Requirements for medical electrical equipment and medical electrical equipment and medical electrical systems used in the home healthcare environment
IEC 60601-2-57 Edition 1.0 2011-01 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic diagnostic monitoring and cosmetic/aesthetic use
{6}
IEC 60601-2-83:2019/AMD1:2022 Medical Electrical Equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
3) Eye Safety
IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems
4) Software Verification and Validation
In this 510(k) submission, the software documentation are the basic documentation. System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
Summary
Based on the above performance as documented in this application, the subject device IPL Hair Removal Device was found to have a safety and effectiveness profile that is similar to the predicate device.
Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the subject device IPL Hair Removal Device (Models: T31A, T32A) is to be concluded substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.