Ice Cooling IPL Hair Removal Device with sapphire treatment window is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Device Story
Over-the-counter, home-use IPL hair removal device; utilizes xenon arc flashlamp to emit light pulses; sapphire treatment window for skin contact; dual-lamp design supports multi-pulse emission. Device features integrated ice cooling for user comfort; skin sensor prevents light emission unless in full contact with skin. Operated by user via finger switch; powered by external adapter. UI20S series includes skin tone recognition. Output intensity adjustable (1-10 levels). Device reduces hair growth by targeting hair follicles below skin surface without cutting/pulling. Benefits include nearly painless hair reduction and long-term regrowth stability.
Clinical Evidence
Bench testing only. Evidence includes biocompatibility (ISO 10993-5, -10, -23), electrical safety/EMC (IEC 60601-1, -1-2, -1-11, -2-57, -2-83), photobiological eye safety (IEC 62471), and usability engineering (IEC 60601-1-6). Software verification and validation confirmed all specifications met and hazards mitigated.
Technological Characteristics
Xenon arc flashlamp IPL source; sapphire treatment window; 550-1200nm wavelength; 1.67-6.67 J/cm² energy density; 1-10 intensity levels; external power adapter (100-240V, 50/60Hz). Includes skin contact sensor and ice cooling system. Biocompatible materials per ISO 10993. Software-controlled pulse delivery.
Indications for Use
Indicated for removal of unwanted hair and permanent reduction in hair regrowth in adults (Fitzpatrick Skin Types I-V).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" being the largest and boldest, followed by "U.S. FOOD & DRUG" and then "ADMINISTRATION" in a smaller font size.
December 13, 2024
Shenzhen Ulike Smart Electronics Co., Ltd. Yang Blue Registration Director 810,Bldg 1,Xunmei Science & Technology Plaza,No. 8 Keyuan Rd Science Park Community, Yuehai Sub-District,Nanshan District Shenzhen, Guangdong 518000 China
# Re: K241998
Trade/Device Name: Ice Cooling IPL Hair Removal Device (U120 DB, U120 RE, U120 GP, U120 PW, UI20S DB, UI20S RE, UI20S PW, UI20S GP,UI20 WH, UI20 GR, UI20 BS, UI20 MP, UI20 BL, UI20 PN, UI20 BR, UI20WG, UI20S WH, UI20S GR, UI20S BS, UI20S MP) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: July 6, 2024 Received: July 9, 2024
Dear Yang Blue:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to. 21 CFR 820.30. Design controls: 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by TANISHA TANISHA L. HITHE -S L. HITHE -> 21:37:46 -05'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241998
#### Device Name
Ice Cooling IPL Hair Removal Device (UI20 DB, U20 RE, UI20 GP, UI20 PW, UI20S RE, UI20S PW, UI20S GP,UI20 WH, UI20 BR, UI20 BS, UI20 MP, UI20 BL, UI20 PN, UI20 BR, UI20WG, UI20S WH, UI20S GR, UI20S MP)
### Indications for Use (Describe)
Ice Cooling IPL Hair Removal Device with sapphire treatment window is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary of K241998
# I. Submitter
Shenzhen Ulike Smart Electronics Co.,Ltd. Address:810, Building 1, Xunmei Science and Technology Plaza, No. 8 Keyuan Road, Science Park Community, Yuehai Sub-District, Nanshan District, Shenzhen 518000, Guangdong, P.R. China Contact person: Blue Yang Email: blue@ulike.com The date the summary was prepared: 12/9/2024
# II. Device
Name of Device: Ice Cooling IPL Hair Removal Device
Model(s): UI20 DB, UI20 RE, UI20 GP, UI20 PW, UI20S DB, UI20S RE, UI20S GP,UI20 WH, UI20 GR, UI20 BS, UI20 MP, UI20 BL, UI20 PN, UI20 BR, UI20WG, UI20S WH, UI20S GR, UI20S BS, UI20S MP,
Common or Usual Name: Light Based Over-The-Counter Hair Removal
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: II
Product Code: OHT
Regulation Number: 21 CFR 878.4810
| Predicate<br>Device | Predicate<br>Device 1 | Predicate<br>Device 2 | Predicate<br>Device 3 | Predicate<br>Device 4 | Predicate<br>Device 5 |
|---------------------|-----------------------------------------------------|----------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------|--------------------------------------------------|
| 510(k) number | K230122 | K130315 | K230739 | K240016 | K242039 |
| Trade Name | IPL Hair<br>Removal<br>Device | iPulse Hair<br>Removal<br>System | Hand-held IPL<br>device (IPL<br>Home Use<br>Hair Removal<br>Device) | IPL Hair<br>Removal<br>Device | Ice Cooling<br>IPL Hair<br>Removal<br>Device |
| Manufacturer | Shenzhen Ulike<br>Smart<br>Electronics<br>Co., Ltd. | CyDen Ltd | Shenzhen<br>IONKA<br>Medical<br>Technology<br>Co., Ltd. | Shenzhen<br>Jianrong<br>Biomedical<br>Electronics<br>Co., Ltd | Shenzhen Ulike Smart<br>Electronics<br>Co., Ltd. |
# III. Predicate Device
{5}------------------------------------------------
# IV. Device Description
Ice Cooling IPL Hair Removal Device is an over-the-counter, home-use and personal device for hair reduction by using Intense Pulsed Light (IPL). It is designed with dual lamps that work together and can emit mutipulses per shot. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with nearly painless pain.
The device is only powered by the external power adapter and its IPL emission activation is by finger switch. This product adopts sapphire treatment window that is suitable for multiple hair removal areas. It has skin-safe detection. It contains a skin sensor to detect appropriate skin contact, if the device is not in full contact with the skin, the device cannot emit the treatment light pulses. Besides, the device has the ice cooling function that will be activated throughout the whole hair removal process to provide users with a more comfortable experience.
Ice Cooling IPL Hair Removal Device includes the following models: UI20 WH, UI20 GR, UI20 BS, UI20 MP, UI20 BL, UI20 PN, UI20 BR, UI20WG, UI20S WH, UI20S GR, UI20S MP, UI20 DB, UI20 RE, UI20 GP, UI20 PW, UI20S DB, UI20S RE, UI20S PW, UI20S GP. Their intended use, performance and operation are basically identical. The model differences are enclosure color, a detachable accessory cover, and skin tone recognition (only for UI20S series with an accessory cover and with skin tone recognition). (U120 series: UI20 WH, U120 GR, UI20 BS, UI20 MP, UI20 BL, UI20 PN, UI20 BR, UI20 WG; UI20S series: UI20 DB, UI20 RE, UI20 GP, UI20 PW, UI20S WH, UI20S GR, UI20S BS, UI20S MP, UI20S DB, UI20S RE, UI20S PW, UI20S GP).
# V. Indications for Use
Ice Cooling IPL Hair Removal Device with sapphire treatment window is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
# VI. Comparison of Technological Characteristics With the Predicate
# Devices
The Ice Cooling IPL Hair Removal Device has the same intended use and similar operational characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate devices for its intended use. Therefore, the IPL Hair Removal Device may be found substantially equivalent to its predicate devices.
Ice Cooling IPL Hair Removal Device is compared with the following Predicate Devices in terms of intended use, design, specifications and performance:
{6}------------------------------------------------
| Comparison Items | Subject Device | Predicate<br>Device 1 | Predicate Device<br>2 | Predicate Device<br>3 | Predicate Device<br>4 | Predicate Device<br>5 | Remark | | | | | | | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|--------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|-------------|--------------|--------------|----------------------------------------------------------------------------------|--------------|-----------|-----------------------------------------------------------------|
| 510(k) number | / | K230122 | K130315 | K230739 | K240016 | K242039 | / | | | | | | | | | |
| Trade Name | Ice Cooling<br>IPL Hair<br>Removal<br>Device | IPL Hair<br>Removal<br>Device | iPulse Hair<br>Removal System | Hand-held<br>device (IPL<br>Home Use Hair<br>Removal Device) | IPL Hair<br>Removal<br>Device | Ice Cooling IPL<br>Hair Removal<br>Device | / | | | | | | | | | |
| Manufacturer | Shenzhen<br>Ulike Smart<br>Electronics<br>Co., Ltd. | Shenzhen Ulike<br>Smart<br>Electronics<br>Co., Ltd. | CyDen Ltd | Shenzhen IONKA<br>Medical<br>Technology Co.,<br>Ltd. | Shenzhen<br>Jianrong<br>Biomedical<br>Electronics<br>Co., Ltd | Shenzhen Ulike<br>Smart<br>Electronics Co.,<br>Ltd. | / | | | | | | | | | |
| Regulation<br>number | 21 CFR<br>878.4810 | 21 CFR<br>878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR<br>878.4810 | 21 CFR<br>878.4810 | Same | | | | | | | | | |
| Product code | OHT | OHT | OHT | OHT | OHT | OHT | Same | | | | | | | | | |
| Device<br>classification | Class II | Class II | Class II | Class II | Class II | Class II | Same | | | | | | | | | |
| Indication for use/<br>Intended use | Ice Cooling<br>IPL Hair<br>Removal<br>Device with<br>sapphire<br>treatment<br>window is<br>indicated for | IPL Hair<br>Removal<br>Device is<br>indicated for<br>the removal of<br>unwanted hair.<br>The device is<br>also indicated | The iPulse Hair<br>Removal System<br>is<br>an<br>Over-the-counter<br>device intended<br>for the removal of<br>unwanted hair. | IPL Home Use<br>Hair Removal<br>Device is<br>indicated for the<br>removal of<br>unwanted hair.<br>The device is also<br>indicated for the | IPL Hair<br>Removal<br>Device is<br>indicated for<br>the removal of<br>unwanted hair.<br>The device is<br>also indicated | Ice Cooling IPL<br>Hair Removal<br>Device with<br>sapphire<br>treatment<br>window is<br>indicated for the<br>removal of | Same | | | | | | | | | |
| the removal of<br>unwanted hair.<br>The device is<br>also indicated<br>for the<br>permanent<br>reduction in<br>hair regrowth,<br>defined as the<br>long-term,<br>stable<br>reduction in the<br>number of<br>hairs regrowing<br>when measured<br>at 6, 9 and 12<br>months after<br>the completion<br>of a treatment<br>regime. | for<br>the<br>permanent<br>reduction in<br>hair regrowth,<br>defined as the<br>long-term,<br>stable reduction<br>in the number<br>of<br>hairs<br>regrowing<br>when measured<br>at 6, 9 and 12<br>months after<br>the completion<br>of a treatment<br>regime. | OTC | for the<br>permanent<br>reduction in hair<br>regrowth, defined<br>as the long-term,<br>stable reduction<br>in the number of<br>hairs regrowing<br>when measured at<br>6, 9 and 12<br>months after the<br>completion of a<br>treatment regime. | OTC | unwanted hair.<br>The device is<br>also indicated for<br>the permanent<br>reduction in hair<br>regrowth,<br>defined as the<br>long-term, stable<br>reduction in the<br>number of hairs<br>regrowing when<br>measured at 6, 9<br>and 12 months<br>after the<br>completion of a<br>treatment<br>regime. | Same | | | | | | | | | | |
| Prescription or OTC | OTC | OTC | OTC | OTC | OTC | Same | | | | | | | | | | |
| Applicable skin | Fitzpatrick<br>Skin Types I-V | Fitzpatrick<br>Skin Types I-V | unknown | Fitzpatrick Skin<br>Types I-V | unknown | Fitzpatrick Skin<br>Types I-V | Same | | | | | | | | | |
| Treatment area | Large areas | Large areas | leg, arm, bikini | unknown | unknown | Large areas (e.g. | Same | | | | | | | | | |
| | (e.g. arms,<br>legs, chest) and<br>small areas<br>(e.g. lip) | (e.g. arms, legs,<br>chest) and<br>small areas<br>(e.g. lip) | and underarm | | | arms, legs, chest)<br>and small areas<br>(e.g. lip) | | | | | | | | | | |
| Device design | | | | | | | | | | | | | | | | |
| Source energy | Supplied by<br>external<br>adapter | Supplied by<br>external<br>adapter | Supplied by<br>external<br>adapter | Supplied by<br>external<br>adapter | Supplied by<br>external<br>adapter | Supplied by<br>external<br>adapter | Same | | | | | | | | | |
| Power supply | 100-240V~,<br>50/60Hz | 100-240V~,<br>50/60Hz | 100-240V~,<br>50/60Hz | 100-240V~,<br>50/60Hz | 100-240V~,<br>50/60Hz | 100-240V~,<br>50/60Hz | Same | | | | | | | | | |
| Dimension | 206.73mm*68.<br>68mm*54.29m<br>m | 60mm x 38mm<br>x<br>170mm | unknown | FZ-608,<br>FZ-608G:<br>98*147*60(mm)<br>FZ-100:<br>198*71*44(mm)<br>FZ-200:<br>216*68*52(mm) | unknown | 206.73mm*68.6<br>8 mm*54.29mm | SE<br>Note 1 | | | | | | | | | |
| Sterilization | Not required | Not required | Not required | Not required | Not required | Not required | Same | | | | | | | | | |
| Output specification | | | | | | | | | | | | | | | | |
| Light source | Intense Pulsed<br>Light | Intense Pulsed<br>Light | Intense Pulsed<br>Light | Intense Pulsed<br>Light | Intense Pulsed<br>Light | Intense Pulsed<br>Light | Same | | | | | | | | | |
| Energy medium | Xenon Arc<br>Flashlamp | Xenon Arc<br>Flashlamp | Xenon Arc<br>Flashlamp | Xenon Arc<br>Flashlamp | Xenon Arc<br>Flashlamp | Xenon Arc<br>Flashlamp | Same | | | | | | | | | |
| Wavelength range | 550-1200mm | 560-1200nm | 530-1100nm | 510nm~1200nm | 550-1200nm | 550-1200nm | SE<br>Note 2 | | | | | | | | | |
| Energy density | $1.67~6.67J/cm^2$ | $2.4-7.2J/cm^2$ | $7-10J/cm^2$ | $1.2-4.3J/cm^2$ | $2.79~6.41J/cm^2$ | SE<br>Note 3 | | | | | | | | | | |
| | | | FZ-608,<br>FZ-608G: 3.33 J/cm2<br>FZ-100: 4.5 J/cm2<br>FZ-200: 5.43 J/cm2 | | | | | | | | | | | | | |
| Output energy | | | $6.5~26J$ | | | | $9.9~19.8J$ | $21~30J$ | $5.4~19.35J$ | $10.9J~25J$ | SE<br>Note 4 | | | | | |
| | | | | | | | | FZ-608,<br>FZ-608G: Level 1: 4.16J Level 2: 4.36J Level 3: 5.1J Level 4: 6.1J Level 5: 6.96J Level 6: 7.96J Level 7: 8.63J Level 8: 9.13J Level 9: 10.0J | | | | | | | | |
| Spot size | | | | | | | | $3.9cm^2$ | | | | $3.3cm^2$ | $3cm^2$ | $4.5cm^2$ | $3.9cm^2$ | SE<br>Note 5 |
| Pulse duration | | | | | | | | $0.88~3.20ms$<br>Multipulse | | | | $1.15-6.2ms$ | Variable -<br>Single pulse 25milliseconds.<br>to Double Pulse 20ms on, 60 ms off | $0.5~0.8 ms$ | unknown | $0.93ms~3.50ms$<br>Single pulse<br>Double pulse<br>Triple pulse |
| Pulsing control | Finger switch | Finger switch | | Finger switch | Finger switch | Finger switch | | Finger switch | | | | Same | | | | |
| Delivery device | Direct<br>illumination to<br>tissue | Direct<br>illumination to<br>tissue | | Direct<br>illumination to<br>tissue | Direct<br>illumination to<br>tissue | Direct<br>illumination to<br>tissue | | Direct<br>illumination to<br>tissue | | | | Same | | | | |
| Output intensity<br>level | 1-10 Levels | 5 Levels | | unknown | FZ-608,<br>FZ-608G: 9 levels<br>FZ-100: 9 levels<br>FZ-200: 6 levels | unknown | | 1-10 Levels | | | | SE<br>Note 7 | | | | |
| Software/Firmware/Microprocess or<br>Control? | Yes | Yes | | Yes | Yes | Yes | | Yes | | | | Same | | | | |
| Additional features | | | | | | | | | | | | | | | | |
| Electrical safety | IEC 60601-1<br>IEC 60601-1-2<br>IEC<br>60601-1-11<br>IEC<br>60601-2-57<br>IEC<br>60601-2-83 | ANSI AAMI<br>ES 60601-1<br>IEC 60601-1-2<br>IEC<br>60601-1-11<br>IEC<br>60601-2-57<br>IEC<br>60601-2-8…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.