The Precision S 4K Sinuscope is intended for use in otolaryngology and head and neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery. The Precision S 4K Sinuscope is also intended for use in minimally invasive cranial neurosurgery in adults and pediatric patients and endonasal skull base surgery in adults and pediatric patients > 6 years of age.
Device Story
Precision S 4K Sinuscope is a rigid, tubular optical instrument for visualization during surgical procedures. Device transmits visible and near-infrared light to illuminate internal anatomy; rigid rod lenses relay images from surgical site to external processing and display systems. Used in OR by surgeons for examination, diagnosis, and therapy. Optional HydroFlow cleaning sheath available for specific lengths. Output provides real-time visualization to assist surgeon in navigating anatomy and performing procedures. Benefits include enhanced visualization during minimally invasive surgeries.
Clinical Evidence
Bench testing only. No clinical or animal data provided. Performance testing included biocompatibility (ISO 10993 series), reprocessing validation (AAMI TIR12, ST98, ST79, ST58, ISO 15883, ISO 17664, ISO 17665, ISO 22441), electrical safety (IEC 60601-1, IEC 60601-2-18), packaging (ASTM D4169), and optical performance (ISO 8600-1). Comparative bench testing against the predicate confirmed equivalent optical characteristics including depth of field, MTF, illumination, and transmission.
Technological Characteristics
Rigid rod lens endoscope; stainless steel working length; distal tip (stainless steel, optical glass, epoxy, glass fibers, solder, copper alloy); PEEK eyepiece. Visible and NIR transmission. Dimensions: 4.0mm/3.1mm OD, 125-180mm WL, 0-70 degree DOV. Reusable; manual/automated cleaning; steam/H2O2 sterilization. Connectivity: optical interface to external display system. No software/algorithm.
Indications for Use
Indicated for otolaryngology, head and neck, rhinology, and endoscopic plastic/reconstructive surgery; minimally invasive cranial neurosurgery in adults and pediatric patients; endonasal skull base surgery in adults and pediatric patients > 6 years of age.
Regulatory Classification
Identification
A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 4, 2026
Stryker Endoscopy
Divya Sekar
Senior Principal, Regulatory Affairs Specialist
5900 Optical Ct.
San Jose, California 95138
Re: K253602
Trade/Device Name: Precision S 4K Sinuscope
Regulation Number: 21 CFR 882.1480
Regulation Name: Neurological Endoscope
Regulatory Class: Class II
Product Code: GWG, EOB
Dated: May 8, 2026
Received: May 8, 2026
Dear Divya Sekar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253602 - Divya Sekar
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JULIA E.
SLOCOMB -S
Digitally signed by
JULIA E. SLOCOMB -S
Date: 2026.06.04
16:29:16 -04'00'
{2}
K253602 - Divya Sekar
Page 3
for Jaime Raben, Ph.D.
Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253602 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Precision S 4K Sinuscope | | |
| Please provide your Indications for Use below. | | ? |
| The Precision S 4K Sinuscope is intended for use in otolaryngology and head and neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery. | | |
| The Precision S 4K Sinuscope is also intended for use in minimally invasive cranial neurosurgery in adults and pediatric patients and endonasal skull base surgery in adults and pediatric patients > 6 years of age. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253602
## 510(k) Summary
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R Part 807.92(c).
### **Submitter:**
| Applicant | Stryker Endoscopy 5900 Optical Court San Jose, CA 95138 |
| --- | --- |
| Contact Person | Divya Sekar Senior Principal, Regulatory Affairs Specialist Email: divya.sekar@stryker.com Phone: 408-754-2473 |
| Date Prepared | November 17, 2025 |
### **Subject Device:**
| Name of Device: | Precision S 4K Sinuscope |
| --- | --- |
| Common or Usual Name: | Sinuscope |
| Classification Name: | Neurological Endoscope^{1} (21 C.F.R. §882.1480) Nasopharyngoscope (flexible or rigid) and accessories^{2} (21 CFR §874.4760) |
| Regulatory Class: | Class II |
| Product Code: | GWG^{1} |
| Subsequent Product Code: | EOB^{2} |
| 510(k) Review Panel: | Neurology |
$^{1}$When intended for use in neurosurgery procedure.
$^{2}$When intended for use in otolaryngology procedure.
### **Predicate Device(s):**
| Precision S 4K Sinuscope | K211202 (primary), K191102 |
| --- | --- |
*NOTE: The predicate device has not been subject to a design-related recall.*
### **Device Description:**
Stryker’s Precision S 4K Sinuscopes are tubular optical instruments used to provide a view of internal patient anatomy for examination, diagnosis, and therapy during surgical procedures. The Precision S 4K Sinuscope transmits visible and near-infrared light to illuminate and image the anatomy, then relaying the image out of the surgical site for processing and display.
### **Indications for Use:**
The Precision S 4K Sinuscope is intended for use in otolaryngology and head and neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery.
Page 1 of 4
{5}
K253602
The Precision S 4K Sinuscope is also intended for use in minimally invasive cranial neurosurgery in adults and pediatric patients and endonasal skull base surgery in adults and pediatric patients > 6 years of age.
# **Comparison of Technological Characteristics with the Predicate Device:**
| Feature | Subject Device | Predicate Device |
| --- | --- | --- |
| | Precision S 4K Sinuscope | Precision S 4K Sinuscope |
| Manufacturer | Stryker Endoscopy | Same as subject device |
| Submission Reference | Current Submission | K211202 (primary), K191102 |
| Intended Use | Endoscopic illumination and imaging during endoscopic procedures. | Same as subject device |
| Indications for Use Statement | The Precision S 4K Sinuscope is intended for use in otolaryngology and head and neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery. The Precision S 4K Sinuscope is also intended for use in minimally invasive cranial neurosurgery in adults and pediatric patients and endonasal skull base surgery in adults and pediatric patients > 6 years of age. | Same as subject device |
| **Design** | | |
| Operating Principles | Transmission of light to illuminate and image procedures, then relays the image out of the surgical site for processing and display. | Same as subject device |
| Image Transmission | Rigid rod lenses | Same as subject device |
| Transmission Spectrum | Visible and Near-infrared | Same as subject device |
| Outer Diameter (OD) | 4.0mm, 3.1mm | Same as subject device |
| Working Length (WL) | 125mm – 180mm | Same as subject device |
| Direction of View (DOV) | 0°-70° | Same as subject device |
| Field of View (FOV) | 80° | 80° for 3.1mm Sinuscope 105° for 4.0mm Sinuscope |
| Depth of Field | 7mm – 35mm | Same as subject device |
| Optional Accessories | Precision S HydroFlow Cleaning Sheath (available for 155mm – 180mm WL Sinuscopes) | N/A – No optional accessory |
Page 2 of 4
{6}
K253602
| Feature | Subject Device | Predicate Device |
| --- | --- | --- |
| | Precision S 4K Sinuscope | Precision S 4K Sinuscope |
| **Materials** | | |
| Working Length (patient-contacting) | Stainless steel | Same as subject device |
| Distal Tip (patient-contacting) | Stainless steel, optical glass, epoxy, glass fibers, solder, copper alloy | Same as subject device |
| Main body (non-patient contacting) | Stainless steel, TiAlOx | Same as subject device |
| Eyepiece (non-patient contacting) | Polyetheretherketone (PEEK) | Same as subject device |
| **Processing** | | |
| Single Use or Reusable | Reusable | Same as subject device |
| Cleaning | Manual and Automated | Same as subject device |
| Disinfection | Automated | Same as subject device |
| Sterilization | Steam (Autoclave), Hydrogen Peroxide (Steris, Sterrad) | Same as subject device |
| Sterility Assurance Level | 10^{-6} | Same as subject device |
### **Performance Data:**
The following performance data were provided in support of the substantial equivalence determination.
| Test | Method | Result |
| --- | --- | --- |
| Biocompatibility | In accordance with FDA-recognized voluntary consensus standard ISO 10993-1:2018 (2-258) | Pass |
| | In accordance with FDA-recognized voluntary consensus standard ISO 10993-5:2009 (2-245) | Pass |
| | In accordance with FDA-recognized voluntary consensus standard ISO 10993-10:2021 (2-296) | Pass |
| | In accordance with FDA-recognized voluntary consensus standard ISO 10993-11:2017 (2-255) | Pass |
| | In accordance with FDA-recognized voluntary consensus standard ISO 10993-23:2021 (2-291) | Pass |
| Cleaning, Disinfection & Sterilization (Reprocessing) | In accordance with AAMI TIR12:2020 /(R)2023 (14-602) | Pass |
| | In accordance with FDA-recognized voluntary consensus standard ANSI AAMI ST98:2022 (14-583) | Pass |
| | In accordance with ISO 15883-1:2006 + A1:2014 | Pass |
| | In accordance with ISO 15883-2:2006 | Pass |
| | In accordance with ISO 15883-5:2021 | Pass |
| | In accordance with FDA-recognized voluntary consensus standard ANSI AAMI ST79:2017 + A1:2020, A2:2020, A3:2020, A4:2020 (14-562) | Pass |
Page 3 of 4
{7}
K253602
| Test | Method | Result |
| --- | --- | --- |
| | In accordance with FDA-recognized voluntary consensus standard ANSI AAMI ST58:2024 (14-607) | Pass |
| | In accordance with FDA-recognized voluntary consensus standard ISO 17664-1:2021 (14-578) | Pass |
| | In accordance with FDA-recognized voluntary consensus standard ISO 17665:2024 (14-601) | Pass |
| | In accordance with FDA-recognized voluntary consensus standard ISO 22441:2022 (14-586) | Pass |
| Electrical Safety | In accordance with FDA-recognized voluntary consensus standard IEC 60601-1:2020 (19-49) | Pass |
| | In accordance with FDA-recognized voluntary consensus standard IEC 60601-2-18:2009 (9-114) | Pass |
| Packaging | In accordance with FDA-recognized voluntary consensus standard ASTM D4169:2022 (14-576) | Pass |
| Optical Performance | In accordance with FDA-recognized voluntary consensus standard ISO 8600-1:2015 (9-110) | Pass |
| Performance – Bench | Sinuscope - Comparative optical performance testing to currently legally marketed predicate device: Depth of Field, Modulated Transfer Function (MTF), Vignetting, Illumination, Optical Transmission, Field of View (FOV), Apparent Field of View (AFOV), Optical Train Concentricity. | Pass |
| | Sinuscope - Fiber Transmission and NIR Wavelength Transmission. | Pass |
| | Sinuscope - Sharp Edge Inspection and Distal Tip Geometry | Pass |
| | Sinuscope - In Focus / Out of Focus Particulates | Pass |
| | Sinuscope – Visual Verification of Transmission Capabilities | Pass |
| | Sinuscope / Cleaning Sheath - Design Validation | Pass |
| | Sinuscope / Cleaning Sheath - Compatibility and Cleaning Performance | Pass |
| | Cleaning Sheath - Fiber transmission, Illumination homogeneity and Bending stiffness | Pass |
*NOTE: Precision S 4K Sinuscope does not require animal or clinical testing to support the determination of substantial equivalence.*
### **Conclusions:**
The Precision S 4K Sinuscope is substantially equivalent in design, intended use, principles of operation, technological characteristics and safety features to the predicate device. There are no new issues of safety and/or effectiveness introduced by the Precision S 4K Sinuscope when used as instructed.
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.