K253261 · Weero Co., Ltd. · PBX · Mar 13, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253261
Device Name
Apollo Quattro (APQ-10M)
Applicant
Weero Co., Ltd.
Product Code
PBX · General, Plastic Surgery
Decision Date
Mar 13, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Q-LF handpiece, Q-FX handpiece, Q-SC handpiece, and Q-BT handpiece in RF mode are intended for: • Relief of minor muscle aches and pain, relief of muscle spasm, and temporary improvement of local blood circulation. Q-EP handpiece in cooling mode is intended for: • Use in dermatologic procedures for the removal of benign lesions of the skin and for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures.
Device Story
Apollo Quattro is a multi-modal stimulator applying radio frequency (RF), low frequency, heating, and cooling energy. System includes main unit and five handpieces (Q-EP, Q-LF, Q-FX, Q-SC, Q-BT). RF/low frequency modes target muscle pain/spasm and circulation; cooling mode (via thermoelectric cooler/TEC and aluminum plate) targets skin lesions and post-surgical inflammation/pain. Operated by healthcare professionals in clinical settings. Device transforms electrical input into thermal or RF energy output delivered via contact applicators. Output affects tissue temperature or provides electrical stimulation to modulate pain and circulation. Benefits include non-invasive pain relief and dermatologic lesion management. Safety and performance validated via bench testing against IEC standards.
Clinical Evidence
Bench testing only. No clinical data. Performance validated via electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and high-frequency surgical equipment standards (IEC 60601-2-2). Biocompatibility confirmed for patient-contacting materials (SUS 304, PC). Cooling performance verified against predicate specifications.
Technological Characteristics
Materials: SUS 304 (electrodes), Polycarbonate (display). Energy: Bipolar RF (1 MHz), Thermoelectric cooling (TEC). Form factor: Main unit with multiple interchangeable handpieces. Connectivity: Standalone. Sterilization: N/A (multiple use, cleaning required). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, ISO 14971, IEC 62304.
Indications for Use
Indicated for patients requiring relief of minor muscle aches, pain, and spasms, or improvement of local blood circulation (RF mode); and for patients undergoing dermatologic procedures for benign skin lesion removal or requiring temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures (cooling mode).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 13, 2026
WEERO Co., Ltd.
Han Moon Young
Regulatory Affairs
A-205, Venture Valley II, 142-10, Saneop-Ro 156 Beon-Gil
Gwonseon-Gu Suwon-Si, Gyeonggi-Do
Suwon, Gyeonggi 16648
Republic Of Korea
Re: K253261
Trade/Device Name: Apollo Quattro (APQ-10M)
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: PBX, QPZ
Dated: February 9, 2026
Received: February 9, 2026
Dear Han Moon Young:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253261 - Han Moon Young
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253261 - Han Moon Young
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin K. Chen -S
Digitally signed by
Colin K. Chen -S
Date: 2026.03.13
07:29:43 -04'00"
Colin Kejing Chen, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
K253261
Page 1 of 1
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253261 | |
| Please provide the device trade name(s). | | |
| Apollo Quattro (APQ-10M) | | |
| Please provide your Indications for Use below. | | |
| Q-LF handpiece, Q-FX handpiece, Q-SC handpiece, and Q-BT handpiece in RF mode are intended for:
• Relief of minor muscle aches and pain, relief of muscle spasm, and temporary improvement of local blood circulation.
Q-EP handpiece in cooling mode is intended for:
• Use in dermatologic procedures for the removal of benign lesions of the skin and for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
K253261
Apollo Quattro
510(k) Summary
# 1. Submitter and US Official Correspondent
Submitter: WEERO Co.,Ltd.
Address: A205 VentureValley II, 142-10, Saneop-ro 156beon-gil, Gwonseon-gu, Suwon-si, Gyeonggi-do, Republic of Korea
Telephone No.: +82-31-427-0787
Official Correspondent: Moon-young Han
Correspondent: WEERO Co.,Ltd.
Address: A205 VentureValley II, 142-10, Saneop-ro 156beon-gil, Gwonseon-gu, Suwon-si, Gyeonggi-do, Republic of Korea
Telephone No.: +82-31-427-0787
Email: ra@weeroweero.com
# 2. Device Information
Trade/Device Name: Apollo Quattro / APQ-10M
Regulation Name:
- Electrosurgical cutting and coagulation device and accessories.
- Contact cooling system for aesthetic use.
Classification Name:
- Massager, Vacuum, Radio Frequency Induced Heat
- Contact Cooling For Skin Lesion Pain Relief
Product Code: PBX, QPZ
Regulatory Class: Class II
Regulation Number: 21CFR878.4400, and 21CFR878.4340
# 3. Predicate Device(Equivalent Legally Marketed Device)
| Manufacturer | Device | 510(k) No. |
| --- | --- | --- |
| Primary Predicate Device | | |
| WEERO Co.,Ltd. | Apollo Duet (Model : APD-4000) | K212253 |
| Secondary Predicate Device | | |
| R2 Technologies, Inc. | Dermal Cooling System | K213294 |
# 4. Description of the Device
Apollo Quattro is a combination stimulator used to relieve pain by applying radio frequency, low frequency, heating, and cooling energy to the affected area.
Apollo Quattro has a Q-EP handpiece, Q-LF handpiece, Q-FX handpiece, Q-SC handpiece, Q-BT handpiece, patch (cleared under K042301), patch cable, funnel hose, and power cable.
The Q-LF, Q-FX, Q-SC, and Q-BT handpieces are used for radio frequency or low frequency output, while the Q-EP handpiece is used for heating or cooling output, or for low frequency when used with the patch and patch cable.
The cooling function of Apollo Quattro is substantially equivalent to the device cleared in K213294. The Apollo Quattro is a cryosurgical device used to cool the skin, without the
Page 1 / 5
{5}
K253261
Page 2 of 5
Apollo Quattro
510(k) Summary
use of cryogenic gases or liquids, for the removal of benign skin lesions and for use when cooling is intended for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures. Surface contact cooling is achieved using a thermoelectric cooler (TEC), with an integrated aluminum plate, to lower the temperature of the skin.
## 5. Indications for use (intended use)
Q-LF handpiece, Q-FX handpiece, Q-SC handpiece, and Q-BT handpiece in RF mode are intended for:
- Relief of minor muscle aches and pain, relief of muscle spasm, and temporary improvement of local blood circulation.
Q-EP handpiece in cooling mode is intended for:
- Use in dermatologic procedures for the removal of benign lesions of the skin and for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures.
Page 2 / 5
{6}
K253261
Page 3 of 5
Apollo Quattro
510(k) Summary
## 6. Substantial Equivalence Discussion
## 1) Comparison Information
(1) Main Product Code: PBX
| Name | Subject device | Predicate device | Comparison |
| --- | --- | --- | --- |
| Device Name | Apollo Quattro (APQ-10M) | Apollo Duet (APD-4000) | |
| Manufacturer | WEERO Co.,Ltd. | WEERO Co.,Ltd. | |
| 510(k) No. | - | K212253 | |
| Product Code, Class | PBX Class II | PBX Class II | Identical |
| Indications for use | Relief of minor muscle aches and pain, relief of muscle spasm, and temporary improvement of local blood circulation. | Relief of minor muscle aches and pain, relief of muscle spasm, and temporary improvement of local blood circulation. | Identical |
| Energy Used | (Non-invasive) Bipolar RF | (Non-invasive) Bipolar RF | Identical |
| Physical specification: Dimensions | Main unit: 15.9(W) x 15.4(D) x 41(H) in / 405(W) x 391(D) x 1041(H) mm
Q-LF handpiece: 1.5(W) x 6.8(D) x 1.5(H) in / 37.3(W) x 172.8(D) x 37.6(H) mm
Q-FX handpiece: 1.1(W) x 5.8(D) x 1.1(H) in / 28.8(W) x 147.1(D) x 28.8(H) mm
Q-SC handpiece: 2.8(W) x 6.8(D) x 2.1(H) in / 71.5(W) x 172.5(D) x 53.1(H) mm
Q-BT handpiece: 2.8(W) x 6.6(D) x 2.8(H) in / 70.8(W) x 168.4(D) x 70.8(H) mm | Main unit: 13.1(W) x 13.1(D) x 7.2(H) in / 334(W) x 334(D) x 183.6(H) mm
RF Handpiece 1.7(W) x 6.4(D) in / 43.2(W) x 162.5(D) mm | Not identical
Although “Dimensions” and “Weight” of subject device are different from the predicate device, they all comply with IEC 60601-1, IEC 60601-2-2 requirements, thus the differences of the function specifications does not raise any safety or effectiveness issue. |
| Weight | Main unit: 20 kg / 44.1 lb
Q-LF handpiece: 78.5 g [0.2 lb]
Q-FX handpiece: 48.5 g [0.1 lb]
Q-SC handpiece: 181.5 g [0.4 lb]
Q-BT handpiece: 138 g [0.3 lb] | Main unit: 5 Kg / 11 lb
RF Handpiece: 200 g [0.44 lb] | |
| Input power | 100-240 VAC, 50/60 Hz | 100-240 VAC, 50/60 Hz | Identical |
| RF Frequency | 1 MHz | 1 MHz | Identical |
| Max. RF Output Power | 42 [W] | 40 [W] | Not identical
Although the “Max. RF Output Power” of the subject device is different than in the predicate device, it was all tested and is compliant with IEC 60601-2-2. Therefore the difference doesn’t impact essential performance, basic safety or substantial equivalence. |
| Patient contact material | Electrode: SUS 304
LED display part: PC | Electrode: SUS 304
LED display part: PC | Identical |
| Standards Met | IEC 60601-1
IEC 60601-1-2
IEC 60601-2-2 | IEC 60601-1
IEC 60601-1-2
IEC 60601-2-2 | Identical |
Page 3 / 5
{7}
K253261
Page 4 of 5
Apollo Quattro
510(k) Summary
(2) Subsequent Product Code: QPZ
The "Apollo Quattro (subject device)" and "Dermal Cooling System (predicate device, K213294)" have the same indications for use as follows,
"Use in dermatologic procedures for the removal of benign lesions of the skin and for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures."
The subject device and predicate device (K213294) also have the same fundamental scientific technological elements.
- comprised of a cooling unit (chiller) with a cold plate/probe applicator used to cool the skin in the area intended for treatment
- same cooling mechanism - thermoelectric cooling
- controlled contact surface cooling at a targeted treatment site on the skin
- metal cooling interface within an applicator
- the lowest treatment temperature (-16 °C)
The key differences that exist between the "Apollo Quattro" and the "Dermal Cooling System" are:
- size and shape of cold plate/probe applicator
The differences do not raise new questions of safety and effectiveness with respect to the proposed indication.
2) Substantial Equivalence Discussion
There are differences on a few things. However, these differences do not affect the significant equivalence of the device and its predicates.
The subject device and predicate devices utilize the same technology, for the same indication for use, and with almost identical design specifications.
All the performance specifications of the subject device are equal or similar to those of its predicate devices. The minor differences in technical specifications should not alter the device safety and effectiveness.
Furthermore, the subject device had underwent the required performance testing and validation testing and demonstrates its conformance with device design requirements and with applicable standards.
The safety features and compliance with safety standards of the subject device are similar to the safety features and compliance with safety standards of the predicate devices. For biocompatibility, all user-contacting materials (i.e., the Stainless Steel 304, and Polycarbonate) are the same as those previously cleared with K212253.
Furthermore, the design and development phases of the subject device were validated throughout a set of performance tests, including software validation testing, electrical and mechanical safety testing according to IEC 60601-1 standard, electromagnetic compatibility testing according to IEC 60601-1-2 standard, safety and essential performance of high frequency surgical equipment and high frequency surgical accessories testing according to IEC 60601-2-2 standard. These performance tests demonstrated that the device specifications meet the system requirements and do not raise new safety or effectiveness concerns.
3) Conclusion
"Apollo Duet (Model : APD-4000)" and "Dermal Cooling System" were chosen as predicate devices for the subject device in consideration of the intended use, indications, performance and principles of operation. Minor differences between the subject device and the predicate devices were found and listed within the above comparison table and discussion. Considerable amount of testing, including electrical safety, electromagnetic compatibility, performance, software verification and validation and usability testing were performed to support the claims of appropriately chosen predicate device. The test results show that the specifications and performance of Apollo Quattro are as safe and effective as legally marketed predicate devices.
Therefore, it is concluded that the Apollo Quattro is substantially equivalent to the legally marketed predicate devices.
Page 4 / 5
{8}
K253261
Page 5 of 5
Apollo Quattro
510(k) Summary
## 7. Non-Clinical (Bench) Performance Data:
As part of demonstrating safety and effectiveness of the Apollo Quattro and in showing substantial equivalence to the predicate device that are subject to this 510(k) submission, we completed a number of non-clinical performance tests against applicable standards.
- Basic safety and essential performance of the Apollo Quattro was tested and evaluated according to the IEC 60601-1:2005/A2:2020.
- Effect to the device by electromagnetic disturbances was tested and evaluated according to the FDA-recognized consensus standard IEC 60601-1-2:2014/A1:2020.
- Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories was tested and evaluated according to the IEC60601-2-2:2017/A1:2023.
- Risk management was recorded by referring to ISO 14971:2019.
- Usability was documented by referring to IEC 60601-1-6:2010/A2:2020.
- For biocompatibility, all user-contacting materials (i.e., Stainless Steel 304 and Polycarbonate) are the same as those previously cleared under K212253.
- Software was tested and evaluated according to IEC 62304:2015
- Content of Premarket Submissions for Device Software Functions
The Apollo Quattro passed all the testing in accordance with internal requirements, national standards, and international standards shown above, to support substantial equivalence of the subject device.
Also, to demonstrate that the Apollo Quattro meets all design specifications and performance requirements, and to measure the accuracy of the output parameters of Apollo Quattro and to compare the output parameters with predicate devices, nonclinical bench testing was performed in accordance with the internal process in compliance with the recommendations of the FDA Guidance for Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery, and Class II Special Controls Guidance Document: Contact Cooling System for Aesthetic Use.
The testing results support that the requirements for performance and electrical safety were met for the acceptance of the device. The Apollo Quattro passed all testing and supports the claims of substantial equivalence to the predicate device.
Regarding the cooling function, performance data is provided in support of the substantial equivalence determination.
Bench testing was completed to demonstrate the ability of the Apollo Quattro to meet performance specifications. No other performance testing was performed for the subject device for this Traditional 510(k) since there was no change to the device design or methods for use required for the range of treatment parameters previously cleared for the Dermal Cooling System (K213294) allows the same cooling characteristics at the skin surface.
No pre-clinical or clinical testing was performed.
## 8. Sterilization/Disinfection/Cleaning/Shelf Life
The Apollo Quattro is intended for multiple use and therefore must be cleaned according to the instructions provided in the device Instructions for Use.
There are no sterilized parts or accessories involved with this device.
## 9. Biocompatibility
| Part | Material | Patient Contact | Duration of Contact by ISO 10993-1 | Bio-compatibility |
| --- | --- | --- | --- | --- |
| Electrodes of handpiece | SUS 304 | Intact Skin | Limited (< 24 hours) | Yes |
| LED Display part | PC(SR3108FM) | | | |
## 10. Conclusion
Based on the comparison with the predicate devices and on the non-clinical performance testing results demonstrating that the Apollo Quattro is as safe and effective as the predicate devices, it can be concluded that the Apollo Quattro is substantially equivalent to the legally marketed predicate devices.
Page 5 / 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.