K212253 · Weero Co. · PBX · Aug 25, 2022 · General, Plastic Surgery
Device Facts
Record ID
K212253
Device Name
Apollo Duet (Model : APD-4000)
Applicant
Weero Co.
Product Code
PBX · General, Plastic Surgery
Decision Date
Aug 25, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Apollo duet(RF Mode) is intended for relief of minor muscle aches and pain, relief of muscle spasm, and temporary improvement of local blood circulation.
Device Story
Apollo Duet is a bipolar radio-frequency (RF) device used in hospital or clinic settings by healthcare professionals. It consists of a main body, power adapter, and an RF handpiece with four electrodes. The device outputs 1MHz bipolar RF energy to the skin to generate heat. An integrated temperature sensor monitors skin surface temperature and automatically terminates output if it exceeds 45°C. The device is used to provide therapeutic heat for muscle pain, spasms, and circulation improvement. Providers apply the handpiece to the patient's skin; the thermal effect is intended to benefit the patient by alleviating musculoskeletal symptoms.
Clinical Evidence
No clinical trials were conducted. Evidence includes bench testing, ex-vivo thermal effect testing on porcine abdominal skin, and a human volunteer study (n=6) measuring skin surface temperature using a thermographic camera to verify the 45°C safety cutoff. Compliance with electrical safety and EMC standards (IEC 60601 series) was demonstrated.
Technological Characteristics
Bipolar RF energy source; 1MHz frequency; 40W max output. Materials: SUS 304 (electrode), PC (display). Integrated temperature sensor for thermal safety cutoff. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 60601-2-10, ISO 14971, IEC 60601-1-6. Biocompatibility per ISO 10993-1 for limited contact (<24 hours).
Indications for Use
Indicated for adults requiring relief of minor muscle aches and pain, relief of muscle spasm, and temporary improvement of local blood circulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 25, 2022
Weero Co. Eunice Cho Quality Management Representative A605 VentureValley II, 142-10, Saneop-ro 156beon-gil, Gwonseon-gu, Suwon-si, Gyeonggi-do, Republic of Korea
Re: K212253
Trade/Device Name: Apollo Duet (Model : APD-4000) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: PBX, GEI Dated: July 15, 2021 Received: July 19, 2021
Dear Eunice Cho:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Colin K. Chen, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212253
Device Name Apollo Duet (Model name: APD-4000)
Indications for Use (Describe)
Apollo duet(RF Mode) is intended for relief of minor muscle aches and pain, relief of muscle spasm, and temporary improvement of local blood circulation.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| | Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K212253 Exhibit 5
Date of Summary Preparation: August. 23, 2022
- 1. Submitter and US Official Correspondent Submitter : WEERO Co. Address: A605 VentureValley II, 142-10, Saneop-ro 156beon-gil, Gwonseon-gu, Suwon-si, Gyeonggi-do, Republic of Korea Telephone No.: +82-31-5182-8588
| Official Correspondent: Eunice cho | |
|------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Correspondent: | WEERO Co. |
| Address: | A605 VentureValley II, 142-10, Saneop-ro 156beon-gil,<br>Gwonseon-gu, Suwon-si, Gyeonggi-do, Republic of Korea |
| Telephone No.: | +82-31-5182-8588 |
| Email: | eunice@weeroweero.com |
| 2. | Device Information | |
|----|----------------------|------------------------------------------------------|
| | Trade/Device Name: | Apollo Duet |
| | Regulation Name: | Electrosurgical cutting and coagulation device and |
| | | accessories |
| | Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| | Product Code: | GEI, PBX |
| | Device Class: | Class II per regulation 21CFR878.4400 |
- Predicate Device(Equivalent Legally Marketed Device) 3.
| Manufacturer: | InMode Ltd. |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | InMode RF Pro System |
| 510(k) Number: | K210492 |
| Classification: | Electrosurgical, Cutting & Coagulation & Accessories<br>: GEI,<br>Massager, Vacuum, Radio Frequency Induced Heat: PBX<br>Massager, Therapeutic, Electric: ISA,<br>Massager, Vacuum, Light Induced Heating: NUV<br>Class II per regulation 21CFR878.4400. |
- 5. Description of the Device
Apollo Duet has RF handpiece attached to the main body, and consists of an electrode cable and a power adapter.
This equipment has RF mode, which is the main function, RF mode is a mode in which Radio-frequency is output in a bipolar method. An RF handpiece is used, and a high frequency is output through 4ea RF handpiece electrodes. In addition, the RF handpiece has a temperature sensor, so the output is automatically turned off when the skin surface temperature exceeds 45 C
- 6. Indications for use (intended use) Apollo duet (RF Mode) is intended for relief of minor muscle aches and pain, relief of muscle spasm, and temporary improvement of local blood circulation.
{4}------------------------------------------------
| Apollo Duet |
|----------------|
| 510(k) Summary |
### Substantial equivalence chart 7.
| Name | Proposed device | Predicate device |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Apollo Duet (RF Mode) | InMode RF Pro System<br>(Applicator: Forma) |
| Product Code<br>Class | GEI<br>Class II | GEI<br>Class II |
| Manufacturer | WEERO Co. | InMode Ltd. |
| 510(k) No. | K212253 | K210492 |
| Indications<br>for use | Apollo duet(RF Mode) is intended for<br>relief of minor muscle aches and pain,<br>relief of muscle spasm, and temporary<br>improvement of local blood circulation. | The InMode RF Pro System with the<br>Non-invasive RF Applicators employs<br>RF energy for various applications:<br>i-Forma, Forma (Plus), Plus (Plus Plus)<br>and Plus90 for relief of minor muscle<br>aches and pain, relief of muscle spasm,<br>and temporary improvement of local<br>blood circulation. |
| Target Population | Adults requiring treatment as<br>specified in the indication for use | Adults requiring treatment as<br>specified in the indication for use |
| Environment Used | Hospital or Clinic setting | Hospital or Clinic setting |
| Input Power | 100-240V~, 50/60Hz | 100-240V~, 50/60Hz |
| Energy Used | (Non-invasive) Bipolar RF | (Non-invasive) Bipolar RF |
| Maximal<br>RF output power | 40[W] | Forma: 50[W] |
| RF Frequency | 1MHz | 1MHz |
| Biocompatibility | Materials are biocompatible. | Materials are biocompatible. |
| Sterility | NA | NA |
| Reprocessing | Applicator to be reprocessed in<br>accordance with user manual<br>instructions. | Applicator to be reprocessed in<br>accordance with user manual<br>instructions. |
The proposed device uses similar or identical technology as the predicate devices and has same intended uses. Based upon the predicted overall performance characteristics for the Apollo Duet, WEERO Co. believes that no significant differences in usage of its underlying technological principles between Apollo Duet and the predicate devices.
{5}------------------------------------------------
# Apollo Duet
# 510(k) Summary
#### 8. Software / Cybersecurity
The Apollo Duet software design and development follows FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Device".
The Apollo Duet cybersecurity follows FDA Guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Device:.
# 9. Electrical Safety, Electromagnetic Compatibility Testing & Performance Data
Bench tests were conducted to verify that the proposed device met all design specifications.
Ex-vivo testing was performed to compare RF ablation thermal effects of subject device and predicate device on porcine abdominal skin tissue.
Skin temperature was measured when using the subject device on 6 healthy volunteers, including 3 people of light skin and 3 people of dark skin, with a thermographic camera at the irradiated areas on the skin.
Safety testing results demonstrated that the proposed device complies with the following standards:
- Basic safety and essential performance of the Apollo Duet is tested and evaluated according to the IEC 60601-1:2005, AMD1:2012.
- Effect to the device by electromagnetic disturbances were tested and evaluated according to the FDA-recognized consensus standard IEC 60601-1-2:2014.
- Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories is tested and evaluated according to the IEC 60601-2-2:2017.
- Particular requirements for the basic safety and essential performance of nerve and muscle stimulators is tested and evaluated according to the IEC 60601-2-10:2012/A1:2016.
- Risk management was recorded by referring to ISO 14971:2019.
- Usability was documented by referring to IEC 60601-1-6:2010(Third Edition) + A1:2013.
10. Biocompatibility
| Part | Material | Patient Contact | Duration of<br>Contact by<br>ISO 10993-1 | Bio-<br>compatibility |
|-------------|---------------|-----------------|------------------------------------------|-----------------------|
| Electrode | SUS 304 | Intact Skin | Limited<br>(< 24 hours) | Yes |
| LED Display | PC (SR3108FM) | Intact Skin | Limited<br>(< 24 hours) | Yes |
# 11. Conclusion
Based on the information provided in this Summary, WEERO Co. the Apollo Duet is as safe and effective as the predicate device as indicated for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.