FIBERGRAFT™ BG Putty GPS Bone Graft Substitute FIBERGRAFT™ BG Putty Bone Graft Substitute FIBERGRAFT™ AERIDYAN™ Matrix Bone Graft Substitute FIBERGRAFT™ BG Matrix Bone Graft Substitute
K253147 · Prosidyan, Inc. · MQV · Oct 22, 2025 · Orthopedic
Device Facts
Record ID
K253147
Device Name
FIBERGRAFT™ BG Putty GPS Bone Graft Substitute FIBERGRAFT™ BG Putty Bone Graft Substitute FIBERGRAFT™ AERIDYAN™ Matrix Bone Graft Substitute FIBERGRAFT™ BG Matrix Bone Graft Substitute
Applicant
Prosidyan, Inc.
Product Code
MQV · Orthopedic
Decision Date
Oct 22, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
FIBERGRAFT™ BG Putty GPS - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT™ BG Putty GPS is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, intervertebral disc space, extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ BG Putty GPS must be used with autograft in the posterolateral spine. When used in intervertebral body fusion procedures, FIBERGRAFT™ BG Putty GPS must be used on its own with an intervertebral body fusion device cleared by FDA for use with a bone void filler. FIBERGRAFT™ BG Putty GPS is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization. FIBERGRAFT™ BG Putty - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT™ BG Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, intervertebral disc space, extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ BG Putty must be used with autograft in the posterolateral spine. When used in intervertebral body fusion procedures, FIBERGRAFT™ BG Putty must be used on its own with an intervertebral body fusion device cleared by FDA for use with a bone void filler. FIBERGRAFT™ BG Putty is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization. FIBERGRAFT™ AERIDYAN™ Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT™ AERIDYAN™ Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the posterolateral spine, extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ AERIDYAN™ Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine. FIBERGRAFT™ AERIDYAN™ Matrix must be hydrated with saline or blood for pelvis and extremity applications. FIBERGRAFT™ BG Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT™ BG Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ BG Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine.
Device Story
Osteoconductive, resorbable bone graft substitutes; used as non-structural scaffolds to fill bony voids/gaps. Composed of 45S5 bioactive glass, boron bioactive glass, and/or bovine type I collagen in an absorbable polymeric carrier. Applied by surgeon via gentle packing into defect sites; may be hydrated with saline, blood, or bone marrow aspirate. Resorbs and is replaced by bone during healing. Intended for use in clinical settings (e.g., OR) by surgeons. No changes to materials, manufacturing, or sterilization from predicate devices; submission solely expands indications to include benign bone cysts and tumors in pediatric patients ≥ 6 years old.
Clinical Evidence
No clinical data provided; submission relies on pre-clinical testing from previous 510(k) clearances.
Technological Characteristics
Resorbable calcium salt bone void fillers. Materials: 45S5 bioactive glass, boron bioactive glass (Borospheres™), bovine type I collagen, and absorbable polymeric carrier. Form factors: Putty (in spindle drive syringe) or Matrix. Non-structural, osteoconductive scaffolds.
Indications for Use
Indicated for bony voids or gaps not intrinsic to skeletal stability in adults and pediatric patients ≥ 6 years old. Used for defects resulting from benign bone cysts/tumors, surgical creation, or traumatic injury in posterolateral spine, intervertebral disc space, extremities, and pelvis. Contraindicated for load-bearing applications.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
MONTAGE Settable, Resorbable Bone Putty, MONTAGE-QS Settable, Resorbable Bone Putty, MONTAGE Flowable Settable, Resorbable Bone Paste, MONTAGE-XT Settable, Resorbable Hemostatic Bone Putty (K243526)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 22, 2025
Prosidyan, Inc.
% Dr. Mehdi Kazemzadeh-Narbat
Director, Regulatory Affairs
MCRA, LLC
803 7th Street NW, Floor 3
Washington, DC 20001 USA
Re: K253417
Trade/Device Name: FIBERGRAFT™ BG Putty GPS Bone Graft Substitute FIBERGRAFT™ BG Putty Bone Graft Substitute FIBERGRAFT™ AERIDYAN™ Matrix Bone Graft Substitute FIBERGRAFT™ BG Matrix Bone Graft Substitute
Regulation Number: 21 CFR 888.3045
Regulation Name: Resorbable Calcium Salt Bone Void Filler Device
Regulatory Class: Class II
Product Code: MQV
Dated: September 25, 2025
Received: September 25, 2025
Dear Dr. Mehdi Kazemzadeh-Narbat:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253417 - Dr. Mehdi Kazemzadeh-Narbat
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253417 - Dr. Mehdi Kazemzadeh-Narbat
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JESSE
MUIR -S
Digitally signed by
JESSE MUIR -S
Date: 2025.10.22
14:18:37 -04'00'
Jesse Muir, Ph.D.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253417 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| FIBERGRAFT™ BG Putty GPS Bone Graft Substitute
FIBERGRAFT™ BG Putty Bone Graft Substitute
FIBERGRAFT™ AERIDYAN™ Matrix Bone Graft Substitute
FIBERGRAFT™ BG Matrix Bone Graft Substitute | | |
| Please provide your Indications for Use below. | | ? |
| FIBERGRAFT™ BG Putty GPS:
“FIBERGRAFT™ BG Putty GPS - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT™ BG Putty GPS is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, intervertebral disc space, extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ BG Putty GPS must be used with autograft in the posterolateral spine. When used in intervertebral body fusion procedures, FIBERGRAFT™ BG Putty GPS must be used on its own with an intervertebral body fusion device cleared by FDA for use with a bone void filler. FIBERGRAFT™ BG Putty GPS is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization.” | | |
| FIBERGRAFT™ BG Putty:
“FIBERGRAFT™ BG Putty - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT™ BG Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, intervertebral disc space, extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ BG Putty must be used with autograft in the posterolateral spine. When used in intervertebral body fusion procedures, FIBERGRAFT™ BG Putty must be used on its own with an intervertebral body fusion device cleared by FDA for use with a bone void filler. FIBERGRAFT™ BG Putty is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization.” | | |
| FIBERGRAFT™ AERIDYAN™ Matrix Bone Graft Substitute:
“FIBERGRAFT™ AERIDYAN™ Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT™ AERIDYAN™ Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the posterolateral spine, extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ AERIDYAN™ Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine. FIBERGRAFT™ AERIDYAN™ Matrix must be hydrated with saline or blood for pelvis and extremity applications.” | | |
| FIBERGRAFT™ BG Matrix Bone Graft Substitute:
“FIBERGRAFT™ BG Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not | | |
{4}
intrinsic to the stability of the bony structure. FIBERGRAFT™ BG Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ BG Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine."
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
{5}
K253147
Prosidyan®
# 510(k) Summary
## Manufacturer:
Prosidyan, Inc.
41 Spring Street Suite 107
New Providence, NJ 07974
## Contact:
Mehdi Kazemzadeh-Narbat, PhD, PMP
Director, Regulatory Affairs
803 7th Street NW, Floor 3
Washington, DC 20001
Office: 202.552.6011
Email: mkazemzadeh@mcra.com
## Prepared By:
MCRA an IQVIA business
Mehdi Kazemzadeh-Narbat, PhD, PMP
Director, Regulatory Affairs
803 7th Street NW, Floor 3
Washington, DC 20001
Office: 202.552.6011
Email: mkazemzadeh@mcra.com
## Date Prepared:
October 7, 2025
## Device Trade Name:
FIBERGRAFT™ BG Putty Bone Graft Substitute
FIBERGRAFT™ BG Putty GPS Bone Graft Substitute
FIBERGRAFT™ AERIDYAN™ Matrix Bone Graft Substitute
FIBERGRAFT™ BG Matrix Bone Graft Substitute
## Device Common Name:
Resorbable calcium salt bone void filler device
## Classification:
21 CFR 888.3045
Class II
## Product Codes:
MQV
{6}
2
# Indications for Use:
"FIBERGRAFT™ BG Putty GPS - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT™ BG Putty GPS is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, intervertebral disc space, extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ BG Putty GPS must be used with autograft in the posterolateral spine. When used in intervertebral body fusion procedures, FIBERGRAFT™ BG Putty GPS must be used on its own with an intervertebral body fusion device cleared by FDA for use with a bone void filler. FIBERGRAFT™ BG Putty GPS is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization."
"FIBERGRAFT™ BG Putty - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT™ BG Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, intervertebral disc space, extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ BG Putty must be used with autograft in the posterolateral spine. When used in intervertebral body fusion procedures, FIBERGRAFT™ BG Putty must be used on its own with an intervertebral body fusion device cleared by FDA for use with a bone void filler. FIBERGRAFT™ BG Putty is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization."
"FIBERGRAFT™ AERIDYAN™ Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT™ AERIDYAN™ Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the posterolateral spine, extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ AERIDYAN™ Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine. FIBERGRAFT™ AERIDYAN™ Matrix must be hydrated with saline or blood for pelvis and extremity applications."
{7}
"FIBERGRAFT™ Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT™ BG Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ BG Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine."
## Predicate Devices:
FIBERGRAFT™ BG Putty GPS (K222276)
FIBERGRAFT™ BG Putty (K222276)
FIBERGRAFT™ AERIDYAN™ Matrix Bone Graft Substitute (K213803)
FIBERGRAFT™ BG Matrix Bone Graft Substitute (K180080)
## Reference Devices:
GlassBone Granules (K242782)
MONTAGE Settable, Resorbable Bone Putty, MONTAGE-QS Settable, Resorbable Bone Putty, MONTAGE Flowable Settable, Resorbable Bone Paste, MONTAGE-XT Settable, Resorbable Hemostatic Bone Putty (K243526)
## Device Description:
FIBERGRAFT™ BG Putty GPS is an osteoconductive, resorbable, biocompatible bone graft substitute to be gently packed into defect sites and to be used as non-structural scaffolds. The FIBERGRAFT™ BG putty GPS is made from 45S5 bioactive glass, where the bioactive glass components are mixed with an absorbable/polymeric carrier to form a cohesive material. FIBERGRAFT™ BG Putty GPS is to be provided in a spindle drive syringe and is compatible with the GPS Cannula product.
FIBERGRAFT™ BG Putty is an osteoconductive, resorbable, biocompatible bone graft substitute to be gently packed into defect sites and to be used as non-structural scaffolds. The FIBERGRAFT™ BG putty is made from 45S5 bioactive glass, where the bioactive glass components are mixed with an absorbable/polymeric carrier to form a cohesive material.
FIBERGRAFT™ AERIDYAN™ Matrix product is composed of 45S5 bioactive glass (M-45 granules), boron bioactive glass (MS-B microspheres/Borospheres™) and bovine type I collagen. After hydration with saline, blood or bone marrow aspirate (BMA), the AERIDYAN™ Matrix can be applied directly to the defect site or molded into the desired shape and gently packed into the defect site as a non-setting putty. In posterolateral spine fusion applications, the product is intended to be hydrated with bone marrow aspirate (BMA) and mixed with autograft in a recommended 1:1 ratio. In pelvis and extremity applications, the product must be hydrated with saline or blood.
{8}
FIBERGRAFT BG Matrix product is composed of 45S5 bioactive glass components (M-45 granules, MS-45 microspheres) and bovine type I collagen. The BG Matrix after hydration with saline, blood, or bone marrow aspirate (BMA) can be applied to the defect site or can be molded into the desired shape and gently packed into the defect site as a non-setting putty. In the posterolateral spine fusion applications, the product is intended to be hydrated with BMA and mixed with autograft in a 1:1 ratio.
## Purpose of this Submission:
The purpose of this Special 510(k) is to expand the indications for use statement (IFUS) of the FIBERGRAFT™ BG Putty and FIBERGRAFT™ BG Putty GPS Bone Graft Substitute (K222276), FIBERGRAFT™ AERIDYAN™ Matrix Bone Graft Substitute (K213803), and FIBERGRAFT™ BG Matrix Bone Graft Substitute (K180080), to include the statement “benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old)” to the current IFUS. Besides the addition of this statement, there is no change in the subject device materials, manufacturing process, sterilization, and packaging process.
## Substantial Equivalence and Predicate Devices:
The subject device is substantially equivalent to its reference devices in terms of added language to the Indications for Use; “benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old)” and identical to predicate devices in terms of materials, manufacturing process, sterilization, and packaging process. All pre-clinical testing conducted to support the previous 510(k) clearances are applicable to the subject device. The only modification proposed in this Special 510(k) is to include the statement “benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old)” to the current IFUS.
## Conclusion:
The subject FIBERGRAFT™ bone substitutes are substantially equivalent to its predicates with respect to intended use, technological characteristics, and performance.
4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.