GlassBone Granules is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. GlassBone Granules resorbs and is replaced with bone during the healing process.
Device Story
GlassBone Granules is a synthetic, biocompatible, osteoconductive bone void filler composed of 45S5 bioactive glass (SiO2, P2O5, CaO, Na2O). The device is supplied as amorphous, non-porous, random-shaped particles in various unit sizes. It is intended for surgical implantation to fill non-structural bony voids or gaps in the extremities and pelvis. Upon implantation, the granules leach ions that react with body fluids, chemically transforming the glass surface into a structure resembling the mineral phase of natural bone. This process facilitates bone replacement during healing. The device is for single-use, gamma-sterilized, and MR safe. It is used by surgeons in clinical settings to treat osseous defects in adults and pediatric patients (≥ 6 years).
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing (pH, solubility, apatite formation, heavy metals, crystallinity, particle size, surface area) and biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, irritation, pyrogenicity, systemic toxicity, implantation effects, genotoxicity).
Technological Characteristics
Material: 45S5 bioactive glass (SiO2, P2O5, CaO, Na2O). Form: Amorphous, non-porous, random-shaped particles. Sizes: 0.5-1mm and 1-3mm. Sterilization: Gamma irradiation (SAL 10^-6). Standards: ASTM F1538, ISO 10993, ISO 11137, ISO 11607. Osteoconductive, MR safe.
Indications for Use
Indicated for filling bony voids or gaps in the skeletal system (extremities and pelvis) resulting from benign bone cysts, tumors, surgical creation, or traumatic injury in adults and pediatric patients ≥ 6 years old. Not for use in defects intrinsic to bony structure stability.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
NovaBone Resorbable Bone Graft Substitute (K052494)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
December 23, 2024
Noraker % Elisa Maldonado-Holmertz RA/QA Consultant Obelix Consulting 806 Jefferson Street Bastrop, Texas 78602
Re: K242782
Trade/Device Name: GlassBone Granules Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: November 12, 2024 Received: November 13, 2024
Dear Ms. Maldonado-Holmertz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by JESSE MUIR JESSE MUIR -S Date: 2024.12.23 09:42:12 -05'00
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K242782
Device Name
GlassBone Granules
#### Indications for Use (Describe)
GlassBone Granules is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. GlassBone Granules resorbs and is replaced with bone during the healing process.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
#### SUBMITTER
#### Submission Sponsor:
Catherine FLACARD Compliance & Innovation Chief Officer Noraker, SAS 60, Avenue Rockefeller 69008 Lyon France Email: c.flacard@noraker.com Tel number: +33 (0)4 78 93 30 92
#### Submission Correspondent:
Elisa Maldonado-Holmertz RA/QA Consultant Obelix Consulting, LLC 806 Jefferson St Bastrop, TX 78602 USA Email: elisamh@obelixconsult.com Tel number: +1 512 431 6069
#### Date Prepared: 13 September 2024
#### SUBJECT DEVICE
Name of Device: GlassBone Granules Common or Usual Name: Filler, Bone Void, Calcium Compound Classification Name: Resorbable calcium salt bone void filler device (21 CFR 888.3045) Regulatory Class: Class II Product Code: MQV
#### PREDICATE DEVICE
BonAlive Granules - K231528
{5}------------------------------------------------
This predicate has not been subject to a design-related recall.
#### REFERENCE DEVICE
NovaBone Resorbable Bone Graft Substitute - K052494 This predicate has not been subject to a design-related recall.
#### DEVICE DESCRIPTION
GlassBone Granules is a synthetic and biocompatible bone substitute device (bioactive glass 4555). GlassBone Granules is composed by weight of 4555 bioactive glass granules (45% SiO2, 6% P2O5, 24.5% CaO and 24.5% Na2O).
It is sterilized with gamma irradiation and is single use and is available in different unit sizes.
#### INDICATIONS FOR USE
GlassBone Granules is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. GlassBone Granules resorbs and is replaced with bone during the healing process.
Rx only.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
GlassBone Granules is substantially equivalent to the following predicate and reference device:
| Manufacturer | Subject Device<br>Noraker SAS | Primary predicate<br>BonAlive Biomaterials, Ltd | Reference Device<br>NovaBone | SIGNIFICANT<br>DIFFERENCES |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Trade Name | GlassBone® Granules | BonAlive® Granules | Resorbable Bone Graft Substitute<br>(Particulate) | None |
| 510(k) Number | - | K231528 | K052494 | - |
| Product Code | MQV | MQV | MQV | None |
| Regulation<br>Number | 888.3045 | 888.3045 | 888.3045 | None |
| Regulation Name | Resorbable calcium salt bone void<br>filler device | Resorbable calcium salt bone void<br>filler device | Resorbable calcium salt bone void<br>filler device | None |
| Indications for<br>Use | GlassBone Granules is an implant<br>intended to fill bony voids or gaps of<br>the skeletal system (i.e., extremities | Bonalive® Orthopedics granules is an<br>implant intended to fill bony voids or<br>gaps of the skeletal system (i.e., | NovaBone Resorbable Bone graft<br>substitute is indicated only for bony | None |
| Manufacturer | Subject Device<br>Noraker SAS | Primary predicate<br>BonAlive Biomaterials, Ltd | Reference Device<br>NovaBone | SIGNIFICANT<br>DIFFERENCES |
| Trade Name | GlassBone® Granules | BonAlive® Granules | Resorbable Bone Graft Substitute<br>(Particulate) | None |
| | and pelvis). These osseous defects<br>may be the result of benign bone<br>cysts and tumors (in adults and<br>pediatric patients ≥ 6<br>years old), are surgically created or<br>the result of traumatic injury to the<br>bone and are not intrinsic to the<br>stability of the<br>bony structure. GlassBone Granules<br>resorbs and is replaced with bone<br>during the healing process. | extremities<br>and pelvis). These osseous defects<br>may be the result of benign bone<br>cysts and tumors (in adults and<br>pediatric patients ≥ 6<br>years old), are surgically created or<br>the result of traumatic injury to the<br>bone and are not intrinsic to the<br>stability of the<br>bony structure. Bonalive®<br>Orthopedics granules resorbs and is<br>replaced with bone during the healing<br>process. | voids or gaps that are not intrinsic to<br>the stability of the bony structure.<br>NovaBone is indicated to be gently<br>packed into bony voids or gaps of the<br>skeletal system (i.e. the extremities,<br>spine and pelvis).<br>These defects may be surgically<br>created osseous defects or osseous<br>defects created from traumatic injury<br>to the bone. The product provides a<br>bone void filler that resorbs and is<br>replaced with bone during the healing<br>process. | |
| Rx or OTC | Rx | Rx | Rx | None |
| Physical Form | Amorphous, non-porous random-<br>shaped particles | Amorphous, non-porous random-<br>shaped particles | Amorphous, non-porous random-<br>shaped particles | None |
| Color | Brown | Brown | White | None |
| Materials<br>Composition | Bioactive glass<br>45S5 Bioglass (SiO₂, P₂O₅, CaO and<br>Na₂O) | Bioactive glass<br>S53P4 glass (SiO₂, P₂O₅, CaO and<br>Na₂O) | Bioactive glass<br>45S5 Bioglass (SiO₂, P₂O₅, CaO and<br>Na₂O) | None |
| Product Sizes | Granule sizes:<br>0.5-1mm<br>1-3mm<br>Product volumes:<br>0.5cc<br>1 cc<br>5 cc<br>10 cc<br>16 cc | Granule sizes:<br>0.5-0.8 mm<br>1.0-2.0 mm<br>Product volumes:<br>1 cc<br>2.5 cc<br>5 cc<br>10 cc | Granule sizes:<br>90-710μm<br>Product volumes:<br>2 cc<br>5 cc<br>10 cc<br>15 cc | Similar without<br>affecting safety or<br>effectiveness |
| Biocompatibility | Biocompatible<br>ISO 10993 | Biocompatible<br>ISO 10993 | Biocompatible<br>ISO 10993 | None |
| Sterilization | Gamma, SAL 10⁻⁶ | Gamma, SAL 10⁻⁶ | EtO, SAL 10⁻⁶ | None |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic | Non-pyrogenic | None |
| Single Use/<br>Reuse | Single use only | Single use only | Single use only | None |
| Mode of action | Works by leaching ions that react<br>with the body fluids transforming the<br>glass surface chemically into one that<br>by its chemical composition and<br>structure resembles the mineral<br>phase found in natural bone. | Works by leaching ions that react<br>with the body fluids transforming the<br>glass surface chemically into one that<br>by its chemical composition and<br>structure resembles the mineral<br>phase found in natural bone. | Works by leaching ions that react<br>with the body fluids transforming the<br>glass surface chemically into one that<br>by its chemical composition and<br>structure resembles the mineral<br>phase found in natural bone. | None |
| Manufacturer | Subject Device<br>Noraker SAS | Primary predicate<br>BonAlive Biomaterials, Ltd | Reference Device<br>NovaBone | SIGNIFICANT<br>DIFFERENCES |
| Trade Name | GlassBone® Granules | BonAlive® Granules | Resorbable Bone Graft Substitute<br>(Particulate) | None |
| Properties | Synthetic<br>Osteoconductive | Synthetic<br>Osteoconductive | Synthetic<br>Osteoconductive | None |
| MR safety | MR Safe | MR Safe | Unknown | None |
{6}------------------------------------------------
{7}------------------------------------------------
### NON CLINICAL PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination:
#### Bench testing:
Bench testing was conducted according to FDA Guidance "Resorbable Calcium Salt Bone Void Filler Device - Class II Special Controls Guidance Document for Industry and FDA Staff" and ASTM F1538.
- . pH
- Solubility .
- Apatite
- Composition - Heavy Metals
- Crystallinity
- Particle Size
- . Surface Area
### Animal testing and biocompatibility:
Biocompatibility testing was conducted in accordance with ISO-10993-1, "Biological evaluation of medical device – Part 1: Evaluation and testing within a risk management process".
- Physico-chemical characterization
- Cytotoxicity
- Sensitization
- Irritation or intracutaneous reactivity
- Material mediated pyrogenicity
- Acute systemic toxicity
- Local effects of implantation
- Genotoxicity
{8}------------------------------------------------
## Sterility and Shelf Life
The following testing was performed on GlassBone Granules:
- ISO 11137 Radiation Sterilization Validation ●
- ISO 11607 Packaging for terminally sterilized medical devices ●
## Clinical Performance Data
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device.
## Statement of Substantial Equivalence
Noraker GlassBone Granules is substantially equivalent to the predicate device when comparing the following characteristics:
- •Indications for Use
- •Principles of Use
- •Fundamental Technology
- •Basic Design
- •Materials Used
- •Where used
- •Standards met
- •Sterilization Method
GlassBone Granules meet all requirements for overall design and confirms that the output meets the design inputs and specifications. The GlassBone Granules passed all testing stated above as shown by acceptable results obtained.
GlassBone Granules complies with the applicable voluntary, national, and international standards and guidance.
Noraker, SAS believes that the bench and non-clinical performance data supports the safety of GlassBone Granules, and the device should perform effectively as intended in the specified use conditions. The non-clinical performance data demonstrates that GlassBone Granules performs comparably to the predicate device that is currently marketed for the same intended use and presents no new concerns about safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.