Tournicare ARMA Automatic Blood Pressure Monitor (ARMA-08)
Applicant
Tournicare Pty, Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Jun 15, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The ARMA automatic Blood Pressure Monitor is indicated for home use for the non-invasive measurement of diastolic and systolic blood pressures and pulse rate of adults by means of an inflatable cuff which is clamped to the upper arm. The intended upper arm circumference is 24 cm to 36 cm (9.4 in to 14.2 in).
Device Story
Automatic upper arm blood pressure monitor; home/OTC use. Input: oscillometric waveform via pressure sensor during brachial artery occlusion. Mechanism: inflatable bladder with clamp-style cuff; rechargeable battery powered. Output: systolic/diastolic blood pressure and pulse rate. User: patient self-use. Benefit: convenient, non-invasive home monitoring of arterial blood pressure and pulse rate.
Clinical Evidence
Clinical testing performed per ISO 81060-2:2018 to validate measurement accuracy. Bench testing included electrical safety (IEC 60601-1, IEC 80601-2-30), EMC (IEC 60601-1-2), battery safety (IEC 62133-2), software lifecycle (IEC 62304), human factors (ISO/IEC 62366-1), transportation (ISTA 3A), and biocompatibility (ISO 10993-1). Data demonstrates safety and effectiveness equivalent to predicate.
Technological Characteristics
Battery-powered non-invasive oscillometric blood pressure monitor. Features clamp-style cuff with inflatable bladder; includes size insert for 24-26 cm circumference. Standards: IEC 60601-1, IEC 80601-2-30, IEC 60601-1-2, IEC 62133-2, IEC 62304, ISO/IEC 62366-1, ISTA 3A, ISO 10993-1.
Indications for Use
Indicated for home use by adults for non-invasive measurement of diastolic/systolic blood pressure and pulse rate via upper arm cuff (circumference 24-36 cm).
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Omron Evolv Model BP7000 Upper Arm Blood Pressure Monitor (K162092)
Submission Summary (Full Text)
{0}

June 15, 2026
Tournicare Pty Ltd.
Roger Gray
VP Qa/ra
Donawa Lifescience
Piazza Albania 10
Rome, 00153
Italy
Re: K253059
Trade/Device Name: Tournicare ARMA Automatic Blood Pressure Monitor (ARMA-08)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: May 15, 2026
Received: May 15, 2026
Dear Roger Gray:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device'
{1}
K253059 - Roger Gray
Page 2
(https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K253059 - Roger Gray
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
STEPHEN C. BROWNING -S
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253059 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Tournicare ARMA Automatic Blood Pressure Monitor (ARMA-08) | | |
| Please provide your Indications for Use below. | | ? |
| The ARMA automatic Blood Pressure Monitor is indicated for home use for the non-invasive measurement of diastolic and systolic blood pressures and pulse rate of adults by means of an inflatable cuff which is clamped to the upper arm. The intended upper arm circumference is 24 cm to 36 cm (9.4 in to 14.2 in). | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
Page 1 of 1
{4}
510(k) #: K253059
510(k) Summary
Prepared on: 2026-06-13
# Contact Details
21 CFR 807.92(a)(1)
Applicant Name
Tournicare Pty Ltd.
Applicant Address
18 Elliot Avenue Balwyn Victoria 3103 Australia
Applicant Contact Telephone
+61 (0) 414 373
Applicant Contact
Mr. Niels van Sparrentak
Applicant Contact Email
nvanspar@tournicare.com
Correspondent Name
Donawa Lifescience
Correspondent Address
Piazza Albania 10 Rome 00153 Italy
Correspondent Contact Telephone
+39 06 578 2665
Correspondent Contact
Mr. Roger Gray
Correspondent Contact Email
rgray@donawa.com
# Device Name
21 CFR 807.92(a)(2)
Device Trade Name
Tournicare ARMA Automatic Blood Pressure Monitor (ARMA-08)
Common Name
Noninvasive blood pressure measurement system
Classification Name
System, Measurement, Blood-Pressure, Non-Invasive
Regulation Number
870.1130
Product Code(s)
DXN
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
Predicate #
Predicate Trade Name (Primary Predicate is listed first)
Product Code
K162092
Omron Evolv Model BP7000 Upper Arm Blood Pressure Monitor
DXN
# Device Description Summary
21 CFR 807.92(a)(4)
The ARMA Blood Pressure Monitor (BPM) device is an automatic upper arm blood pressure monitor intended for non-invasive measurement or monitoring of adults' arterial blood pressure and pulse rate. The device is powered by a rechargeable battery and is designed for home (OTC) use. ARMA BPM operates by occluding the brachial artery with an inflatable bladder. A pressure sensor detects and generates an oscillometric waveform from which systolic and diastolic blood pressure values and the pulse rate are extrapolated. The physical architecture of the device is different from classic blood pressure measurement cuffs which feature a full wrap-around cuff and bladder, whereas the ARMA uses a clamp mechanism and bladder. A size insert is also supplied with the device for users with a small arm circumference 24 cm to 26 cm.
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The ARMA automatic Blood Pressure Monitor is indicated for home use for the non-invasive measurement of diastolic and systolic blood
K253059
Page 1 of 2
{5}
pressures and pulse rate of adults by means of an inflatable cuff which is clamped to the upper arm. The intended upper arm circumference is 24 cm to 36 cm (9.4 in to 14.2 in).
## Indications for Use Comparison
21 CFR 807.92(a)(5)
Indications for use are substantially similar. The subject device does not include irregular heartbeat detection, which is included in the predicate device. The upper arm circumference range is different, with the subject device having a smaller arm circumference range. These differences do not constitute a new intended use.
## Technological Comparison
21 CFR 807.92(a)(6)
The subject and predicate devices are both battery-powered and use the same principle of operation (oscillometric method, automatic inflation, deflation and measurement) to measure the user's blood pressure and pulse rate.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Non-clinical testing included demonstration of compliance with applicable international, FDA-recognized, standards including:
- Electrical safety in accordance with IEC 60601-1 and IEC 80601-2-30
- EMC (Electromagnetic Compatibility) in accordance with IEC 60601-1-2
- Lithium-ion battery safety in accordance with IEC 62133-2
- Software life cycle compliance in accordance with IEC 62304
- Human factors in accordance with ISO/IEC 62366-1
- Transportation simulation in accordance with ISTA 3A
- Biocompatibility of patient-contacting parts in accordance with FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" - Guidance for Industry and Food and Drug Administration Staff", September 8, 2023
Clinical testing was carried out in accordance with international, FDA-recognized, standard ISO 81060-2:2018, "Non-invasive sphygmomanometers, Part 2: Clinical investigation of intermittent automated measurement type", to evaluate the accuracy of the ARMA device.
ARMA BPM has similar indications for use, technological characteristics, and principles of operation to the predicate device cleared under 510(k) submission K162092. Clinical and bench test performance data demonstrate that the ARMA BPM is as safe and as effective as the predicate and does not raise any new safety and effectiveness concerns. Clinical performance has been validated in accordance with an FDA-recognized standard.
K253059
Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.