Evolv Model BP7000 Upper Arm Blood Pressure Monitor
K162092 · Omron Healthcare, Inc. · DXN · Nov 22, 2016 · Cardiovascular
Device Facts
Record ID
K162092
Device Name
Evolv Model BP7000 Upper Arm Blood Pressure Monitor
Applicant
Omron Healthcare, Inc.
Product Code
DXN · Cardiovascular
Decision Date
Nov 22, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with arm circumference ranging from 9 inches to 17 inches (22cm to 42cm). The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.
Device Story
Battery-powered automatic non-invasive blood pressure monitor for home use; utilizes oscillometric method. Device inflates cuff via integral pump; monitors cuff pressure; extracts pulse waveform data. Software analyzes waveform to determine systolic/diastolic blood pressure and pulse rate. User operates device by applying cuff, powering on/off, and initiating measurement. Optional smartphone app (Omron connect) displays trend graphs; app does not perform diagnostics or medical analysis. Device stores up to 100 readings. Benefits include convenient home monitoring of blood pressure and pulse, with irregular heartbeat detection to alert users.
Clinical Evidence
Clinical investigation validated BP7000 accuracy against auscultation method using a calibrated sphygmomanometer by trained medical staff. Results demonstrated equivalent performance to the auscultation reference. Nonclinical testing included comparative bench testing, biocompatibility (ISO 10993-1), electrical safety, EMC, and usability testing in simulated home environments.
Technological Characteristics
Oscillometric blood pressure measurement; cuff pressure range 0-299mmHg; pulse rate range 40-180 beats/min. Battery-powered. One-piece cuff/display design. Biocompatible materials per ISO 10993-1. Connectivity via optional Bluetooth to smartphone app. Software-based analysis of pulse waveforms.
Indications for Use
Indicated for adult patients with arm circumference 22-42cm for measurement of blood pressure and pulse rate, including detection of irregular heartbeats.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Omron HEM-7320 Noninvasive blood pressure measurement system (K133383)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The seal is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 22, 2016
Omron Healthcare, Inc. % Ronald Warren Senior Director, Regulatory Affairs Experien Group 755 N. Mathilda Ave. Suite 100 Sunnyvale, California 94085
Re: K162092
Trade/Device Name: Evolv Model Bp7000 Upper Arm Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: October 11, 2016 Received: October 13, 2016
Dear Ronald Warren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerel
Mark Jellman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162092
Device Name Evolv Model BP7000 Upper Arm Blood Pressure Monitor
Indications for Use (Describe)
The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with arm circumference ranging from 9 inches (22cm to 42cm). The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY ...
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## OMRON EVOLV MODEL BP7000 UPPER ARM BLOOD PRESSURE MONITOR 510(k) SUMMARY
# 510(k) Notification K162092
#### GENERAL INFORMATION [807.92(a)(1)]
#### Applicant:
Omron Healthcare, Inc. 1925 West Field Court Lake Forest, IL 60045 USA Phone: 847-247-5626 FAX: 847-680-6269
Contact Person: Ronald S. Warren Senior Director, Regulatory Affairs Experien Group, LLC 224 Airport Parkway, Suite 250 San Jose, CA 95110 USA Phone: 408-505-3926 FAX: 408-400-0856
Date Prepared: October 7, 2016
#### DEVICE INFORMATION
#### Trade Name:
Evolv Model BP7000 Upper Arm Blood Pressure Monitor
## Generic/Common Name:
Noninvasive blood pressure measurement system
#### Classification:
Class II per 21CFR870.1130
Product Code: DXN
{4}------------------------------------------------
# PREDICATE DEVICE(S)
The Evolv Model BP7000 Upper Arm Blood Pressure Monitor is substantially equivalent to the Omron HEM-7320 Noninvasive blood pressure measurement system (K133383).
# DEVICE DESCRIPTION
The Evolv Model BP7000 Upper Arm Blood Pressure Monitor ("BP7000") is a battery powered automatic non-invasive blood pressure system intended for home use. The device inflates a cuff with an integral controllable pump, then deflates the cuff via an electric valve. During inflation the cuff pressure is monitored and pulse waveform data is extracted. The extracted pulse waveform data is then analyzed by software which determines pulse rate, as well as systolic and diastolic blood pressure. The systolic and diastolic blood pressures are measured using the oscillometric method. The cuff pressure range is 0 to 299mmHg and the pulse rate range is 40 to 180 beats/min.
BP7000 is intended to be used for arms ranging from 22 to 42cm in circumference. The cuff is not replaceable. The device also detects the appearance of irregular heartbeats during measurement. The device displays the latest blood pressure reading, while up to 100 readings can be stored in memory. The operation of the device is intended for home use. Functions and other features that are controlled by the end user include applying the arm cuff to the upper arm, powering on/off the system, starting or stopping the BP and pulse measurement cycle, and installing and changing the batteries as needed. As an optional feature, the user can also pair the BP7000 to a smartphone when employing the "Omron connect" app. This app is an optional feature and is only intended to display trend graphs of measured systolic and diastolic blood pressure, and pulse rate. This app does not provide any diagnostic or measurement functions, and does not interpret or analyze the data for medical decision making. Unlimited readings can be stored in the app for archiving and review by the user. Aside from this optional app for smartphones, BP7000 does not connect with other collateral devices.
## INDICATIONS FOR USE
The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with arm circumference ranging from 9 inches to 17 inches (22cm to 42cm). The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
The proposed BP7000 has similar technological characteristics as compared to the predicate HEM-7320 device. Both devices are intended for home use and employ the cuff oscillometric method for measuring blood pressure and pulse rate. Both devices have the same cuff pressure range of 0 to 299mmHg and pulse rate range is 40 to 180 beats/min. Both devices are intended for an arm circumference range of 22 to 42mm. The accuracy of pressure reading is ±3mmHg for both devices, and accuracy of pulse rate is ±5% in both devices. There are minor difference between BP7000 and HEM-7320 but these relate to convenience factors for home use. The cuff and display for BP7000 are provided in a one-piece configuration, while there is a separate cuff
{5}------------------------------------------------
## OMRON EVOLV MODEL BP7000 UPPER ARM BLOOD PRESSURE MONITOR 510(k) SUMMARY
and stand-alone desk top display for HEM-7320. BP7000 is battery powered while HEM-7320 is battery powered or AC powered. HEM-7320 has the option to store BP readings for up to two users. BP7000 stores data for one user. BP7000 has slightly smaller dimensions as weighs slightly less than HEM7320. HEM-7320 cuff can be deflated with automatic pressure release or controlled deflation settings. BP7000 cuff is deflated with automatic pressure release only.
# SUBSTANTIAL EQUIVALENCE
The indications for use for the predicate device is substantially equivalent to the proposed indications for use for BP7000. Comparison testing demonstrated that the proposed device is equivalent to the predicate device with regard to measurement of blood pressure in a pulse wave generator test. Minor differences in technological features relate to convenience considerations for home use but do not impact safety or performance of blood pressure or pulse rate measurements. Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. Thus, BP7000 is substantially equivalent to the predicate HEM-7320 device.
# PERFORMANCE DATA [807.92(b)]
All necessary bench and clinical testing was conducted on BP7000 to support a determination of substantial equivalence to the predicate device.
# Nonclinical Testing Summary:
The nonclinical, bench testing included:
- Comparative blood pressure and pulse rate testing to the predicate device
- Performance verification testing of BP7000 to confirm acceptable performance of device features and functions
- Cleaning verification testing to confirm device retains its performance when cuff is ● exposed to surfactants as may be experienced in home use environment
- . Usability testing with a representative population of study participants in a simulated home use environment
Other nonclinical safety testing included:
- Biocompatibility of patient-contacting materials per ISO 10993-1 requirements
- Electrical safety, electromagnetic compatibility, and electrostatic discharge testing
- o Software verification and validation
The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of BP7000 meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that BP7000 does not raise new questions of safety or effectiveness for measurement of blood pressure and pulse in a home use environment when compared to the predicate device.
{6}------------------------------------------------
# OMRON EVOLV MODEL BP7000 UPPER ARM BLOOD PRESSURE MONITOR 510(k) SUMMARY
### Clinical Testing Summary:
A clinical investigation was conducted with the objective of validating the accuracy of blood pressure measurements by BP7000 based on an oscillometric method as compared to an auscultation method using a calibrated sphygmomanometer by trained medical staff. The results demonstrated that BP7000 performed equivalently to the auscultation method.
### Conclusions
Based on the results from the nonclinical and clinical tests performed in support of BP7000, it is concluded that that the proposed device is safe, as effective, and performs at least as safely and effectively as the legally marketed predicate device.
#### SUMMARY
BP7000 and HEM-7320 are designed for the measurement of blood pressure, pulse rate and detection of irregular pulses in adult population for home use. These devices have the same indications for use. Performance characteristics related to BP measurement and pulse rate are comparable between the two devices. The minor differences in labeling and technological characteristics between the proposed device and the predicate devices have been evaluated and determined to not raise any new issues of safety or effectiveness. As such, the proposed BP7000 is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.