K252974 · Encore Medical L.P. · JWH · Dec 19, 2025 · Orthopedic
Device Facts
Record ID
K252974
Device Name
EMPOWR Knee
Applicant
Encore Medical L.P.
Product Code
JWH · Orthopedic
Decision Date
Dec 19, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Enovis™ knee devices are intended for treatment of patients who are candidates for knee arthroplasty per the Indication for use. While total knee replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients.
Device Story
EMPOWR Knee system; total knee prosthesis for arthroplasty. Submission introduces EMPOWR Dynamic Natural Alignment (DNA) surgical technique; kinematic alignment strategy preserving native joint line. No design changes to implants. Used by orthopedic surgeons in clinical settings. Implants restore mobility and reduce pain in patients with joint disability. Cemented or cementless application depending on specific component (e.g., Porous Knee variants).
Clinical Evidence
No clinical data submitted. Evidence consists of bench testing, including initial stability (micromotion) testing and cadaveric assessment of the kinematic surgical technique.
Technological Characteristics
Total knee prosthesis; metal/polymer/metal components. Materials and design remain unchanged from predicate devices. No software or electronic components. Surgical technique is kinematic alignment (DNA).
Indications for Use
Indicated for patients with disability due to degenerative, post-traumatic, or rheumatoid arthritis; avascular necrosis of femoral condyle; post-traumatic loss of joint configuration (with patellofemoral erosion/dysfunction or prior patellectomy); moderate valgus, varus, or flexion deformities; or unmanageable fractures. Also indicated for salvage of failed surgical attempts. Mechanical alignment approach covers all listed conditions; EMPOWR Dynamic Natural Alignment (DNA) approach covers arthritis, avascular necrosis, and moderate deformities.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
December 19, 2025
Encore Medical, L.P.
Morgan Paronish
Senior Regulatory Affairs Specialist
9800 Metric Boulevard
Austin, Texas 78758
Re: K252974
Trade/Device Name: EMPOWR Knee
Regulation Number: 21 CFR 888.3560
Regulation Name: Knee Joint Patellofemorotibial Polymer/Metal/Polymer Semi-Constrained Cemented Prosthesis
Regulatory Class: Class II
Product Code: JWH, MBH, OIY
Dated: September 17, 2025
Received: September 17, 2025
Dear Morgan Paronish:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}
K252974 - Morgan Paronish
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K252974 - Morgan Paronish
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lixin Liu -S
Lixin Liu, Ph.D
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
EMPOWR Knee
Page 7 of 42
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252974 | ? |
| Please provide the device trade name(s). | | ? |
| EMPOWR Knee | | |
| Please provide your Indications for Use below. | | |
| When a mechanical alignment approach is utilized, these devices are indicated for patients suffering from
disability due to:
• degenerative, post-traumatic or rheumatoid arthritis;
• avascular necrosis of the femoral condyle;
• post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or
prior patellectomy;
• moderate valgus, varus or flexion deformities;
• treatment of fractures that are unmanageable using other techniques.
These devices may also be indicated in the salvage of previously failed surgical attempts.
When the EMPOWR® Dynamic Natural Alignment (DNA) approach is utilized, these devices are indicated
for patients with disability due to:
• degenerative, post-traumatic or rheumatoid arthritis;
• avascular necrosis of the femoral condyle;
• moderate valgus, varus or flexion deformities.
All EMPOWR devices except for the EMPOWR Revision Knee™, EMPOWR PS Knee™, and EMPOWR
Partial Knee™ may be implanted using the EMPOWR® Dynamic Natural Alignment (DNA) surgical
technique.
All devices are intended for cemented applications except for the EMPOWR Porous® Knee Femur,
EMPOWR Porous® Knee Tibia, and Porous Patella which are intended for cementless applications.
While knee replacements are not intended to withstand activity levels and loads of normal healthy bone,
they are a means of restoring mobility and reducing pain for many patients. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
enovis
K252974 - Page 1 of 3
510(k) Summary
I. SUBMITTER
Encore Medical, L.P.
9800 Metric Blvd.
Austin, TX 78758
Contact Person: Morgan Paronish
Email: morgan.paronish@enovis.com
Phone: 317-519-5611
Date: December 3, 2025
II. DEVICE
Name of Device: EMPOWR Knee
Common or Usual Name: Total Knee Prosthesis
Classification Name: Knee joint patellofemorotibial polymer/metal/polymer semiconstrained cemented prosthesis.
Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis
Regulation: 21 CFR 888.3560, 21 CFR 888.3565
Regulatory Class: II
Product Codes: JWH, MBH, OIY
III. PREDICATE / REFERENCE DEVICES
| Predicates | EMPOWR 3D Knee - Primary Predicate | K143242 |
| --- | --- | --- |
| | EMPOWR CR Tibial Insert | K170573 |
| | EMPOWR Porous Knee | K171991 |
| | EMPOWR Universal Tibial Baseplate | K173723 |
| | EMPOWR Porous Femur with HAnano Surface | K210308 |
| | Domed Tri-Peg Patella, Highly Cross Linked Domed Tri-Peg Patella With VE | K121835 |
| | Porous Patella | K212941 |
| References | Medacta GMK Sphere – Kinematic Alignment | K173890 |
| | Zimmer Persona Personalized Knee System | K172524 |
IV. DEVICE DESCRIPTION
The purpose of this submission is to obtain clearance for the EMPOWR® Dynamic Natural Alignment (DNA) surgical technique. The proposed technique is a kinematic alignment (KA) strategy, providing an alternative alignment approach to the current mechanical alignment (MA) strategy. The purpose of the KA technique is to restore normal knee function by preserving the patient's native joint line. The technique may be used to implant all EMPOWR devices except for the EMPOWR Revision Knee™ and the EMOWR PS Knee™. This is a labeling change only. No design changes are being introduced to the implant systems.
Page 1 of 3
{5}
enovis
K252974 - Page 2 of 3
## V. INDICATIONS FOR USE / INTENDED USE
### Intended Use:
Enovis™ knee devices are intended for treatment of patients who are candidates for knee arthroplasty per the Indication for use. While total knee replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients.
### Indications for Use:
When a mechanical alignment approach is utilized, these devices are indicated for patients suffering from disability due to:
- degenerative, post-traumatic or rheumatoid arthritis;
- avascular necrosis of the femoral condyle;
- post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy;
- moderate valgus, varus or flexion deformities;
- treatment of fractures that are unmanageable using other techniques.
These devices may also be indicated in the salvage of previously failed surgical attempts.
When the EMPOWR® Dynamic Natural Alignment (DNA) approach is utilized, these devices are indicated for patients with disability due to:
- degenerative, post-traumatic or rheumatoid arthritis;
- avascular necrosis of the femoral condyle;
- Moderate valgus, varus, or flexion deformities.
All EMPOWR devices except for the EMPOWR Revision Knee™, EMPOWR PS Knee™, and EMPOWR Partial Knee™ may be implanted using the EMPOWR® DNA surgical technique.
All devices are intended for cemented applications except for the EMPOWR Porous® Knee Femur, EMPOWR Porous® Knee Tibia, and Porous Patella which are intended for cementless applications.
While knee replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring and reducing pain for many patients.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
No design changes are being introduced with this submission. Thus, there are no differences in technological characteristics between the subject and predicate devices. The addition of a kinematic alignment surgical technique for use with existing devices does not affect the materials, design features, packaging, or sterilization of the implant systems. Performance testing demonstrates substantial equivalence between the subject and predicate devices.
{6}
enovis
K252974 - Page 3 of 3
## Performance Testing
Risks were identified based on the proposed surgical technique and testing was conducted to mitigate those risks. Based on the risk analysis, the following test was conducted according to internal protocols.
- initial stability (micromotion) testing
- cadaveric assessment of the kinematic surgical technique
## Animal Studies
No animal data submitted.
## Clinical Studies
No clinical data submitted.
## VIII. CONCLUSIONS
Performance testing demonstrates that the devices implanted using the EMPOWR® DNA surgical technique are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.