ELEOS™ Limb Salvage System with NanoCept® Technology
K252920 · Onkos Surgical, Inc. · QZZ · Oct 15, 2025 · Orthopedic
Device Facts
Record ID
K252920
Device Name
ELEOS™ Limb Salvage System with NanoCept® Technology
Applicant
Onkos Surgical, Inc.
Product Code
QZZ · Orthopedic
Decision Date
Oct 15, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3900
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ELEOS™ and ELEOS™ Limb Salvage System with NanoCept® Technology is indicated for resection and replacement of the proximal femur, intercalary portion of the femur, total femur, distal femur, and proximal tibia in skeletally mature patients with the following conditions: 1) Non-inflammatory degenerative joint disease such as osteoarthritis, traumatic arthritis, avascular necrosis, ankylosis, protrusion acetabuli, and painful hip dysplasia; 2) Inflammatory degenerative joint disease such as rheumatoid arthritis; 3) Correction of functional deformity; 4) Revision procedures where other treatments or devices have failed; and, 5) Treatment of fractures that are unmanageable using other techniques. The ELEOS™ and ELEOS™ Limb Salvage System with NanoCept® Technology is also indicated for procedures where resection and replacement of the proximal femur, intercalary portion of the femur, total femur, distal femur, and proximal tibia is required with the following conditions: 1) Patients suffering from severe arthropathy of the hip and/or knee that does not respond to any conservative therapy or better alternative surgical treatment; 2) Surgical intervention for severe trauma, revision hip or knee arthroplasties, and/or Oncology indications; 3) Metastatic diseases The ELEOS™ Limb Salvage System with NanoCept® Technology MDPB coating, where applied, is intended to reduce bacterial contamination prior to implantation resulting from deposition in the operating room on the surface of the device components. The clinical impact associated with the MDPB coating, including prevention of infection or reduction of infection risk in patients, has not been evaluated in human clinical trials. The MDPB coating is not intended to treat existing infections and does not act within or on the body.
Device Story
Limb salvage system for joint reconstruction; consists of cobalt-chromium (CoCr) implants; features NanoCept® technology, a covalently bound 12-Methacryloyloxydodecyl Pyridinium Bromide (MDPB) quaternary ammonium compound coating. Coating applied to proximal tibia components; functions to reduce bacterial contamination on device surface prior to implantation by killing bacteria deposited from operating room environment. Not intended to act in or on patient body; does not treat existing infections. Used by orthopedic surgeons in surgical settings for limb salvage and joint replacement. Benefits patient by reducing pre-implantation bacterial load on implant surface.
Clinical Evidence
No clinical data provided. Performance supported by fretting and corrosion engineering analysis, coating integrity rationale, and biocompatibility risk assessment.
Technological Characteristics
Cobalt-chromium (CoCr) substrate; surface modified with 12-Methacryloyloxydodecyl Pyridinium Bromide (MDPB) quaternary ammonium compound coating. Mechanical limb salvage system; non-active. Sterilization and packaging consistent with predicate devices.
Indications for Use
Indicated for skeletally mature patients requiring resection and replacement of proximal femur, intercalary femur, total femur, distal femur, or proximal tibia due to non-inflammatory/inflammatory degenerative joint disease, functional deformity, failed prior procedures, unmanageable fractures, severe arthropathy, severe trauma, revision arthroplasty, oncology indications, or metastatic disease.
Regulatory Classification
Identification
A limb and joint salvage device with quaternary ammonium compound coating is a metallic implant with or without polymer bearing for bone and joint replacement. Implants are for resection and replacement of an extremity bone (including the entire bone, epiphyseal bone, metaphyseal bone, or diaphyseal bone), or an extremity bone and the surrounding joint(s) in a skeletally mature patient. The device includes a quaternary ammonium compound coating that is covalently bonded to the device, intended to reduce microbial contamination on the surface of the device prior to implantation.
Special Controls
In combination with the general controls of the FD&C Act. the limb and joint salvage device with quaternary ammonium compound coating is subject to the following special controls:
- (1) Animal performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Animal testing must include imaging, histology. and histomorphometry to assess bone formation, healing, and tissue response at relevant timepoints over the course of healing.
- (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and include the following:
- Evaluation of the static and dynamic performance of the implant: (i)
- Evaluation of coated implant initial fixation: (ii)
- Evaluation of coating integrity: (iii)
- Evaluation of range of motion; (iv)
- Evaluation of articulating bearing wear and bearing material; (v)
- Evaluation of implant disassembly: (vi)
- (vii) Evaluation of fretting and corrosion;
- (viii) Evaluation of antimicrobial performance with clinically relevant microbial species; and
- (ix) Coating characterization, including a detailed description of the substrate morphology and coating process and an evaluation of coating physicochemical properties such as density, thickness, chemistry, and uniformity.
- (3) An analysis must be provided that identifies and evaluates any contribution to the development and spread of antimicrobial resistance.
- (4) An analysis or information must be provided to support that the antimicrobial does not act within or on the body.
- (5) The patient-contacting components of the device must be demonstrated to be biocompatible.
- (6) Performance data must support the sterility and pyrogenicity of the device components intended to be provided sterile.
- Performance data must validate the reprocessing instructions for the reusable (7) instrumentation to be used with the device.
- Performance data must support the shelf life of the device by demonstrating continued (8) sterility, package integrity, and device functionality over the identified shelf life.
- (9) Labeling must include the following:
- Identification of device materials; (i)
- A shelf life: and (ii)
- (iii) Instructions for removal/revision procedures.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 15, 2025
Onkos Surgical, Inc.
Elizabeth Rose
Regulatory Affairs Contractor
77 East Halsey Road
Parsippany, New Jersey 07054
Re: K252920
Trade/Device Name: ELEOS™ Limb Salvage System with NanoCept® Technology
Regulation Number: 21 CFR 888.3900
Regulation Name: Limb And Joint Salvage Device With Quaternary Ammonium Compound Coating
Regulatory Class: Class II
Product Code: QZZ, KRO
Dated: September 12, 2025
Received: September 12, 2025
Dear Elizabeth Rose:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252920 - Elizabeth Rose
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252920 - Elizabeth Rose
Page 3
Sincerely,
RYAN TROMBETTA -S
For: Limin Sun, Ph.D.
Assistant Director
DHT6A: Division of Joint
Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252920 | |
| Device Name ELEOS™ Limb Salvage System with NanoCept® Technology | |
| Indications for Use (Describe) The ELEOS™ and ELEOS™ Limb Salvage System with NanoCept® Technology is indicated for resection and replacement of the proximal femur, intercalary portion of the femur, total femur, distal femur, and proximal tibia in skeletally mature patients with the following conditions: 1) Non-inflammatory degenerative joint disease such as osteoarthritis, traumatic arthritis, avascular necrosis, ankylosis, protrusion acetabuli, and painful hip dysplasia; 2) Inflammatory degenerative joint disease such as rheumatoid arthritis; 3) Correction of functional deformity; 4) Revision procedures where other treatments or devices have failed; and, 5) Treatment of fractures that are unmanageable using other techniques. The ELEOS™ and ELEOS™ Limb Salvage System with NanoCept® Technology is also indicated for procedures where resection and replacement of the proximal femur, intercalary portion of the femur, total femur, distal femur, and proximal tibia is required with the following conditions: 1) Patients suffering from severe arthropathy of the hip and/or knee that does not respond to any conservative therapy or better alternative surgical treatment; 2) Surgical intervention for severe trauma, revision hip or knee arthroplasties, and/or Oncology indications; 3) Metastatic diseases The ELEOS™ Limb Salvage System with NanoCept® Technology MDPB coating, where applied, is intended to reduce bacterial contamination prior to implantation resulting from deposition in the operating room on the surface of the device components. The clinical impact associated with the MDPB coating, including prevention of infection or reduction of infection risk in patients, has not been evaluated in human clinical trials. The MDPB coating is not intended to treat existing infections and does not act within or on the body. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K252920 Page 1 of 3
# 510(k) Summary
Date Prepared: October 15, 2025
Applicant: Onkos Surgical, Inc
77 East Halsey Road
Parsippany, NJ 07054
Establish Registration Number: 2184009
Contact Person: Arley Perez
Senior Director, Product Development
Phone: 239-216-0701
Email: APerez@onkossurgical.com
Elizabeth Rose
Regulatory Affairs
Phone: 423-252-9924
Email: erose@onkossurgical.com
Trade Name: ELEOS™ Limb Salvage System with NanoCept® Technology
Common Name: Limb and joint salvage device with quaternary ammonium compound coating
Classification Name: Limb And Joint Salvage Device with Coating for Bacteria Reduction
Classification: Class II (with special controls)
Regulation Number: 21 CFR 888.3900
Product Code: QZZ, KRO
Primary Predicate: ELEOSx™ Limb Salvage System
Submission Number: DEN210058
Additional Predicate: ELEOS™ Limb Salvage System
Submission Number: K161520
# Device Description
The Onkos Surgical ELEOS™ Limb Salvage System, a reconstruction implant system intended for joint replacement, limb salvage, and restoration of limb function. The subject technology consists of 510(k)-cleared limb salvage systems (i.e., ELEOS™ Limb Salvage System) and applies an antibacterial coating to certain cobalt chromium (CoCr) components.
{5}
K252920 Page 2 of 3
The ELEOS™ Limb Salvage System with NanoCept® Technology includes components that have been modified with an antibacterial coating, which includes a covalently bound Quaternary Ammonium Compound (QAC), specifically 12-Methacryloyloxydodecyl Pyridinium Bromide (MDPB). The purpose of the antibacterial coating is to reduce bacterial contamination on the surface of the device prior to implantation, by killing bacteria external to the patient which deposit onto the device surface from the operating environment. This coating is not intended to act in or on the body of the patient but rather reduces contamination on the surface of the device prior to implantation.
The purpose of this 510(k) submission is to introduce the ELEOS™ Limb Salvage System with NanoCept® Technology (Proximal Tibia component of the Proximal Tibia assembly) manufactured with cobalt-chromium (CoCr) substrate material cleared in the original predicate device K161520 with an added MDPB coating used in the predicate ELEOSx™ Limb Salvage System DEN210058. This modification aims to expand the product portfolio by adding a CoCr proximal tibia with MDPB coating.
## Indications for Use
The ELEOS™ and ELEOS™ Limb Salvage System with NanoCept® Technology is indicated for resection and replacement of the proximal femur, intercalary portion of the femur, total femur, distal femur, and proximal tibia in skeletally mature patients with the following conditions:
1) Non-inflammatory degenerative joint disease such as osteoarthritis, traumatic arthritis, avascular necrosis, ankylosis, protrusion acetabuli, and painful hip dysplasia;
2) Inflammatory degenerative joint disease such as rheumatoid arthritis;
3) Correction of functional deformity;
4) Revision procedures where other treatments or devices have failed; and,
5) Treatment of fractures that are unmanageable using other techniques.
The ELEOS™ and ELEOS™ Limb Salvage System with NanoCept® Technology is also indicated for procedures where resection and replacement of the proximal femur, intercalary portion of the femur, total femur, distal femur, and proximal tibia is required with the following conditions:
1) Patients suffering from severe arthropathy of the hip and/or knee that does not respond to any conservative therapy or better alternative surgical treatment;
2) Surgical intervention for severe trauma, revision hip or knee arthroplasties, and/or Oncology indications;
3) Metastatic diseases
The ELEOS™ Limb Salvage System with NanoCept® Technology MDPB coating, where applied, is intended to reduce bacterial contamination prior to implantation resulting from deposition in the operating room on the surface of the device components. The clinical impact associated with the MDPB coating, including prevention of infection or reduction of infection risk in patients, has not been evaluated in human clinical trials. The MDPB coating is not intended to treat existing infections and does not act within or on the body.
## Technological Comparison to Predicate
The subject device is substantially equivalent to the predicate device as the subject device shares the same intended use—limb salvage and joint reconstruction—and operating principle, design,
{6}
K252920 Page 3 of 3
materials, geometry, manufacturing, sterilization and packaging. The modification to apply the MDPB coating to the proximal tibia component of the proximal tibia assembly does not introduce new risks related to mechanical performance, coating characterization, or antibacterial efficacy.
## Performance Data
No performance testing was conducted; however, this submission includes supporting information that demonstrates the subject device's performance:
- Fretting and Corrosion engineering analysis
- Coating integrity (handling) rationale
- Biocompatibility risk assessment
## Conclusion
The information included in this submission demonstrates that the ELEOS Limb Salvage System with NanoCept® Technology is substantially equivalent to the legally marketed predicate devices, ELEOSx™ Limb Salvage System (DEN210058) and ELEOS™ Limb Salvage System (K161520).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.