Compass Steerable Needle with Adapters SRA-1-01 and SRA-2-01
K252874 · Serpex Medical, Inc. · KTI · Oct 8, 2025 · Ear, Nose, Throat
Device Facts
Record ID
K252874
Device Name
Compass Steerable Needle with Adapters SRA-1-01 and SRA-2-01
Applicant
Serpex Medical, Inc.
Product Code
KTI · Ear, Nose, Throat
Decision Date
Oct 8, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.4680
Device Class
Class 2
Indications for Use
The Compass Steerable Needle is a steerable transbronchial biopsy needle intended to be used through a compatible working channel bronchoscope or Medtronic Extended Working Channel (EWC) for the collection of tissue from the intrapulmonary regions.
Device Story
Compass Steerable Needle (CSN) is a sterile, single-use, 22-gauge transbronchial biopsy needle featuring a unidirectional, steerable distal tip. Device facilitates tissue acquisition from intrapulmonary regions. Subject device adds two adapters (SRA-1-01, SRA-2-01) to existing Model CSN1002 to enable compatibility with robotic bronchoscopy platforms (Ion Endoluminal System, Monarch Platform, Galaxy System). Device is used by clinicians during bronchoscopic procedures. Needle is inserted through bronchoscope working channel; clinician steers distal tip to target site; tissue is collected via aspiration. Output is tissue sample for diagnostic evaluation. Benefits include improved access to intrapulmonary regions for biopsy.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included sterilization, packaging/labeling, dimensional analysis, functional pre- and post-simulated use, and tensile testing. All design verification tests met pre-defined acceptance criteria.
Technological Characteristics
22-gauge transbronchial needle; unidirectional steerable distal tip; sterile, single-use. Adapters SRA-1-01 and SRA-2-01 facilitate connection to robotic bronchoscopes. Sterilization via Ethylene Oxide (EtO) per ANSI AAMI ISO 11135:2014. Packaging/testing standards include ISO 11607-2, ASTM D4169-23, ASTM D4332-22, ASTM F1886, ASTM F1980-21, ASTM F2096-11, and ASTM F88/F88M-23. Biocompatibility per ISO 10993-1 and ISO 10993-7.
Indications for Use
Indicated for patients requiring transbronchial biopsy for tissue collection from intrapulmonary regions using a compatible working channel bronchoscope or Medtronic Extended Working Channel.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 8, 2025
Serpex Medical, Inc.
Diane Horwitz
Regulatory Consultant
3350 Scott Blvd.
Suite 37B
Santa Clara, California 95054
Re: K252874
Trade/Device Name: Compass Steerable Needle with Adapters SRA-1-01 and SRA-2-01
Regulation Number: 21 CFR 874.4680
Regulation Name: Bronchoscope (flexible or rigid) and accessories
Regulatory Class: Class II
Product Code: KTI
Dated: September 9, 2025
Received: September 10, 2025
Dear Diane Horwitz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
2K252874 - Diane Horwitz
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
3K252874 - Diane Horwitz
Page 3
Sincerely,
Ethan L. Nyberg -S
Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252874 | ? |
| Please provide the device trade name(s). | | ? |
| Compass Steerable Needle with Adapters SRA-1-01 and SRA-2-01 | | |
| Please provide your Indications for Use below. | | ? |
| The Compass Steerable Needle is a steerable transbronchial biopsy needle intended to be used through a compatible working channel bronchoscope or Medtronic Extended Working Channel (EWC) for the collection of tissue from the intrapulmonary regions. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
serpex
# 510(k) SUMMARY
## 1. SUBMITTER INFORMATION
Submitter:
Serpex Medical, Inc.
Neil Zimmerman
3350 Scott Blvd, Suite 37B
Santa Clara, CA 95054
Email: nzimmerman@ajaxhealth.com
Official Correspondent:
Diane Horwitz, Ph.D.
Consultant
3350 Scott Blvd, Suite 37B
Santa Clara, CA 95054
Telephone: 703.307.2921
Email: dmh@mandellhorwitzconsulting.com
Date Prepared: September 9, 2025
## 2. DEVICE INFORMATION
Proprietary Name: Compass Steerable Needle with Adapters SRA-1-01 and SRA-2-01
Common/Usual Name: Aspiration Needle
Classification Name: Bronchoscope Accessory
Regulatory Class: Class II
Product Code: KTI
Regulation Number: 21 CFR 874.4680
## 3. Predicate Device Information
Proprietary Name: Compass Steerable Needle
Common/Usual Name: Aspiration Needle
Classification Name: Bronchoscope Accessory
510K Number: K221206
## 4. Device Description
The Compass Steerable Needle (CSN) is sterile, single use, 22-gauge transbronchial needle with a unidirectional, steerable distal tip for the acquisition of tissue from the intrapulmonary regions. The Compass Steerable Needle include two model numbers, Model CSN1001 and Model CSN1002. Model CSN1001 includes two accessory adapters for attachment to bronchoscopes.
The subject of this 510(k) is the addition of two new adapters to Model CSN1002. Adapter SRA-1-01 connects the Model CSN1002 to the Ion™ Endoluminal System, and Adapter SRA-2-01 connects the Model CSN1002 to the Monarch™ Platform or Galaxy System™.
Serpex Medical, Inc. Compass Steerable Needle with Adapters SRA-1-01 and SRA-2-01
{5}
serpex
# 5. Intended Use/Indication for use
The Compass Steerable Needle is a steerable transbronchial biopsy needle intended be used through a compatible working channel bronchoscope or Medtronic Extended Working Channel (EWC) for the collection of tissue from the intrapulmonary regions.
# 6. Comparison of Technological Characteristics
The Compass Steerable Needle with Adapters SRA-1-01 and SRA-2-01 is substantially equivalent to the Compass Steerable Needle cleared under K221206. The adapters to the Model CSN1002 (SRA-1-01 and SRA-2-01) connect the CSN1002 to robotic bronchoscopes to obtain tissue from intrapulmonary regions. The Compass Steerable Needle device is unchanged.
| | Subject Device Compass Steerable Needle with Adapters SRA-1-01 and SRA-2-01 | Predicate Device Compass Steerable Needle K221206 | Same or Different |
| --- | --- | --- | --- |
| Intended Use | To collect tissue from the intrapulmonary regions. | To collect tissue from the intrapulmonary regions. | Same |
| Indications for use | The Compass Steerable Needle is a steerable transbronchial biopsy needle intended be used through a compatible working channel bronchoscope or Medtronic Extended Working Channel (EWC) for the collection of tissue from the intrapulmonary regions. | The Compass Steerable Needle is a steerable transbronchial biopsy needle intended be used through a compatible working channel bronchoscope or Medtronic Extended Working Channel (EWC) for the collection of tissue from the intrapulmonary regions. | Same |
| Connecting Devices | Unchanged previous connecting device under K221206 CSN1002 Added two new adapters to connect to the following bronchoscopes: • Ion™ Endoluminal System bronchoscope • Monarch™ Platform bronchoscope • Galaxy System™ bronchoscope | CSN1001 Olympus 190 or Pentax bronchoscope 600mm length 2.0mm working channel CSN1002 Medtronic Illumisite EWC 1070mm length, 2.0mm working channel | Similar. Though the connecting bronchoscopes are different, the Compass Steerable Needle serves the same functional role - to obtain tissue samples using the same method. Connections have been demonstrated to meet specifications in bench testing. |
| Sterilization | EtO Adapters in sterile single-use pouches; shelf box contains 10 pouches | EtO Compass Steerable Needle and accessory adapters in sterile single use shelf tray. | Same; both have single use sterile package |
Serpex Medical, Inc. Compass Steerable Needle with Adapters SRA-1-01 and SRA-2-01
{6}
serpex
7. SUMMARY OF NON-CLINICAL TESTING
Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971: 2019. The design verifications tests and their acceptance criteria were identified and performed with passing results.
Bench testing was performed with the Compass Steerable Needles with Adapters SRA-1-01 and SRA-2-01 using sterile finished devices at t=0 and aged time points. Testing supported that all product requirements and specifications were met and demonstrated substantial equivalence. Testing included:
- Sterilization
- Packaging and labeling
- Dimensional
- Functional pre- and post-simulated use
- Tensile
The following standards have been applied to the Compass Steerable Needle Adapters:
- ISO 14971 Third Edition 2019-12 Medical devices - Application of risk management to medical devices
- ISO 11607-2 Packing for Terminally Sterilized Medical Devices
- ASTM D4169-23 Standard Practice for Performance Testing of Shipping Containers and Systems
- ASTM D4332-22 Standard Practice for Conditioning Containers, Packages, or Packaging Components for Testing
- ANSI AAMI ISO 11135:2014 Sterilization of health-care products — Ethylene oxide — Requirements for development, validation, and routine control of a sterilization process for medical devices
- ANSI/AAMI ST72: 2019 Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing
- USP-NF <151> Pyrogen Test
- ISO 10993-7 Biological evaluation of medical devices – Part 7: Ethylene oxide sterilization residuals
- ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- ASTM F1886/ FM1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- ASTM F1980-21 Accelerated aging
- ASTM F2096-11 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
- ASTM F88/F88M-23 Standard Test Method for Seal Strength of Flexible Barrier Materials
Testing demonstrated that all endpoints were met.
8. CONCLUSION
The Compass Steerable Needle with Adapters SRA-1-01 and SRA-2-01 are substantially equivalent to the Compass Steerable Needle (K221206) in intended use, indications for use, technological characteristics, and performance testing. Substantial equivalence has been demonstrated.
Serpex Medical, Inc. Compass Steerable Needle with Adapters SRA-1-01 and SRA-2-01
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.