K221206 · Serpex Medical, Inc. · KTI · Aug 25, 2022 · Ear, Nose, Throat
Device Facts
Record ID
K221206
Device Name
Compass Steerable Needle
Applicant
Serpex Medical, Inc.
Product Code
KTI · Ear, Nose, Throat
Decision Date
Aug 25, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Indications for Use
The Compass Steerable Needle is a steerable biopsy needle intended be used through a compatible working channel bronchoscope or Medtronic Extended Working Channel (EWC) for the collection of tissue from the intrapulmonary regions.
Device Story
Compass Steerable Needle (CSN) is a sterile, single-use, 22-gauge biopsy needle for intrapulmonary tissue collection. Device comprises handle, shaft, and needle; handle provides control of rotation, extension, retraction, and distal tip articulation (70°±10° unidirectionally). Clinician inserts device through compatible bronchoscope or Medtronic Illumisite EWC working channel; advances to target using translation arm; articulates distal shaft via plunger; extends needle to obtain sample. Luer connector allows stylet or syringe aspiration. Used in clinical bronchoscopy suites. Provides tissue samples for diagnostic evaluation; enables access to difficult-to-reach lung regions via steerable tip. Benefits include improved biopsy targeting and diagnostic yield.
Clinical Evidence
No clinical trials performed. Evidence consists of bench testing (dimensional, mechanical, durability, torque, corrosion) and simulated-use validation in human cadavers. Human factors usability studies (formative and limited summative) confirmed device safety and effectiveness for intended users in simulated bronchoscopy environments. All performance endpoints met.
Technological Characteristics
22G needle with Franseen tip; materials include Pebax, Vestamid, stainless steel, PET, and Nitinol. Distal tip articulates 70°±10° unidirectionally. EO sterilized. Single-use. Compatible with 2.0mm working channels. Complies with ISO 10993-1 (biocompatibility), ISO 11607-1 (packaging), and ISO 11135 (sterilization).
Indications for Use
Indicated for clinicians trained in bronchoscopy to collect tissue from intrapulmonary regions in patients requiring biopsy via flexible bronchoscope or Medtronic EWC.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's full name on the right. The symbol on the left is a stylized representation of a human figure, while the right side features the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 25, 2022
Serpex Medical, Inc. % Laurie Lewandowski Vice President Honkanen Consulting, Inc. 738 Saddle Wood Drive Eagan, Minnesota 55123
Re: K221206
Trade/Device Name: Compass Steerable Needle Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: KTI Dated: July 22, 2022 Received: July 25, 2022
Dear Laurie Lewandowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Brandon Blakely, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221206
Device Name Compass Steerable Needle
Indications for Use (Describe)
The Compass Steerable Needle is a steerable biopsy needle intended be used through a compatible working channel bronchoscope or Medtronic Extended Working Channel (EWC) for the collection of tissue from the intrapulmonary regions.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" width="12" x="0" y="0"></rect> <path d="M2,2 L10,10 M2,10 L10,2" stroke="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" width="12" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|| | Prescription Use (Part 21 CFR 801 Subpart D)
|_ | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
#### 1. SUBMITTER INFORMATION
Submitter: Serpex Medical, Inc. Sasha Schrode 3350 Scott Blvd, Suite 37B Santa Clara, CA 95054 sschrode(@serpexmedical.com Email:
Primary Contact: Laurie Lewandowski Consultant, Honkanen Consulting, Inc. 738 Saddle Wood Drive Eagan, MN 55123 Telephone: 612-770-4038 (cell) Email: llewandowski(@serpexmedical.com
DATE PREPARED:
August 25, 2022
#### 2. DEVICE INFORMATION
| Proprietary Name: |
|----------------------|
| Common/Usual Name: |
| Classification Name: |
| Regulatory Class: |
| Product Code: |
| Regulation Number: |
Compass Steerable Needle Aspiration Needle Bronchoscope Accessory Class II KTI 21 CFR 874.4680
#### 3. PREDICATE AND REFERENCE DEVICE INFORMATION
#### PRIMARY PREDICATE DEVICE:
| Proprietary Name: | PeriView Flex |
|----------------------|------------------------|
| Common/Usual Name: | Aspiration Needle |
| Classification Name: | Bronchoscope Accessory |
| 510K Number: | K171232 |
#### SECONDARY PREDICATE DEVICE:
Proprietary Name: Common/Usual Name: Classification Name: 510K Number:
Ion™ Endoluminal System; Flexision™ Biopsy Needle Aspiration Needle Bronchoscope (Flexible or Rigid) and Accessories K182188
{4}------------------------------------------------
### REFERENCE DEVICE:
Proprietary Name: Common/Usual Name: Classification Name: 510K Number:
Morrison Steerable Needle Biopsy Needle Instrument, Biopsy K151396
## 4. DEVICE DESCRIPTION
The Compass Steerable Needles (CSN) are sterile, single use, 22-gauge needles with a unidirectional, steerable distal tip for the acquisition of tissue from the intrapulmonary regions.
The Steerable Needle consists of a handle, shaft, and needle. The handle provides the user with control of device rotation, extension, retraction, distal tip articulation of 70°±10° unidirectionally within a plane and a sampling mechanism to extend and retract the needle out of the shaft to obtain tissue samples. A Luer connector on the proximal end of the device provides the connection for the stylet or a syringe for aspiration during sampling. There are two models of the Compass Steerable Needle.
Model CSN1001 can be coupled to Olympus® 190 or Pentax® bronchoscopes with a 2.0 working channel and 600 mm working length. It is packaged with a stylet, and adapters.
Model CSN1002 can be coupled to the Medtronic Illumisite™ Extended Working Channel (EWC) with a 2.0 mm working channel. It is packaged with a stylet.
The Compass Steerable Needles with stylet are inserted and coupled to either a bronchoscope or a Medtronic Illumisite EWC. The translation arm advances the device into the lung. Depressing the plunger articulates the distal end of the shaft. The sampling mechanism is depressed extending the needle to obtain a sample.
## 5. INTENDED USE/INDICATION FOR USE
The Compass Steerable Needle is a steerable biopsy needle intended be used through a compatible working channel bronchoscope or Medtronic Extended Working Channel (EWC) for the collection of tissue from the intrapulmonary regions.
## 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The Compass Steerable Needles have the same basic technological characteristics as the primary predicate, PeriView Flex, cleared under K171232 with a Franseen shaped needle and the ability to articulate and couple to bronchoscopes. The secondary predicate Ion™ Endoluminal System; Flexision™ Biopsy Needle (K182188) and the reference device, the Morrison Biopsy Needle (K151396) were used to support articulation and needle size. The subject and predicate devices all operate in the same manner to obtain tissue during sampling in the lungs.
{5}------------------------------------------------
A comparison of the Compass Steerable Needles and the PeriView Flex are shown in the following table with a discussion following.
| Attribute | Proposed Device<br>Compass Steerable Needle | PeriView Flex<br>K171232<br>Primary Predicate | Difference and<br>Impact on<br>Substantial<br>Equivalence |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Intended Use | To collect tissue from the<br>intrapulmonary regions. | Identical | SE |
| Indications<br>for use | The Compass Steerable Needle is a<br>steerable biopsy needle intended be<br>used through a compatible working<br>channel bronchoscope or Medtronic<br>Extended Working Channel (EWC)<br>for the collection of tissue from the<br>intrapulmonary regions. | This device is intended to<br>be used through a<br>compatible bronchoscope<br>for the collection of tissue<br>from the intrapulmonary<br>regions. Do not use for any<br>purpose other than its<br>intended use. | SE<br>Similar to primary<br>predicate with minor<br>wording differences |
| | Users | Clinicians trained in bronchoscopy<br>devices, accessories, and procedures | Identical |
| Anatomic Site | Lung | Identical | SE |
| Method of<br>Introduction | Endobronchial delivered to the target<br>through a flexible endoscope or<br>EWC. | Endobronchial delivered to<br>the target through a<br>flexible endoscope or<br>EWC. | SE |
| Required<br>Working<br>Channel | 2mm | 2mm | SE |
| General<br>Design | Handle,<br>Shaft,<br>Needle,<br>Stylet,<br>Adapters | Handle,<br>Sheath,<br>Needle,<br>Stylet | SE<br>Adapters to couple the<br>device. |
| Needle Size | 22G | 21G | SE |
| Needle Tip | Franseen | Chiba (angled beveled) | See discussion |
| Needle<br>Protrusion<br>Length | 15 mm | 20 mm | See discussion |
| Coupling | Coupled to bronchoscope | N/A | See discussion |
| Mode of<br>Sampling<br>Action | Single / multiple puncture and<br>aspirate manually controlled | Single / multiple puncture<br>and aspirate<br>manually controlled | SE |
| Angle of<br>Articulation | Shaft articulates 70° in a single<br>direction | N/A | See discussion |
| Dimensions | Shaft OD ≤ 1.93mm<br>Working Length<br>CSN1001 CSN1002<br>715- 1035-<br>800mm 1085mm | Shaft OD = 1.5mm<br>Total Length = 115cm | SE<br>All work within a 2.0<br>bronchoscope or<br>working channel |
| Stylet OD | 0.43mm | 0.47 mm | SE |
| Attribute | Proposed Device<br>Compass Steerable Needle | PeriView Flex<br>K171232<br>Primary Predicate | Difference and<br>Impact on<br>Substantial<br>Equivalence |
| | | | Both stylets fit inside |
| | | | needle |
| Radiopaque | Yes | Yes | SE |
| Life/Sterility | Single Use, EO sterilized | Identical | SE |
| Patient<br>Contacting<br>Materials | Pebax<br>Vestamid<br>Stainless steel<br>PET Heat Shrink<br>Nitinol | Stainless Steel,<br>PTFE,<br>Pebax,<br>Nitinol | SE<br>All devices comprised<br>of standard medical<br>device materials |
| Biocompatible | Yes, conforms to 10993-1 | Yes, conforms to 10993-1 | SE |
{6}------------------------------------------------
Discussion: The Compass Steerable Needles are the same as the primary predicate in terms with the exception of the needle gauge, needle tip, protrusion, articulation and coupling to a bronchoscope.
The needle size is substantially equivalent to the secondary predicate, the Flexision with needle sizes of 23G, 21G and 19G and reference device, the Morrison needle at 19G.
Franseen needle tips have been used for lung biopsy devices for > 20 years as supported by devices K770523 Needle Biopsy, Boyd Lung, and the Merit Medical Temno Elite Biopsy System (K201166).
The protrusion length is of the Compass Steerable Needle is within the same range as the secondary predicate, Flexision, of 1-3 cm.
The Compass Steerable Needle articulates the distal shaft 70° ± 10° in a single direction. The secondary predicate articulates the catheter 180° in all directions using electromechanical controls while the reference device articulates in two directions using manual controls.
Coupling of devices to a bronchoscope has previously been cleared in ultrasound bronchoscopes as supported by Olympus ViziShot 2 EBUS-TBNA (K193527).
The different technological characteristics of the new device do not raise different questions of safety and effectiveness.
{7}------------------------------------------------
## 7. PERFORMANCE DATA
Bench Testing - pre and post aging
| Test | Method | Results |
|------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Visual | Inspect with the naked eye or under a X scope | Devices met acceptance criteria |
| Dimensional | Measure: OD, ID, Length, Needle Throw,<br>Articulation Length, Stylet Protrusion Length,<br>Stylet OD, Travel, Retraction | Devices met acceptance criteria |
| Force - Insertion, plunger<br>and sampling mechanism | Measure insertion force into the scope; force to<br>depress plunger/sampling mechanism with<br>Instron | Devices met acceptance criteria |
| Articulation | Measure angle, length, curve profile, stability<br>and planarity | Devices met acceptance criteria |
| Tensile | Tensile test all joints and coupling | Devices met acceptance criteria |
| Adapter<br>Vacuum/pneumostasis | Ability to hold a vacuum of 20 Hg mercury | Devices met acceptance criteria |
| Needle Leak | Flush 5ml saline, inspect for leak | Devices met acceptance criteria |
| Durability | Condition device simulating use during a<br>procedure, visual, articulation and functional | Devices met acceptance criteria |
| Bending/Buckling | Measure bending and buckling force | Devices met acceptance criteria |
| Torque Transmission /<br>Integrity | Measure peak torque at 90° rotation / no joint<br>separation after 720° rotation | Devices met acceptance criteria |
| Lumen patency needle | Ensure stylet can be inserted | Devices met acceptance criteria |
| Corrosion resistance | Test per ISO 10555-1:2013 | Devices met acceptance criteria |
| Packaging | Distribution per ASTM D4169:2016<br>Aging per ASTM F1980:2016,<br>Packaging per ISO 11607-1:2019<br>Visual per ASTM F1886/F1886M-16<br>Bubble Leak per ASTM F2096-11:2019 | Packages met accept criteria |
The test results demonstrated that differences in device characteristics between the subject device and predicate device do not raise any new questions of safety or effectiveness.
### Validation testing
Validation testing was performed demonstrating the Compass Steerable Needles user needs and intended use were met. Testing under simulated use conditions was performed in accordance with the Instructions for Use. Human cadavers were used to demonstrate
Serpex Medical, Inc. Compass Steerable Needle
{8}------------------------------------------------
clinical performance within the human lung anatomy. Cadaver testing demonstrated the Compass Steerable Needles Instructions for Use are understandable, and the device meets the intended use no new risks identified.
### Human Factors Usability
Formative studies were conducted to identify and minimize use errors related to the use of the Compass Steerable Needles. Studies were conducted by intended user groups in a simulated bronchoscopy suite and involved preoperative preparation and simulated procedures. Since the user interface of the device is similar to the predicate device and the formative studies did not identify any new critical tasks unique to the device, a limited Summative Validation with study was conducted. The Instructions for Use and Reference Guide were assessed in the usability study. The Compass Steerable Needles has been assessed and found to be safe and effective for its intended use, by the intended users, in its intended use environment.
### Biocompatibility
Biocompatibility testing was performed in accordance with:
- . ISO 10993-1:2018 for a limited (<24 hour) tissue contacting device and
- FDA Guidance: "Use of International Standard ISO-10993, "Biological Evaluation of . Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process," September 2020
Testing demonstrated that all endpoints were met.
### Sterilization:
Sterilization (ethylene oxide) and packaging of the Compass Steerable Needles were validated using the following standards:
- ANSI/AAMI/ISO 11607-1:2019 Packaging for terminally sterilized medical . devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
- . ANSI/AAMI/ISO 11135:2014 Sterilization of health-care products ethylene oxide – Requirements for the development, validation and routine control of a sterilization process for medical devices
- FDA Guidance: Submission and Review of Sterility Information in Pre-Market Notification 510(k) Submissions for Device Labeled as Sterile (January 21, 2016) Testing demonstrated that all endpoints were met.
## 8. CONCLUSION
The Compass Steerable Needles are substantially equivalent in terms of the indications for use, technological characteristic, performance testing and comparison to the cited predicates, and do not present any new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.