K252870 · Align Technology, Inc. · NXC · Oct 9, 2025 · Dental
Device Facts
Record ID
K252870
Device Name
Invisalign Specifix Attachment System
Applicant
Align Technology, Inc.
Product Code
NXC · Dental
Decision Date
Oct 9, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Invisalign Specifix Attachment System is a directly fabricated accessory indicated to bond attachments and/or engagement features to apply force for tooth movement during orthodontic treatment of malocclusion, and/or for retention of Align orthodontic appliances.
Device Story
Invisalign Specifix Attachment System is a 3D-printed, doctor-prescribed accessory used in orthodontic clinics. It functions as a template to bond attachments and engagement features onto a patient's teeth. The device is custom-designed based on the patient's unique anatomical tooth surface and the practitioner's prescription. Once seated on the dentition, it facilitates the precise placement and bonding of attachments or engagement features (which interact with orthodontic elastics and ligatures) to assist in tooth movement or appliance retention. The template is removed after the bonding process is complete. By enabling the attachment of these features, the device supports the overall orthodontic treatment plan, allowing for the application of forces necessary for tooth movement and the secure retention of Align orthodontic appliances.
Clinical Evidence
No clinical data. Bench testing only. Evaluations included biocompatibility, flexural strength, wear resistance, water sorption/solubility, shear bond strength, force system analysis, packaging validation, stability, volume precision, and fatigue testing. All tests met pre-defined acceptance criteria.
Technological Characteristics
3D-printed orthodontic accessory. Materials differ from predicate. Functions as a bonding template for attachments and engagement features. Designed via patient-specific anatomical data. Sterilization/biocompatibility per FDA additive manufacturing guidance. Connectivity: N/A (custom-manufactured device).
Indications for Use
Indicated for patients undergoing orthodontic treatment of malocclusion requiring bonded attachments or engagement features for tooth movement or retention of Align orthodontic appliances.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
Predicate Devices
Invisalign System, Pre-formed Attachment System (K222894)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 9, 2025
Align Technology, Inc.
Niya Su
Sr. Manager, Regulatory Affairs
2820 Orchard Parkway
San Jose, CA 95134
Re: K252870
Trade/Device Name: Invisalign Specifix Attachment System
Regulation Number: 21 CFR 872.5470
Regulation Name: Orthodontic Plastic Bracket
Regulatory Class: Class II
Product Code: NXC, DYT
Dated: September 9, 2025
Received: September 9, 2025
Dear Niya Su:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252870 - Niya Su
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252870 - Niya Su
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252870 | ? |
| Please provide the device trade name(s). | | ? |
| Invisalign Specifix Attachment System | | |
| Please provide your Indications for Use below. | | ? |
| Invisalign Specifix Attachment System is a directly fabricated accessory indicated to bond attachments and/or engagement features to apply force for tooth movement during orthodontic treatment of malocclusion, and/or for retention of Align orthodontic appliances. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
| 510(k) #: K252870 | 510(k) Summary | Prepared on: 2025-10-07 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Align Technology, Inc. | |
| Applicant Address | 2820 Orchard Parkway San Jose CA 95134 United States | |
| Applicant Contact Telephone | 206-779-7758 | |
| Applicant Contact | Ms. Niya Su | |
| Applicant Contact Email | n.su@aligntech.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Invisalign Specifix Attachment System | |
| Common Name | Orthodontic plastic bracket | |
| Classification Name | Aligner, Sequential | |
| Regulation Number | 872.5470 | |
| Product Code(s) | NXC, DYT | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K222894 | Invisalign System, Pre-formed Attachment System | NXC |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| Align Technology, Inc. is submitting this Special 510(k) to modify Align's Invisalign Pre-Formed Attachment System (PFA) cleared in 510(k) K222894. The brand name of subject (modified) device is Invisalign Specifix Attachment System. Invisalign Specifix Attachment System is a doctor prescribed 3D-printed device intended to bond 3D-printed attachments and/or engagement features to the patient's teeth. Invisalign Specifix attachments apply force for tooth movement along with Invisalign System aligner during the orthodontic treatment of malocclusion, and/or supports retention of Align orthodontic appliances. The engagement features are intended to engage with commercially available orthodontic elastics and ligatures. Invisalign Specifix Attachment System is designed based on the doctor/dental practitioner's prescription and patient's anatomical tooth surface. The dentition-matched Invisalign Specifix Attachment System allows dental practitioner to seat the device on the patient's unique dentition, and facilitates placement of attachments and/or engagement features during bonding and are removable post bonding. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| Invisalign Specifix Attachment System is a directly fabricated accessory indicated to bond attachments and/or engagement features to apply force for tooth movement during orthodontic treatment of malocclusion, and/or for retention of Align orthodontic appliances. | | |
| Indications for Use Comparison | | 21 CFR 807.92(a)(5) |
| The indications for use of the subject device is different than the predicate Invisalign Pre-Formed Attachment System with the addition of engagement features and retention of additional Align orthodontic appliance. The addition of engagement feature does not constitute a new intended use of the device, which is to apply force for tooth movement during treatment of malocclusion. The predicate device provides retention of Invisalign System aligner. The subject device provides retention of additional Align | | |
{5}
orthodontic appliance. This change does not affect the intended use of the device which provides retention of orthodontic appliance.
## Technological Comparison
21 CFR 807.92(a)(6)
The subject device and predicate device have the following similarities:
- Same Intended Use and similar Indications for Use;
- Same principles of operation;
- Similar technological characteristics and functional characteristics.
The subject device and predicate device have the following differences:
- Subject device has different material than the predicate device;
- Subject device has engagement features to engage with orthodontic accessories, such as elastics and ligatures;
- Subject device can provide retention for Align orthodontic appliances, whereas predicate device provides retention for Invisalign System aligner only;
- Subject device has modified packaging to enhance package integrity during transportation. The number of devices in the package is not changed.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
The subject device, Invisalign Specifix Attachment System, was evaluated through a complete set of biocompatibility and bench testing including flexural strength, wear by toothbrushing, water sorption and solubility, shear bond strength, force system, packaging validation, stability, volume precision, button retention and fatigue test, button shear bond force, and retention force test with additional Align orthodontic appliance. The testing performed was in alignment with the recommendations in the FDA guidances titled "Technical Considerations for Additive Manufactured Medical Devices" and "Denture Base Resins - Performance Criteria for Safety and Performance Based Pathway." All testing passed acceptance criteria and demonstrated that the subject/modified device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.