K222894 · Align Technology, Inc. · NXC · Dec 22, 2022 · Dental
Device Facts
Record ID
K222894
Device Name
Invisalign System, Pre-Formed Attachment System
Applicant
Align Technology, Inc.
Product Code
NXC · Dental
Decision Date
Dec 22, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Invisalign System is intended for the orthodontic treatment of malocclusion.
Device Story
Invisalign System uses removable clear plastic aligners to move teeth incrementally. Subject device adds Pre-Formed Attachment (PFA) System as an optional alternative to traditional composite-filled templates. Inputs: dental scans or PVS impressions. Processing: proprietary ClinCheck 3D software creates treatment plans; CAD designers and doctors review/approve plans; PFA manufactured via additive manufacturing. Output: customized aligners and pre-formed attachments. Used in dental clinics by orthodontists/dentists. PFA resin bonds to tooth surface to assist aligner retention and optimize force systems. Benefits: provides alternative attachment delivery mechanism, improving clinical workflow for attachment placement.
Clinical Evidence
No clinical or animal testing was performed. Substantial equivalence is supported by bench testing, including bond strength, stain resistance, wear, force measurement, volume precision, and usability testing. Biocompatibility was verified per ISO 10993-1 and ISO 7405.
Technological Characteristics
Aligners/templates: thermoplastic polymer. PFA system: photo-polymerizable methacrylate-based resin. Additive manufacturing used for PFA. Software: 3D treatment planning (ClinCheck). Connectivity: digital scan/impression input. Sterilization: not specified (device is patient-specific).
Indications for Use
Indicated for orthodontic treatment of malocclusion in children, adolescents, and adults. Treats dental and skeletal malocclusion including Class I, II, and III, A-P, vertical (open/deep bite), transverse (narrow arch/crossbite), and inter-arch (spacing/crowding) at all severity levels. Mandibular advancement feature indicated for Class II malocclusions in mixed and permanent dentition.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the agency's name, "U.S. Food & Drug Administration."
December 22, 2022
Align Technology, Inc. John Ray Principal, Regulatory Affairs 2820 Orchard Parkway San Jose. California 95134
Re: K222894
Trade/Device Name: Invisalign System, Pre-Formed Attachment System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: September 23, 2022 Received: September 23, 2022
Dear John Ray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael E. Adjodha -S
Michael E. Adjodha, M. ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222894
Device Name
Invisalign System with Pre-Formed Attachment System
Indications for Use (Describe) The Invisalign System is intended for the orthodontic treatment of malocclusion.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 7: 510(K) Summary Align Technology's Invisalign System with Pre-Formed Attachment System
#### General Information
| 510(k) Sponsor | Align Technology, Inc. |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address | 2820 Orchard Parkway<br>San Jose, CA 95134 |
| FDA Registration Number | 2953749 |
| Contact Person | Manjunath Bisalehalli<br>Manager, Regulatory Affairs<br>Align Technology, Inc.<br><br>cc: Shweta Daga<br>Director, Regulatory Affairs<br>Align Technology Inc. |
| Contact Information | Email: mbisalehalli@aligntech.com<br>Phone: +1 408-789-1743<br>Cell Phone: +1 412-298-4604<br>Fax: +1408-470-1010 |
| Date Prepared | September 23, 2022 |
#### Name of Modified Device and Name/Address of Sponsor
| Name of Device | Invisalign System with Pre-Formed Attachment System |
|-------------------------|--------------------------------------------------------------------|
| Name/Address of Sponsor | Align Technology, Inc.<br>2820 Orchard Parkway, San Jose, CA 95134 |
| Trade/Proprietary Name | Invisalign System<br>Pre-Formed Attachment System |
| Common Name | Aligner, Sequential |
| Classification Name | Orthodontic Plastic Bracket |
| Regulation Number | 21 CFR 872.5470 |
| Product Code | NXC |
| Regulatory Class | II |
#### Predicate Device
| Name of Device | Invisalign System |
|-------------------------|--------------------------------------------------------------------|
| Name/Address of Sponsor | Align Technology, Inc.<br>2820 Orchard Parkway, San Jose, CA 95134 |
| Trade/Proprietary Name | Invisalign System |
| Common Name | Aligner, Sequential |
| Classification Name | Orthodontic Plastic Bracket |
| Regulation Number | 21 CFR 872.5470 |
| Product Code | NXC |
| Regulatory Class | II |
{4}------------------------------------------------
## Purpose of the Traditional 510(k) notice
The purpose of this Traditional 510(k) notice is to request clearance for the addition of Pre-Formed Attachment (PFA) System to the Invisalign System's existing components. The PFA system is an optional alternative attachments bonding mechanism for Invisalign treatment.
Additionally, the Traditional 510(k) introduces the proprietary, 3D device-generation manufacturing software, labeling documentation associated with the PFA system, and the nonsignificant changes to the Invisalign System implementation since the last 510(k) clearance.
#### Intended Use
The Invisalign System with the Pre-Formed Attachment System (Modified device) has the same intended use as the cleared Predicate device, the Invisalign System (K220287), which is as follows:
The Invisalign System is indicated for the orthodontic treatment of malocclusion.
# Device Description
The subject device, Invisalign System with Pre-Formed Attachments (PFA) System, is a modification to the predicate device, the Invisalign System (K220287, cleared on April 7th, 2022). The Invisalign System with PFA system consists of removable orthodontic appliances (aligners), proprietary treatment planning 3D software, attachment templates and/or PFA system.
Like the predicate device, the Invisalign system with PFA (subject device) consists of a series of doctor prescribed, customized, thin, clear plastic aligners that gently move the patient's teeth in small increments from their original state to a more optimal, treated state to address dental malocclusion. The Invisalian system aligners are used in patients with primary, mixed (primary and permanent), and permanent dentition. The Invisalign system aligners with mandibular advancement feature(s) are used in patients with mixed and permanent dentition to correct Class II malocclusions. The Invisalign system is intended to treat dental and skeletal malocclusion in patients with Class I, II and III including Antero-Posterior (A-P), vertical (open bite, deep bite), transverse (narrow arch or jaw, crossbite), or inter-arch (spacing and crowding) correction at all severity levels. In some cases, surgical intervention, or addition of other appliances in addition to Invisalign treatment may be needed at the doctor's discretion.
The proprietary ClinCheck (CC) 3-D software is inclusive of both dental practitioner-facing functions and internal Align personnel-facing functions. The Align facing software enables Align's computer-aided design (CAD) designers to create a treatment plan based on a doctor's prescription and inputs (e.g., dental scans). The treatment plans created using the Align personnel-facing software are then reviewed and approved by doctors before the aligners are manufactured. The doctors' reviews and approvals are exchanged via ClinCheck software
{5}------------------------------------------------
(dental practitioner-facing software). Once the treatment plan is approved by the dental practitioner, the information is sent to the manufacturing facility for creation of the aligners.
When attachments are prescribed as part of the treatment plan, the subject device has the option of delivering them preformed with PFA system or with a disposable attachment template. The attachment templates are disposable polyurethane appliances which match the patient's existing dentition and include wells for the placement of dental composite (sold separately). During the first visit, the dental practitioner uses the template in bonding dental composite to the tooth surface to create attachments on the teeth. These attachments help create forces on the tooth which can assist aligner retention or optimized aligner force system for tooth movement.
The PFAs are made of a fully cured proprietary photo-polymerizable methacrylate-based resin (similar to dental composite material). PFA system is intended to enable correct placement of attachments and bond attachments to the tooth surface for aligner retention and to optimize aligner force system for tooth movement.
The patient-customized PFA are manufactured using additive manufacturing technologies and configured to individual patient's treatment plans and teeth anatomy using a proprietary internal facing 3D device-generation manufacturing software.
{6}------------------------------------------------
## Comparison with Predicate Device
In accordance with 21 CFR 807.92(a)(6) a summary of the technological characteristics' comparison of the proposed modified device to the predicate Device is provided below.
#### Technological Characteristics comparison with the Predicate Device
The Invisalign System with Pre-Formed Attachment (PFA) System (subject device) introduces an alternative attachment delivery mechanism to Invisalign treatment. The proposed change does not alter the aligners' technological characteristics and does not impact the current safety and effectiveness profile of the Invisalign System.
| | Invisalign System with PFA System<br>Subject/Modified Device | Invisalign System (K220287)<br>Predicate/ Existing Device | Comparison<br>Assessment |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use/<br>Indication for<br>Use | The Invisalign System is<br>intended for the orthodontic<br>treatment of malocclusion. | The Invisalign System is<br>intended for the orthodontic<br>treatment of malocclusion. | Same as predicate device |
| Patient<br>Population | Children, Adolescents and<br>Adults | Children, Adolescents and<br>Adults | Same as predicate device |
| Use Location | Dental intraoral Devices | Dental intraoral Devices | Same as predicate device |
| OTC or<br>Prescription<br>(Rx) Device | Rx only | Rx only | Same as predicate device |
| | Technological Characteristics | | |
| Principle of<br>Operation | Aligners: Sequential aligners<br>apply continuous gentle force<br>to the teeth and/or position<br>mandible forward. | Aligners: Sequential aligners<br>apply continuous gentle force<br>to the teeth and position<br>mandible forward | Same as predicate device |
| | 3D Software: Produces 3D-<br>model file of the PVS<br>impression or the digital scan.<br>Identifies the individual teeth<br>that requires treatment (i.e.,<br>repositioning). Creates a<br>treatment plan (i.e., 3-D<br>models that represent the<br>treatment plan) which is<br>reviewed by the treating dental<br>practitioner using ClinCheck's<br>doctor facing function to reject<br>or request modifications to the<br>set-up prior to approval. | 3D Software: Produces 3D-<br>model file of the PVS<br>impression or the digital scan.<br>Identifies the individual teeth<br>that requires treatment (i.e.,<br>repositioning). Creates a<br>treatment plan (i.e., 3-D<br>models that represent the<br>treatment plan) which is<br>reviewed by the treating dental<br>practitioner using ClinCheck's<br>doctor facing function to reject<br>or request modifications to the<br>set-up prior to approval. | Same as predicate device |
| | Delivery Mechanism of Attachment | | |
| | 1) Attachment Template -<br>Bonds dental composite<br>(material provided by doctor)<br>to the tooth surface to create<br>attachments used for aligner<br>retention or optimized aligner | 1) Attachment Template -<br>Bonds dental composite<br>(material provided by doctor)<br>to the tooth surface to create<br>attachments used for aligner<br>retention or optimized aligner | Same as predicate device |
| | Invisalign System with PFA<br>System<br>Subject/Modified Device | Invisalign System (K220287)<br>Predicate/ Existing Device | Comparison<br>Assessment |
| | force system for tooth<br>movement.<br>Delivery Mechanism: formed<br>by dental practitioner who fills<br>and packs wells in the<br>Template tray with<br>commercially available dental<br>composite. The final shape &<br>configuration is cured when | force system for tooth<br>movement.<br>Delivery Mechanism: formed<br>by dental practitioner who fills<br>and packs wells in the<br>Template tray with<br>commercially available dental<br>composite. The final shape &<br>configuration is cured when | |
| | tray is seated on dentition.<br>2) PFA System –<br>enables correct placement of<br>attachments and bond<br>attachments to the tooth<br>surface for aligner retention<br>and to optimize aligner force<br>for tooth movement.<br>Delivery of Mechanism: the<br>attachment shape and<br>configuration defined by the<br>treatment plan, manufactured<br>together with positioner. The<br>shapes are cured to the tooth<br>when the positioner is seated<br>on dentition | tray is seated on dentition.<br>2) Not applicable | Similar to predicate<br>device<br>The attachment<br>configuration prescribed<br>by treatment is identical to<br>the predicate device,<br>however the mechanism<br>of delivery attachments<br>onto the dentition is<br>different. Therefore, the<br>subject device is<br>substantially equivalent to<br>the predicate device |
| Material | Aligners and Attachment<br>Templates - Thermoplastic<br>polymer | Aligners and Attachment<br>Templates - Thermoplastic<br>polymer | Same as predicate device |
| | PFA System - Photo-<br>polymerizable methacrylate-<br>based resin | Not applicable | Similar to predicate<br>device<br>Although the material is<br>different when the<br>alternative attachment<br>delivery mechanism, pre-<br>formed attachments, is<br>prescribed by the Doctor.<br>The pre-formed<br>attachments material is<br>biocompatible and has<br>demonstrated that it can<br>deliver the force system<br>with the Invisalign System. |
| Lower Jaw<br>Adjustment<br>Mechanism | Invisalign System aligners with<br>mandibular advancement<br>feature (MAF) such as<br>enhanced precision wings with<br>a curved design. For Invisalign<br>system with standard aligners<br>this is not applicable. | Invisalign System aligners with<br>mandibular advancement<br>feature (MAF) such as<br>enhanced precision wings with<br>a curved design. For Invisalign<br>system with standard aligners<br>this is not applicable. | Same as predicate device |
{7}------------------------------------------------
{8}------------------------------------------------
# Performance Data
The existing Invisalign System's components: aligners, 3D software and attachment template, remain unchanged from the predicate device. The newly added PFA system as an optional alternative attachment delivery mechanism, underwent a complete set of functional and performance testing; including, but not limited to bond strength, stain resistance, wear, force measurement, volume precision, and usability testing. Based on the performance testing outcomes, the proposed addition of PFA system does not impact the current safety and effectiveness profile of the Invisalign System and is substantially equivalent to the predicate device.
No animal or clinical testing was required to validate these modifications implemented to the Invisalign System.
# Biocompatibility
The existing Invisalign System's components: aligners, 3D software and attachment template, remain unchanged from the predicate device. Biocompatibility testing was completed for the proposed PFA system and assessed in accordance with ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, and ISO 7405, Dentistry - Evaluation of biocompatibility of medical devices used in dentistry. The results demonstrate that the PFA system does not pose any significant biologicals risks and is considered safe for its intended use in humans.
# Software Testing
In accordance with IEC 62304. Medical device software - Software life cycle processes; successful software verification and validation (V&V) testing at the unit, integration, and system level was performed to qualify the newly added software with the subject device.
## Substantial Equivalence
The subject device is a modification to the predicate device, Invisalign System (K220287), which was cleared on April 7th, 2022. The subject device and the previously cleared predicate device are similar in that they have:
- same intended use,
- same indications for use,
- . similar principles of operation, and
- similar technological characteristics
The minor differences in the principles of operation and technological characteristics between the subject device and predicate device do not raise different questions of safety or efficacy. Therefore, the subject device is found to be substantially equivalent to the legally marketed predicate device, Invisalign System (K220287).
{9}------------------------------------------------
# Conclusion
Align Technology's Invisalign system with the PFA system has the same intended use and indications for use as the previously cleared Invisalign system (K220287). The minor principles of operation and technological differences between the subject device and its predicate device do not raise new issues of safety or effectiveness. Performance data and biocompatibility testing demonstrate that the subject device is as safe and effective as the predicate Invisalign System. Thus, the modified Invisalign system with PFA system is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.