K252786 · Shenzhen Suyzeko Limited. · OLI · Dec 1, 2025 · General, Plastic Surgery
Device Facts
Record ID
K252786
Device Name
Led Light Phototherapy Bed (Pro 450/ GY-680A)
Applicant
Shenzhen Suyzeko Limited.
Product Code
OLI · General, Plastic Surgery
Decision Date
Dec 1, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Led Light Phototherapy Bed is indicated for use as a non-invasive dermatological aesthetic treatment for the reduction of circumference of hips, waist, and thighs. Led Light Phototherapy Bed is indicated to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, and pain and stiffness associated with arthritis, for promoting relaxation of the muscle tissue, and to temporarily increase local blood circulation.
Device Story
Low-intensity LED phototherapy bed; utilizes red and infrared light (630nm, 660nm, 855nm, 940nm) to deliver topical heating. System comprises two adjustable LED panels, power adapter, and software-controlled console. Operated by professionals in clinical settings; includes emergency stop button for safety. Device delivers light energy to patient; operator selects application modes via console. Heating effect promotes muscle relaxation, increased local blood circulation, and temporary pain relief; aesthetic mode targets circumference reduction. Benefits include non-invasive pain management and body contouring.
Clinical Evidence
Bench testing only. Evidence includes electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2, IEC 60601-2-57), photobiological safety (IEC 62471:2006), and software verification/validation. Performance testing confirmed specifications for irradiance, timing, and uniformity. No clinical data presented.
Indicated for adults seeking non-invasive aesthetic reduction of hip, waist, and thigh circumference, or temporary relief of minor muscle/joint pain, muscle spasms, and arthritis-related stiffness/pain, and to increase local blood circulation.
Regulatory Classification
Identification
A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 1, 2025
Shenzhen Suyzeko Limited.
Meng Liu
General manager
Room 301, Building N, No.17 Innovation Industrial Park
Xintian Community Guanhu Street, Longhua District
Shenzhen, Guangdong 518110
China
Re: K252786
Trade/Device Name: Led Light Phototherapy Bed (Pro 450/ GY-680A)
Regulation Number: 21 CFR 878.5400
Regulation Name: Low Level Laser System For Aesthetic Use
Regulatory Class: Class II
Product Code: OLI, ILY
Dated: September 2, 2025
Received: September 2, 2025
Dear Meng Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252786 - Meng Liu
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252786 - Meng Liu
Page 3
Sincerely,
YAN FU -S
Digitally signed by YAN FU -S
Date: 2025.12.01
10:12:41 -05'00'
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252786 | |
| Device Name Led Light Phototherapy Bed (Pro 450/ GY-680A) | |
| Indications for Use (Describe) Led Light Phototherapy Bed is indicated for use as a non-invasive dermatological aesthetic treatment for the reduction of circumference of hips, waist, and thighs. Led Light Phototherapy Bed is indicated to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, and pain and stiffness associated with arthritis, for promoting relaxation of the muscle tissue, and to temporarily increase local blood circulation. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) #: K252786
510(k) Summary
Prepared on: 2025-12-01
| Contact Details | 21 CFR 807.92(a)(1) |
| --- | --- |
| Applicant Name | Shenzhen Suyzeko Limited. |
| Applicant Address | Room 301, Building N, No.17 Innovation Industrial Park, Xintian Community Guanhu Street, Longhua District Shenzhen Guangdong 518110 China |
| Applicant Contact Telephone | 86-755-29500180 |
| Applicant Contact | Mr. Meng Liu |
| Applicant Contact Email | jiang13620586569@126.com |
| Device Name | 21 CFR 807.92(a)(2) |
| Device Trade Name | Led Light Phototherapy Bed ( Pro 450/ GY-680A) |
| Common Name | Low level laser system for aesthetic use |
| Classification Name | Fat Reducing Low Level Laser |
| Regulation Number | 878.5400 |
| Product Code(s) | OLI, ILY , 21 CFR 890.5500 |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K202955 | Contour Light CL-100 | OLI |
| K153399 | LightStim Professional Led Bed | ILY |
**Device Description Summary** 21 CFR 807.92(a)(4)
LED phototherapy bed is a low-intensity LED light source (red and infrared LED) system that uses 630nm light to reduce the circumference of the hips, waist, and thighs for aesthetic benefits, and temporarily increases skin temperature using 630nm, 660nm, 855nm, and 940nm light.
The device consists of two LED panels, a software control console for selecting and maintaining LED light applications, a power adapter, and cables connected to the energy source. The LED panels can be adjusted to roughly fit the external dimensions of the patient being treated. The device is powered by connecting it to an 220-240V AC, at 4000VA, 50-60Hz power source.
The equipment is controlled by professional operators. The operator control device includes additional safety measures, including an emergency stop button. If activated, the emergency stop will stop the program and shut down, delivering power to the treated person through the LED light.
**Intended Use/Indications for Use** 21 CFR 807.92(a)(5)
Led Light Phototherapy Bed is indicated for use as a non-invasive dermatological aesthetic treatment for the reduction of circumference of hips, waist, and thighs.
Led Light Phototherapy Bed is indicated to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, and pain and stiffness associated with arthritis, for promoting relaxation of the muscle tissue, and to temporarily increase local blood circulation.
{5}
| Indications for Use Comparison 21 CFR 807.92(a)(5) |
| --- |
| The indications for use are the same between proposed device and predicate device. |
| Technological Comparison 21 CFR 807.92(a)(6) |
| The wavelength of Slimming mode of the proposed device (630nm) is similar to the predicate K202955 (635nm); The wavelength of the pain relief mode of the proposed device is same as the predicates K153399 (630nm, 660nm, 855nm, and 940nm).
The Output Intensity/Irradiance of the Slimming mode of the proposed device (4.1 mW/cm2) is same to the predicates K202955; The Output Intensity/ Irradiance of the Pain relief mode of the proposed device (58-60 mW/cm2) is similar to the predicates K153399 (60 mW/cm2).
The treatment time (30 min) is the same as K202955 and within range of K153399. The structure of the proposed device is same to the predicate K153399. |
| Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) |
| Non-Clinical Tests Submitted:
The following non-clinical testing was provided in this 510(k):
Electrical Safety and Electromagnetic Compatibility Testing – Led Light Phototherapy Bed has been tested and meets the following standard requirements of medical equipment:
● IEC60601-1: 2005+2005+CORR.1:2006+CORR.2:2007+A1:2012 Medical electrical equipment-Part 1: General requirements for basic safety and essential performance.
● IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility - Requirements and tests.
● IEC 60601-2-57:2011 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility - Requirements and tests.
● Photobiological Safety Testing – The Led Light Phototherapy Bed have been tested and comply with IEC 62471:2006, Photobiological safety of lamps and lamp systems, 1st edition. This IEC standard incorporates the principles of the following ANSI IESNA recommended practices.
Software Verification and Validation – Software documentation consistent with enhanced documentation level was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
Equipment performance testing and verification– The LED Light Phototherapy Bed Pro 450/ GY-680A were subjected to equipment performance testing (testing items include: Appearance, Effective irradiance, Timing and functions, Effective irradiation area, Effective red irradiance Uniformity, Safety test, Dielectric strength, Packaging inspection);
The device has passed all the tests mentioned above, and based on these test results, the manufacturer believes that the LED Light Phototherapy Bed Pro 450/ GY-680A are essentially equivalent to the device without causing new safety and effectiveness issues.
After analyzing non-clinical testing data, the intended use and supporting data can conclude that the device in the submission is substantially equivalent to the predicate device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.