K153399 · Led Intellectual Properties, LLC · ILY · Jul 5, 2016 · Physical Medicine
Device Facts
Record ID
K153399
Device Name
LightStim Professional LED Bed
Applicant
Led Intellectual Properties, LLC
Product Code
ILY · Physical Medicine
Decision Date
Jul 5, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LightStim Professional LED Bed is an over-the-counter device intended to emit energy in the visible and IR spectrum intended to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain or muscle spasm, the temporary increase in local blood circulation, and the temporary relaxation of muscles.
Device Story
LightStim Professional LED Bed; tanning-bed-style device (6.5ft x 3ft); 18,240 LEDs emitting visible and IR spectrum (630nm, 660nm, 855nm, 940nm). Device provides topical heating to elevate tissue temperature. Operated by trained personnel; includes mandatory one-on-one training and instruction manual. Features AC to DC power supply and 30-minute auto-off timer. Used for temporary relief of minor muscle/joint pain, stiffness, arthritis, and muscle spasms; increases local blood circulation. Benefits patient via non-invasive thermal therapy.
Clinical Evidence
Bench testing only. Skin temperature testing conducted on participants at six body locations over 30 minutes. Results showed temperature fluctuations between 40-41.3°C, consistent with FDA guidelines for ILY category and identical to predicate device performance. Electrical safety and electromagnetic compatibility testing performed per IEC/EN 60601-1 (3rd Ed) and IEC/EN 60601-1-2.
Technological Characteristics
18,240 LEDs; wavelengths 630nm, 660nm, 855nm, 940nm; output 60mW/cm2. Dimensions approx 6.5ft x 3ft. AC to DC power supply. Standards: IEC/EN 60601-1 (3rd Ed) for basic safety/performance, IEC/EN 60601-1-2 for EMC, IEC-62471 for radiation safety. 30-minute auto-off timer.
Indications for Use
Indicated for temporary relief of minor muscle and joint pain/stiffness, minor arthritis pain, muscle spasm, temporary increase in local blood circulation, and temporary muscle relaxation. Over-the-counter use.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three faces in profile, stacked on top of each other. The figure is positioned to the right of the text, which reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 5, 2016
Led Intellectual Properties LLC Mr. Steve Marchese CEO 16552 Von Karman Ave Irvine, California 92606
Re: K153399
Trade/Device Name: LightStim Professional Led Bed Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: ILY Dated: June 14, 2016 Received: June 14, 2016
Dear Mr. Marchese:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use Statement
| 510(k) Number (if known) | K153399 |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | LightStim Professional LED Bed |
| Indications for Use | The LightStim Professional LED Bed is an over-the-counter device intended to emit energy in the visible and IR spectrum intended to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain or muscle spasm, the temporary increase in local blood circulation, and the temporary relaxation of muscles. |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use_ (Per 21 CFR 801. 109) OR
X Over-The-Counter Use
{3}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness
LED Intellectual Properties, LLC.
### Device: LightStim Professional LED Bed
#### 1. General Information
Submitter: LED Intellectual Properties, LLC 16552 Von Karman Ave. Irvine, Ca. 92606
Contact Person: Steve Marchese & Chase Marchese Phone: Steve (949) 502-4088 Chase (714) 924-0492
Emails: Steve@lightstim.com & Chase@lightstim.com
#### 2. Device name and code
Device Proprietary Name: LightStim Professional LED Bed
Classification Code and name: ILY - Infrared Lamp, Therapeutic Heating
#### 3. Predicate Devices
LED Intellectual Properties LLC: Pain Therapy Light (K083580, LightStim Professional 2-Panel Light (K150098), LightStim LED Belt (K151333)
#### 4. Device Description
The LightStim Professional LED Bed is a device that is similar in design to the bottom half of a tanning bed. It is approximately 6.5 feet in length and 3 feet wide. It has 18,240 LEDs. There is an on/off switch that turns off after 30 minutes of use, and an AC to DC power supply.
The LightStim Professional LED Bed is intended to be operated only by a person who has been personally trained by Company personnel. Each LED Bed includes a detailed Instruction Manual and includes one-on-one training and instruction by Company training staff. That training includes a thorough review of the Instruction Manual for the device and actual one-on-one training on the operation of the LED Bed with the end user.
#### 5. Indications for Use
The LightStim Professional LED Bed is an over-the-counter device intended to emit energy in the visible and IR spectrum intended to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain or muscle spasm, the temporary increase in local blood circulation, and the temporary relaxation of muscles.
{4}------------------------------------------------
#### 6. Substantial Equivalency
| Characteristic | LightStim for Pain<br>(K083580) | LightStim LED Belt<br>(K151333) | LightStim Professional 2-<br>Panel Light (K150098) | LightStim Professional LED<br>Bed |
|------------------------------------------------------|------------------------------------|---------------------------------|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Output in milliwatts | 60mW cm2 | 60mW cm2 | System #2 = 60mW cm2 | 60mW cm2 |
| Treatment Time | 5-30 minutes | 5-30 minutes | System #2 = 5-30 minutes | 5-30 minutes |
| Wavelengths (nm) | 630nm, 660nm, 855nm,<br>940nm | 630nm, 660nm, 855nm,<br>940nm | System #2 = 630nm, 660nm,<br>855nm, 940nm | 630nm, 660nm, 855nm, 940nm<br>Procured from the same<br>manufacturer with the exact same<br>specifications as all three of the |
| Manufacturer | LED Intellectual Properties | LED Intellectual Properties | LED Intellectual Properties | LED Intellectual Properties |
| Basic Safety+Performance<br>Electronic Compatibility | IEC/EN 60601-1<br>IEC/EN 60601-1-2 | IEC/EN 60601-1 3rd<br>edition | IEC/EN 60601-1 3rd Edition<br>IEC/EN 60601-1-2 | IEC/EN 60601-1 3rd Edition<br>IEC/EN 60601-1-2 |
| Radiation Safety | IEC-62471 | IEC-62471 | IEC-62471 | IEC-62471 |
| FDA Clearance | OTC | OTC | OTC | OTC |
Device Technological Characteristics Comparrison Table
The Conclusion drawn by LED intellectual Properties, LLC based on the table above, is that the LightStim Professional LED Bed is substantially equivalent to the predicates devices and raises of safety or effectiveness.
#### 7. Performance Testing
The LightStim Professional LED Bed device was tested using the same skin temperature testing as all of the predicate devices. The participants skin temperature was measured on six different locations of the body prior to using the device, then each body location was rechecked four times at 3-minute intervals and once at the end of the 30-minute test period. All of the participants skin temperature fluctuated between 40-41.3 degrees centigrade during the 30-min temperature test, which is within the FDA guidelines for the IL Y cateqory, and those readings were almost identical to the readings of the predicate devices, which used the same procedure. Temperature readings were measured using the same device that was used for the predicate devices, a "Mastercool, Infrared Laser Thermometer".
As the same temperature testing procedure was conducted, using the same thermometer, and achieving almost identical results as all three predicate devices, the LightStim Professional LED Bed raises no new issues of safety or effectiveness.
The LightStim Professional LED Bed device results in patient contact with an electrically powered component therefore; it was tested for conformance to IEC / EN 60601-1 3° Edition: Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance
The LightStim Professional LED Bed device includes an electronic component, and it was tested for conformance to IEC / EN 60601-1-2 3ª Edition Medical Electrical Equipment - Part 1-2: General Requirements for Safety and Essential Performance-Collateral standard: Electromagnetic compatibility-Requirements and tests.
{5}------------------------------------------------
Therefore, taking into consideration the Table for Substantial Equivalence, an analysis of safety, indications, intended uses, performance, and technological properties, the LightStim Professional LED Bed raises no new issues of safety or effectiveness and has been found to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.