K252603 · Shenzhen Kaiyan Medical Equipment Co., Ltd. · OHS · Jan 28, 2026 · General, Plastic Surgery
Device Facts
Record ID
K252603
Device Name
HIGHERDOSE Body Sculptor (Model: GS-03)
Applicant
Shenzhen Kaiyan Medical Equipment Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Jan 28, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HIGHERDOSE BODY SCULPTOR (Model: GS-03) is an over-the-counter device that is intended to improve the appearance of wrinkles.
Device Story
Handheld, home-use LED phototherapy device; emits 650nm red light and 850nm near-infrared light to improve wrinkle appearance. Device constructed of ABS/PC plastic and silicone; powered by rechargeable lithium battery via charging disk. User operates via power button to toggle device and LED therapy. Device uses internal timer and software to control treatment duration and LED intensity. No measurements or diagnostic conclusions performed. Benefits include non-invasive wrinkle reduction via light therapy. Used by consumers in home settings.
Clinical Evidence
No clinical data. Safety and effectiveness supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), battery safety (IEC 62133-2), and biocompatibility (ISO 10993-5, -10, -23).
Indicated for over-the-counter use to improve the appearance of wrinkles in the general population.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 28, 2026
Shenzhen Kaiyan Medical Equipment Co., Ltd.
Alain Dijkstra
Building #3, and Building#5, 40th of Fuxin Street, Huaide
Community, Fuyong Town, Baoan District
Shenzhen, Guangdong 518103
China
Re: K252603
Trade/Device Name: HIGHERDOSE Body Sculptor (Model: GS-03)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS
Dated: July 26, 2025
Received: August 18, 2025
Dear Alain Dijkstra:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252603 - Sophia Zhou
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252603 - Sophia Zhou
Page 3
Sincerely,
YAN FU-S
Digitally signed by YAN FU-S
Date: 2026.01.28 19:38:14
-05'00"
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252603 | |
| Device Name HIGHERDOSE Body Sculptor (Model:GS-03) | |
| Indications for Use (Describe) The HIGHERDOSE BODY SCULPTOR (Model: GS-03) is an over-the-counter device that is intended to improve the appearance of wrinkles. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.” | |
{4}
510(k) Summary
K252603
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
1. Submitter's Information
Sponsor
Sponsor Name: Shenzhen Kaiyan Medical Equipment Co., Ltd.
Establishment Registration Number: 3011644607
Address: Building #3, and Building#5, 40th of Fuxin Street, Huaide Community, Fuyong Town, Baoan District, Shenzhen, Guangdong, 518103, China
Contact Person (including title): Alain Dijkstra (CEO)
Tel: 0755-82129361
Fax: 0755-25024651
E-mail: alaindijkstra@kaiyanmedical.com
Distributor
Company Name: HIGHERDOSE LLC
Address: 42 Broadway, 12th Floor, #212, NewYork, NY 10004
Establishment Registration Number: 3019734322
Contact Person: Sumish Khadka
E-mail: sumish@higherdose.com
Manufacturer:
Manufacturer Name: Shenzhen Kaiyan Medical Equipment Co., Ltd.
Establishment Registration Number: 3011644607
Address: Building #3, and Building#5, 40th of Fuxin Street, Huaide Community, Fuyong Town, Baoan District, Shenzhen, Guangdong, 518103, China
Contact Person (including title): Alain Dijkstra (CEO)
Tel: 0755-82129361
Fax: 0755-25024651
E-mail: alaindijkstra@kaiyanmedical.com
2. Date of the summary prepared: January 27, 2026
3. Subject Device Information
Classification Name: Light Based Over-the-Counter Wrinkle Reduction (OHS)
Trade Name: HIGHERDOSE Body Sculptor (Model: GS-03)
Review Panel: General & Plastic Surgery
Product Code: OHS
Regulation Number: 878.4810
Page 1 / 8
{5}
Regulatory Class: II
## 4. Predicate Device Information
### Predicate Device 1 Information
Sponsor: Light Tree Ventures Europe B.V.
Trade Name: Infrared Heat, (Model: E0221)
Classification Name: Light Based Over The Counter Wrinkle Reduction
510(K) Number: K223893
Review Panel: General & Plastic Surgery
Product Code: OHS
Regulation Number: 878.4810
Regulation Class: II
### Predicate Device 2 Information
Sponsor: Shenzhen Kaiyan Medical Equipment Co., Ltd
Trade Name: Solawave 2-in-1 Skincare Mini (61043)
Classification Name: Light Based Over The Counter Wrinkle Reduction
510(K) Number: K250532
Review Panel: General & Plastic Surgery
Product Code: OHS
Regulation Number: 878.4810
Regulation Class: II
### Predicate Device 3 Information
Sponsor: Dongguan Laiguang Electronic Technology Co., Ltd.
Trade Name: Bestqool LED therapy device
Classification Name: Light Based Over The Counter Wrinkle Reduction, Over-The-Counter Powered Light Based Laser For Acne
510(K) Number: K242789
Review Panel: General & Plastic Surgery
Product Code: OHS, OLP
Regulation Number: 878.4810
Regulation Class: II
## 5. Device Description
The HIGHERDOSE Body Sculptor (Model: GS-03) is a home-use light-emitting diode phototherapy device with two proven wavelengths of light 650nm Red light and 850nm Near infrared red light, both of these lights are known to reducing the wrinkle.
The main device is made of black plastic (ABS+PC) and silicone button cup, and contain light emitting diode (LED) which will emit red light and infrared light. The device is a handheld device.
Page 2 / 8
{6}
The device contains a rechargeable Lithium battery, which can be charged by charging disk. The HIGHERDOSE Body Sculptor cannot be operated while charging.
Press and hold the power button to turn the product on/off. Press the power button briefly to turn the LED light therapy on/off.
The device is not used to make measurements of any sort, or to draw any conclusions regarding the indication to treat. The device does not require checks on the light output as the LEDs do not dim with age to any practical extent.
## 6. Intended Use / Indications for Use
The HIGHERDOSE BODY SCULPTOR (Model: GS-03) is an over-the-counter device that is intended to improve the appearance of wrinkles.
Page 3 / 8
{7}
Page 4 / 8
# 7. Comparison to predicate device and conclusion
Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise and new questions of safety or effectiveness.
| Elements of Comparison | Subject Device | Predicate Device 1 K223893 | Predicate Device 2 K250532 | Predicate Device 3 K242789 | Remark |
| --- | --- | --- | --- | --- | --- |
| Company | Shenzhen Kaiyan Medical Equipment Co.,Ltd | Light Tree Ventures Europe B.V. | Shenzhen Kaiyan Medical Equipment Co.,Ltd | Dongguan Laiguang Electronic Technology Co.,Ltd. | -- |
| Trade Name | HIGHERDOSE Body Sculptor | Infrared Heat (Model: E0221) | Solawave 2-in-1 Skincare Mini (61043) | Bestqool LED therapy device Model: LMK-001LMK-004LMK-005LMK-006 | -- |
| Classification Name | Light Based Over the Counter Wrinkle Reduction | Light Based Over The Counter Wrinkle Reduction | Light Based Over the Counter Wrinkle Reduction | Light Based Over the Counter Wrinkle Reduction, Over-The-Counter Powered Light-Based Laser for Acne | -- |
| 510(k) Number | K252603 | K223893 | K250532 | K242789 | -- |
| Product Code | OHS | OHS | OHS | OHS, OLP | Same |
| FDA Device Classification | Class II | Class II | Class II | Class II | Same |
| Intended Use / Indications for Use | Light Therapy: The HIGHERDOSE BODY SCULPTOR (Model: GS-03) is an over-the-counter device that is intended to | The infrared Heat (Model:E0221) is intended to emit energy in the red and infrared spectrum for use in the treatment of full face | The Solawave 2-in-1 Skincare Mini (Model: 61043) is an over-the-counter device that emits energy in the | This device produces light in the blue light (415nm) is intended to reduce mild to moderate inflammatory acne | Same |
{8}
Page 5 / 8
| | improve the appearance of wrinkles. | wrinkles. | red and infrared spectrum for treating wrinkles on the face and decolletage. | vulgaris. The red light (635nm) in combination with near-infrared light (850nm) is intended to improve the appearance of wrinkles. | |
| --- | --- | --- | --- | --- | --- |
| Intended Loaction Use | Body | Face | Face | Face and Body | |
| Energy Type | Light emitting diodes | Light emitting diodes | Light emitting diodes | Light emitting diodes | Same |
| Wavelengths | Red: 650nm±10nm, Near-infrared :850nm±10nm | Red: 630nm±20nm Near-infrared :830±20nm | Red: 630nm ± 10nm Near-infrared: 830nm ±10nm | Red :635nm±10nm Near-infrared :850nm±10nm | Similar Note 1 |
| Total Intensity (mW/cm²) | Red+Near-infrared : 650nm+850nm Total:40±10 mW/cm² | Red+Near-infrared : 630nm + 830nm Total: 30mW/cm² | Red+Near-infrared : 630nm+830nm Total: 40-55 mW/cm² | Red+Near-infrared : 635nm+850nm Total: 7.5mW/cm² ± 25% | Similar Note 2 |
| Treatment Time | 10 mins | 10 minutes per day, 2-5 times per week | 3 minutes per treatment area | 4 times a week 30-60 minutes session 4 weeks | Same |
| Software controller | Device uses a timer and software to control treatment duration and LED’s intensity | Device uses a timer and software to control treatment duration and LED’s intensity | Device uses a timer and software to control treatment duration and LED’s intensity | Device uses a timer and software to control treatment duration | Same |
Note1:
The subject device’s wavelength is similar with the predicate devices 1, 2, 3 (K223893, K250532, K242789). The wavelengths of these lights are all within the spectral range of red and near-infrared light. Therefore, the subject device and predicate device provide the same treatment benefit.
Additionally, the subject device has already passed the testing regarding to standard IEC 60601-2-57 and LED material has passed IEC 62471 standards. In conclusion, the device is safe and effective. The slight differences between the subject device and the predicate devices will not raise any safety or effectiveness issues.
Note 2:
The subject device’s intensity is similar with the predicate devices 1, 2 (K223893, K250532). The intensity of these lights are all within the spectral range of red and near-infrared light. Therefore, the subject device will not raise any safety or effectiveness issues.
{9}
The subject device is not higher than Predicate Device 2 (55mW/cm², indicating the safety of the device, and no lower than Predicate Device 3(K242789) 7.5mW/cm² +/- ± 25%, indicating the effectiveness of the device. The subject device's intensity is between with rage of 3 Predicate Devices.
Additionally, the subject device has already passed the testing regarding to standard IEC 60601-1 and IEC 60601-2-57. In conclusion, the device is safe and effective. The slight differences between the subject device and the predicate devices will not raise any safety or effectiveness issues.
# Final Conclusion:
The subject device is the same or similar to the legally marketed predicate device K223893, K250532, K242789.
# 8. Test Summary
# 8.1 Summary of Non-Clinical Performance Testing
# 1) Performance Testing Summary
HIGHERDOSE Body Sculptor (Model: GS-03) has been evaluated the safety and performance by lab bench testing as following:
| Title of the test | Test Method/Applicab le Standards | Acceptance criteria | Unexpected Results/Sig nificant Deviations | Test results |
| --- | --- | --- | --- | --- |
| General requirements for basic safety and essential performance | IEC 60601-1:2005 /AMD1:2012/AMD 2:2020 | The test is carried out under the test method specified in the standard, and the test result is within the test acceptance range of the standard. | NA | Pass |
| Electromagnetic disturbances | IEC 60601-1-2:20 14+A1:2020 | No degradation of performance was found during test or Lower than limits of measurement | NA | Pass |
| Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. | IEC 60601-1-11:2 015/AMD1:2020 | The device operates normally, and can provide basic safety and essential performance. | NA | Pass |
| Particular Requirements for The Basic Safety and Essential Performance of Non-Laser Light Source Equipment Intended for Therapeutic, Diagnostic, Monitoring and Cosmetic/Aesthetic Use | IEC 60601- 2-57:2 011 | The test is carried out under the test method specified in the standard, and the test result is within the test acceptance range of the standard. | NA | Pass |
{10}
| Title of the test | Test Method/Applicable Standards | Acceptance criteria | Unexpected Results/Significant Deviations | Test results |
| --- | --- | --- | --- | --- |
| Photobiological safety of lamps and lamp systems. | IEC 62471:2006 | The test is carried out under the test method specified in the standard, and the test result is within the test acceptance range of the standard.
Purchase the products from the supplier, and the test is conducted by the supplier. | NA | Pass |
| Li-ion Battery safety | IEC 62133-2:2017 | The test is carried out under the test method specified in the standard, and the test result is within the test acceptance range of the standard.
Purchase the products from the supplier, and the test is conducted by the supplier. | NA | Pass |
| Performance Test | The Performance Test Report performs the following tests on the finished product: Power Density Test; Leakage current test. | The device can meet the requirement of the performance test, Power Density test and Leakage current test. | NA | Pass |
Page 7 / 8
{11}
Page 8 / 8
2) Biocompatibility testing
The applied part of subject device, gel and spray will have direct contact with the human body.
The Biocompatibility testing was conducted on GS-03. The subject device complies with the biocompatibility requirements of ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Sensitization), and ISO 10993-23 (Irritation).
3) Usability Testing
Usability testing was conducted on GS-03, the device complies with IEC 62366-1 and IEC 60601-1-6.
4) Software verification and validation testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA'S Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions: Guidance for Industry and Food and Drug Administration Staff"
8.2 Clinical Performance
Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively.
9. Final Conclusion:
The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate devices K223893, K250532, K242789.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.