K250532 · Shenzhen Kaiyan Medical Equipment Co., Ltd. · OHS · May 22, 2025 · General, Plastic Surgery
Device Facts
Record ID
K250532
Device Name
Solawave 2-in-1 Skincare Mini (Model: 61043)
Applicant
Shenzhen Kaiyan Medical Equipment Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
May 22, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Solawave 2-in-1 Skincare Mini (Model: 61043) is an over-the-counter device that emits energy in the red and infrared spectrum for treating wrinkles on the face and decolletage.
Device Story
Hand-held, battery-powered device; emits red (630nm) and infrared (830nm) LED light for wrinkle reduction; includes non-medical warming function. Used by consumers in home settings. User applies device to face and decolletage for 3 minutes per treatment area. Output consists of light energy; healthcare providers do not interact with device output; device intended for aesthetic benefit to patient.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including electrical safety, electromagnetic compatibility, photobiological safety, and biocompatibility (ISO 10993-5, -10, -23).
Technological Characteristics
Hand-held LED-based light therapy device. Wavelengths: 630nm (red) and 830nm (infrared). Power: 3.7V, 450mAh Lithium-Ion battery. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, IEC 62471, IEC 62133-2. Biocompatibility: ISO 10993-5, ISO 10993-10, ISO 10993-23.
Indications for Use
Indicated for over-the-counter treatment of wrinkles on the face and decolletage using red and infrared light energy.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 22, 2025
Shenzhen Kaiyan Medical Equipment Co., Ltd
Alain Dijkstra
Official Correspondent
Building#3 and Building#5, 40th of Fuxin Street,
Huaide Community Fuyong Town, Baoan District
Shenzhen, Guangdong 518103
China
Re: K250532
Trade/Device Name: Solawave 2-in-1 Skincare Mini (61043)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS
Dated: February 22, 2025
Received: February 24, 2025
Dear Alain Dijkstra:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250532 - Alain Dijkstra
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K250532 - Alain Dijkstra
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.05.22
15:29:33 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250532 | |
| Device Name Solawave 2-in-1 Skincare Mini (Model: 61043) | |
| Indications for Use (Describe) The Solawave 2-in-1 Skincare Mini (Model: 61043) is an over-the-counter device that emits energy in the red and infrared spectrum for treating wrinkles on the face and decolletage. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250532
510(k) Summary
This summary of 510(k) is being submitted in accordance with the requirement of 21 CFR 807.92.
1. Submitter's Information
Sponsor Name: Shenzhen Kaiyan Medical Equipment Co., Ltd
Establishment Registration Number: 3011644607
Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China
Contact Person (including title): Alain Dijkstra (Manager)
Tel: +86-135-10378748
Fax: +86-755-25024651
E-mail: registrar01@kaiyanmedical.com
Distributor:
Company: Solawave Inc.
Address: 3633 Lenawee Avenue Suite 100 Los Angeles, CA 90016 USA
Application Correspondent:
Contact Person: Alain Dijkstra
Company: Shenzhen Kaiyan Medical Equipment Co., Ltd
Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China
Tel: +86 755 82129361
Fax: +86 755 25024651
Email: registrar01@kaiyanmedical.com
2. Subject Device Information:
Common/Trade Name: Solawave 2-in-1 Skincare Mini (Model: 61043)
Classification Name: Light Based Over the Counter Wrinkle Reduction
Review Panel: General & Plastic Surgery
Product Code: OHS
Regulation Number: 21 CFR 878.4810
Regulation Class: II
3. Predicate and Reference Device Information
Predicate Device (K213841)
Sponsor: Premier North America Inc
Trade Name: ENEO TOTALE
Classification Name: Light Based Over the Counter Wrinkle Reduction
Review Panel: General & Plastic Surgery
Product Code: OHS
Regulation Number: 21 CFR 878.4810
Regulation Class: II
K250532
{5}
K250532
2 / 6
Reference Device 1 (K241102)
Sponsor: FOREO, Inc.
Trade Name: Luna™ 4 plus
Classification Name: Light Based Over the Counter Wrinkle Reduction; Transcutaneous electrical nerve stimulator for pain relief.
Review Panel: General & Plastic Surgery
Product Code: OHS,NFO
Regulation Number: 21 CFR 878.4810, 21 CFR 882.5890
Regulation Class: II
Reference Device 2 (K202055)
Sponsor: Heat In A Click LLC
Trade Name: Looper
Classification Name: Light Based Over the Counter Wrinkle Reduction; over-the-counter powered light based laser for acne
Review Panel: General & Plastic Surgery
Product Code: OHS,OLP
Regulation Number: 21 CFR 878.4810
Regulation Class: II
Reference Device 3 (K242382)
Sponsor: ISMART Developments Ltd
Trade Name: décoLITE
Classification Name: Light Based Over the Counter Wrinkle Reduction;
Review Panel: General & Plastic Surgery
Product Code: OHS
Regulation Number: 21 CFR 878.4810
Regulation Class: II
4. Device Description
The Solawave 2-in-1 Skincare Mini (Model: 61043) is a hand-held, battery-powered device that reduce wrinkles by emitting LED red light (630nm) and infrared light (830nm). The device is powered by a Lithium-Ion rechargeable battery, and it comes with a charging cable, storage bag, and instruction manual.
The Solawave 2-in-1 Skincare Mini is a revolutionary device that offers two simultaneous approaches for your skin routine in one handheld device:
1. Red and Near-Infrared Light Therapy
2. Warming function (This function is not for medical purpose.)
5. Intended Use / Indications for Use
The Solawave 2-in-1 Skincare Mini (Model: 61043) is an over-the-counter device that emits energy in the red and infrared spectrum for treating wrinkles on the face and decolletage.
{6}
# 6. Comparison to predicate and reference devices
Compare with the predicate devices, the subject device is very similar in design principle, intended use, indications for use, functions and the applicable standards.
| Elements of Comparison | Subject device | Predicate device (K213841) | Reference device 1 (K241102) | Reference device 2 (K202055) | Reference device 3 (K242382) | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| Manufacturer | Shenzhen Kaiyan Medical Equipment Co., Ltd | Premier North America Inc | FOREO, Inc. | Heat In A Click LLC | ISMART Developments Ltd | -- |
| 510 (K) Number | K250532 | K213841 | K241102 | K202055 | K242382 | -- |
| Device Name | Solawave 2-in-1 Skincare Mini | ENEO TOTALE | Luna™ 4 plus | Looper | décoLITE | -- |
| Model | 61043 | unknown | Unknown | ZX-579S | / | -- |
| OTC/Rx | OTC | OTC | OTC | OTC | OTC | Same |
| Regulation Class | Class II | Class II | Class II | Class II | Class II | Same |
| Product Code | OHS | OHS | OHS,NFO | OHS, OLP | OHS | Same |
| Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810
21 CFR 882.5890 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Indications for Use / Intended use | The Solawave 2-in-1 Skincare Mini (Model: 61043) is an over-the-counter device that emits energy in the red and infrared spectrum for treating wrinkles on the face and decolletage. | ENEO TOTALE is an over the counter device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles. | 1. The Red + IR light is intended to treat periorbital wrinkles
2. The microcurrent is intended to treat facial stimulation | Looper (Model: ZX-579S) is a hand-held device for over-the-Counter aesthetic purposes. The Photon mode red light is indicated for the use in Treating wrinkles on the face. | The décoLITE LED device is an over-the-counter device that is intended for the use in the treatment of wrinkles in the decolletage area. | Similar (Note 1) |
| Intended Location Use | Face, decolletage | Face | Face | Face | Decolletage (upper chest) | Similar (Note 2) |
K250532
{7}
| Elements of Comparison | Subject device | Predicate device (K213841) | Reference device 1 (K241102) | Reference device 2 (K202055) | Reference device 3 (K242382) | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| Power Source | Lithium battery: DC 3.7V, 450mAh, 1.665Wh | Adaptor:100~240V AC 50/60Hz 2.4A Lithium battery: 3.75Vdc, 3000 mAh | Internal rechargeable Lithium battery | DC 3.7V 1000mA Li battery | Li battery | Similar (Note 3) |
| Irradiance source | LED | LED | LED, micro current | LED | LED | Same |
| Handheld | Yes | Yes | Yes | Yes | No(Mask type) | Similar |
| Dose/per time(J/cm²) | 7.2~9.9 | 37.5~52.5 | 3.78~15.12 | 9.9 | 18 | Similar (Note 4) |
| LED wavelength | Red: 630nm ± 10nm Infrared: 830nm ± 10nm | Red: 633nm± 5nm Infrared:830nm± 5nm | RED light:633nm± 10nm Red+IR: 633+10nm/8 50nm±10nm | Red: 630±00nm | Red: 630nm ± 10nm NIR: 830nm ± 10nm | Similar (Note 4) |
| Irradiances | RED+IR: 40-55 mW/cm² | 69mW/cm² for 633nm 56mW/cm² for 830nm | RED+IR: 63 mW/cm² | Red light: 55mW/cm²±10 % | 30mW/cm2 total | Similar (Note 4) |
| Treatment Time | 3 minutes per treatment area | 5-7 minutes on each treatment zone For the first month (4 weeks)treatment should be performed 3 times a week for 15-20 minutes each time | 1-4 min (adjustable) per treatment | 3 minutes per target area | 600 seconds (10 minutes) 5× weekly, 6 weeks | Similar (Note 4) |
K250532
4 / 6
{8}
| Elements of Comparison | Subject device | Predicate device (K213841) | Reference device 1 (K241102) | Reference device 2 (K202055) | Reference device 3 (K242382) | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| Electrical Safety | Compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57 IEC 62471 | Compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62471 | Compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57 IEC60601-2-10 | Compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57 | IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57 IEC 60601-2-83 IEC 62471 | Same |
| Biocompatibility | ISO 10993-1 ISO10993-5 ISO10993-10 ISO 10993-23 | ISO10993-5 ISO10993-10 | ISO10993-5 ISO10993-10 ISO 10993-23 | ISO10993-5 ISO10993-10 | unknown | Same |
Comparison in Detail(s):
Note 1:
The Intended use of subject device is to treat wrinkle on the face, which is slightly different from the predicate device and reference device 1 (which treat periorbital wrinkles); However, it is the same as reference device 2, which is indicated for the entire face wrinkle treatment and both of them are hand held LED device.
Note 2:
The intended location use of the subject device is for face and decolletage, which is same as the predicate device, reference device 1, reference device 2 (for face) and reference device 3 (for decolletage).
Note 3:
The power supply for the subject device is a little different from that of the predicate device, however the lithium battery of the subject device has been tested under standard IEC 62133-2, so this difference should not raise any safety/effectiveness issues.
Note 4:
The subject device is a little different from the predicate device in terms of "Treatment Time", "Dose/per time" and "Irradiances", however they have the same LED wavelengths; And the "Treatment time", "Dose/per time" and "Irradiances" of the subject device are very close to those of reference devices 1 and 2; so it can be concluded that the subject device can achieve the same treatment effect (wrinkle treatment) as the predicate and reference devices. In addition, all of them have passed the tests of IEC 60601-1, IEC 60601-1-11 and IEC 60601-2-57, these differences will not raise any new safety or effectiveness issues.
7. Test Summary
7.1 Non-Clinical Tests Performed
1) Electrical safety, and electromagnetic compatibility Test
K250532
{9}
Non-clinical tests were performed on the subject device to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility:
- IEC 60601-1 2020-08 Edition 3.2 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-11 Edition 2.1 2020-07 Medical Electrical Equipment --Part 1: General Requirements for Basic Safety and Essential Performance --Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment.
- IEC 60601-2-57 Edition 2.0 2023 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.
- IEC 60601-1-2 Edition 4.1 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
- IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems.
- IEC 62133-2:2017+AMD1:2021 Edition1.1 Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems.
## 2) Biocompatibility Test
The component of the Solawave 2-in-1 Skincare Mini (Model: 61043) has been conformed to ISO 10993-5, ISO 10993-10 and ISO 10993-23.
## 3) Software verification and validation
Software verification and validation testing was conducted and documentation provided as recommended by FDA's Guidance for Industry and FDA Staff "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device is considered as a "moderate" level concern, since a malfunction of or a latent design flaw in the Software Device leads to an erroneous diagnosis or a delay in the delivery of appropriate medical care that would likely lead to Minor Injury.
## 7.2 Summary of Clinical Performance
Clinical testing was not needed for this 510(k).
## 8. Date Prepared: May 22, 2025
## 9. Final Conclusion
The subject device is as safe, as effective, and performs as well as the legally marketed predicate device, cleared under K213841, Class II (21 CFR 878.4810), product code OHS.
K250532
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.