Facial & Body Beauty Device (INIA-BD001, INIA-BD002, INIA-BD003, INIA-BD004, F2C15, F3222 Pro, F5516, F3606, F2210 Pro, F5808, INIA-ED001, INIA-ED002, INIA-BLD001, E1507)

K252553 · Shenzhen Jianchao Intelligent Technology Co., Ltd. · NFO · Oct 9, 2025 · Neurology

Device Facts

Record IDK252553
Device NameFacial & Body Beauty Device (INIA-BD001, INIA-BD002, INIA-BD003, INIA-BD004, F2C15, F3222 Pro, F5516, F3606, F2210 Pro, F5808, INIA-ED001, INIA-ED002, INIA-BLD001, E1507)
ApplicantShenzhen Jianchao Intelligent Technology Co., Ltd.
Product CodeNFO · Neurology
Decision DateOct 9, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Facial & Body Beauty Device is a hand-held device for over-the counter aesthetic purposes. The Microcurrent stimulation mode is indicated for facial skin stimulation and body skin stimulation.

Device Story

Hand-held, portable, reusable aesthetic device; utilizes microcurrent stimulation for facial and body skin. Device consists of main unit and internal rechargeable lithium battery; user places electrode head on skin to deliver stimulation. Operation controlled by internal microprocessor; features automatic shut-off after 10-minute treatment cycle. Used in home environment by consumers for aesthetic purposes. Output is bi-phase square-wave pulse; device provides regulated voltage. Healthcare providers do not operate; output is intended for consumer aesthetic benefit. Safety features include low battery indicator and on/off status indicator.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10), battery safety (IEC 62133-2), and biocompatibility (ISO 10993-5, 10993-10, 10993-23).

Technological Characteristics

Hand-held stimulator; materials include PC, PC+ABS, and Zinc alloy. Sensing/actuation via microcurrent electrical stimulation; energy source is internal 3.7V Li-ion battery. Output: 1 channel, regulated voltage, bi-phase square-wave pulse. Connectivity: standalone (charging cable provided). Sterilization: non-sterile. Software: microprocessor-controlled with automatic shut-off.

Indications for Use

Indicated for facial and body skin stimulation for over-the-counter aesthetic purposes in general adult populations.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION October 9, 2025 Shenzhen Jianchao Intelligent Technology Co., LTD % Riley Chen RA Specialist Feiying Drug &amp; Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90 Qianhai Road Shenzhen, Guangdong 518052 China Re: K252553 Trade/Device Name: Facial &amp; Body Beauty Device (INIA-BD001, INIA-BD002, INIA-BD003, INIA-BD004, F2C15, F3222 Pro, F5516, F3606, F2210 Pro, F5808, INIA-ED001, INIA-ED002, INIA-BLD001, E1507) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: August 12, 2025 Received: August 13, 2025 Dear Riley Chen: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of {1} K252553 - Riley Chen Page 2 Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {2} K252553 - Riley Chen Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Tushar Bansal -S Tushar Bansal, PhD Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252553 | ? | | Please provide the device trade name(s). | | ? | | Facial & Body Beauty Device (INIA-BD001, INIA-BD002, INIA-BD003, INIA-BD004, F2C15, F3222 Pro, F5516, F3606, F2210 Pro, F5808, INIA-ED001, INIA-ED002, INIA-BLD001, E1507) | | | | Please provide your Indications for Use below. | | ? | | Facial & Body Beauty Device is a hand-held device for over-the counter aesthetic purposes. The Microcurrent stimulation mode is indicated for facial skin stimulation and body skin stimulation. | | | | Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D) | ? | | | ✗ Over-The-Counter Use (21 CFR 801 Subpart C) | | {4} Shenzhen Jianchao Intelligent Technology Co., LTD 510(k)s – 510(k) Summary # 510(k) Summary - K252553 "510(k) Summary" as required by 21 CFR Part 807.92. ## I. Submitter Shenzhen Jianchao Intelligent Technology Co., Ltd. Rm301, Bldg.25, No.68 Hexiu West Road, Fuhai St., Baoan, Shenzhen, CN Post code: 518107 Fax: +86-0755-86961489 Fred Li Title: Director Tel.: +86 18675507170 Email: fred@vellcolife.com Date of preparation: 2025-10-09 ## II. Device Information Name of Device: Facial &amp; Body Beauty Device Models: INIA-BD001, INIA-BD002, INIA-BD003, INIA-BD004, F2C15, F3222 Pro, F5516, F3606, F2210 Pro, F5808, INIA-ED001, INIA-ED002, INIA-BLD001, E1507 Common or Usual Name: Stimulator, Transcutaneous Electrical, Aesthetic Purposes Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: II Product Code: NFO Regulation Number: 21 CFR 882.5890 ## III. Predicate Device | Predicate Device | Manufacturer | 510(k) Number | Product code | | --- | --- | --- | --- | | BEAGANK 4T PLUS | Belega Inc. | K233010 | NFO | | Microcurrent Facial Device Models: CEC101, EE0101, EEI101 | Shenzhen Dachi communication Co.,Ltd. | K244004 | NFO | ## IV. Device Description Facial &amp; Body Beauty Device is portable, non-sterile and reusable device, which is designed to achieve the aesthetic effect. The device mainly consists of a main unit and charging cable, and it is supplied by internal rechargeable lithium battery, which can be recharged by external charger through the provided charging cable, but the device can not be used when charging. To use the device, user should place the electrode head on the face and body. The device will automatically shut down after treatment time is over. {5} Shenzhen Jianchao Intelligent Technology Co., LTD 510(k)s – 510(k) Summary ## V. Indications for Use Facial &amp; Body Beauty Device is a hand-held device for over-the counter aesthetic purposes. The Microcurrent stimulation mode is indicated for facial skin stimulation and body skin stimulation. ## VI. Comparison of Technological Characteristics With the Predicate Device The Facial &amp; Body Beauty Device has the same intended use as the predicate devices. The technological characteristics, features, specifications, design and intended use are similar to the predicate devices. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate devices for its intended use. Page 2 of 9 {6} Shenzhen Jianchao Intelligent Technology Co., LTD 510(k)s - 510(k) Summary | Comparison Elements | Subject Device | | Primary predicate device | Secondary predicate device | Remark | | --- | --- | --- | --- | --- | --- | | K number | K252553 | | K233010 | K244004 | / | | Trade name/ Model | Facial & Body Beauty Device, Model: INIA-BD001, INIA-BD002, INIA-BD003, INIA-BD004, F2C15, F3222 Pro, F5516, F3606, F2210 Pro, F5808 | Facial & Body Beauty Device, Model: INIA-ED001, INIA-ED002, INIA-BLD001, E1507 | BEAGANK 4T PLUS | Microcurrent Facial Device Models: CEC101, EE0101, EEI101 | / | | Device class | Class II | Class II | Class II | Class II | Same | | Product code | NFO | NFO | NFO | NFO | Same | | Indication for use/Intended use | Facial & Body Beauty Device is a hand-held device for over-the counter aesthetic purposes. The Microcurrent stimulation mode is indicated for facial skin stimulation and body skin stimulation. | Facial & Body Beauty Device is a hand-held device for over-the counter aesthetic purposes. The Microcurrent stimulation mode is indicated for facial skin stimulation and body skin stimulation. | The BEAGANK 4T PLUS is a handheld portable device for over-the-counter aesthetic use including facial and neck stimulation or body skin stimulation. | CEC101 and EE0101: Microcurrent Facial Device is intended for facial stimulation for over-the-counter aesthetic use. EEI101: Microcurrent Facial Device is intended for facial, neck and body skin stimulation for over-the-counter aesthetic use. | Same | | OTC or prescription | OTC | OTC | OTC | OTC | Same | | Treatment site | Face and body | Face and body | Face, neck and body | EEI101: Facial, neck and body skin CEC101: Facial skin EE0101: Facial skin | Same | | Power Supply | Internal Li-ion battery: 3.7Vd.c. 1100mAh | Internal Li-ion battery: 3.7Vd.c. 300mAh | Internal rechargeable Lithium-ion battery | Input:5V 1A; Output:3.7V CEC101: 3.7V/2600mAh EE0101: 3.7V/800mAh EEI101: 3.7V/2600mAh | Different Note 1 | Page 3 of 9 {7} Shenzhen Jianchao Intelligent Technology Co., LTD 510(k)s – 510(k) Summary | Comparison Elements | | Subject Device | | Primary predicate device | Secondary predicate device | Remark | | --- | --- | --- | --- | --- | --- | --- | | K number | | K252553 | | K233010 | K244004 | / | | Handheld design | | Yes | Yes | Yes | Yes | Same | | Software/Firmware/Microprocessor Control? | | Yes | Yes | Yes | Yes | Same | | Average DC current through electrodes when device is on but no pulses are being applied | | 0A (<1μA) | 0A (<1μA) | <1μA | <1μA | Same | | Output channels | | 1 output channel | 1 output channel | 1 output channel | 1 output channel | Same | | Regulated Current or Regulated Voltage? | | Regulated voltage | Regulated voltage | Voltage Control | Voltage Control | Same | | Automatic Overload Trip? | | No | No | Not required due to circuit design | No | Same | | Automatic Shut Off | | Yes | Yes | Yes | Yes | Same | | User Override Control? | | Yes | Yes | Yes | Yes | Same | | Indicator | On/Off Status | Yes | Yes | Not publicly available | Yes | Same | | | Low Battery | Yes | Yes | Yes | Yes | Same | | | Voltage/Current Level | Yes | Yes | Not publicly available | Yes | Same | | Waveform Type | | Bi-phase square-wave pulse | Bi-phase square-wave pulse | Rectangle, biphasic asymmetric | Bi-phase square-wave pulse | Same | Page 4 of 9 {8} Shenzhen Jianchao Intelligent Technology Co., LTD 510(k)s – 510(k) Summary | Comparison Elements | Subject Device | | Primary predicate device | Secondary predicate device | Remark | | --- | --- | --- | --- | --- | --- | | K number | K252553 | | K233010 | K244004 | / | | Maximum Output Voltage | Mode 1: 16V@500Ω; 21V@2KΩ; 22.5V@10KΩ Mode 2: 15V@500Ω; 22V@2KΩ; 23V@10KΩ | Mode 1: 16V@500Ω; 18V@2KΩ; 22.5V@10KΩ Mode 2: 16V@500Ω; 18V@2KΩ; 20V@10KΩ | Mode 1: 0.12V@ 500 Ω 0.51V@ 2K Ω 1.83V@ 10K Ω Mode 2: 0.11V@ 500 Ω 0.50V@ 2K Ω 1.87V@ 10K Ω Mode 3: 16.6V@ 500 Ω 21.0V@ 2K Ω 22.5V@ 10K Ω Mode 4: 15.3V@ 500 Ω 21.2V@ 2K Ω 22.9V@ 10K Ω | CEC101: 3.77 V±10% @ 500 Ω 11.85 V±10% @ 2K Ω 33.2 V±10% @ 10K Ω EE0101: 3.57 V±10%@ 500 Ω 11.25 V±10% @ 2K Ω 31.25 V±10% @ 10K Ω EEI101: 3.54 V±10%@ 500 Ω 12.3 V±10% @ 2K Ω 34 V±10% @ 10K Ω | Similar Note 2 | | Maximum Output Current | Mode 1: 32mA@500Ω; 10.5mA@2KΩ; 2.25mA@10KΩ Mode 2: 30mA@500Ω; 11mA@2KΩ; 2.3mA@10KΩ | Mode 1: 32mA@500Ω; 9mA@2KΩ; 2.25mA@10KΩ Mode 2: 32mA@500Ω; 9mA@2KΩ; 2mA@10KΩ | Mode 1: 0.19mA@ 500 Ω 0.18mA@ 2K Ω 0.16mA@ 10K Ω Mode 2: 0.45mA@ 500 Ω 0.39mA@ 2K Ω 0.24mA@ 10K Ω Mode 3: 30.8mA@ 500 Ω 8.52mA@ 2K Ω 2.26mA@ 10K Ω Mode 4: 31.4mA@ 500 Ω | CEC101: 7.54 mA±10% @ 500 Ω 5.93 mA±10% @ 2K Ω 3.32 mA±10% @ 10K Ω EE0101: 7.14 mA±10%@ 500 Ω 5.63 mA±10%@ 2K Ω 3.13 mA±10%@ 10K Ω EEI101: 7.08 mA±10%@ 500 Ω 6.15 mA±10%@ 2K Ω 3.4 mA±10%@ 10K Ω | Similar Note 2 | Page 5 of 9 {9} Shenzhen Jianchao Intelligent Technology Co., LTD 510(k)s – 510(k) Summary | Comparison Elements | Subject Device | | Primary predicate device | Secondary predicate device | Remark | | --- | --- | --- | --- | --- | --- | | K number | K252553 | | K233010 | K244004 | / | | | | | 8.60mA@ 2K Ω 2.15mA@ 10K Ω | | | | Pulse Period (Pulse Width) | Mode 1: P:84μs, N:84μs Mode 2: P:112μs, N:76μs | 80μs | Mode 1: 130μs Mode 2: 130μs Mode 3: 80μs Mode 4: 92μs | CEC101: 174μs EE0101: 174μs EEI101: 145μs | Different **Note 2** | | Output Frequency (Hz) | Mode 1: 3.6kHz Mode 2: 1.6kHz | Mode 1: 3.4kHz Mode 2: 1.4kHz | Mode 1: 3.80kHz Mode 2: 3.80kHz Mode 3: 3.65kHz Mode 4: 1.64kHz | 40Hz, 97.8Hz | Similar **Note 2** | | Net Charge | 0 μC @ 500Ω | 0 μC @ 500Ω | Mode 1: 0.025μC@500Ω Mode 2: 0.025μC@500Ω Mode 3: -1.3μC @ 500Ω Mode 4: 0μC @ 500Ω | 0 μC @ 500Ω | Same | | Maximum current density | Mode 1: 0.099mA/cm² Mode 2: 0.11mA/cm² | Mode 1: 0.20mA/cm² Mode 2: 0.18mA/cm² | Mode 1: 0.15~1.36 mA/cm²@ 500Ω Mode 2: 0.10~3.21 mA/cm²@500Ω Mode 3: 4.26~220 mA/cm²@500Ω Mode 4: 3.13~224.3 mA/cm²@500Ω | CEC101: 0.09mA/cm²@500Ω EE0101: 0.45mA/cm²@500Ω EEI101: 0.15mA/cm²@500Ω | Similar **Note 2** | | Maximum Power Density | Mode 1: 0.026mW/cm² Mode 2: 0.031mW/cm² | Mode 1: 0.026mW/cm² Mode 2: 0.024mW/cm² | Mode 1: 8.51~11.39 μW/cm²@500Ω Mode 2: 17.46~33.87 μW/cm²@500Ω Mode 3: 1.124~106.6 mW/cm²@500Ω Mode 4: 0.56~51.78 mW/cm²@500Ω | CEC101: 0.045mW/cm²@500Ω EE0101: 0.21mW/cm²@500Ω EEI101: 0.065mW/cm²@500Ω | Similar **Note 2** | Page 6 of 9 {10} Shenzhen Jianchao Intelligent Technology Co., LTD 510(k)s – 510(k) Summary | Comparison Elements | Subject Device | | Primary predicate device | Secondary predicate device | Remark | | --- | --- | --- | --- | --- | --- | | K number | K252553 | | K233010 | K244004 | / | | Treatment recommendation | 10 minutes | 10 minutes | 5 minutes | CEC101: 3-8 minutes EE0101: 5-8 minutes EEI101: 5-8 minutes | Similar Note 3 | | Main Materials | PC, PC+ABS, Zinc alloy | PC, ABS, Stainless Steel, Zinc alloy, Silica gel | ABS | ABS, Stainless steel | Different, but solved by biocompatibility tests | | Dimension | 52(L)×48(W)×188(H)mm | 164.5(L)×31(W)×17.3(H) mm | Not publicly available | CEC101:193.68*68*60mm EE0101: 90*65*39mm EEI101: 202.6*69.32*62mm | Different Note 4 | | Net Weight | 220g | 95.3g | 136g | CEC101: 266g EE0101: 120g EEI101: 230g | Different Note 4 | Comparison in Detail(s): Note 1: Although the “Power source” is different from the primary predicate device, they are both powered by the internal lithium battery. The lithium battery of the subject device complies with the IEC 62133-2 standard, and the device complies with IEC 60601-1 and IEC 60601-1-2 requirements, so this difference will not raise any safety or effectiveness issue. Note 2: - The “Maximum Output Voltage” and “Maximum Output Current” of the subject device (both model INIA-BD001 and INIA-ED001) are very close to that of the mode 3 and mode 4 of the primary predicate device, thought there is still a little difference, the setting of the subject device is within the maximum and minimum range of predicate devices; - The “Pulse Width” and “Output frequency” of subject device (both model INIA-BD001 and INIA-ED001) are similar with the settings of mode 3 and mode 4 of the primary predicate device, thought there is still a little difference, the setting of the subject device is within the {11} Shenzhen Jianchao Intelligent Technology Co., LTD 510(k)s – 510(k) Summary maximum and minimum range of predicate devices; &gt; Although the “Maximum Current Density” and “Maximum Power Density” of subject device are little different from the primary predicate device, there are in the maximum and minimum range of predicate devices; We also have tested the product waveform parameters according to FDA guidance and tested the device according to the requirements of IEC 60601-2-10, the tests are all passed, so these differences do not raise safety and effectiveness problems. **Note 3:** The treatment recommendation of time of the subject device is 10 minutes, it is longer than that of the primary predicate device, but is very close to the predicate device 1, which is 8 minutes. The subject device complies with IEC 60601-1 and IEC 60601-2-10 requirements, so this difference will not raise any safety or effectiveness issue. **Note 4:** Though the dimension and weight of the subject device are different from the predicate devices, this difference is insignificant and do not raise any safety or effectiveness problems. **VII. Non-Clinical Testing** The following performance data were provided in support of the substantial equivalence determination. **1) Electrical Safety and EMC Safety** - IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests - IEC 60601-1-11 Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment - IEC 60601-2-10 Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators - IEC 62133-2 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems **2) Biocompatibility Safety** Page 8 of 9 {12} Shenzhen Jianchao Intelligent Technology Co., LTD 510(k)s – 510(k) Summary - ISO 10993-5: 2009, Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity - ISO 10993-10: 2021, Biological evaluation of medical devices – Part 10: Tests for skin sensitization - ISO 10993-23: 2021, Biological evaluation of medical devices – Part 23: Tests for irritation ## VIII. Clinical Testing Not applicable. ## IX. Conclusions Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device is as safe, as effective, and performs as well as the legally marketed predicate devices. Page 9 of 9
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...