K252547 · Boston Scientific Corporation · LLZ · Oct 28, 2025 · Radiology
Device Facts
Record ID
K252547
Device Name
TheraSphere 360™ Y-90 Management Platform
Applicant
Boston Scientific Corporation
Product Code
LLZ · Radiology
Decision Date
Oct 28, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The TheraSphere 360™ Y-90 Management Platform includes Treatment Planning and Activity Calculation functionalities as optional interactive tools intended for calculating the activity of TheraSphere Microspheres required at treatment time based upon the desired dose, lung shunt fraction, anticipated residual waste, and liver mass. The Treatment Planning and Activity Calculation functionalities include features to aid in TheraSphere Microspheres dose vial selection. Additionally, the TheraSphere 360 Platform includes optional post-treatment analysis functionalities to be used following treatment with TheraSphere Microspheres. For post-TheraSphere Microspheres treatment, the TheraSphere 360 Platform should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose or for pre-treatment planning when there is a need for retreatment using TheraSphere Microspheres.
Device Story
TheraSphere 360 Y-90 Management Platform is a browser-based, end-to-end software platform supporting TheraSphere Microspheres radioembolization. Input data includes manual physician-entered parameters: desired dose, lung shunt fraction, anticipated residual waste, and liver mass. The device transforms these inputs using single-compartment MIRD schema dosimetry calculations to determine required microsphere activity and aid in dose vial selection. Used in clinical settings by authorized users; the platform provides output for treatment planning and retrospective post-treatment dose evaluation. It benefits patients by facilitating accurate dosimetry and management of Y-90 microsphere therapy. Unlike predicate devices that process digital medical images (CT, MRI, PET, etc.) for segmentation and fusion, TheraSphere 360 relies on manual data entry via a web portal, streamlining the workflow for authorized clinicians.
Clinical Evidence
No clinical data. Performance validated via bench testing, including full system and software verification (unit and integration testing), cybersecurity evaluation, and design validation to ensure user needs are met. Usability evaluation confirmed that medical device functions and algorithms meet requirements.
Technological Characteristics
Web-based, browser-accessible software platform. Dosimetry principle: single-compartment MIRD schema. Connectivity: online/cloud-hosted portal. Software: standalone medical device functionality for radioembolization activity calculation and vial selection.
Indications for Use
Indicated for use by trained medical professionals as an optional tool for calculating TheraSphere Microspheres activity for treatment planning and retrospective post-treatment dose determination. Not for prospective dose calculation or pre-treatment planning in cases requiring retreatment.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 28, 2025
Boston Scientific Corporation
Amy Mckinney
Fellow, Regulatory Affairs
300 Boston Scientific Way
Malborough, Massachusetts 01752
Re: K252547
Trade/Device Name: TheraSphere 360™ Y-90 Management Platform
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: August 12, 2025
Received: August 12, 2025
Dear Amy Mckinney:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252547 - Amy Mckinney
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252547 - Amy Mckinney
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lora D. Weidner, Ph.D.
Assistant Director
Radiation Therapy Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252547 | |
| Device Name TheraSphere 360™ Y-90 Management Platform | |
| Indications for Use (Describe) The TheraSphere 360™ Y-90 Management Platform includes Treatment Planning and Activity Calculation functionalities as optional interactive tools intended for calculating the activity of TheraSphere Microspheres required at treatment time based upon the desired dose, lung shunt fraction, anticipated residual waste, and liver mass. The Treatment Planning and Activity Calculation functionalities include features to aid in TheraSphere Microspheres dose vial selection. Additionally, the TheraSphere 360 Platform includes optional post-treatment analysis functionalities to be used following treatment with TheraSphere Microspheres. For post-TheraSphere Microspheres treatment, the TheraSphere 360 Platform should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose or for pre-treatment planning when there is a need for retreatment using TheraSphere Microspheres. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Page 1 of 5
K252547
510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Submitter: | Boston Scientific Corp. |
| --- | --- |
| Company Contact Person: | Ms. Amy E McKinney, MS, RAC
Fellow, Regulatory Affairs
Boston Scientific Corp
Phone: 651-287-5096
Email: Amy.mckinney@bsci.com |
| Alternate Contact Person: | Ms. Janet Bukovcan
Director, Regulatory Affairs
Boston Scientific Corp.
Phone: 613-801-1848
Email: Janet.bukovcan@bsci.com |
| Device Name: | TheraSphere 360™ Y-90 Management Platform |
| Device Classification Name: | Medical image management and processing system (LLZ)
21 CFR 892.2050 |
| Review Panel: | Radiology |
| Predicate Devices / Reference 510(k): | Simplicit^{90}Y (K182016) |
| Date of Preparation: | October 16, 2025 |
Device Description:
The TheraSphere 360 Y-90 Management Platform is an end-to-end, browser-based platform that will host a wide range of resources (e.g. radioembolization activity calculations, ordering, tracking, and education) that support Authorized Users of TheraSphere Microspheres.
The TheraSphere 360 Platform includes treatment planning functionality, activity calculation functionality, vial selection and ordering, and post-treatment analysis functionality. The treatment planning functionality and the activity calculation functionality include an interactive tool intended for calculating the activity of TheraSphere Microspheres required at the treatment time based upon the desired dose, lung shunt fraction, anticipated residual waste, and liver mass.
The Vial Selector function allows users to select and order TheraSphere Microspheres dose vials from inventory that match desired results.
The post-treatment analysis functionality is intended as an optional tool for post-treatment evaluation following TheraSphere Microspheres treatment.
{5}
K252547 - Page 2 of 5
# Indications For Use:
The TheraSphere 360™ Y-90 Management Platform includes Treatment Planning and Activity Calculation functionalities as optional interactive tools intended for calculating the activity of TheraSphere Microspheres required at treatment time based upon the desired dose, lung shunt fraction, anticipated residual waste, and liver mass.
The Treatment Planning and Activity Calculation functionalities include features to aid in TheraSphere Microspheres dose vial selection.
Additionally, the TheraSphere 360 Platform includes optional post-treatment analysis functionalities to be used following treatment with TheraSphere Microspheres. For post-TheraSphere Microspheres treatment, the TheraSphere 360 Platform should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose or for pre-treatment planning when there is a need for retreatment using TheraSphere Microspheres.
# Technological Comparison:
The technological characteristics of the TheraSphere 360 Platform medical device functions are similar to those of the predicate device and do not raise issues of safety or effectiveness. The key differences are related to how the software is accessed and how data is input into the device. TheraSphere 360 is accessed online via a web-based portal and uses manual physician data input, whereas the predicate device is installed on a user's computing platform and incorporates the use of digital medical images and tools to register and segment the images. TheraSphere 360 and the predicate have the same principles of operation, and both utilize single-compartment MIRD schema for the dosimetry calculations, in accordance with the TheraSphere Microspheres System labeling.
The table below provides a summary comparison of the submitted device compared to the predicate device:
| Characteristic | TheraSphere™ 360 Y-90 Management Platform Indications for Use | Predicate Device: Simplicit^{90}Y (K182016) Indications for Use | Comment |
| --- | --- | --- | --- |
| Indications for Use | The TheraSphere 360 Y-90 Management Platform includes Treatment Planning and Activity Calculation functionalities as optional interactive tools intended for calculating the activity of TheraSphere Microspheres required at treatment time based upon the desired dose, lung shunt fraction, anticipated residual waste, and liver mass.
The Treatment Planning and Activity Calculation functionalities include features to aid in TheraSphere Microspheres dose vial selection. | Simplicit^{90}Y is a standalone software device that is used by trained medical professionals as a tool to aid in evaluation and information management of digital medical images.
Simplicit^{90}Y supports the reading, rendering and display of a range of DICOM compliant imaging and related formats including but not limited to CT, PT, NM, SPECT, MR, SC, RTSS. Simplicit^{90}Y enables the saving of sessions in a proprietary format as well as the export of formats including CSV and PDF files.
Simplicit^{90}Y is indicated, as an accessory to TheraSphere, to | Similar.
The indications of the TheraSphere 360 Y-90 Management Platform are encompassed within the indications of Simplicit^{90}Y and represent a narrower scope. |
{6}
K252547 - Page 3 of 5
| Characteristic | TheraSphere™ 360 Y-90 Management Platform Indications for Use | Predicate Device: Simplicit^{90}Y (K182016) Indications for Use | Comment |
| --- | --- | --- | --- |
| | Additionally, the TheraSphere 360 Platform includes optional post-treatment analysis functionalities to be used following treatment with TheraSphere Microspheres. For post-TheraSphere Microspheres treatment, the TheraSphere 360 Platform should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose or for pre-treatment planning when there is a need for retreatment using TheraSphere Microspheres. | provide pre-treatment dosimetry planning support including Lung Shunt Fraction estimation (based on planar scintigraphy) and liver single-compartment MIRD schema dosimetry, in accordance with TheraSphere labelling.
Simplicit^{90}Y provides tools to create, transform, and modify contours/Regions of Interest for calculation of Lung Shunt Fraction and Perfused Volume. Simplicit^{90}Y includes features to aid in TheraSphere dose vial selection, dose vial ordering and creation of customizable reports.
Simplicit^{90}Y is indicated for post-treatment dosimetry and evaluation following Yttrium-90 (Y-90) microsphere treatment. Simplicit^{90}Y provides tools to create, transform, and modify contours/Regions of Interest for the user to define objects in medical image volumes to support TheraSphere post-Y-90 treatment calculation and evaluation. The objects include, but are not limited to, tumors and normal tissues, and liver volumes. Simplicit^{90}Y is indicated for registration, fusion display and review of medical images allowing medical professionals to incorporate images, such as CT, MRI, PET, CBCT and SPECT in TheraSphere Yttrium-90 (Y-90) microspheres pre-treatment planning and post-Y-90 treatment evaluation.
For post-Yttrium-90 (Y-90) treatment, Simplicit^{90}Y should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose or for the case where there is a need for retreatment using Y-90 microspheres. | |
| Software Access Method | Multi-function standalone software that is accessed online. | Standalone software that is installed on the user's computing platform. | Different.
Both the predicate and |
{7}
K252547 - Page 4 of 5
| Characteristic | TheraSphere™ 360 Y-90 Management Platform Indications for Use | Predicate Device: Simplicit^{90}Y (K182016) Indications for Use | Comment |
| --- | --- | --- | --- |
| | | | the proposed device are standalone software; however, the proposed device is available online and does not require installation. The access method of the device does not impact the substantial equivalence of the medical device functions of TheraSphere 360 to Simplicit^{90}Y. |
| Principles of Operation | Medical Internal Radiation Dose (MIRD) schema for dosimetry calculations | Medical Internal Radiation Dose (MIRD) schema for dosimetry calculations | Identical. |
| Device Functionalities | Contains medical device functionalities which utilize radioembolization activity calculations to determine the dose of TheraSphere Microspheres required at treatment time. | Contains medical device functionalities which utilize radioembolization activity calculations to determine the dose of TheraSphere Microspheres required at treatment time.
Additionally,
Simplicit^{90}Y is a medical image and information management system that is intended to receive, transmit, store, retrieve, display and process digital medical images, as well as create, display and print reports from those images.
Simplicit^{90}Y provides the user with the means to display, register and fuse medical images from multiple modalities.
Simplicit^{90}Y provides tools to create, transform, and modify contours for the user to define objects in medical image volumes for use in TheraSphere® pre-treatment dosimetry planning and for post-treatment dosimetry. The objects include, but are not limited to, tumors and normal tissues. | Similar.
TheraSphere 360 and Simplicit^{90}Y differ in that TheraSphere 360 contains a subset of the device functionality of Simplicit^{90}Y. |
{8}
K252547 – Page 5 of 5
# Summary of Performance Data
Performance testing confirmed that the TheraSphere 360 Platform performs safety and effectively and does not introduce any new or different safety risks. The TheraSphere 360 Platform medical device functions underwent full system and software verification testing, including unit and integration testing and cybersecurity evaluation, to ensure that the product meets the defined system and software requirements. A design validation evaluation was conducted to ensure that the platform meets defined user needs. Verification and validation testing, including a usability evaluation, successfully demonstrated that the medical device functions and algorithms meet requirements and user needs, confirming safety and effectiveness of TheraSphere 360 and substantial equivalence to the predicate device.
Clinical testing was not required to validate the TheraSphere 360 Y-90 Management Platform medical device functions.
# Conclusion:
In conclusion, the TheraSphere 360 Y-90 Management Platform medical device functions were evaluated against the predicate device and found to be substantially equivalent on the basis that 1) the indications for use of the TheraSphere 360 Platform consist of a narrower scope than those of the predicate device, and 2) the technology of the TheraSphere 360 Platform medical device functions is similar to the predicate device and the differences do not raise any new or different issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.