Simplicit90Y is a standalone software device that is used by trained medical professionals as a tool to aid in evaluation and information management of digital medical images. Simplicit90Y supports the reading and display of a range of DICOM compliant imaging and related formats including but not limited to CT, PT, NM, SPECT, MR, SC, RTSS. Simplicit90Y enables the saving of sessions in a proprietary format as well as the export of formats including CSV and PDF files. Simplicit90Y is indicated, as an accessory to TheraSphere®, to provide pre-treatment dosimetry planning support including Lung Shunt Fraction estimation (based on planar scintigraphy) and liver single-compartment MIRD schema dosimetry, in accordance with TheraSphere® labelling. Simplicit90Y provides tools to create, transform, and modify contours/Regions of Interest for calculation of Lung Shunt Fraction and Perfused Volume. Simplicit90Y includes features to aid in TheraSphere® dose vial selection, dose vial ordering and creation of customizable reports. Simplicit90Y is indicated for post-treatment dosimetry and evaluation following Yttrium-90 (Y-90) microsphere treatment. Simplicit90Y provides tools to create, transform, and modify contours/Regions of Interest for the user to define objects in medical image volumes to support TheraSphere® post-Y-90 treatment calculation. The objects include, but are not limited to, tumors and normal tissues, and liver volumes. Simplicit90Y is indicated for registration, fusion display and review of medical mages allowing medical professionals to incorporate images, such as CT, MRI, PET, CBCT and SPECT in TheraSphere® Yttrium-90 (Y-90) microspheres pretreatment planning and post-Y-90 treatment evaluation. For post-Yttrium-90 (Y-90) treatment, Simplicit90Y should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose or for the case where there is a need for retreatment using Y-90 microspheres.
Device Story
Standalone software for Windows; aids clinicians in TheraSphere® Y-90 microsphere dosimetry. Inputs: DICOM images (CT, PT, NM, SPECT, MR, SC, RTSS). Processing: manual/semi-automated/automated image registration (rigid/deformable) and segmentation of tumors/tissues; calculates Lung Shunt Fraction and MIRD-based dosimetry (single-compartment pre-treatment; multi-compartment/voxel-wise post-treatment). Outputs: dosimetry reports, dose volume histograms, isodose contour plans. Used by physicians/clinicians in clinical settings to support treatment planning and efficacy assessment. Benefits: standardized dosimetry workflow, improved accuracy in dose estimation, and quality assurance for Y-90 therapy.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation performed via functional and algorithmic testing to confirm compliance with user needs and intended use.
Indicated for trained medical professionals to perform pre-treatment dosimetry planning and post-treatment dosimetry evaluation for patients undergoing TheraSphere® Yttrium-90 (Y-90) microsphere treatment. Includes Lung Shunt Fraction estimation, liver single-compartment MIRD dosimetry, and post-treatment multi-compartment/voxel-wise MIRD dosimetry. Contraindicated for prospective dose calculation or retreatment planning in post-Y-90 treatment scenarios.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 6, 2018
Mirada Medical Ltd. % Ms. Sarah Bond VP Product Management Oxford Center for Innovation, New Road OXFORD, OX1 1BY OXFORDSHIRE GREAT BRITAIN
Re: K182016
Trade/Device Name: Simplicit90Y Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: LLZ Dated: October 22, 2018 Received: October 25, 2018
Dear Ms. Bond:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hse 2. Nils
for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182016
Device Name Simplicit90Y
#### Indications for Use (Describe)
Simplicit90Y is a standalone software device that is used by trained medical professionals as a tool to aid in evaluation and information management of digital medical images.
Simplicit90Y supports the reading and display of a range of DICOM compliant imaging and related formats including but not limited to CT, PT, NM, SPECT, MR, SC, RTSS. Simplicit90Y enables the saving of sessions in a proprietary format as well as the export of formats including CSV and PDF files.
Simplicit90Y is indicated, as an accessory to TheraSphere®, to provide pre-treatment dosimetry planning support including Lung Shunt Fraction estimation (based on planar scintigraphy) and liver single-compartment MIRD schema dosimetry, in accordance with TheraSphere® labelling. Simplicit90Y provides tools to create, transform, and modify contours/Regions of Interest for calculation of Lung Shunt Fraction and Perfused Volume. Simplicit90Y includes features to aid in TheraSphere® dose vial selection, dose vial ordering and creation of customizable reports.
Simplicit90Y is indicated for post-treatment dosimetry and evaluation following Yttrium-90 (Y-90) microsphere treatment. Simplicit90Y provides tools to create, transform, and modify contours/Regions of Interest for the user to define objects in medical image volumes to support TheraSphere® post-Y-90 treatment calculation. The objects include, but are not limited to, tumors and normal tissues, and liver volumes.
Simplicit90Y is indicated for registration, fusion display and review of medical mages allowing medical professionals to incorporate images, such as CT, MRI, PET, CBCT and SPECT in TheraSphere® Yttrium-90 (Y-90) microspheres pretreatment planning and post-Y-90 treatment evaluation.
For post-Yttrium-90 (Y-90) treatment, Simplicit90Y should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose or for the case where there is a need for retreatment using Y-90 microspheres.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Mirada, a company that specializes in imaging software. The logo is composed of the word "MIRADA" in a bold, sans-serif font, with the "M" in a darker blue and the "I" in a lighter blue. Below the word "MIRADA" is the tagline "The Imaging Software People" in a smaller, lighter blue font. The logo is simple and modern, and it effectively communicates the company's focus on imaging software.
# 510(k) Summary of Safety and Effectiveness
(The following information is in conformance with 21 CFR 807.92)
| Date of summary: | 21st July 2018 |
|--------------------------|-------------------------------------------------------------------------------------------|
| Submitter's name: | Mirada Medical Ltd |
| Submitter's address: | Oxford Centre for Innovation, New Road, Oxford.<br>Oxfordshire,<br>OX1 1BY United Kingdom |
| Submitter's contact: | Sarah Bond |
| Telephone number: | +44 (0)1865 261410 |
| Device Proprietary Name: | Simplicit90Y |
| Device Common Name(s): | Simplicit90Y |
| Classification Name: | Class II: Picture Archiving and Communications System |
| | (892.2050) Product Code: LLZ |
### Predicate Device
| 510(k) Number | Trade Name | Manufacturer |
|-------------------|---------------------|---------------------|
| K172218 | MIM – Y90 Dosimetry | MIM Software Inc. |
| Reference Devices | | |
| 510(k) Number | Trade Name | Manufacturer |
| K102687 | Mirada RT | Mirada Medical Ltd. |
| K130393 | Mirada RTx | Mirada Medical Ltd. |
| K101228 | Mirada XD | Mirada Medical Ltd. |
### Intended Use
Simplicit®Y is intended to be used by trained medical professionals for TheraSphere® pre-treatment dosimetry planning and post-treatment dosimetry evaluation following Y-90 treatment.
Simplicit®Y is a medical image and information management system that is intended to receive, transmit, store, retrieve, display and process digital medical images, as well as create, display and print reports from those images. The medical modalities of these medical imaging systems include, but are not limited to, CT, MRI, SPECT and PET.
Simplicit®Y provides the user with the means to display, register and fuse medical images from multiple modalities.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Mirada. The logo consists of the word "MIRADA" in a bold, sans-serif font. The "M" and the rest of the letters are in dark blue, while the "I" is in a lighter blue. Below the word "MIRADA" is the tagline "The Imaging Software People" in a smaller, lighter blue font.
Mirada Medical Ltd, Oxford Centre for Innovation, New Road, Oxford, OX1 1BY, UK Tel: +44 (0)1865 261410 Registered in England Company Number 6515901 VAT NO: 939 6451 81
Simplicit®Y provides tools to create, transform, and modify contours for the user to define objects in medical image volumes for use in TheraSphere® pre-treatment dosimetry planning and for posttreatment dosimetry. The objects include, but are not limited to, tumors and normal tissues.
For post-Yttrium-90 (Y-90) treatment, Simplicit®Y should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose or for the case where there is a need for retreatment using Y-90 microspheres.
## Device Description
Simplicit®Y is a software device which provides features and tools for use in pre-treatment dosimetry planning of TheraSphere® Y-90 microspheres treatment and post-treatment evaluation of Y-90 microspheres treatment and operates on Windows computer systems.
Simplicit®Y pre-treatment dosimetry planning features include Lung Shunt Fraction estimation (based on planar scintigraphy) and liver single-compartment MIRD schema in accordance with TheraSphere® labelling. Simplicit®Y additionally provides tools to support TheraSphere® dose vial selection, dose vial ordering and includes the creation of customizable reports.
After administration of Y-90 microspheres, Simplicit®V provides post-treatment dosimetry evaluation with multi-compartment and voxel-wise MIRD techniques applying the Local Deposition Model with scaling for known injected activity to assist the clinician in performing assessment of treatment efficacy and quality assurance, including assessment of absorbed dose to structures such as liver, lung and tumors.
Simplicit®Y provides 2D and 3D image display and advanced dosimetry analysis tools including isodose contour line plan and dose volume histograms.
Simplicit®9Y provides tools for multi-modal image fusion using rigid and deformable registration capable of manual, semi-automated and fully automated operation. Simplicit90Y includes evaluation tools for assessment of registration quality.
Simplicit®9Y provides semi-automated and automated tools for segmentation of Region/Volume Of Interest on multi-modal images.
Simplicit®Y supports the reading, rendering and display of a range of DICOM compliant imaging and related formats including but not limited to CT, PT, NM, SC, RTSS. Simplicit90Y enables the saving of sessions in a proprietary format as well as the export of RTSS, CSV and PDF files.
Simplicit®Y should not be used to change a treatment plan after treatment has been delivered with Yttrium-90 (Y-90) microsphere implants.
## Indications for Use
Simplicit®9Y is a standalone software device that is used by trained medical professionals as a tool to aid in evaluation and information management of digital medical images.
Simplicit®9Y supports the reading, rendering and display of a range of DICOM compliant imaging and related formats including but not limited to CT, PT, NM, SPECT, MR, SC, RTSS. Simplicit®Y enables the saving of sessions in a proprietary format as well as the export of formats including CSV and PDF files.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Mirada. The logo is in blue and consists of the word "MIRADA" in large, bold letters. Below the word "MIRADA" is the phrase "The Imaging Software People" in a smaller font.
Simplicit®Y is indicated, as an accessory to TheraSphere®, to provide pre-treatment dosimetry planning support including Lung Shunt Fraction estimation (based on planar scintigraphy) and liver single-compartment MIRD schema dosimetry, in accordance with TheraSphere® labelling. Simplicit®Y provides tools to create, transform, and modify contours/Regions of Interest for calculation of Lung Shunt Fraction and Perfused Volume. Simplicit®Y includes features to aid in TheraSphere® dose vial selection, dose vial ordering and creation of customizable reports.
Simplicit®Y is indicated for post-treatment dosimetry and evaluation following Yttrium-90 (Y-90) microsphere treatment. Simplicit90Y provides tools to create, transform, and modify contours/Regions of Interest for the user to define objects in medical image volumes to support TheraSphere® post-Y-90 treatment calculation and evaluation. The objects include, but are not limited to, tumors and normal tissues, and liver volumes.
Simplicit®Y is indicated for registration, fusion display and review of medical images allowing medical professionals to incorporate images, such as CT, MRI, PET, CBCT and SPECT in TheraSphere® Yttrium-90 (Y-90) microspheres pre-treatment planning and post-Y-90 treatment evaluation.
For post-Yttrium-90 (Y-90) treatment, Simplicit®ºY should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose or for the case where there is a need for retreatment using Y-90 microspheres.
## Substantial Equivalence
The proposed device was evaluated against the predicate device and found to be substantially equivalent on the basis that:
- The indications of the proposed device consist of a narrower scope than those of the predicate device.
- The technology of the proposed device is similar to the predicate device and the differences do not raise any new or different issues of safety and effectiveness.
Both devices provide similar tools with which to register and segment images for post-treatment dosimetry of Y-90 microspheres.
The key differences between the predicate device and the proposed device is that the proposed device is indicated for a narrower set of clinical applications than the predicate device as it is intended for Y-90 dosimetry only. The features and indications for the proposed device are substantially equivalent to the applicable features and indications of the predicate device.
Additionally, the proposed device is indicated for pre-treatment dosimetry for Y-90 as an accessory to TheraSphere®, where the predicate device is only intended for pre-treatment dosimetry for general purpose brachytherapy planning. Although the clinical context for pre-treatment dosimetry is different – the broad purpose of pre-treatment dosimetry is substantially the same.
The pre-treatment dosimetry features of the proposed device augment the use of TheraSphere® within the TheraSphere® labelling, perform in accordance with its intended use, and do not result in any new potential safety risks.
We conclude that the proposed device is substantially equivalent to the predicate device and performs as well as the applicable sub-set of the intended use of the predicate device and poses no unanswered questions with regard to safety and efficacy.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for Mirada, a company that specializes in imaging software. The logo features the company name in a bold, sans-serif font, with the first letter "M" in a lighter shade of blue. Below the company name is the tagline "The Imaging Software People" in a smaller font, also in blue. The logo is simple and modern, and it effectively communicates the company's focus on imaging software.
Mirada Medical Ltd, Oxford Centre for Innovation, New Road, Oxford, OX1 1BY, UK Tel: +44 (0)1865 261410 Registered in England Company Number 6515901 VAT NO: 939 6451 81
## Testing
Simplicit®Y is validated and verified against its user needs and intended use by the successful execution of planned performance, functional and algorithmic testing included in this submission. The results of performance, functional and algorithmic testing demonstrate that Simplicit®Y meets the user needs and requirements of the device, which are demonstrated to be substantially equivalent to those of the listed predicate device.
Verification and Validation for Simplicit®Y has been carried out in compliance with the requirements of CFR 21 Part 820 and in adherence to the DICOM standard.
## Conclusion
In conclusion, performance testing demonstrates that Simplicit®Y is substantially equivalent to, and performs at least as safely and effectively as the listed predicate device. Simplicit®0Y meets requirements for safety and effectiveness and does not introduce any new potential safety risks.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.