Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W
K252541 · Spectrum Medical S.R.L. · DTR · Mar 12, 2026 · Cardiovascular
Device Facts
Record ID
K252541
Device Name
Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W
Applicant
Spectrum Medical S.R.L.
Product Code
DTR · Cardiovascular
Decision Date
Mar 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4240
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration with a maximum flow rate of 3.0 l/min. Device is intended for pediatric patients (BSA according following characteristics: 0.6 m2 < BSA < 1 m2)
Device Story
Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W is a single-use, sterile heat exchanger for cardiopulmonary bypass (CPB) procedures; manages blood temperature via external heater-cooler systems using water or glycol-based fluids. Device inputs: circulating blood from CPB circuit; heat transfer fluid from external system. Device outputs: temperature-regulated blood returned to patient. Used in clinical settings (OR) by perfusionists; maintains normothermia or facilitates hypothermia/rewarming during surgery. Benefits: precise thermal control for pediatric patients (BSA 0.6-1.0 m2) during bypass. Healthcare providers monitor blood temperature and adjust heater-cooler settings based on surgical requirements; device ensures thermal stability throughout procedure.
Clinical Evidence
No clinical data. Bench testing only, including hemolysis, pressure drop, and heat transfer efficiency tests. Biocompatibility per ISO 10993-1, sterilization validation per ISO 11135, and packaging validation per ISO 11607-1.
Technological Characteristics
Materials: Polycarbonate, stainless steel, epoxy resin, phosphorylcholine coating. Principle: Heat exchange via stainless steel interface with water/glycol-based fluid. Form factor: 1/4" barbed connectors. Connectivity: Standalone (integrated into CPB circuit). Sterilization: EtO. Standards: ISO 7199, ISO 10993-1, ISO 11135, ISO 11607-1.
Indications for Use
Indicated for pediatric patients (0.6 m2 < BSA < 1 m2) undergoing cardiopulmonary bypass (CPB) procedures requiring extracorporeal circulation for up to 6 hours with a maximum blood flow rate of 3.0 l/min.
Regulatory Classification
Identification
A cardiopulmonary bypass heat exchanger is a device, consisting of a heat exchange system used in extracorporeal circulation to warm or cool the blood or perfusion fluid flowing through the device.
Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 (K223879)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 12, 2026
Spectrum Medical S.r.l.
Raffaella Tommasini
Spectrum Medical Group Regulatory and Quality Director
via di Mezzo, 23
Mirandola, Modena 41037
Italy
Re: K252541
Trade/Device Name: Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W
Regulation Number: 21 CFR 870.4240
Regulation Name: Cardiopulmonary Bypass Heat Exchanger
Regulatory Class: Class II
Product Code: DTR
Dated: January 28, 2026
Received: January 28, 2026
Dear Raffaella Tommasini:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252541 - Raffaella Tommasini
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252541 - Raffaella Tommasini
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Meaghan Erlewein -S
For Nicole Gillette
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W
Page 7 of 34
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252541 | ? |
| Please provide the device trade name(s). | | ? |
| Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W | | |
| Please provide your Indications for Use below. | | ? |
| Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration with a maximum flow rate of 3.0 l/min.
Device is intended for pediatric patients (BSA according following characteristics: 0.6 m2 < BSA < 1 m2) | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K252541
Spectrum Medical
Spectrum Medical S.r.l.
Via di Mezzo, 23
41037 Mirandola (MO)
Tel: +39 0535 1803050
FAX: +39 0535 1803051
www.spectrummedical.com
# 510(K) SUMMARY
## I. SUBMITTER
Submitter Name: Spectrum Medical S.r.l.
Submitter Address: Via di Mezzo, 23 41037 Mirandola (MO) Italy
Contact Person: Raffaella Tommasini, Spectrum Medical Group Regulatory and Quality Director
Phone: +39 0535 1803050
e-mail: raffaella.tommasini@spectrummedical.com
Fax: +39 0535 1803051
Date Summary Prepared: March 6th, 2026
## II. DEVICE
Proprietary Name: Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W
Common Name: Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W
Classification Name: 21 CFR 870.4240 Cardiopulmonary bypass heat exchanger
Regulatory Class: II
Product Code: DTR
Panel: Cardiovascular Devices, Office of Health Technology 2 (OHT2) / Division of Health Technology 2 B (Circulatory Support, Structural and Vascular Devices)
## III. PREDICATE DEVICES
### PRIMARY PREDICATE
Proprietary Name: AMG PMP Pediatric
Common Name: AMG PMP Pediatric
Classification Name: 21 CFR 870.4350 Cardiopulmonary Bypass Oxygenator
Regulatory Class: II
Product Code: DTZ
Panel: Cardiovascular Devices, Office of Health Technology 2 (OHT2) / Division of Health Technology 2 B (Circulatory Support, Structural and Vascular Devices)
510(k) Number: K202206
Spectrum Medical S.r.l. - Via di Mezzo, 23 - 41037 Mirandola (MO)
P.IVA/VAT 03663360364 - Email: info.ITA@spectrummedical.com
{5}
Spectrum Medical
Spectrum Medical S.r.l.
Via di Mezzo, 23
41037 Mirandola (MO)
Tel: +39 0535 1803050
FAX: +39 0535 1803051
www.spectrummedical.com
# PREDICATE/REFERENCE DEVICE
Proprietary Name: Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4
Common Name: Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W
Classification Name: 21 CFR 870.4240 Cardiopulmonary bypass heat exchanger
Regulatory Class: II
Product Code: DTR
Panel: Cardiovascular Devices, Office of Health Technology 2 (OHT2) / Division of Health Technology 2 B (Circulatory Support, Structural and Vascular Devices)
510(k) Number: K223879
# IV. DEVICE DESCRIPTION
Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W is designed to manage the temperature of blood during surgical procedures requiring cardiopulmonary bypass (CPB) for periods lasting less than 6 hours.
The device is designed to:
- keep circulating blood at a specific temperature, depending on the type of surgery being performed;
- maintain blood/patient thermoregulation during the CPB;
- rewarm blood at the conclusion of the CPB in order to restore normothermic patient condition.
Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W has been designed to be powered by heater-cooler systems that use
- Water, or
- Glycol-based solution
as Heat Transfer Fluid (HTF).
# V. INTENDED USE / INDICATIONS FOR USE
Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration with a maximum flow rate of 3.0 l/min.
Device is intended for pediatric patients (BSA according following characteristics: 0.6 m² < BSA < 1 m²).
Spectrum Medical S.r.l. - Via di Mezzo, 23 - 41037 Mirandola (MO)
P.IVA/VAT 03663360364 - Email: info.ITA@spectrummedical.com
{6}
Spectrum Medical
Spectrum Medical S.r.l.
Via di Mezzo, 23
41037 Mirandola (MO)
Tel: +39 0535 1803050
FAX: +39 0535 1803051
www.spectrummedical.com
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Device | Proposed Device – Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W | Predicate Device – AMG PMP Pediatric (K202206) | Reference Device - Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 (K223879) |
| --- | --- | --- | --- |
| Name | Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W | AMG PMP Pediatric | Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 |
| 510(k) Number | K252541 | K202206 | K223879 |
| Regulation # | 870.4240 | 870.4350 | 870.4240 |
| Regulation Name | Cardiopulmonary bypass heat exchanger | Cardiopulmonary Bypass Oxygenator | Cardiopulmonary bypass heat exchanger |
| Product Code | DTR | DTZ | DTR |
| Classification | II | II | II |
| Indication for Use | Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration with a maximum flow rate of 3.0 l/min. | The device is indicated for patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation for six hours or less with a maximum blood flow rate of 4 liters/minute. | The Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration. |
Spectrum Medical S.r.l. - Via di Mezzo, 23 - 41037 Mirandola (MO)
P.IVA/VAT 03663360364 - Email: info.ITA@spectrummedical.com
{7}
Spectrum Medical
Spectrum Medical S.r.l.
Via di Mezzo, 23
41037 Mirandola (MO)
Tel: +39 0535 1803050
FAX: +39 0535 1803051
www.spectrummedical.com
| Device | Proposed Device – Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W | Predicate Device – AMG PMP Pediatric (K202206) | Reference Device - Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 (K223879) |
| --- | --- | --- | --- |
| Target population | Pediatric patients (Body Surface Area according following characteristics: 0.6 m² < BSA < 1m²). | Pediatric / small adults. | Adults |
| Target User | Perfusionist. | Perfusionist. | Perfusionist. |
| Biocompatibility requirements | In compliance with ISO 10993 series. | In compliance with ISO 10993 series. | In compliance with ISO 10993 series. |
| Main Contacting Materials | - Polycarbonate;
- Stainless Steel;
- Epoxy resin;
- Phosphorylcholine coating. | For Heat Exchanger integrated in the Oxygenator:
- Polycarbonate;
- Stainless Steel;
- Polyurethane resin;
- Phosphorylcholine coating. | - Polycarbonate;
- Stainless Steel;
- Epoxy resin;
- Phosphorylcholine coating. |
| Type of conductive material | Stainless Steel | Stainless Steel | Stainless Steel |
| Cooling/heating fluid | Water or glycol-based solution | Water | Water or glycol-based solution |
| Priming Volume | 24 ± 2.5 ml | 190 ml (Oxygenator + Heat Exchanger) | 24 ± 2.5 ml |
| Max Flow Rate | 3.0 l/min | 4.0 l/min | 3.0 l/min |
| Inlet/Outlet Blood Connector | 1/4" barbed (6.35 mm)
1/4" barbed (6.35 mm) | 3/8" barbed connectors | 1/4" barbed (6.35 mm)
1/4" barbed (6.35 mm) |
Spectrum Medical S.r.l. - Via di Mezzo, 23 - 41037 Mirandola (MO)
P.IVA/VAT 03663360364 - Email: info.ITA@spectrummedical.com
{8}
Spectrum Medical
Spectrum Medical S.r.l.
Via di Mezzo, 23
41037 Mirandola (MO)
Tel: +39 0535 1803050
FAX: +39 0535 1803051
www.spectrummedical.com
| Device | Proposed Device – Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 W | Predicate Device – AMG PMP Pediatric (K202206) | Reference Device - Quantum PureFlow Standard Heat Exchanger Medium Flow 1/4 (K223879) |
| --- | --- | --- | --- |
| Inlet/Outlet Cooling/Heating Fluid Connector | CPC HFC profile | 1/2" Hansen quick couplings | CPC HFC profile |
| Single-use | Yes. | Yes. | Yes. |
| Sterile Condition | Sterile. | Sterile. | Sterile. |
| Sterilization Method | EtO sterilization | EtO sterilization | EtO sterilization |
Table 1 – Comparative Data with Predicate and Reference Devices
Spectrum Medical S.r.l. - Via di Mezzo, 23 - 41037 Mirandola (MO)
P.IVA/VAT 03663360364 – Email: info.ITA@spectrummedical.com
{9}
Spectrum Medical
Spectrum Medical S.r.l.
Via di Mezzo, 23
41037 Mirandola (MO)
Tel: +39 0535 1803050
FAX: +39 0535 1803051
www.spectrummedical.com
# VII. PERFORMANCE DATA
## NON-CLINICAL TESTING
As subject device is substantially equivalent to reference device (K223879) except for target population, all previously available data about non-clinical testing has been considered as still valid, in terms of:
- Performance evaluation according to ISO 7199:2016 [Recognition Nr. 3-150] (only for sections applicable to Heat Exchangers);
- Validation of the EtO Sterilization process, according to ISO 11135:2014 [Recognition Nr.: 14-529],
- Packaging Validation tests according to ISO 11607-1:2019 [Recognition Nr. 14-530];
- Biocompatibility of the finished product (worst case condition), according to International Standard ISO 10993-1:2018 [Recognition Nr. 2-258] and FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
The following tests for evaluation of devices' performances have been performed:
- Hemolysis test;
- Pressure drop test;
- Heat transfer efficiency.
In addition, updated data supporting claimed shelf-life have been included.
## Animal Study
No animal studies have been performed to support changes object of the present 510(k).
## CLINICAL TESTING
No clinical testing has been performed to support changes object of the present 510(k).
# VIII. CONCLUSIONS
Based on evaluation of indications for use and technological characteristics, the proposed device has been demonstrated to be appropriate for its intended use and is considered substantially equivalent to predicate and reference devices (K202206 and K223879 respectively).
Spectrum Medical S.r.l. - Via di Mezzo, 23 - 41037 Mirandola (MO)
P.IVA/VAT 03663360364 - Email: info.ITA@spectrummedical.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.