The device is indicated for patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation for six hours or less with a maximum blood flow rate of 4 liters/minute.
Device Story
Oxygenator for gas exchange and temperature control during open-heart surgery; utilizes PMP hollow fiber membrane and integrated heat exchanger. Blood flows through stainless steel tubes for temperature regulation via external water circulation; gas (O2-rich) flows through microporous PMP fibers. Used in clinical settings by perfusionists/surgeons. Provides extracorporeal circulation support; enables hypothermia/normothermia maintenance. Benefits patient by facilitating gas exchange and thermal management during bypass.
Clinical Evidence
Bench testing only. Verification and validation performed per ISO 7199:2016 and FDA guidance for cardiopulmonary bypass oxygenators. Testing included blood cell damage, gas transfer rates, pressure drop, pathway integrity (blood/gas/water), priming volume, residual volume, and connector tensile strength.
Indicated for pediatric and small adult patients undergoing cardiopulmonary bypass surgery requiring extracorporeal circulation for ≤6 hours at max blood flow of 4 L/min.
Regulatory Classification
Identification
A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”
{0}------------------------------------------------
April 1, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
Eurosets S.r.l. Katia Vescovini RA/QA/CQ Manager Strada Statale 12, nº 143 Medolla. Modena i-41036 Italy
Re: K202206
Trade/Device Name: AMG PMP Pediatric Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: March 1, 2021 Received: March 4, 2021
# Dear Katia Vescovini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K202206
Device Name AMG PMP PEDIATRIC
Indications for Use (Describe)
The device is indicated for patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation for six hours or less with a maximum blood flow rate of 4 liters/minute. PATIENT POPULATION: Pediatric / small adults.
Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
### 1. General Information
#### Submitter :
#### EUROSETS Sri
Strada Statale 12, n° 143 41036 Medolla (MO) — ITALY Tel.: +39 0535 660311 Fax +39 0535 51248
#### Establishment Registration Number: 3003752502
| Contact: | Dr. Katia Vescovini |
|----------|------------------------------------------|
| | Tel.: +39 0535 660311 |
| | Email: kvescoviniregulatory@eurosets.com |
| Summary Preparation Date: | March 31. 2021. |
|---------------------------|-----------------|
|---------------------------|-----------------|
#### 2. Name & Classification
Device Name: Regulation Name: Regulation Number: Product Code: CLASS:
AMG PMP PEDIATRIC Cardiopulmonary bypass oxygenator 870.4350 DTZ ll
# 3. Predicate Devices
The AMG PMP PEDIATRIC device is substantially equivalent to the following devices:
| Applicant | Device name | 510(k) Number |
|------------------------------|-----------------------------------|-----------------------------------|
| Sorin Group Italia<br>S.r.l. | EOS PMP, EOS PMP Integrated | K150489<br>Primary predicate Dev. |
| Eurosets | A.M.G. MODULE PMP NO T.P. STERILE | K141492 |
# 4. Indications for Use
The device is indicated for patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation for six hours or less with a maximum blood flow rate of 4 liters/minute. PATIENT POPULATION: Pediatric / small adults.
# 5. Device Description
AMG PMP PEDIATRC is an oxygenator used to exchange gases between blood and gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery. It is composed by PMP (Polymethylpentene) hollow fiber membrane with an integrated heat exchanger. It is provided together with the accessories that are the gas line and the convenience kit.
Oxygenator module, AMG PMP PEDIATRIC consist of three path, blood path, water path. These three paths, thanks to the particular configuration, allow the bloodtemperature control and gas exchange. The device center consists stainless steel tubes that allow the control of the temperature of the blood. The exchange of heat is between the water (with controlled temperature) that flows outside of steel tubes and the blood that flows inside the steel tubes. The device is surrounded by an outer compartment that contains a microporous membrane of Polymethy|pentene (PMP) consisting of capillary allow fibers that allows gas exchange. The air from the gas mixer is rich in O2 and follows the gas path. It enters through the gas inlet port on the top of the device, goes through microporous PMP fibers and exits from gas escape port, at the same time blood flows outside the microporous fibers. The design of heat exchangers for cooling and rewarming blood in the oxygenator utilizes a biologically inert surface to achieve the desired rate of heat exchange without producing any localized overheating of the blood.
{4}------------------------------------------------
# 6. Comparison with the predicate devices
| | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | SECONDARY PREDICATE DEVICE |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MODEL NAME: | AMG PMP PEDIATRIC | K150489<br>EOS PMP | K141492<br>A.M.G. MODULE PMP NO T.P. STERILE |
| APPLICANT: | Eurosets S.r.l. | Sorin Group Italia S.r.l. | Eurosets S.r.l. |
| CLASSIFICATION | | | |
| CLASS: | II | II | II |
| REGULATION<br>NUMBER: | 870.4350 | 870.4350 | 870.4350 |
| PRODUCT CODE | DTZ | DTZ | DTZ |
| REGULATION<br>NAME: | Cardiopulmonary Bypass Oxygenator | Cardiopulmonary Bypass Oxygenator | Cardiopulmonary Bypass Oxygenator |
| | INDICATIONS FOR USE & PATIENT POPULATION | | |
| | | The device is intended for use in | |
| INDICATIONS<br>FOR USE: | The device is indicated for patients<br>who undergo cardiopulmonary bypass<br>surgery requiring extracorporeal<br>circulation for six hours or less with a<br>maximum blood flow rate of 4<br>liters/minute. | patients who undergo<br>cardiopulmonary bypass surgery<br>requiring extracorporeal circulation<br>with a maximum blood flow rate 5<br>liters/minute. It provides oxygenation<br>and carbon dioxide removal from<br>venous blood. The integrated heat<br>exchanger provides blood<br>temperature control and allows the<br>use of hypothermia or aids in the<br>maintenance of normothermia during<br>surgery. The device is intended to be<br>used for 6 hours or less. | Is intended in surgical procedure<br>requiring extracorporeal gas exchange<br>support and blood temperature control<br>for periods of up to 6 hours.<br>The advanced Membrane Gas Exchange<br>for extracorporeal circulation is a<br>microporous hollow-fiber oxygenator<br>with an integral heat exchanger used to<br>perform cardiopulmonary bypass. It<br>includes a detachable 4.5 liter blood<br>reservoir. |
| PATIENT<br>POPULATION: | Pediatric / small adults | Pediatric / small adults | Adults |
| | | TECHNICAL FEATURES | |
| MIN<br>BLOOD<br>FLOW: | 0,5l/min | 0,5l/min | 1l/min |
| MAX<br>BLOOD<br>FLOW: | 4l/min | 5l/min | 7l/min |
| KIND OF FIBER<br>OXYGENATOR: | Polymethylpentene (PMP) | Polymethylpentene (PMP) | Polymethylpentene (PMP) |
| MEMBRANE<br>SURFACE AREA: | 1,35m² | 1,4m² | 1,81m² |
| HEAT<br>EXCAHNGER: | integrated | integrated | integrated |
| HEAT EXHANGE<br>SURFACE AREA: | 0,08m² | 0,15m² | 0,08m² |
| STATIC PRIMING<br>VOLUME: | 190ml | ≤200ml | 220ml |
| COATING: | Phosphorylcholine | Phosphorylcholine | Phosphorylcholine |
| Materials | Polycarbonate (PC)<br>Polyurethane resin<br>Stainless Steel<br>Silicone<br>Polyvinyl Chloride (PVC)<br>Polypropylene (PP)<br>High Density Polyethylene (HDPE)<br>Low Density Polyethylene (LDPE)<br>Thermoplastic Elastomer - Styrene-<br>Ethylene-Butylene-Styrene (SEBS)<br>Hydrophobic Acrylic Copolymer<br>Acrylonitrile-Butadiene-Styrene (ABS)<br>PolyTetraFluoroEthylene (PTFE) | Unknown | Polycarbonate (PC)<br>Polyurethane resin<br>Stainless Steel<br>Silicone<br>Polyvinyl Chloride (PVC)<br>Polypropylene (PP)<br>High Density Polyethylene (HDPE)<br>Low Density Polyethylene (LDPE)<br>Thermoplastic Elastomer - Styrene-<br>Ethylene-Butylene-Styrene (SEBS)<br>Hydrophobic Acrylic Copolymer<br>Acrylonitrile-Butadiene-Styrene (ABS)<br>PolyTetraFluoroEthylene (PTFE) |
| OXYGENATOR CONNECTIONS | | | |
| VENOUS INLET: | 3/8" (9,53mm) | 3/8" (9,53mm) | 3/8" (9,53mm) |
| ARTERIAL<br>OUTLET: | 3/8" (9,53mm) | 3/8" (9,53mm) | 3/8" (9,53mm) |
| GAS INLET: | 1/4" (6,35mm) | 1/4" (6,35mm) | 1/4" (6,35mm) |
| | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | SECONDARY PREDICATE DEVICE |
| MODEL NAME: | AMG PMP PEDIATRIC | K150489<br>EOS PMP | K141492<br>A.M.G. MODULE PMP NO T.P. STERILE |
| STANDARDS | | | |
| BIOLOGICAL<br>STANDARDS: | ISO 10993-1 Biological Evaluation of<br>Medical Devices Part 1: Evaluation and<br>Testing; | ISO 10993-1 Biological Evaluation of<br>Medical Devices Part 1: Evaluation and<br>Testing; | ISO 10993-1 Biological Evaluation of<br>Medical Devices Part 1: Evaluation and<br>Testing; |
| PRODUCT<br>SPECIFIC<br>STANDARDS: | Guidance for Cardiopulmonary Bypass<br>Oxygenators 510(k) Submissions; Final<br>Guidance for Industry and FDA Staff,<br>November 13, 2000;<br>ISO 7199:2016 Cardiovascular implants<br>and artificial organs - Blood-gas<br>exchangers (oxygenators); | Guidance for Cardiopulmonary Bypass<br>Oxygenators 510(k) Submissions; Final<br>Guidance for Industry and FDA Staff,<br>November 13, 2000 | Guidance for Cardiopulmonary Bypass<br>Oxygenators 510(k) Submissions; Final<br>Guidance for Industry and FDA Staff,<br>November 13, 2000; |
{5}------------------------------------------------
As can be seen from the table in the underlined parts, the Indications for Use of the AMG PMP PEDIA TRC and the predicate devices (K150489, K141492) are fundamentally the same with only some additional details provided for the predicate devices that do not affect the intended use of the subject device.
# 7. Performance Data
A program of design verification and validation testing was performed according to the standards "Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions; Final Guidance for Industry and FDA Staff, November 13, 2000"; and ISO 7199:2016 "Cardiovascular implants and artificial organs - Blood-gas exchangers (oxygenators)".
This design verification and validation testing and includes the following tests:
- Blood cell damage ●
- . Gas transfer rate & pressure drop
- . Blood pathway integrity
- . Heat exchanger fluid pathway integrity
- Gas pathway integrity
- Blood volume capacity (static priming volume)
- . Residual blood volume
- . Blood pathway connectors (tensile strength test)
- . Heat exchanger fluid pathway connectors
- . Heat exchanger performance evaluation
- Gas pathway connectors integrity (tensile strength test) .
# 8. Summary
Results of these performance tests allow to demonstrate that the subject device AMG PMP PEDIATRIC met the safety and performance requirements as per its indication for use and that AMG PMP PEDIATRIC subject device is substantially equivalent with the EOS PMP predicate devices, provingthat the AMG PMP PEDIATRIC is as safe, as effective, and performs as well as the EOS PMP.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.