Ryurei (1.5 mm) is indicated for: • Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. • Balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction. Ryurei (2.0–4.0 mm) is indicated for: • Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. • Balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction. • Balloon post-deployment expansion of balloon expandable stents (bare metal and drug-eluting).
Device Story
Ryurei is a percutaneous transluminal coronary angioplasty (PTCA) dilatation catheter; used by physicians in clinical settings to improve myocardial perfusion or treat acute myocardial infarction. Device consists of a double-lumen shaft with a distal inflatable balloon and hub for inflation/deflation. Operates manually; physician advances catheter over a guidewire to the target lesion; inflates balloon to specific diameter/pressure to dilate stenosis or expand stents. Radiopaque markers facilitate fluoroscopic positioning; hydrophilic coating reduces friction. Output is mechanical dilation of the vessel or stent. Benefits include restored blood flow and improved myocardial perfusion.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including dimensional verification, balloon rated burst pressure, fatigue testing (with and without stents), compliance, inflation/deflation time, bond strength, flexibility, torque, radiopacity, coating integrity, and particulate evaluation. Biocompatibility testing was performed per ISO 10993-1 for limited contact (<24 hours) blood-contacting devices.
Technological Characteristics
Double-lumen PTCA catheter; materials include Nylon 12-polytetramethylene glycol copolymer, Nylon 12, denatured polyethylene, stainless steel, dimethyl acrylamide-glycidyl methacrylate copolymer, platinum-iridium, polycarbonate, and urethane acrylate. Features hydrophilic coating, radiopaque markers, and depth markers. Sterilized via Ethylene Oxide (ISO 11135). Manual operation; no software or electronic components.
Indications for Use
Indicated for patients with coronary ischemia or acute myocardial infarction requiring balloon dilatation of stenotic coronary arteries or bypass grafts; 2.0–4.0 mm models also indicated for post-deployment expansion of balloon-expandable stents.
Regulatory Classification
Identification
Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.
Special Controls
*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.
Predicate Devices
Sapphire II PRO Balloon Dilatation Catheter (K180921)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 27, 2026
Terumo Corporation
Kirsten Steinmann
Senior Regulatory Affairs Specialist
Terumo Medical Corporation
265 Davidson Avenue
Suite 320
Somerset, New Jersey 08873
Re: K252295
Trade/Device Name: Ryurei
Regulation Number: 21 CFR 870.5100
Regulation Name: Percutaneous Transluminal Coronary Angioplasty (Ptca) Catheter
Regulatory Class: Class II
Product Code: LOX
Dated: July 23, 2025
Received: July 23, 2025
Dear Kirsten Steinmann:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252295 - Kirsten Steinmann
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252295 - Kirsten Steinmann
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jenny R. Katsnelson -S
Digitally signed by Jenny R. Katsnelson -S
Date: 2026.03.27 16:30:01
-04'00'
for Lydia Glaw
Assistant Director
DHT2C: Division of Coronary and Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252295 | |
| Device Name Ryurei | |
| Indications for Use (Describe) Ryurei (1.5 mm) is indicated for: • Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. • Balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction. Ryurei (2.0–4.0 mm) is indicated for: • Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. • Balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction. • Balloon post-deployment expansion of balloon expandable stents (bare metal and drug-eluting). | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K252295
# 510(k) Summary
## A. SUBMITTER INFORMATION (807.92(a)(1))
**Prepared by:** Kirsten Steinmann
Senior Regulatory Affairs Specialist
Terumo Medical Corporation
Tel. (908) 381-5634
**Prepared for:** *Owner/Operator*
Terumo Corporation
44-1, 2-Chome, Hatagaya
Shibuya-Ku, Tokyo
Japan 151-0072
Owner/Operator Number: 8010026
*Manufacturer and Sterilization Facility (Applicant)*
Ashitaka Factory of Terumo Corporation
150 Maimaigi-cho
Fujinomiya, Shizuoka 418-0015, Japan
Registration Number: 9681834
**Contact Person:** Kirsten Steinmann
Senior Regulatory Affairs Specialist
Terumo Medical Corporation
265 Davidson Ave, Suite 320
Somerset, NJ 08873
Tel. (908) 381-5634
E-mail: kirsten.steinmann@terumomedical.com
**Date prepared:** February 24, 2026
{5}
K252295
## B. DEVICE NAME (807.92(a)(2))
Proprietary Name: Ryurei
Percutaneous Transluminal Coronary Angioplasty (PTCA)
Common Name: Catheter
Catheters, transluminal coronary angioplasty, percutaneous
Classification Name: Cardiovascular
Classification Panel: 21 CFR 870.5100
Regulation: LOX
Product Code: Class II
Classification:
## C. PREDICATE DEVICE (807.92(a)(3))
The legally marketed device(s) to which substantial equivalence is/are claimed is:
Predicate Device:
K180921 Sapphire II PRO Balloon Dilatation Catheter, manufactured by OrbusNeich Medical Trading Inc.
Reference Devices:
- K220629 Emerge Monorail PTCA Dilatation Catheter, manufactured by Boston Scientific
- K141025 Artimes Balloon Dilatation Catheter, manufactured by BrosMed Medical
- K141236 NC Emerge PTCA Dilatation Catheter by Boston Scientific Corporation
- K163372 RX Takeru Balloon Dilatation Catheter by Kaneka Corporation
## D. REASON FOR 510(k) SUBMISSION
This traditional 510(k) for Ryurei - PTCA Dilatation Catheter (RX) is being submitted for a new device for the purposes of establishing substantial equivalence to a legally marketed predicate device. The differences lie in the specifications, additional sizes, and the indication for balloon post-deployment expansion of balloon expandable stents.
{6}
K252295
## E. INDICATIONS FOR USE (807.92(a)(5))
Ryurei (Ø1.5 mm) is indicated for:
- Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion.
- Balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction.
Ryurei (Ø2.0–4.0 mm) is indicated for:
- Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion.
- Balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction.
- Balloon post-deployment expansion of balloon expandable stents (bare metal and drug-eluting).
## F. SUBSTANTIAL EQUIVALENCE COMPARISON (807.92(a)(6))
Ryurei, the subject of this 510(k), is substantially equivalent to the Sapphire II Pro Balloon Dilatation Catheter manufactured by OrbusNeich Medical.
The Ryurei is substantially equivalent to the predicate device, given that:
- It has the same intended use.
- It uses the same operating principle.
- It incorporates the same basic design/construction.
- It is sterilized by the same method (Ethylene Oxide).
- It uses the same fundamental scientific technology.
- Performance (Bench), Biocompatibility and Packaging integrity test data establishes the substantial equivalency to the predicate device.
In addition to the above-listed predicate, Terumo has identified the reference device Emerge Monorail PTCA Dilatation Catheter for the indication of balloon post-deployment expansion of balloon expandable stents (bare metal and drug-eluting).
A comparison of the technological characteristics is summarized in Table 3 below.
{7}
K252295
Table 3: Summary of Comparative Information
| Device Characteristic | Subject Device: Ryurei | Predicate: Sapphire II PRO Balloon Dilatation Catheter | Reference: Emerge Monorail PTCA Dilatation Catheters | Reference: Artimes Balloon Dilatation Catheter | Reference: NC EMERGE PTCA DILATATION CATHETER | Reference: RX Takeru Balloon Dilatation Catheter |
| --- | --- | --- | --- | --- | --- | --- |
| Manufacturer | Terumo Corporation | OrbusNeich Medical | Boston Scientific | BrosMed Medical | Boston Scientific Corporation | Kaneka Corporation |
| Intended Use | For balloon dilatation in a coronary artery or bypass graft. | Same | Same | Same | Same | Same |
| Indications for Use | Ryurei (@1.50 mm) is indicated for:
• Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion.
• Balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction.
Ryurei (@2.00–4.00 mm) is indicated for:
• Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion.
• Balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction.
• Balloon post-deployment expansion of balloon expandable stents (bare metal and drug-eluting). | The Sapphire II Pro Balloon Dilatation Catheter (@1.0–1.25mm configurations) is indicated for:
Balloon pre-dilatation of a stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis) for the purpose of improving myocardial perfusion.
The Sapphire II Pro Balloon Dilatation Catheter (@1.5–4.0mm configurations) is indicated for:
Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion.
Balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction.
The Sapphire II Pro Balloon Dilatation is also indicated for:
Percutaneous transluminal angioplasty in the peripheral vasculature, including renal, femoral, infra-popliteal, tibial, and peroneal arteries. | The Emerge Monorail PTCA Dilatation Catheters (1.20 mm balloon models) are indicated as pre-dilatation catheters in the stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis).
The Emerge Monorail PTCA Dilatation Catheters (balloon models 1.50–5.00 mm) are indicated for the balloon catheter dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. | The Artimes Balloon Dilatation Catheter is indicated for:
Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion
Balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction | The NC Emerge OTW and NC Emerge MR PTCA Dilatation Catheters are indicated for the balloon catheter dilatation of the stenotic portion of a native coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion in patients with atherosclerosis.
The NC Emerge OTW and NC Emerge MR PTCA Dilatation Catheters (balloon models 2.00 mm–5.00 mm) are also indicated for the post-delivery expansion of balloon expandable stents (bare metal and drug-eluting). | The RX Takeru PTCA Balloon Dilatation Catheter is indicated for balloon dilatation of the stenotic portion in the coronary artery or bypass graft stenosis for the purpose of myocardial perfusion.
This product (balloon models 2.0–5.0 mm) is also indicated for the post-delivery expansion of balloon expandable stents. |
| Operation Principle | Manual | Manual | Manual | Manual | Same | Same |
| Design / Construction | The tip of the catheter is equipped with a balloon inflatable to a specific diameter and length at recommended pressures. | Same | Same | Same | Same | Same |
| | One or two radiopaque markers inside the balloon. | Same | Same | Same | Same | Same |
| | On the shaft, there are two depth markers. | Same | Same | Same | Same | Same |
| | Double-lumen structure, with a guidewire lumen and an inflation lumen. | Same | Same | Same | Same | Same |
{8}
K252295
| Device Characteristic | Subject Device: Ryurei | Predicate: Sapphire II PRO Balloon Dilatation Catheter | Reference: Emerge Monorail PTCA Dilatation Catheters | Reference: Artimes Balloon Dilatation Catheter | Reference: NC EMERGE PTCA DILATATION CATHETER | Reference: RX Takeru Balloon Dilatation Catheter |
| --- | --- | --- | --- | --- | --- | --- |
| | A hub is attached to the balloon for inflation/deflation. | Same | Same | Same | Same | Same |
| | The surface of catheter is coated with hydrophilic polymer coating. | Same | Same | Same | Same | Same |
| Materials | • Nylon 12-polytetramethylene glycol copolymer
• Nylon 12
• Denatured polyethylene
• Stainless steel
• Dimethyl acrylamide-glycidyl methacrylate copolymer
• Platinum-Iridium
• Polycarbonate
• Urethane acrylate resin
• Pigment | Information not publicly available. | Information not publicly available. | Information not publicly available. | Information not publicly available. | Information not publicly available. |
| | Presence of hydrophilic coating: Yes | Yes
Hydrophilic coating | Yes
Hydrophilic coating | Yes
Hydrophilic coating | Yes
Hydrophilic coating | Yes
Hydrophilic coating |
| Package Materials | • Individual package
-Individual package
Film: Polyethylene-polyester laminated film
Paper: Polyethylene
-Individual package Label: Paper
• Unit box
-Unit box: Coated cardboard
-Unit box Label: Paper
• Shipping carton
-Shipping carton: Corrugated cardboard
-Shipping carton Label: Paper | Information not publicly available. | Information not publicly available. | Information not publicly available. | Information not publicly available. | Information not publicly available. |
| Package | Individual package on which the product label and the peel-off labels are attached 1 unit per package. | Same | Same | Same | Same | Same |
| Specifications | Catheter effective length: 145 cm | Catheter effective length: 140 cm | Catheter effective length: 144 cm | Catheter effective length: 140 cm | Catheter effective length: 143 cm | Catheter effective length: 145 cm |
| | Balloon length: 5, 10, 15, 20, 30, 40 mm | Balloon length: 5, 8, 10, 12, 15, 20, 30 mm | Balloon length: 8, 12, 15, 20, 30, 40 mm | Balloon length: 5, 8, 10, 12, 15, 20, 25, 30 mm | Balloon length: 6, 8, 12, 15, 20, 30 mm | Balloon length: 6, 8, 12, 15, 20 mm |
| | Balloon diameter: 1.5, 2.0, 2.25, 2.75, 3.0, 3.25, 3.5, 3.75, 4.0 mm | Balloon diameter: 1.0, 1.25, 1.5, 1.75, 2.0, 2.25, 2.5, 2.75, 3.0, 3.25, 3.5, 4.0 mm | Balloon diameter: 1.2, 1.5, 2.0, 2.25, 2.5, 2.75, 3.0, 3.25, 3.5, 3.75, 4.0, 4.5, 5.0 mm | Balloon diameter: 1.0, 1.25, 1.5, 1.75, 2.0, 2.25, 2.5, 2.75, 3.0, 3.25, 3.5, 3.75, 4.0 mm | Balloon diameter: 2.0, 2.25, 2.5, 2.75, 3.0, 3.25, 3.5, 3.75, 4.0, 4.5, 5.0, 5.5, 6.0 mm | Balloon diameter: 1.5, 2.0, 2.25, 2.5, 2.75, 3.0, 3.25, 3.5, 3.75, 4.0 mm |
| | Nominal pressure: 6 atm (608 kPa) | Same | Same | Same | Nominal pressure: 12 atm (1215 kPa) | Same |
| | Rated burst pressure: #1.50–3.00 mm: 14 atm (1419 kPa)
#3.25–4.00 mm: 12 atm (1216 kPa) | Rated burst pressure: 14 atm (1419 kPa) | Rated burst pressure: #1.2 mm: 18 atm (1824 kPa)
#1.5–3.25 mm: 14 atm (1419 kPa)
#3.5–5.0 mm: 12 atm (1216 kPa) | Rated burst pressure: 14 atm (1419 kPa) | Rated burst pressure: #2.0–4.0 mm: 20 atm (2024 kPa)
#4.5–6.0 mm: 18 atm (1824 kPa) | Rated burst pressure: 14 atm (1419 kPa) |
{9}
K252295
| Device Characteristic | Subject Device: Ryurei | Predicate: Sapphire II PRO Balloon Dilatation Catheter | Reference: Emerge Monorail PTCA Dilatation Catheters | Reference: Artimes Balloon Dilatation Catheter | Reference: NC EMERGE PTCA DILATATION CATHETER | Reference: RX Takeru Balloon Dilatation Catheter |
| --- | --- | --- | --- | --- | --- | --- |
| | Radiopaque marker type: ø1.50 mm: Single marker
ø2.00–4.00 mm: Double marker | Radiopaque marker type: ø1.0–1.5 mm: Single marker
ø1.75–4.0 mm: Double marker | Radiopaque marker type: ø1.2–1.5 mm: Single marker
ø2.0–5.0 mm: Double marker | Radiopaque marker type: ø1.0–1.75 mm: Single marker
ø2.0–4.0 mm: Double marker | Radiopaque marker type: Double marker | Radiopaque marker type: ø1.5 mm: Single marker
ø2.0–5.0 mm: Double marker |
| *Sterilization* | Ethylene oxide | Same | Same | Same | Same | Same |
| *Shelf Life* | 36 months after sterilization | Information not publicly available. | 24 months after sterilization | Information not publicly available. | Information not publicly available. | 36 months after sterilization |
{10}
K252295
## G. NON CLINICAL TESTS (807.92(b)(1))
### Performance
Performance testing was conducted to ensure the operability and functionality of Ryurei throughout the shelf life, verify conformity to the applicable external and internal standards, and demonstrate substantial equivalence to the predicate device. Except for the Radio-detectability and Simulated Use tests, which were only performed on non-aged samples, the following performance tests were conducted on both non-aged and aged samples.
The below table provides a list of performance tests that were performed on Ryurei.
| Test Item |
| --- |
| Dimensional Verification |
| Balloon Rated Burst Pressure |
| Balloon Fatigue (Repeat Balloon Inflations) |
| Balloon Compliance (Diameter vs. Pressure) |
| Balloon Inflation and Deflation Time |
| Catheter Bond Strength |
| Tip Pull Test |
| Flexibility and Kink Test |
| Torque Strength |
| Radiopacity |
| Coating Integrity |
| Particulate Evaluation |
| Balloon Rated Burst Pressure (in Stent) |
| Balloon Fatigue (Repeat Balloon Inflations; in Stent) |
| Surface |
| Corrosion Resistance |
| Hub |
| Sliding resistance |
| Simulated use test
(Balloon Preparation, Deployment and Retraction) |
### Biocompatibility
In accordance with ISO 10993-1, the Ryurei is classified as: Externally Communicating Device, Circulating Blood, Limited Contact (<24 hours). The finished device's blood/body contacting parts were tested in accordance with the tests recommended in the FDA Guidance for Industry and Food and Drug Administration Staff - Use of International Standard ISO-10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." Screening tests were performed on aged devices to show that the biocompatibility is maintained throughout the shelf life of the product. Results of the testing
{11}
K252295
demonstrate that the device is biocompatible throughout the shelf life of the product.
| Non-aged, sterile, whole device |
| --- |
| Cytotoxicity |
| Sensitization |
| Intracutaneous Reaction |
| Systemic Toxicity (Acute) |
| Pyrogenicity |
| Hemolysis |
| Complement Activation |
| In vivo Thrombogenicity |
| Physicochemical |
| FT-IR |
| Aged (3 years), sterile, whole device |
| Cytotoxicity |
| Hemocompatibility |
| Physicochemical |
| FT-IR |
Results of the testing demonstrate that the device is biocompatible throughout the shelf life of the product.
## Sterilization
The sterility of the device is assured using a sterilization method validated in accordance with ISO 11135:2014/Amd 1:2018, Sterilization of Health Care Products – Ethylene Oxide – Requirements for the development, validation and routine control of a sterilization process for medical devices, to provide a Sterility Assurance Level (SAL) of $10^{-6}$.
## H. CLINICAL TESTS (807.92(b)(2))
This $510(\mathrm{k})$ does not include data from clinical tests.
## I. CONCLUSION (807.92(b)(3))
In summary, Ryurei, subject of this $510(\mathrm{k})$, is substantially equivalent in its intended use/, technology/principal of operation, materials, and performance to the Sapphire II Pro Balloon Dilatation Catheter manufactured by OrbusNeich Medical Trading Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.