K220629 · Boston Scientific · LOX · Mar 24, 2022 · Cardiovascular
Device Facts
Record ID
K220629
Device Name
Emerge Monorail PTCA Dilatation Catheter
Applicant
Boston Scientific
Product Code
LOX · Cardiovascular
Decision Date
Mar 24, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Emerge Over-The-Wire and Emerge Monorail PTCA Dilatation Catheters (1.20 mm balloon models) are indicated as pre-dilatation catheters in the stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis). The Emerge Over-The-Wire (balloon models 1.50-4.00 mm) and Emerge Monorail PTCA Dilatation Catheters (balloon models 1.50-5.00 mm) are indicated for the balloon catheter dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The Emerge Over-The-Wire (balloon models 2.00-4.00 mm) and Emerge Monorail PTCA Dilatation Catheters (balloon models 2.00-5.00 mm) are also indicated for the postdelivery expansion of balloon expandable stents (bare metal and drug-eluting).
Device Story
Emerge Monorail PTCA Dilatation Catheter is a sterile, single-use intravascular device; used by physicians in clinical settings to widen coronary artery blockages or dilate stents. Device consists of a shaft with a Pebax balloon at the distal tip. Operation involves inserting the catheter into the coronary artery/bypass graft; inflating the balloon to dilate the stenosis or stent; and deflating for removal. Output is mechanical dilation of the vessel or stent, improving myocardial perfusion. Benefits include restored blood flow and optimized stent apposition. This submission extends the existing balloon size matrix to include 4.50mm and 5.00mm diameters and 40mm lengths.
Clinical Evidence
Bench testing only. Performance testing included effective length, crossing profile, balloon preparation/deployment/retraction, withdrawal, fatigue, inflation/deflation, coating integrity, and particulate evaluation. Additional data leveraged from predicate/reference devices included shaft dimensions, rated burst pressure, compliance, bond strength, flexibility, torque, radiopacity, shelf life, and biocompatibility.
Technological Characteristics
Intravascular PTCA dilatation catheter; Pebax balloon; shaft construction. Sterilization method and materials are identical to predicate. Dimensions: 1.20-5.00 mm balloon diameters; 8-40 mm lengths. Connectivity: None (mechanical device).
Indications for Use
Indicated for patients with coronary artery or bypass graft stenosis (≥70% stenosis) requiring pre-dilatation, dilatation to improve myocardial perfusion, or post-delivery expansion of bare metal or drug-eluting stents.
Regulatory Classification
Identification
Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.
Special Controls
*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 24, 2022
Boston Scientific Corporation Ms. Aoife O'Flaherty Senior Regulatory Affairs Specialist Ballybrit Business Park Galway, Ireland
### Re: K220629
Trade/Device Name: Emerge Monorail PTCA Dilatation Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Regulatory Class: Class II Product Code: LOX Dated: March 3, 2022 Received: March 4, 2022
Dear Ms. O'Flaherty:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220629
Device Name Emerge Monorail PTCA Dilatation Catheter
#### Indications for Use (Describe)
The Emerge Over-The-Wire and Emerge Monorail PTCA Dilatation Catheters (1.20 mm balloon models) are indicated as pre-dilatation catheters in the stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis).
The Emerge Over-The-Wire (balloon models 1.50-4.00 mm) and Emerge Monorail PTCA Dilatation Catheters (balloon models 1.50-5.00 mm) are indicated for the balloon catheter dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.
The Emerge Over-The-Wire (balloon models 2.00-4.00 mm) and Emerge Monorail PTCA Dilatation Catheters (balloon models 2.00-5.00 mm) are also indicated for the postdelivery expandable stents (bare metal and drug-eluting)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary per 21 CFR §807.92
| Sponsor | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, Massachusetts 01752<br>USA |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Contact Name and<br>Information | Aoife O'Flaherty<br>Ballybrit Business Park,<br>Galway, Ireland.<br>Phone: 353-91-517299<br>Email: Aoife.OFlaherty@bsci.com |
| Date Prepared | March 3, 2022 |
| Proprietary Name | Emerge™ Monorail™ PTCA Dilatation Catheter |
| Common Name | PTCA Dilatation Catheter |
| Product Code | LOX |
| Classification | Class II, 21 CFR Part 870.5100 |
| Predicate Device | Emerge™ Monorail™ PTCA Dilatation Catheter, K163174,<br>cleared December 14, 2016 |
| Reference Device | Apex™ PTCA Dilatation Catheter, P860019/S208, approved<br>November 7, 2008 |
#### Device Description
The Emerge™ Monorail™ (MR) PTCA Dilatation Catheter is a sterile, single-use, intravascular medical devices used to widen coronary artery blockages and/or dilate coronary artery stents upon stent placement. The catheter consists of a shaft with a Pebax balloon near the distal tip and is available in a range of balloon sizes.
#### Indications for Use / Intended Use
The Emerge Over-The-Wire and Emerge Monorail PTCA Dilatation Catheters (1.20 mm balloon models) are indicated as pre-dilatation catheters in the stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis).
The Emerge Over-The-Wire (balloon models 1.50-4.00 mm) and Emerge Monorail PTCA Dilatation Catheters (balloon models 1.50-5.00 mm) are indicated for the balloon catheter dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.
The Emerge Over-The-Wire (balloon models 2.00-4.00 mm) and Emerge Monorail PTCA Dilatation Catheters (balloon models 2.00-5.00 mm) are also indicated for the postdelivery expansion of balloon expandable stents (bare metal and drug-eluting).
{4}------------------------------------------------
## Comparison of Technological Characteristics
The proposed Emerge™ Monorail™ PTCA Dilatation Catheter sizes are substantially equivalent to the predicate Emerge™ Monorail™ PTCA Dilatation Catheter, cleared by FDA under premarket notification K163174, December 14, 2016. The proposed 4.50mm and 5.00mm diameter (for 8 - 20mm lengths) and 40mm length (for 2.00, 2.50, 3.00, 3.50, and 4.00mm diameter) Emerge™ Monorail™ balloon sizes serve as an extension to the cleared Emerge™ Monorail™ matrix of balloons 1.20 - 4.00mm in diameter and 8 - 30mm in length.
The proposed and predicate devices share the following technological characteristics:
- . Fundamental device design and intended use
- . Device materials
- Packaging materials and design ●
- Sterilization method ●
- Fundamental manufacturing processes .
The proposed and predicate devices differ in the following technological characteristic:
- . Balloon lengths and diameters
{5}------------------------------------------------
### Non-clinical Performance Data
Bench testing was performed to support a determination of substantial equivalence. The results of these tests provide reasonable assurance that the proposed Emerge MR device has been designed and tested to assure conformance to the requirements for its intended use.
The following performance tests were completed on the proposed Emerge MR devices:
- Effective Length .
- Crossing Profile ●
- Balloon Preparation, Deployment and Retraction ●
- Withdrawal into a Guide Catheter ●
- Balloon Fatigue ●
- Balloon Inflation and Deflation ●
- Balloon Coating Integrity
- Particulate Evaluation .
The following tests were leveraged from the predicate/reference devices:
- Shaft Inner and Outer Diameter
- Balloon Rated Burst Pressure ●
- Balloon Rated Burst Pressure (in Stent) ●
- Balloon Fatique (select balloon sizes only) ●
- Balloon Fatigue (Repeat Balloon Inflations; in Stent) ●
- Balloon Compliance ●
- Catheter Bond Strength ●
- Tip Pull Test
- Flexibility and Kink Test ●
- Torque Strength
- Radiopacity .
- Shaft Coating Integrity ●
- Shelf Life
- Biocompatibility ●
- Sterilization
### Conclusion
Based on the indications for use, technological characteristics, and performance testing, the proposed Emerge™ Monorail™ PTCA Dilatation Catheter has been shown to be appropriate for its intended use and is considered to be substantially equivalent to Emerge™ Monorail™ PTCA Dilatation Catheter (K161374).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.