Sterile Hypodermic Syringe for Single Use (20ml, 30ml, 50ml, 60ml)
Applicant
Chirana T.Injecta,A.S.
Product Code
FMF · General Hospital
Decision Date
Apr 17, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Sterile Hypodermic Syringes for Single Use are intended to be used for medical purposes for injection or withdraw fluids from body.
Device Story
Sterile single-use piston syringe; consists of barrel, piston, and gasket. Manual operation by clinician; plunger movement within barrel facilitates fluid injection or withdrawal. Luer or Luer-Lock connector (EN ISO 80369-7:2021) allows attachment to needles or extension lines. Barrel features graduated scale for volume control. Used in clinical settings for fluid management. Benefits include sterile, single-use delivery of fluids and precise volume measurement.
Clinical Evidence
Bench testing only. Testing included air/liquid leakage, piston operation forces, seal strength, sterile barrier integrity, residual volume, nominal volume, Luer connector performance, and comprehensive biocompatibility (cytotoxicity, sensitization, irritation, reactivity, pyrogenicity, systemic toxicity, hemocompatibility) per ISO 10993 and ISO 7886-1 standards.
Technological Characteristics
Materials: Polypropylene barrel/piston, synthetic rubber gasket, medical grade silicone lubricant. Sizes: 20mL, 30mL, 50mL, 60mL. Connector: Luer/Luer-Lock per EN ISO 80369-7:2021. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Manual operation.
Indications for Use
Indicated for medical purposes to inject or withdraw fluids from the body in adult and pediatric populations. Prescription use only.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Sterile Hypodermic Syringes for Single Use (K213811)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 17, 2026
CHIRANA T. Injecta, A.S.
Rastislav Broska
Quality Manager
Nám.Dr.A.Schweitzera 194
Stará Turá, 916 01
Slovakia
Re: K252279
Trade/Device Name: Sterile Hypodermic Syringe for Single Use (20ml, 30ml, 50ml, 60ml)
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston Syringe
Regulatory Class: Class II
Product Code: FMF
Dated: February 20, 2026
Received: March 13, 2026
Dear Rastislav Broska:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252279 - Rastislav Broska
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252279 - Rastislav Broska
Page 3
Sincerely,
Shruti N. Mistry -S
Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{3}
Sterile Hypodermic Syringe for Single Use
Page 8 of 43
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252279 | ? |
| Please provide the device trade name(s). | | ? |
| Sterile Hypodermic Syringe for Single Use (20ml, 30ml, 50ml, 60ml) | | |
| Please provide your Indications for Use below. | | ? |
| Sterile Hypodermic Syringes for Single Use are intended to be used for medical purposes for
injection or withdraw fluids from body. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
1
# K252279 510(K) SUMMARY
| Submitter's Name: | CHIRANA T. Injecta, A.S. |
| --- | --- |
| Submitter's Address: | Nám. Dr. Schweitzera 194
916 01 Stará Turá
Slovak Republic |
| Submitter's Telephone: | 421-32-770-9972 |
| Contact Person: | Rastislav Broska
00421 910 955 937
rbroska@vitrexmg.com |
| Date Summary was Prepared: | April 17, 2026 |
| | |
| Trade or Proprietary Name: | Sterile Hypodermic Syringe for Single Use
(20ml, 30ml, 50ml, 60ml) |
| Common or Usual Name: | Piston Syringe |
| Classification: | Class II per 21 CFR §880.5860 |
| Product Code: | FMF |
| Classification Panel: | Division of Drug Delivery, General Hospital Devices, and Human Factors |
# DESCRIPTION OF THE DEVICES SUBJECT TO PREMARKET NOTIFICATION:
The Sterile Hypodermic Syringes for Single Use type consist of piston, gasket and barrel with printed scale which is graduated in milliliters. Material of barrel allows to control the position of the piston according to the scale. Syringes with Luer or Luer-Lock are manufactured according to EN ISO 80369-7:2021. The cone of all Luer-Lock syringes is centered on the barrel axis. The Luer-Lock connector ensures the connection of syringe with needle or extension line.
# INDICATIONS FOR USE
Sterile Hypodermic Syringes for Single Use are intended to be used for medical purposes for injection or withdraw fluids from body.
# TECHNOLOGICAL CHARACTERISTICS
## Materials:
The materials of the syringes include polypropylene (barrel, piston), latex free synthetic rubber gasket and medical grade silicone.
## Sizes:
The Sterile hypodermic syringe includes size of 20mL, 30mL, 50ml and 60 mL.
## Principles of Operation:
The Sterile Hypodermic Syringe by plunger movement in the barrel inject or withdraw fluids from body.
The subject and predicate device has nearly identical technological characteristics and the minor differences do not raise any new or different questions of safety and effectiveness when compared
{5}
to the predicate device. Specifically, the following characteristics are identical between the subject and predicate:
- Indications for Use
Sizes
Materials of manufacture
Structural support mechanism
The subject Sterile Hypodermic Syringes for Single Use are nearly identical to the previously cleared predicate device - Sterile Hypodermic Syringes for Single Use, 510(k) number K213811 in intended use, design, sizes, materials, technological characteristics, biocompatibility, and sterilization.
Table 3-1: Predicate Comparison Table
| Item | | Subject Device | Predicate Device K213811 | Remark |
| --- | --- | --- | --- | --- |
| Device Name | | Sterile Hypodermic Syringe for Single Use | Sterile Hypodermic Syringes for Single Use | |
| Product Code | | FMF | FMF | Same |
| Regulation No. | | 21 CFR 880.5860 | 21 CFR 880.5860 | Same |
| Class | | Class II | Class II | Same |
| Indications for Use | | Sterile hypodermic syringe for single use is intended to be used for medical purposes for injection or withdraw fluids from body. | The Sterile Hypodermic Syringes for Single Use are intended to be used for medical purposes on adult and pediatric population to inject fluids into or withdraw fluids from the body. | Different Comment 1 |
| Design | Design | barrel, plunger, piston | barrel, plunger, piston | Same |
| | Connector Types | Luer Lock/ Luer Slip | Luer Lock/ Luer Slip | Same |
| | Cone Placement | Luer Lock: centric Luer Slip (50 ml): eccentric | Luer Lock: centric Luer Slip: centric and eccentric (≥ 5 mL) | Different Comment 2 |
| | Sizes | 20ml, 30ml, 50ml, 60ml | 10ml, 20ml, 30ml, 50mL, 60ml | Different Comment 3 |
| Materials | Barrel | Polypropylene (PP) | Polypropylene (PP) | Same |
| | Plunger | Polypropylene (PP) | Polypropylene (PP) | Same |
| | Gasket | Polyisoprene rubber | Polyisoprene rubber | Same |
| | Lubricants | Polydimethylsiloxane | Polydimethylsiloxane | Same |
| Use | | Prescription use | Prescription use | Same |
| Single use | | Yes | Yes | Same |
| Syringe Performance | | Complied with ISO 7886-1:2017 | Complied with ISO 7886-1:2017 | Same |
| Biocompatibility | | Conforming to ISO 10993-1:2018 | Conforming to ISO 10993-1:2018 | Same |
| Shelf-life | | 5 years | 5 years | Same |
| Sterilization | | EO (ethylene gas) to SAL=10-6 | EO (ethylene gas) to SAL=10-6 | Same |
Comment 1.
The omission of explicit patient population in the subject device indications for use does not limit
{6}
or expand the intended use compared to the predicate device. The subject device is not intended for a different patient population, and no changes in design, materials, or performance are associated with this wording difference.
## Comment 2.
The cone placement of the subject device syringes is identical to that of the corresponding predicate device syringes of the same size. Specifically, both the subject and predicate Luer Lock syringes have a centric cone placement. For Luer Slip syringes, predicate devices above 5 mL (including 50 mL) have an eccentric cone placement, which is consistent with the subject device 50 mL syringe. These configurations are consistent with established design conventions for syringes of similar type and volume and do not introduce any new questions of safety or effectiveness when compared to the predicate device.
## Comment 3.
The subject device product range does not include a 10 mL syringe size, whereas the predicate device (K213811) includes a 10 mL configuration. This difference represents a reduced size offering rather than a modification of design, materials, or performance characteristics. The omission of the 10 mL size does not impact the technological characteristics, intended use, or fundamental operating principles of the subject device. Therefore, this difference does not raise new questions of safety and effectiveness when compared to the predicate device.
## PERFORMANCE DATA
The subject device, Sterile Hypodermic Syringes for Single Use, have been tested in the following test modes:
- Air leakage past syringe piston during aspiration (EN ISO 7886-1:2018, art.13.2, Annex. B)
- Liquid leakage at syringe piston under compression (EN ISO 7886-1:2018, art.13.2, Annex. D)
- Forces required to operate the piston (EN ISO 7886-1:2018, Annex. E)
- Seal strength paper - foil (EN 868-5:2018)
- Sterile barrier test (EN ISO 11607-2:2020)
- Residual (dead space) volume test (EN ISO 7886-1:2018, 13.1, Table 1, Annex C)
- Nominal volume test (EN ISO 7886-1:2018, Table No.1)
- Luer connector tests (ISO 80369-7:2021)
- Cytotoxicity (ISO 10993-5)
- Skin sensitization (ISO 10993-10)
{7}
- Skin Irritation (ISO 10993-10)
- Intracutaneous reactivity (ISO 10993-10)
- Material mediated pyrogenicity (ISO 10993-11/USP 151)
- Acute systemic toxicity (ISO 10993-11)
- Hemocompatibility (ISO 10993-4:2017)
The results of this non-clinical testing show that the performance of the subject Hypodermic Syringes is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
## CONCLUSION
The subject Sterile Hypodermic Syringe for Single Use is substantially equivalent to the predicate device in intended use, operating principle, technology, design, materials and performance.
4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.