K213811 · Berpu Medical Technology Co., Ltd. · FMF · May 12, 2022 · General Hospital
Device Facts
Record ID
K213811
Device Name
Sterile Hypodermic Syringes for Single Use
Applicant
Berpu Medical Technology Co., Ltd.
Product Code
FMF · General Hospital
Decision Date
May 12, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Sterile Hypodermic Syringes for Single Use are intended to be used for medical purposes on adult and pediatric population to inject fluids into or withdraw fluids from the body.
Device Story
Sterile hypodermic syringe for single use; consists of barrel, plunger, and piston. Available in various volumes (10ml-60ml) with luer slip or luer lock connectors for attachment to hypodermic needles. Operated manually by clinicians to inject or withdraw fluids. Device is sterile, single-use, and intended for medical purposes. Benefits include accurate fluid delivery or withdrawal. Biocompatible materials ensure safety during limited contact with blood path.
Clinical Evidence
Bench testing only. Conformance to ISO 7886-1:2017 (manual syringes) and ISO 80369-7:2016 (connectors). Biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, reactivity, systemic toxicity, hemocompatibility, pyrogenicity). Particulate matter testing per USP <788>. Sterile barrier integrity verified via ASTM F1886, F88, and F1929. Sterilization validated per ISO 11135. Shelf-life of 5 years supported by accelerated aging and stability studies.
Technological Characteristics
Barrel/plunger: Polypropylene (PP). Piston: Polyisoprene rubber. Manual operation. Luer slip/lock connectors. EO sterilization (SAL 10^-6). Biocompatibility per ISO 10993. Standards: ISO 7886-1:2017, ISO 80369-7:2016, ISO 10993-7:2008, USP <85>, USP <788>, ASTM F1886/F1886M-16, ASTM F88/F88-15, ASTM F1929-15.
Indications for Use
Indicated for adult and pediatric populations for the injection of fluids into or withdrawal of fluids from the body.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Sterile syringe for single use with/without needle (K210227)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 12, 2022
BERPU MEDICAL TECHNOLOGY CO., LTD Buxin Yu Management Representative No.14 Xingii Road, Yongxing Street Longwan District, 325000, Wenzhou, Zhejiang Province
Re: K213811
Trade/Device Name: Sterile Hypodermic Syringes For Single Use Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: Class II Product Code: FMF Dated: February 2, 2022 Received: April 4, 2022
Dear Buxin Yu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213811
Device Name
Sterile Hypodermic Syringe for Single Use
Indications for Use (Describe)
The Sterile Hypodermic Syringes for Single Use are intended to be used for medical purposes on adult and pediatric population to inject fluids into or withdraw fluids from the body.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K213811 510(k) Summary
### 1. Date of Preparation: May 7, 2022
### 2. Sponsor Identification
BERPU MEDICAL TECHNOLOGY CO., LTD NO.14 Xingji Road, Yongxing Street, Longwan District, 325000, Wenzhou, Zhejiang Province
Establishment Registration Number: 3004496829
Contact Person: Buxin Yu Position: Management Representative Tel: +86-577-86651999 Fax: +86-577-86630389 Email: zhc@berpu.com
### 3. Identification of Proposed Device
Trade Name: Sterile Hypodermic Syringe for Single Use
Regulatory Information Classification Name: Piston Syringe Classification: II Product Code: FMF Regulation Number: 21 CFR 880.5860 Review Panel: General Hospital
### 4. Identification of Predicate Devices
Predicate Device 510(k) Number: K210227 Product Name: Sterile syringe for single use with/without needle
### 5. Device Description
The Sterile Hypodermic Syringe for Single Use is intended for single use only, which consists of barrel, plunger and piston. The proposed device is available in a variety syringe volume. The syringe is available in luer slip and luer lock two connector types which are intended to be connected with a hypodermic needle.
{4}------------------------------------------------
| Characteristic | Predicate Device | Subject Device |
|------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterile syringe for single use<br>with/without needle<br>K210227 | Sterile Hypodermic Syringes for Single<br>Use<br>K213811 | |
| Indication for Use | The Sterile Syringes for Single Use<br>with/without needle are intended to be<br>used for medical purposes on adult<br>and inject inject fluids into or withdraw<br>fluids from the body. | The Sterile Hypodermic Syringes for<br>Single Use are intended to be used for<br>medical purposes on adult and pediatric<br>population to inject fluids into or<br>withdraw fluids from the body. |
### 6. Indication for Use
The indications for use of the subject device are slightly different from the predicate device. The difference is the intended population is included in the indications of subject device. This difference is does not affect safety and effectiveness of the proposed device.
### Substantially Equivalent (SE) Comparison 7.
Table 2 Comparison of Technology Characteristics of Sterile Hypodermic Syringes for Single Use
| Item | Proposed Device<br>K213811 | Predicate Device<br>K210227 | Remark |
|-------------------------------|------------------------------------------------------------|---------------------------------------------------------------|------------|
| Product name | Sterile Hypodermic Syringes<br>for Single Use | Sterile syringe for single use<br>with/without needle | - |
| Product Code | FMF | FMF | same |
| Regulation No. | 21 CFR 880.5860 | 21 CFR 880.5860 | Same |
| Class | CLASS II | CLASS II | same |
| Syringe volume | 10ml, 20ml, 30ml, 50mL, 60ml | 1ml, 2ml, 3ml, 5ml, 10ml,<br>20ml, 30ml, 50ml, 60ml,<br>100ml | Analysis 1 |
| Nozzle type | Luer slip; Luer lock | Luer slip; Luer lock | same |
| Configuration and<br>material | Barrel<br>Polypropylene<br>(PP) | Barrel<br>Polypropylene<br>(PP) | |
| | Plunger<br>Polypropylene<br>(PP) | Plunger<br>Polypropylene<br>(PP) | Analysis 2 |
| | Piston<br>Polyisoprene<br>Rubber | Piston<br>Silicone<br>Rubber | |
| Operation Mode | For manual use only | For manual use only | same |
| Syringe Performance | Complied with ISO<br>7886-1:2017 | Complied with ISO<br>7886-1:2017 | same |
| Biocompatibility | Comply with ISO 10993 series<br>standards, which includes: | Comply with ISO 10993 series<br>standards, which includes: | same |
{5}------------------------------------------------
| | ISO 10993-5: 2009; | ISO 10993-5: 2009; | |
|--------------------|----------------------------------|----------------------------------|------------|
| | ISO 10993-10: 2010 | ISO 10993-10: 2010 | |
| | ISO 10993-4: 2017 | ISO 10993-4: 2017 | |
| | ISO 10993-11: 2017 | ISO 10993-11: 2017 | |
| OTC use | No | No | same |
| Single for Use | Yes | Yes | same |
| Patient population | Adult and children | unknown | Analysis 3 |
| Shelf-life | 5 years | 3 years | Analysis 4 |
| Sterilization | EO Sterilization | EO Sterilization | same |
| SAL | 10-6 | 10-6 | same |
| Label/Labeling | Complied with 21 CFR part<br>801 | Complied with 21 CFR part<br>801 | same |
# Analysis 1- Syringe volume
The proposed has less specification of syringe volume compared to predicate device. However, the different syringe volume will be selected by physician per injection requirement and this difference does not affect intended use. In addition, the syringe volume for proposed device is covered by predicate device. The performance of the proposed devices has been performed on the final finished device. The test results shows passed the requirements of standard of ISO 7886-1. Therefore, this difference is not considered to affect the Substantially Equivalency (SE) between the proposed and predicate devices.
# Analysis 2-Configuration and material
The configurations for both proposed device and predicate device are similar, the difference is just in component material. However, the biocompatibility test result does not show any adverse effect which can demonstrate the safety of proposed device. Therefore, this difference is not considered to affect substantially equivalence.
Analysis 3-Paptient population
This difference does not alter suitability of the proposed device for its intended use
# Analysis 4-Shelf-life
The shelf life of the proposed device is 5 years, which is longer than the predicated device. The product performance after 5 years has been determined based on accelerated aging study. The packaging integrity test of proposed device was conducted on the proposed device. The results can demonstrate that the packaging was able to maintain sterility of the sterilized finished device during its shelf life of 5 years.
#### 8. Non-Clinical Test Performance Testing
The Sterile Hypodermic Syringe for Single Use described in this summary were tested and demonstrated to be in conformance with the following FDA recognized standards:
- . ISO 7886 -1:2017 Sterile hypodermic syringes for single use -- Part 1: Syringes for manual use
{6}------------------------------------------------
- . ISO 80369-7: 2016 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
# Biocompatibility Testing:
In accordance with ISO 10993-1, the syringe is classified as: Externally communicating Device. Blood Path Indirect, Limited Contact (<24hours). The proposed device was evaluated for the following tests. The results for the biocompatibility testing showed that there are no negative impacts from the materials that are used in the proposed device.
- . Cytotoxicity
- . Sensitization
- Intracutaneous Reactivity Test .
- Acute Systemic Toxicity .
- Hemocompatibility
- . Pyrogenicity
Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injections and met the USP acceptance criteria.
# Sterile Barrier Packaging Test
Sterile barrier packaging testing were performed on the proposed device, which include visual inspection (ASTM F1886/F1886M-16), seal strength (ASTM F88/F88-15) and dye penetration test (ASTM F1929-15). The test result showed that the device package can maintain its integrity.
# Sterilization and Shelf Life Test
The sterilization method has been validated per ISO 11135, which has thereby determined the routine control and monitoring parameters. The shelf life of the Sterile Hypodermic Syringe for Single Use is 5 years, determined based on stability studies which includes accelerated aging.
Sterilization and shelf life testing listed were performed on the proposed device.
| Item | Standard |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| EO residue | ISO 10993-7:2008 |
| ECH residue | ISO 10993-7:2008 |
| Bacteria Endotoxin Limit | USP <85> |
| Shelf Life Evaluation | Physical, Mechanical, Chemical, Package Tests were performed on<br>aging samples to verify the claimed shelf life of the device |
#### 9. Conclusion
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The Sterile Hypodermic Syringes for Single Use is substantially equivalent to the Sterile Syringe for Single Use with/without Needle with respect to the indications for use, target populations, treatment method, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.