K252189 · Dongguan Yanxi Technology Co., Ltd. · OHS · Sep 12, 2025 · General, Plastic Surgery
Device Facts
Record ID
K252189
Device Name
LED Light Therapy Mask (2311, 2344, 2457, 2458)
Applicant
Dongguan Yanxi Technology Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Sep 12, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Red light: Treatment of full-face wrinkles(suitable for model 2344 and model 2457). Amber light: Treatment of full-face wrinkles. Blue light: Treatment of mild to moderate inflammatory acne. Red+Infrared Light: Treatment of full-face wrinkles.
Device Story
LED Light Therapy Mask (models 2311, 2344, 2457, 2458) uses LEDs to emit red (630nm), blue (460nm), amber (605nm), and infrared (850nm) light for skin therapy. Device consists of mask with internal LEDs and external controller for mode, energy level, and time selection. User places mask over face; controller manages treatment duration. Integrated protective eye-shield prevents direct ocular exposure. Device automatically powers off after session. Intended for OTC home use. Output light energy targets wrinkles and acne; clinical benefit derived from photobiological effects on skin tissue.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness supported by biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-11, -2-83), EMC (IEC 60601-1-2), and photobiological eye safety (IEC 62471) testing.
Technological Characteristics
LED-based light therapy mask. Wavelengths: Red (630nm), Blue (460nm), Amber (605nm), Infrared (850nm). Power: DC 5V, lithium polymer batteries (1000mAh-4000mAh). Connectivity: Wired controller. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-83, IEC 62471, ISO 10993. Software: Embedded firmware for timing and mode control.
Indications for Use
Indicated for treatment of full-face wrinkles (Red, Amber, Red+Infrared light) and mild to moderate inflammatory acne (Blue light) in adults.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 12, 2025
Dongguan Yanxi Technology Co., Ltd.
% Bing Huang
Registration Engineer
Feiying Drug & Medical Consulting Technical Service Group
Rm 2401 Zhenye International Business Center, No. 3101-90
Qianhai Road
Shenzhen, Guangdong 518082
China
Re: K252189
Trade/Device Name: LED Light Therapy Mask (2311, 2344, 2457, 2458)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, OLP
Dated: July 14, 2025
Received: July 14, 2025
Dear Bing Huang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}
K252189 - Bing Huang
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-
{2}
K252189 - Bing Huang
Page 3
devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -
Date: 2025.09.12
10:23:36 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252189 | ? |
| Please provide the device trade name(s). | | ? |
| LED Light Therapy Mask (2311, 2344, 2457, 2458) | | |
| Please provide your Indications for Use below. | | ? |
| Red light: Treatment of full-face wrinkles(suitable for model 2344 and model 2457).
Amber light: Treatment of full-face wrinkles.
Blue light: Treatment of mild to moderate inflammatory acne.
Red+Infrared Light: Treatment of full-face wrinkles. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
Dongguan Yanxi Technology Co., LTD. 510(k)s -510(k) Summary
# 510(k) Summary
"510(k) Summary" as required by 21 CFR Part 807.92.
K252189
Date: 2025-09-03
## I. Submitter
Dongguan Yanxi Technology Co., LTD.
Room 501, No 19, Guanglong Road, Tianxin Village, Huangjiang Town, Dongguan City, Guangdong Province
Post code: 523763
Tel.: +86 18608323806
Lai xing liang
Title: Quality Supervisor
Tel.: +86 13144884788
Email: 342308911@qq.com
## II. Device
Name of Device: LED Light Therapy Mask
Model(s): 2311, 2344, 2457, 2458
Common or Usual Name: Light Based Over The Counter Wrinkle Reduction Over-The-Counter Powered Light Based For Acne
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: II
Product Code: OHS, OLP
Regulation Number: 21 CFR 878.4810
## III. Predicate Device & Reference Device
Predicate device:
| Manufacturer | Predicate Device | 510(k) Number | Cleared Date |
| --- | --- | --- | --- |
| Guangzhou Ahead
Intelligent Technology Co., Ltd. | Light Therapy System
(Model: M500) | K251042 | June 9, 2025 |
| Shenzhen SUNGPO HI-TECH Electronic Co.,Ltd | LED Facial Mask/ MZ-01,
NEWKEY-01, SP-FM-01 | K230351 | May 5, 2023 |
| Shenzhen Siken 3D
Technology Development Co., Ltd. | LED Light Therapy Mask/Model: SKB-1818P,SKB-1918,SKB-1918P,SKB-1918PLUS,SKB- | K243040 | December 20, 2024 |
{5}
Dongguan Yanxi Technology Co., LTD. 510(k)s -510(k) Summary
| | 2318L,SKB-2318P,SKB-2318PRO,IN-FM002,SKB-2418 | | |
| --- | --- | --- | --- |
Reference device:
| Manufacturer | Reference Device | 510(k) Number | Cleared Date |
| --- | --- | --- | --- |
| Dongguan Boyuan Intelligent Technology Co.,Ltd | LED Light Therapy Device (Models: KFB290, KFB291, KFB265, KFB293) | K241857 | October 11, 2024 |
| Galactic Beauty, LLC | MMSphereTM | K190443 | June 24, 2019 |
## IV. Device Description
The LED Light Therapy Mask adopts light emitting diodes (LED) in the red (630nm±5nm), blue (460nm±5nm), amber (605nm±5nm) and infrared (850nm±10nm) spectrum to irradiate on the face to realize its therapeutic effect. The LED Light Therapy Mask adopts the form of a mask that contains LEDs on the inner surface of the main unit. A controller is connected to the main unit to control the device, such as turn on/off the device, switch mode (LED color), select energy level and adjust time. To use the device, user should place the mask over the face and use the controller to operate. The device will automatically turn off after each treatment. To prevent irradiation of LED lights to eyes during the treatment, LED Light Therapy Mask has incorporated protective eye-shield which blocks light energy from LEDs.
## V. Indications for Use
Red light: Treatment of full-face wrinkles(suitable for model 2344 and model 2457).
Amber light: Treatment of full-face wrinkles.
Blue light: Treatment of mild to moderate inflammatory acne.
Red+Infrared Light: Treatment of full-face wrinkles.
## VI. Comparison of Technological Characteristics With the Predicate Device
LED Light Therapy Mask is compared with the following Predicate Devices and Reference Devices in terms of intended use, design, specifications, and performance:
{6}
Dongguan Yanxi Technology Co., LTD.
510(k)s -510(k) Summary
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Reference Device 1 | Reference Device 2 | Remark |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 510(k) Number | K252189 | K251042 | K230351 | K243040 | K241857 | K190443 | / |
| Trade name | LED Light Therapy Mask (Models:2311, 2344, 2457,2458) | Light Therapy System (Model: M500) | LED Facial Mask/ MZ-01, NEWKEY-01, SP-FM-01 | LED Light Therapy Mask/Model: SKB-1818P,SKB-1918,SKB-1918P,SKB-1918PLUS,SKB-2318L,SKB-2318P,SKB-2318PRO,IN-FM002,SKB-2418 | LED Light Therapy Device (Models: KFB290, KFB291, KFB265, KFB293) | MMSphere™ | / |
| Manufacturer | Dongguan Yanxi Technology Co.,Ltd | Guangzhou Ahead Intelligent Technology Co., Ltd. | Shenzhen SUNGPO HI-TECH Electronic Co.,Ltd | Shenzhen Siken 3D Technology Development Co., Ltd. | Dongguan Boyuan Intelligent Technology Co.,Ltd | Galactic Beauty, LLC | / |
| Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Product code | OHS, OLP | OHS, OLP | OHS, OLP | OHS, OLP | OHS, OLP, ILY | OHS, OLP | Same |
| Device classification | Class II | Class II | Class II | Class II | Class II | | Same |
| Indication for use/ Intended use | 2311 and model 2458: Red+Infrared Light: Treatment of full-face | Blue light: Treatment of mild to moderate inflammatory acne. | LED Facial Mask is an over the counter device that is intended to use | LED Light Therapy Mask is an over the counter device that is | KFB290, KFB291: Red light: Treatment of full-face wrinkles. | MMSphereTM Light Therapy Device emits energy in the | Same |
Page 3 of 9
{7}
Dongguan Yanxi Technology Co., LTD.
510(k)s -510(k) Summary
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Reference Device 1 | Reference Device 2 | Remark |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | wrinkles.
Amber light: Treatment of full-face wrinkles.
Blue light: Treatment of mild to moderate inflammatory acne.
2344 and model 2457: Red light: Treatment of full-face wrinkles.
Red+Infrared Light: Treatment of full-face wrinkles.
Amber light: Treatment of full-face wrinkles.
Blue light: Treatment of mild to moderate inflammatory acne. | Red+IR light: Treatment of facial wrinkles.
Amber light: Treatment of facial wrinkles.
Mixed light: Treatment of mild to moderate inflammatory acne. | LED light for the treatment of wrinkles and mild to moderate acne. | intended to use LED light for the treatment of wrinkles and mild to moderate acne. | Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm;
Relieving stiffness; Promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
Red+Infrared Light: Treatment of full-face wrinkles.
Amber light: Treatment of full-face wrinkles.
Blue light: Treatment of mild to moderate inflammatory acne.
Mixed light (Red+Blue+Infrared): Treatment of mild to moderate inflammatory acne. | red, blue and amber regions of the spectrum, specifically indicated to treat wrinkles and/or mild to moderate acne. The MMSphereTM is designed to be used for 20 minute treatments three to seven times per week. | |
| Location for use | Face | Face | Face | Face | Face | Face | Same |
Page 4 of 9
{8}
Dongguan Yanxi Technology Co., LTD.
510(k)s -510(k) Summary
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Reference Device 1 | Reference Device 2 | Remark |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Prescription or OTC | OTC | OTC | OTC | OTC | OTC | OTC | Same |
| Power supply | Input: DC 5V/1A
2311 and 2344:
Lithium polymer
Battery 2000mAh
2457:
Lithium polymer
Battery 4000mAh
2458:
Lithium polymer
Battery 1000mAh | External adapter
100 -240 V ~ 50 / 60 Hz | An external adapter
Input: AC 100-240V
50-60Hz 0.2A
Output: DC 12V
0.5A | Input: AC 100-240V 50-60Hz
0.3A
Output: DC 5V 1A
3.7V 650mAh
Li-ion Battery
3.7V1000mAh
Li-ion Battery | Input:100-240V, 50/60Hz
Output: 5V, 1A
Lithium ion battery: 1300mAh | Unknown | Same |
| Light source | LEDs | LEDs | LEDs | LEDs | LEDs | LEDs | Same |
| Wavelength | Red : 630nm±5nm
Blue : 460nm±5nm
Amber : 605nm±5nm
Infrared : 850nm±10nm | Blue: 460nm±5nm
Red: 630nm±5nm
Amber: 605nm±5nm
Near-infrared: 850nm±5nm | Blue: 465nm±5nm
Red: 625nm±5nm
Amber: 605nm±5nm | SKB-1818P,SKB-1918,SKB-1918P,SKB-1918PLUS:
Blue: 460nm±10nm
Red: 620nm±10nm
SKB-2318L,SKB-2318P,SKB-2318PRO,IN-FM002,SKB-2418:
Blue: 460nm±10nm
Red: 620nm±10nm
Infrared: 850nm±10nm | KFB290, KFB291:
635nm±5nm visible red light;
850nm±5nm Invisible red light;
465nm±5nm blue light;
605nm±5nm amber light. | 605nm
625nm
465nm | Same |
Page 5 of 9
{9}
Dongguan Yanxi Technology Co., LTD.
510(k)s -510(k) Summary
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Reference Device 1 | Reference Device 2 | Remark |
| --- | --- | --- | --- | --- | --- | --- | --- |
| LED Intensity (mW/cm²) | 2311:
Red+IR: 15~22
Blue: 16~29
Amber: 6~10
2344:
Red: 21
Blue: 31
Amber: 15
Red+IR: 22
2457:
Red: 11~14
Blue:20~27
Amber: 6~9
Red+IR: 9~14
2458:
Red+IR: 16
Blue: 20
Amber: 11 | M500:
Blue: 40mW/cm²
Amber: 10mW/cm²
Red+IR: 20mW/cm²
Mixed mode: 12mW/cm² | Blue: 15~63
Red: 31~75
Amber:15~42 | SKB-1818P:
Mode 1: Red: 5.5
Mode 2: Blue: 9
SKB-1918:
Mode 1: Red: 5
Mode 2: Blue: 6
SKB-1918P:
Mode 1: Red: 6
Mode 2: Blue: 6.5
SKB-1918PLUS:
Mode 1: Red: 3.5
Mode 2: Blue: 4.0
SKB-2318L,SKB-2318P,SKB-2318PRO,IN-FM002,SKB-2418:
Mode 1: Red: 3.5~10
Mode 2: Blue: 2.5~8
Mode 3: Red+Infrared Light: 6~27 | KFB290, KFB291:
Red: 25
IR: 3
Red+IR: 30
Blue: 18
Amber: 20
Mixed light: 9 | Red 2.45mW/cm²
Blue 1.33mW/cm²
Amber: unpublished | Similar Note 1 |
| Treatment time | 10, 15, 20, 25, 30 | 10 minutes each time, 3~5 times a week | 10 minutes each session. | It is recommended to use it | For red, blue and red+infrared: 10, 20, 30 minutes | 20mins/day, 120 days | Same |
Page 6 of 9
{10}
Dongguan Yanxi Technology Co., LTD.
510(k)s -510(k) Summary
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Reference Device 1 | Reference Device 2 | Remark |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | minutes. | | It is recommended to use | for 10 minutes a day, 3 times per week. | For infrared, amber light and mixed light: 10, 20 minutes. | | |
| Electrical safety | IEC 60601-1
IEC 60601-1-11
IEC 60601-1-2
IEC 60601-2-83
IEC 62133-2
IEC 62471 | IEC60601-1-2
IEC60601-1
IEC60601-1-11
IEC60601-2-83
IEC 62471 | IEC60601-1-2
IEC60601-1
IEC60601-1-11
IEC60601-2-57
IEC 62471 | EN/IEC 60601-1
EN/IEC 60601-1-2
EN/IEC 60601-11
EN/IEC 62471
IEC 60601-2-57
IEC 62366-1
EN/IEC 60601-1-6 | IEC 60601-1;
IEC 60601-1-2;
IEC 60601-1-11;
IEC 60601-2-83;
IEC 62471 | ANSI/AAMI ES 60601-1
EN 60601-1-2
IEC 60601-1-6
IEC 60601-1-11
IEC 60601-2-57
IEC 62133-2
IEC 62471 | Same |
| Biocompatibility feature | ISO 10993-5,
ISO10993-10,
ISO 10993-23 | ISO 10993-5
ISO 10993-10
ISO 10993-23 | Comply with ISO10993-1,
ISO10993-5 and ISO10993-10 | ISO 10993-1 | ISO 10993-5
ISO 10993-10
ISO 10993-23 | ISO 10993-1 | Same |
Page 7 of 9
{11}
Dongguan Yanxi Technology Co., LTD. 510(k)s -510(k) Summary
## VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### 1) Biocompatibility Safety
The materials of the patient-directly contacting components of the subject device is performed the biocompatibility evaluation in accordance with the “Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020”, as recognized by FDA. The following testing was performed to, and passed, including:
- ISO 10993-5: 2009, Biological evaluation of medical devices –Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10: 2021, Biological evaluation of medical devices – Part 10: Tests for skin sensitization
- ISO 10993-23: 2021, Biological evaluation of medical devices – Part 23: Tests for irritation
### 2) Electrical Safety and EMC Safety
Electrical safety and Eye safety testing was performed to, and passed, the following standards:
- IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION, Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance –Collateral standard: electromagnetic compatibility
- IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-11 Edition 2.1 2020-07 CONSOLIDATED VERSION, Medical Electrical Equipment –Part 1: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- IEC 60601-2-83:2022, Medical Electrical Equipment – Part 2-83: Particular Requirements For The Basic Safety And Essential Performance Of Home Light Therapy Equipment
- IEC 62133-2 Edition 5.0 2021-09, Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems.
### 3) Eye Safety
- IEC 62471:2006, Photobiological safety of lamps and lamp systems
### 4) Software Verification and Validation
Software documentation consistent with *Basic Documentation* of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
## Summary
Based on the above performance as documented in this application, the subject device was found to have a safety and effectiveness profile that is similar to the predicate devices and reference devices.
{12}
Dongguan Yanxi Technology Co., LTD. 510(k)s -510(k) Summary
## VIII. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, it can be concluded that the LED Light Therapy Mask is as safe, as effective, and performs as well as the legally marketed predicate devices.
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.