K190443 · Galactic Beauty, LLC · OHS · Jun 24, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190443
Device Name
MMSphere
Applicant
Galactic Beauty, LLC
Product Code
OHS · General, Plastic Surgery
Decision Date
Jun 24, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MMSphere™is an OTC, multi-use Light Therapy Device using LED light therapy technology for the treatment of mild to moderate wrinkles and/or acne.
Device Story
MMSphere™ is an OTC LED light therapy device for treating mild to moderate acne and wrinkles. Device utilizes red (625nm), blue (465nm), and amber (605nm) light. Operated by the user in a home setting, the device can be used as a handheld unit or placed on a countertop. Users wear provided goggles during treatment. Device features a control knob for power and setting selection. Output is light energy delivered to the skin. Intended to improve skin condition through regular 20-minute sessions. Benefits include non-invasive, self-administered light therapy for cosmetic and dermatological concerns.
Clinical Evidence
Bench testing only. Device conformity demonstrated via electrical safety (IEC 60601-1), EMC (EN 60601-1-2), photobiological safety (IEC 62471), and battery safety (IEC 62133-2). A usability study with 35 subjects evaluated device operation (powering on/off, changing settings, recharging). Usability findings identified issues with knob resistance and charging port accessibility.
Technological Characteristics
LED-based light therapy system emitting red (625nm), blue (465nm), and amber (605nm) light. Form factor supports both handheld and stationary (countertop) use. Powered by internal secondary lithium cells. Complies with IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57, and IEC 62471 standards. Connectivity includes a charging port for battery maintenance.
Indications for Use
Indicated for the treatment of wrinkles and/or mild to moderate acne in patients using the device for 20-minute sessions, three to seven times per week.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Galactic Beauty, LLC Ellen Marmur President/CEO 12 E 87 St., Ste 1A New York, New York 10128
June 24, 2019
## Re: K190443
Trade/Device Name: MMSphere Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHS. OLP Dated: Mav 28, 2019 Received: May 29, 2019
Dear Ellen Marmur:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Neil R.P. Ogden, MS Acting Assistant Director, THT4A3 DHT4A: Division of General Surgery Devices OHT5: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
Indications for Use (Describe)
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary MMSphere™ Light Therapy System
This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR § 878.4810.
Submission Date: February 22, 2018
| Submitter Information | Galactic Beauty, LLC<br>Attn: Ellen Marmur, MD<br>12 E 87 ST STE 1A<br>New York, New York 10128<br>Tel: 212-996-6900<br>emarmur@gmail.com |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Specification Developer | Galactic Beauty, LLC<br>Attn: Ellen Marmur, MD<br>12 E 87 ST STE 1A<br>New York, New York 10128<br>Tel: 212-996-6900<br>emarmur@gmail.com |
#### 2. General Information
- Classification Name: Light Based Over The Counter Acne and Wrinkle Reduction 2.1
- Common/Usual Name: Acne and Wrinkle Light Therapy System 2.2
- Proprietary Names: MMSphere™ 2.3
- 2.4 Classification: Class II
- 2.5 Classification Number: 878.4810
- 2.6 Product Code: OHS/OLP
- Review Panel: General & Plastic Surgery 2.7
### 3. Device Description
MMSphere™is an OTC, multi-use Light Therapy Device using LED light therapy technology for
the treatment of mild to moderate wrinkles and/or acne.
The device uses a combination of red light (625nm), blue light (465nm), and amber (605nm) to create different treatment settings.
The user will wear provided goggles for treatments. Device can be used with handheld option or placed on a countertop. See Appendix: 11-1-Device Drawing and 11-3 System Level Schematic
{4}------------------------------------------------
### 4. Indications for Use
MMSphere™Light Therapy Device emits energy in the red, blue and amber regions of the spectrum, specifically indicated to treat wrinkles and/or mild to moderate acne. The MMSphere™is designed to be used for 20 minute treatments three to seven times per week.
#### 5. Predicate Device
This device is substantially equivalent to the following predicates, which are currently cleared under product codes OHS/OLP:
- 1. K120775 LightStim For Wrinkles
- 2. K180847 Neutrogena Light Therapy Acne Mask+
- 3. K180856 Neutrogena Light Therapy Aging Mask+
## 6. Comparison of Technological Characteristics with The Predicate Device
LightStim For Wrinkles K120775, Neutrogena Light Therapy Acne Mask K180847 and Neutrogena Light Therapy Aging Mask K180856 to the MMSphere™ Light Therapy System with respect to intended use, technological characteristics, principles of operation and performance data.
{5}------------------------------------------------
# Device Comparison Table
| | MMSphere™ | Neutrogena Light<br>Therapy Acne<br>Mask | Neutrogena Light<br>Therapy Aging<br>Mask | LightStim For<br>Wrinkles |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| 510K Number | K190443 | K180847 | K180856 | K120775 |
| Product Code | OHS, OLP | OLP | OHS | OHS |
| Classification<br>Name | OTC Powered<br>Light for<br>Wrinkle<br>Reduction and<br>Acne | Over the Counter<br>powered Light<br>Based Laser for<br>Acne | Light-based<br>over-the-counter<br>wrinkle reduction | Light-based<br>over-the-counter<br>wrinkle reduction |
| Regulatory Class | Class II | | | |
| Classification<br>Regulation | 21 CFR 878.4810 | | | |
| Classification<br>Panel | General and Plastic Surgery | | | |
| Indications for<br>Use | OTC Light<br>Therapy Device<br>using LED light<br>for the<br>treatment of<br>wrinkles and<br>mild to<br>moderate acne. | The Light<br>Therapy Acne<br>Mask + is<br>intended to emit<br>energy in the red<br>and blue region<br>of the<br>spectrum,<br>specifically<br>indicated to treat<br>mild to moderate<br>acne on the face. | The Neutrogena<br>Light Therapy<br>Aging<br>Mask+ is an over<br>the<br>counter device<br>that is<br>indicated for the<br>treatment of full<br>face<br>wrinkles. | The Light for<br>Wrinkles is an<br>Over-The-Counter<br>handheld device<br>intended for use in<br>the treatment of<br>full-face wrinkles. |
| Handheld or<br>stationary | Both | Mask | Mask | Handheld |
| Irradiance Source | LEDs | LEDs | LEDs | LEDs |
| Light Color and<br>Wavelengths | 605nm<br>625nm<br>465nm | Red 630nm±5nm<br>Blue 440nm±5nm | Red 620-640nm<br>IR 820-880nm | 605nm,<br>630nm,<br>660nm,<br>855nm |
| Total Energy<br>Dose J/cm² * | Blue- 24.48<br>Red- 44 | 38.38 | 47.58 | 25.92 |
| Power Density | Red 2.45<br>mW/cm²<br>Blue 1.33<br>mW/cm² | 1.07 mW/cm² | 1.32 mW/cm² | 2.4 mW/cm² |
| Treatment<br>Regimen | 20mins/day,<br>120 days | 10mins/day, 60<br>days | 10mins/day, 60<br>days | 3mins/day, 60days |
{6}------------------------------------------------
# Conclusion
After an analysis of the safety indications, intended uses, performance, design materials, power output, technological characteristics, treatment areas, and treatment regimes, the Sponsor believes that no significant differences exist between the new device and the predicate devices and no new issues of safety or effectiveness are raised. Therefore substantial equivalency has been demonstrated.
### 7. Performance Testing
Bench performance testing was undertaken to demonstrate that the MMSphere™is safe and effective and substantially equivalent to the predicate devices. The following are applicable consensus standards.
- 1. ISO 10993-1: Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process
- 2. ANSI/AAMI ES 60601-1:2005+A2 (R2012) +A1: Medical electrical equipment—Part 1: General requirements for basic safety and essential performance
- 3. EN 60601-1-2:2015: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances - Requirements and tests
- 4. IEC 60601-1-6:2010+A1: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- 5. IEC 60601-1-11:2015: Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- 6. IEC 60601-2-57:2011: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
- 7. IEC 62133-2:2017: Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary lithium cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
- 8. IEC 62471:2006: Photobiological safety of lamps and lamp systems
{7}------------------------------------------------
#### Non-Clinical Testing 8.
This device is in conformity with IEC 60601 electrical safety testing, IEC 60601 EMC testing and IEC 62471 photobiological safety testing.
## Usability Testing
35 subjects identified with wrinkles and/or acne were given the device, the user manual and a charging cable. They were asked to turn the device on and off, change the device settings, and demonstrate how to recharge the device. After this exercise, they were asked to complete a usability survey list of 10 questions indicating how easy it was to follow instructions and use the device.
Out of 35 users participating in the Usability Study:
## Addressing Usability Findings
During the course of administering the Usability Test, two specific concerns were discovered:
- . The design of the knob that controls the ON and OFF positions as well as the light settings were found to be too tight to turn comfortably by the operator.
- The second concern was discovered when the operators were asked to charge . the device. The access port to plug in the charging cable on the base of the device was found to be too recessed and not easily identifiable or accessible.
{8}------------------------------------------------
### 9. Substantial Equivalence
Based upon the analysis of the overall performance characteristics the MMSphere™ device has the same intended use as the predicate devices. The device also has similar technological characteristics to the predicate devices. The differences between the devices do not pose any safety risks to the user.
We have shown by the data contained in this 510(k) submission that the Sponsor has found no significant differences between the MMSphere™ and the predicate devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.