VITA Multi-Function Head Brush (TB-2443F, TB-2442AF, TB-2343F, TB-2442F)
K252053 · Touchbeauty Beauty & Health ( Shenzhen ) Co., Ltd · OHS · Oct 27, 2025 · General, Plastic Surgery
Device Facts
Record ID
K252053
Device Name
VITA Multi-Function Head Brush (TB-2443F, TB-2442AF, TB-2343F, TB-2442F)
Applicant
Touchbeauty Beauty & Health ( Shenzhen ) Co., Ltd
Product Code
OHS · General, Plastic Surgery
Decision Date
Oct 27, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
VITA Multi-Function Head Brush is an Over-The-Counter (OTC) device. Head Mode: The device is indicated to promote hair growth in male with androgenic alopecia who have Norwood-Hamilton classifications of IIa-V and females with androgenic aplopeia who have Ludwig-Savin Classification of I-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV. Face Mode: LED therapy function: The device is indicated for the treatment of full face wrinkles and/or mild to moderate inflammatory acne.
Device Story
Hand-held multi-function device; consists of host, removable treatment head, and charging cable. Head mode uses low-level laser diodes (650nm) and red LEDs (660nm) to promote hair growth; face mode uses LEDs (660nm/465nm) for wrinkle reduction and acne treatment. Includes micro-motor for facial skin vibration; liquid reservoir for hair care application. Operated by user via single button; automatic 10-minute shut-off. Powered by rechargeable lithium battery. Used in home environment. Provides light-based therapy to stimulate hair follicles and treat skin conditions; benefits include hair growth promotion and reduction of acne/wrinkles.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness supported by biocompatibility testing (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-2, -1-11, -2-83), laser safety (IEC 60825-1), photobiological safety (IEC 62471), and software verification/validation.
Technological Characteristics
Hand-held device; rechargeable lithium battery. Emits light via 8 laser diodes (650nm) and 12 red LEDs (660nm) for hair growth; LEDs (660nm/465nm) for skin therapy. Includes micro-vibration motor. Materials evaluated per ISO 10993. Software classified as Basic Documentation of concern. Compliance with IEC 60601-1, IEC 60601-2-83, and IEC 62471.
Indications for Use
Indicated for males with androgenic alopecia (Norwood-Hamilton IIa-V) and females with androgenic alopecia (Ludwig-Savin I-4, II-1, II-2, or frontal) with Fitzpatrick Skin Phototypes I-IV to promote hair growth. Also indicated for treatment of full face wrinkles and/or mild to moderate inflammatory acne.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 27, 2025
Touchbeauty Beauty & Health (Shenzhen) Co., Ltd
% Bing Huang
RA Engineer
Feiying Drug & Medical Consulting Technical Service Group
Rm 2401 Zhenye International Business Center
No. 3101-90, Qianhai Road
Shenzhen, Guangdong 518052
China
Re: K252053
Trade/Device Name: VITA Multi-Function Head Brush (TB-2443F, TB-2442AF, TB-2343F, TB-2442F)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, OLP, OAP, ISA
Dated: July 1, 2025
Received: July 1, 2025
Dear Bing Huang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252053 - Bing Huang
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K252053 - Bing Huang
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.10.27 21:28:32
-04'00"
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252053 | |
| Device Name VITA Multi-Function Head Brush Model(s): TB-2443F, TB-2442AF, TB-2343F, TB2442F | |
| Indications for Use (Describe) VITA Multi-Function Head Brush is an Over-The-Counter (OTC) device. Head Mode: The device is indicated to promote hair growth in male with androgenic alopecia who have Norwood-Hamilton classifications of IIa-V and females with androgenic aplopeia who have Ludwig-Savin Classification of I-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV. Face Mode: LED therapy function: The device is indicated for the treatment of full face wrinkles and/or mild to moderate inflammatory acne. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s -510(k) Summary
# 510(k) Summary
"510(k) Summary" as required by 21 CFR Part 807.92.
K252053
Date: 2025-10-27
# I. Submitter
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
29/F, Block B, Tiley Central Plaza, No.195 Haide 3rd Road, Nanshan District, Shenzhen, China.
Post code: 518064
Tel.: +86 400-1191-180
Zhou Elin
Title: CEO
Tel.: +86 755 3366 2222 ext.666
Email: chow@touchbeauty.com
# II. Device
Name of Device: VITA Multi-Function Head Brush
Model(s): TB-2443F, TB-2442AF, TB-2343F, TB2442F
Common or Usual Name: Light Based Over The Counter Wrinkle Reduction(OHS)
Over-The-Counter Powered Light Based For Acne(OLP)
Laser, Comb, Hair (OAP)
Therapeutic electric massager(ISA)
Classification Name: Laser surgical instrument for use in general and plastic surgery and in
dermatology(OHS, OLP, OAP)
Massager, Therapeutic, Electric(ISA)
Regulatory Class: II
Product Code: OHS, OLP, OAP, ISA
Regulation Number: 21 CFR 878.4810; 21CRF 890.5500; 890.5660
# III. Predicate Device & Reference Device
Predicate device:
| Manufacturer | Predicate Device | 510(k) Number | Cleared Date |
| --- | --- | --- | --- |
| YIBIN YINGTONG
INTELLIGENT
TECHNOLOGY CO.,LTD. | Intelligence LaserComb
(YTLC001-W) | K213789 | June 24, 2022 |
| Light Tree Ventures Europe
B.V. | Aduro Comb
(Model: SZ-22A) | K230597 | April 28, 2023 |
| Galactic Beauty, LLC | MMSphere™ | K190443 | June 24, 2019 |
{5}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s -510(k) Summary
Reference device:
| Manufacturer | Predicate Device | 510(k) Number | Cleared Date |
| --- | --- | --- | --- |
| Shenzhen Sungrow LED Technology Co.,Ltd | LED Light Therapy Mask (Models:G15, G15P, G15K, G11P, G11, G10, G13, G14, G17, VISO, PRANA, Chin2Chest, BBL-FACEMASK) | K250830 | June 9, 2025 |
## IV. Device Description
The VITA Multi-Function Head Brush is a hand-held device. It consists of a host, a removable treatment head, and a charging cable. The host has a button, indicator lights, and the device’s power-on/off, treatment mode switch can be achieved by pressing the button. The removable treatment head has lights row, stainless steel bristles and liquid reservoir (applicable for model TB-2443F/ TB-2442AF), of which the liquid reservoir is used for the hair care after treatment.
The LEDs and low-level lasers in the removable treatment head emit light on the treatment area. In addition, the device is also equipped with a built-in micro motor to generate micro-vibration to relax the facial skin. The device has a head mode and face mode, and the device will automatically shut down after each mode completes treatment within 10 minutes. The device is powered by a built-in rechargeable lithium battery.
## V. Indications for Use
VITA Multi-Function Head Brush is an Over-The-Counter (OTC) device.
Head Mode: The device is indicated to promote hair growth in male with androgenic alopecia who have Norwood-Hamilton classifications of IIa-V and females with androgenic aplopecia who have Ludwig-Savin Classification of I-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV. Face Mode:
LED therapy function: The device is indicated for the treatment of full face wrinkles and/or mild to moderate inflammatory acne.
## VI. Comparison of Technological Characteristics With the Predicate Device
VITA Multi-Function Head Brush is compared with the following Predicate Devices and Reference Devices in terms of intended use, design, material, specifications, and performance:
{6}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s -510(k) Summary
Table 1: Comparison between subject device and the FDA cleared OAP device.
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- |
| 510(k) Number | K252053 | K213789 | K230597 | / |
| Trade name | VITA Multi-Function Head Brush (Model(s): TB-2443F, TB-2442AF, TB-2343F, TB-2442F) | Intelligence LaserComb (YTLC001-W) | Aduro Comb (Model: SZ-22A) | / |
| Product code | OAP | OAP | OAP | Same |
| Device classification | II | II | II | Same |
| Prescription or OTC | OTC | OTC | OTC | Same |
| Indication for use/ Intended use | VITA Multi-Function Head Brush is an Over-the-Counter (OTC) device. Head Mode: The device is indicated to promote hair growth in male with androgenic alopecia who have Norwood-Hamilton classifications of IIa-V and females with androgenic aplopecia who have Ludwig-Savin Classification of I-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV. | The Intelligence LaserComb (Model: YTLC001-W) is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classifications of IIa-V or females with androgenic alopecia who have Ludwig-Savin Classifications of I-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV. | The Aduro Comb (Model: SZ-22A) is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood- Hamilton Classifications of IIa - V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I-IV. | Same |
| Type of Laser | Low-level laser diodes and light emitting diodes | Visible red light-emitting diodes | Light emitting diodes | Same |
| Wavelength | Laser: 650nm±10nm Red light LED: 660nm±10nm | 655nm±10nm | 650±10nm | Same |
| Amounts of laser | Laser: 8pcs Red light LED: 12pcs | 10 | 22 LEDs | Similar |
| Energy of Per Laser Diode | ≤5mW | ≤5mW | Unknown | Same |
{7}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s -510(k) Summary
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- |
| Power Intensity | 3.0mW/cm2 | 2.5 mW/cm2 | 2.77 mW/cm2 | Similar |
| Dosage Intensity | Laser:1.8J/cm2 | 1.2 J/cm2 | 4.98 J/ cm2 | Similar |
| Treatment Time | 10 minutes per treatment | 8 minutes per treatment | Each Treatment: 30 min | Same |
| Treatment Frequency | 3 times per week | 3 times per week (every other day) | every other day, for 16 weeks | Same |
| Applicable People | Norwood-Hamilton IIa-V (males) Ludw ig-Savin I-4, II-1, II-2, or f rontal (females) | Norwood-Hamilton IIa-V (males) Ludw ig-Savin I-4, II-1, II-2, or f rontal (females) | Norwood-Hamilton IIa-V (males) Ludw ig-Savin I-4, II-1, II-2, or f rontal (females) | Same |
| Applicable Skin | Fitzpatrick Skin Phototypes I-IV | Fitzpatrick Skin Phototypes I-IV | Fitzpatrick Skin Phototypes I-IV | Same |
| Appearance Design | Comb | Comb | Comb | Same |
Table 2: Comparison between subject device and the FDA cleared OHS and OLP device.
| Comparison Elements | Subject Device | Predicate Device 1 | Reference Device 1 | Remark |
| --- | --- | --- | --- | --- |
| 510(k) Number | K252053 | K190443 | K250830 | / |
| Trade name | VITA Multi-Function Head Brush (Model(s): TB-2443F, TB-2442AF, TB-2343F, TB-2442F) | MMSphere™ | LED Light Therapy Mask (Models:G15, G15P, G15K, G11P, G11, G10, G13, G14, G17, VISO, PRANA, Chin2Chest, BBL-FACEMASK) | / |
| Product code | OHS, OLP | OHS, OLP | OHS, OLP | Same |
| Device classification | II | II | II | Same |
| Prescription or OTC | OTC | OTC | OTC | Same |
| Indication for use/ Intended use | VITA Multi-Function Head Brush is an Over-the-Counter (OTC) device. | MMSphereTM Light Therapy Device emits energy in the red, blue and amber regions of the spectrum, specifically | Red light: Treatment of full-face wrinkles. Yellow light: Treatment of full-face wrinkles. Red+Infrared light: | Same |
{8}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s -510(k) Summary
| Comparison Elements | Subject Device | Predicate Device 1 | Reference Device 1 | Remark |
| --- | --- | --- | --- | --- |
| | **Face mode:**
LED therapy function: The device is indicated for the treatment of full face wrinkles and/or mild to moderate inflammatory acne. | indicated to treat wrinkles and/or mild to moderate acne. The MMSphereTM is designed to be used for 20 minute treatments three to seven times per week. | Treatment of full-face wrinkles.
Blue light: Treatment of mild to moderate inflammatory acne.
Mixed light (Red+Blue+Infrared): Treatment of mild to moderate inflammatory acne. | |
| Handheld or stationary | Hand-held Type | Both | Mask | Same |
| Irradiance source | LEDs | LEDs | LEDs | |
| Anatomical sites | Entire Face | Entire Face | Whole face | |
| Wavelength | 660nm+/-10nm
465nm+/-10nm | 605nm
625nm
465nm | Blue: 415 nm+/-10nm
Red: 660 nm+/-10nm
Yellow: 590nm+/-10nm
Near-Infrared: 850nm+/-10nm | Same |
| Power Intensity | Red: 2.9mW/cm²
Blue:2.8mW/cm² | Red 2.45mW/cm²
Blue 1.33mW/cm² | Mode 1: Red:15-22
Mode 2: Yellow: 20-30
Mode 3: Red-near infrared: 20-30
Mode 4: Blue-red-near infrared: 25-45
Mode 5: Blue: 13-23
Mode 6: Yellow: 20-30, red-near infrared: 20-30, blue: 13-23 | Similar |
Page 5 of 7
{9}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s -510(k) Summary
## VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### 1) Biocompatibility Safety
The materials of the patient-directly contacting components of the subject device is performed the biocompatibility evaluation in accordance with the “Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020”, as recommended by FDA. The following testing was performed to, and passed, including:
- ISO 10993-5: 2009, Biological evaluation of medical devices –Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10: 2021, Biological evaluation of medical devices – Part 10: Tests for skin sensitization
- ISO 10993-23: 2021, Biological evaluation of medical devices – Part 23: Tests for irritation
### 2) Electrical Safety and EMC Safety
Electrical safety and Eye safety testing was performed to, and passed, the following standards:
- IEC 60601-1-2:2014 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance –Collateral standard: electromagnetic compatibility
- IEC 60601-1:2005 Medical electrical equipment –Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-11:2015 Medical Electrical Equipment –Part 1: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- IEC 60601-2-83:2019 Medical Electrical Equipment – Part 2-83: Particular Requirements For The Basic Safety And Essential Performance Of Home Light Therapy Equipment
- IEC 62133-2:2017 Secondary cells and batteries containing alkaline or other non-acid electrolytes –Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems.
- IEC 602825-1:2014 Safety of laser products – Part 1: Equipment classification and requirements.
### 3) Eye Safety
- IEC 62471:2006 Photobiological safety of lamps and lamp systems
### 4) Software Verification and Validation
Software documentation consistent with **Basic Documentation** of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
## Summary
Based on the above performance as documented in this application, the subject device was found to have a safety and effectiveness profile that is similar to the predicate devices.
Page 6 of 7
{10}
TOUCHBEAUTY BEAUTY & HEALTH ( SHENZHEN ) CO., LTD.
510(k)s –510(k) Summary
## VIII. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, it can be concluded that the VITA Multi-Function Head Brush is as safe, as effective, and performs as well as the legally marketed predicate devices.
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.