The subject device is intended for use in primary and revision total knee arthroplasty to alleviate pain and restore function.
Device Story
Triathlon® Gold Femoral Components are orthopedic implants designed to resurface the distal femur in total knee arthroplasty (TKA). The device articulates against a tibial component and optional patellar component to restore joint function and alleviate pain. Components are additively manufactured from Ti-6Al-4V powder via Electron Beam Melting (EBM) and finished with a Titanium Nitride (TiN) coating applied via Physical Vapor Deposition (PVD). The bone-facing surfaces feature a Tritanium® porous structure to facilitate fixation. Available in 16 configurations (8 sizes, left/right), the devices are sterile, single-use implants intended for surgical implantation by orthopedic surgeons in a hospital or surgical center setting. The device replaces damaged articular surfaces, allowing for improved patient mobility and pain reduction.
Clinical Evidence
No clinical data was required or provided to demonstrate substantial equivalence. Evidence is based on non-clinical laboratory testing and engineering analyses, including physical, chemical, and mechanical characterization of porous and solid surfaces, TiN coating adhesion, femoral fatigue testing, wear testing, MRI safety, and biocompatibility evaluation.
Indicated for patients with painful, disabling knee joint disease (noninflammatory degenerative joint disease, osteoarthritis, traumatic arthritis, avascular necrosis, rheumatoid arthritis, post-traumatic arthritis); post-traumatic loss of knee joint configuration/function; moderate varus, valgus, or flexion deformity with stable ligamentous structures; revision of unsuccessful knee replacement; or distal femur/proximal tibia fractures not manageable by standard techniques. Indicated for cemented or uncemented use.
Regulatory Classification
Identification
A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.
Trident® II Tritanium® Acetabular Shells and 6.5mm Low Profile Hex Screw (K161569)
Smith & Nephew Legion Primary Knee System (K093746)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 24, 2025
Howmedica Osteonics Corp. dba Stryker Orthopaedics
Shing Jen Tai
Principal Regulatory Affairs Specialist
325 Corporate Drive
Mahwah, New Jersey 07430
Re: K252044
Trade/Device Name: Triathlon® Total Knee System - Triathlon® Gold Femoral Components
Regulation Number: 21 CFR 888.3565
Regulation Name: Knee Joint Patellofemorotibial Metal/Polymer Porous-Coated Uncemented Prosthesis
Regulatory Class: Class II
Product Code: MBH, JWH
Dated: June 30, 2025
Received: September 25, 2025
Dear Shing Jen Tai:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252044 - Shing Jen Tai
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252044 - Shing Jen Tai
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lixin Liu, Ph.D
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Triathlon® Total Knee System - Triathlon® Gold Femoral Components
Page 20 of 53
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252044 | ? |
| Please provide the device trade name(s). | | ? |
| Triathlon® Total Knee System - Triathlon® Gold Femoral Components | | |
| Please provide your Indications for Use below. | | ? |
| General Total Knee Arthroplasty (TKR) Indications: | | |
| - Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis. | | |
| - Post-traumatic loss of knee joint configuration and function. | | |
| - Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. | | |
| - Revision of previous unsuccessful knee replacement or other procedure. | | |
| - Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture-management techniques | | |
| The Triathlon® Gold Femoral Components are indicated for both uncemented and cemented use. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ? |
| | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
K252044 Page 1 of 4
# 510(k) Summary
Sponsor
Stryker Orthopaedics
325 Corporate Drive
Mahwah, NJ 07430
Contact Person
Shing Jen Tai
Principal Regulatory Affairs Specialist
Howmedica Osteonics Corp
325 Corporate Drive
Mahwah, NJ 07430
269-800-2088
Shingjen.tai@stryker.com
Alternate Contact
Margaret Klippel
Chief Regulatory Affairs Specialist
Howmedica Osteonics Corp
325 Corporate Drive
Mahwah, NJ 07430
201-831-5559
Margaret.klippel@stryker.com
Date Prepared: June 30, 2025
Proprietary Name: Triathlon® Total Knee System – Triathlon® Gold Femoral Components
Common Name: Total Knee Joint Replacement Prosthesis
Classification Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis (21 CFR 888.3565)
Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis (21 CFR 888.3560)
Product Codes: MBH; JWH
{5}
K252044 Page 2 of 4
# Legally Marketed Primary and Additional Predicate Devices to Which Substantial Equivalence is Claimed:
- Triathlon™ Total Knee System - K051380 (Primary Predicate)
- Omni TiN Coated Apex Knee™ System - K191765 (Additional predicate)
# Legally Marketed Reference Devices Used to Support Substantial Equivalence:
- Triathlon® Tritanium® Tibial Baseplates – K123486
- Exactech Truliant® Porous Tibial Tray and Alteon Bone Screws – K182346
- Biocore9 Acetabular Cup System - K212761
- Endotec B-P Hemispherical Acetabular Component - K963101
- Trident® II Tritanium® Acetabular Shells and 6.5mm Low Profile Hex Screw - K161569
- Smith & Nephew Legion Primary Knee System - K093746
# Reason for 510(k) Submission:
The purpose of this Traditional 510(k) Premarket Notification is to introduce the new Triathlon® Gold Femoral Components as a line extension of the previously cleared Triathlon® cruciate retaining (CR) femoral components within the Triathlon® Total Knee System. This submission proposes adding the Triathlon® Gold Femoral Components to the system as an additional CR femoral component option that is 1) manufactured from titanium alloy (Ti-6Al-4V) and fully coated with Titanium Nitride (TiN), and 2) for uncemented and cemented use. Addition of these femoral components does not change the intended use, indications for use, operational principles, or fundamental scientific technology of the Triathlon® Total Knee System.
# Device Description:
The Triathlon® Total Knee System is a total knee prosthesis designed to resurface the articulating surface of the femoral and tibial bones. In a total knee replacement, Triathlon® Gold Femoral Components are intended to replace the articulating surface of the distal femur and articulate against an articular surface of the tibial component as well as an optional patellar component. Triathlon® Gold Femoral Components are additively manufactured from Ti-6Al-4V powder, using Electron Beam Melting (EBM) technology, and then fully coated with titanium nitride (TiN), using Physical Vapor Deposition (PVD). In addition to TiN coating, the bone cut facing surfaces of these femoral components contain the Tritanium® porous structure. Indicated for both uncemented and cemented use, the subject femoral components are provided in right and left configurations and eight proportional sizes (sizes 1-8) for a total of 16 unique components to accommodate differences in patient anatomy. The subject femoral components are sterile, single use devices.
{6}
K252044 Page 3 of 4
## Intended Use:
The subject device is intended for use in primary and revision total knee arthroplasty to alleviate pain and restore function.
## Indications for Use:
**General Total Knee Arthroplasty (TKR) Indications:**
- Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
- Post-traumatic loss of knee joint configuration and function.
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- Revision of previous unsuccessful knee replacement or other procedure.
- Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture-management techniques
The Triathlon® Gold Femoral Components are indicated for both uncemented and cemented use.
## Summary of Technological Characteristics:
The device comparisons and performance testing show that the subject Triathlon® Total Knee System – Triathlon® Gold Femoral Components are substantially equivalent to the predicate devices based on intended use, indications for use, design, materials, technical and performance characteristics, and operational principles that do not raise different questions of safety and effectiveness.
## Non-Clinical Testing:
The following non-clinical laboratory testing and engineering analyses were performed to determine substantial equivalence:
- Physical characterization of the porous surfaces
- Chemical characterization of the porous surfaces
- Mechanical characterization of the porous surfaces
- Chemical characterization of the solid surface
- Mechanical characterization of the solid surface
- Characterization of TiN coating
- Femoral fatigue testing
- Coating adhesion verification testing
- Tibial-femoral and patella-femoral engineering analyses
- Femoral backside engineering analysis
- Clean wear testing
- Third body wear testing
{7}
K252044 Page 4 of 4
- MRI Safety testing
- Biocompatibility evaluation
## Clinical Testing:
Clinical testing was not required to demonstrate substantial equivalence.
## Conclusion:
The subject Triathlon® Total Knee System – Triathlon® Gold Femoral Components are substantially equivalent to the primary predicate Triathlon™ Total Knee System (K051380) and additional predicate Omni TiN Coated Apex Knee™ System (K191765) based on intended use, indications for use, design, material, technological characteristics, operational principles, and non-clinical performance data that do not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.