The Trident® II Tritanium® Acetabular Shell is intended for use in total hip arthroplasty and is intended for either primary or revision Total Hip Arthroplasty. Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis. Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure. Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results. Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum. When used with MDM Liners: Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. Dislocation risks. When used with Constrained Liner: The Trident Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability. The Trident® II Tritanium Acetabular Shells are indicated for cementless use only.
Device Story
Cementless acetabular shell and bone screws for primary or revision total hip arthroplasty. Shells manufactured from Ti-6Al-4V ELI alloy using Laser Rapid Manufacturing (LRM) additive manufacturing to create solid and porous structures (Tritanium In-Growth Technology). Three shell designs: Solidback, Clusterhole, Multihole. Compatible with existing Trident polyethylene inserts, MDM liners, and surgical instruments. Screws feature hex geometry, manufactured from wrought Ti-6Al-4V ELI. Used by orthopedic surgeons in clinical/OR settings. Provides stable acetabular fixation to restore hip function and reduce pain. Benefits include improved bone ingrowth potential and compatibility with established liner systems for complex revision cases.
Clinical Evidence
No clinical data. Substantial equivalence based on non-clinical bench testing including ASTM F1820 (push-out, lever-out, torque-out), fatigue testing, fretting evaluation, ASTM F543 (bone screw testing), plastic deformation, material chemistry, and porous surface characterization per FDA guidance. Bacterial endotoxin testing performed per ANSI/AAMI ST72:2011.
Indicated for patients with painful, disabling hip joint disease (degenerative/rheumatoid/posttraumatic arthritis, avascular necrosis), failed prior hip procedures, or poor bone stock. MDM liner use indicated for proximal femur fractures (nonunion, femoral neck, trochanteric) and dislocation risks. Constrained liner use indicated for primary/revision patients at high risk of dislocation due to prior history, bone loss, laxity, neuromuscular disease, or instability.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three figures in profile facing right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 14, 2016
Stryker Orthopaedics Valerie Giambanco Staff Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430
Re: K161569 Trade/Device Name: Trident® II Tritanium® Acetabular Shells and 6.5 mm Low Profile Hex Screws Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, KWZ, LZO, JDI, MEH Dated: August 31, 2016 Received: September 1, 2016
Dear Valerie Giambanco:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
### 510(k) Number (if known): K161569
Device Name: Trident® II Tritanium® Acetabular Shells and 6.5mm Low Profile Hex Screws Indications for Use:
Indications for Use
· Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
· Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
When used with MDM Liners
• Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
• Dislocation risks
When used with Constrained Liner:
· The Trident Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.
The Trident® II Tritanium Acetabular Shells are indicated for cementless use only.
Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## 510(k) Summary
| Sponsor | Stryker Orthopaedics<br>325 Corporate Drive<br>Mahwah, NJ 07430 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Valerie Giambanco<br>Staff Regulatory Affairs Specialist<br>Stryker Orthopaedics<br>Valerie.giambanco@stryker.com<br>Ph: 201-831-6275<br>Fax: 201-831-3275 |
| Date Prepared: | October 14, 2016 |
| Proprietary Name: | Trident® II Tritanium® Acetabular Shells and 6.5 mm Low Profile Hex<br>Screws |
| Common Name: | Total Hip Joint Replacement |
| Classification Name: | Hip joint metal/polymer/metal semi-constrained porous-coated<br>uncemented prosthesis. (888.3358) |
| | Hip joint metal/polymer constrained cemented<br>or uncemented prosthesis. (888.3310) |
| | Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis (888.3353) |
| | Hip joint metal/polymer semi-constrained cemented prosthesis<br>(888.3350) |
| Product Codes: | LPH, KWZ, LZO, JDI, MEH |
### Legally Marketed Predicate Devices to Which Substantial Equivalence is Claimed:
- Tritanium® Acetabular Shell System (K081171) [Shell] ●
- Restoration Acetabular Cup (K943549) [Screw] .
#### Legally Marketed Reference Devices Used to Support Substantial Equivalence:
- Tritanium® PL Cage (K152304) ●
- . Trident® Tritanium® PST® Acetabular Shells (K142606)
- Vitallium Cancellous Bone Screw (K944213) .
{4}------------------------------------------------
- Osteolock Acetabular Cup & Bone Screw (K903362)
- . Osteonics® HA Generation II Acetabular Component System (K983382)
- Restoration Anatomic Shell (K151264)
### Device Description:
The Trident® II Tritanium® Acetabular Shell and 6.5mm Low Profile Hex Screws are sterile, single-use devices that are intended for cementless fixation into a prepared acetabulum for either primary or revision Total Hip Arthroplasty. The Trident® II Tritanium® Acetabular Shell is an extension of the Trident System product line and features the same locking mechanism as the current Trident® Tritanium® product line. The Trident® II Tritanium® Acetabular Shell is intended to be used with existing Trident® polyethylene inserts, MDM® liners and existing surgical instruments.
The subject device is manufactured from Ti-6Al-4V ELI alloy. The implant consists of a unique configuration of both solid and porous structures that are simultaneously built using a Laser Rapid Manufacturing (LRM) method of additive manufacturing, applying Stryker's proprietary Tritanium® In-Growth Technology.
There are three designs of Trident® II Tritanium® Acetabular Shells:
- Solidback (sizes 42A-66H)
- Clusterhole (sizes 42A-66H)
- Multihole (sizes 42A-72J) ●
The new compatible 6.5mm Low Profile Hex Screws feature a hex geometry and are manufactured from wrought Ti-6Al-4V ELI. The new screws range in lengths from 15-80mm.
#### Intended Use:
The Trident® II Tritanium® Acetabular Shell is intended for use in total hip arthroplasty and is intended for either primary or revision Total Hip Arthroplasty.
#### Indications:
The Trident® II Tritanium® Acetabular Shells have the same Indications for Use as the Tritanium® Acetabular Shells – with the additional specific indications noted for MDM® and Constrained Insert compatibility, as identical to the recently cleared Restoration® Anatomic Shell (K151264). The optional compatible bone screws share indications as the mating acetabular shell.
#### Indications for Use:
• Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
• Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
{5}------------------------------------------------
• Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
• Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
### When used with MDM Liners:
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
- Dislocation risks
# When used with Constrained Liners:
• The Trident Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.
The Trident II Tritanium Acetabular Shells are indicated for cementless use only.
Summary of Technological Characteristics: Device comparisons and performance testing show that the Trident® II Tritanium® Acetabular Shell and 6.5mm Low Profile Hex Screw are substantially equivalent to the predicates in terms of intended use, indications, design, materials, performance characteristics and operational principles.
Non-Clinical Testing: The following non-clinical laboratory testing was performed to determine substantial equivalence:
- . Push out (ASTM F1820)
- Lever out (ASTM F1820)
- Torque out (ASTM F1820)
- Range of Motion Analysis
- Fatigue
- Fretting Evaluation of the MDM liner
- Bone Screw Testing (ASTM F543)
- Plastic Deformation
- Material Chemistry
- . Characterization of the Physical and Mechanical Properties of the porous surface –this testing established that the porous surface meets the requirements outlined in the FDA guidance documents, "Guidance Document for Testing Orthopedic Implants With Modified Metallic Surfaces Apposing Bone Or Bone Cement", April 28, 1994
{6}------------------------------------------------
- Bacterial endotoxin testing (BET) was used for pyrogenicity testing to achieve an . acceptable endotoxin limit as specified in ANSI/AAMI ST72:2011.
Clinical Testing: Clinical testing was not required as a basis for substantial equivalence.
Conclusion: Based upon a comparison of intended use, materials, summary of technological characteristics, and preclinical testing, the Tritanium® Acetabular Shell and the 6.5mm Low Profile Hex Screw are substantially equivalent to the predicate devices identified in this premarket notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.