K251844 · Kerecis Limited · OXH · Jul 15, 2025 · General, Plastic Surgery
Device Facts
Record ID
K251844
Device Name
Kerecis SurgiBind (50241)
Applicant
Kerecis Limited
Product Code
OXH · General, Plastic Surgery
Decision Date
Jul 15, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair, or reinforcement in plastic or reconstructive surgery.
Device Story
Kerecis SurgiBind is a lyophilized, terminally sterilized surgical mesh derived from Wild North Atlantic Cod fish skin. It functions as a biocompatible, resorbable scaffold for soft tissue reinforcement. The device is intended for single-use in plastic or reconstructive surgery. Before implantation, the mesh is rehydrated using sterile saline, Ringer’s solution, or autologous body fluids (e.g., blood). The surgeon applies the rehydrated mesh to the site of soft tissue weakness to provide structural support. The device is physically identical to the predicate Kerecis Reconstruct; the only modification is the expanded list of approved rehydration fluids in the labeling. The device benefits patients by providing a biological matrix that supports tissue repair.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including rehydration and suture retention tests, comparing the device performance when rehydrated with autologous body fluids versus saline/Ringer’s solution.
Technological Characteristics
Material: Lyophilized Wild North Atlantic Cod fish skin. Form factor: Surgical mesh. Energy source: None (passive implant). Sterilization: Ethylene Oxide. Connectivity: None. Software: None.
Indications for Use
Indicated for patients requiring soft tissue repair or reinforcement in plastic or reconstructive surgery where soft tissue weakness exists.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 15, 2025
Kerecis Limited
Skuli Magnusson
Vice President (Regulatory Affairs, Quality & Clinical Affairs)
Sundastraeti 38
P.O. Box 151
Isafjordur, 400
Iceland
Re: K251844
Trade/Device Name: Kerecis SurgiBind (50241)
Regulation Number: 21 CFR 878.3300
Regulation Name: Surgical Mesh
Regulatory Class: Class II
Product Code: OXH
Dated: June 12, 2025
Received: June 16, 2025
Dear Skuli Magnusson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251844 - Skuli Magnusson
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K251844 - Skuli Magnusson
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TEK N. LAMICHHANE -S
Tek N Lamichhane, PhD
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PBC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251844 | |
| Device Name Kerecis SurgiBind (50241) | |
| Indications for Use (Describe) For implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair, or reinforcement in plastic or reconstructive surgery. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary - K251844
Premarket Notification Submission (510(k) Summary)
## Submitter Information
Sponsor Name: Kerecis Limited
Sponsor Address: Eyrargata 2 – PO Box 151, 400 Isafjordur, Iceland
Sponsor Telephone: +354-419-8000
Establishment Registration: 301060025
Applicant: Skuli Magnusson,
Title: VP of Regulatory Affairs, QA & Clinical Affairs
Email Direct: sm@kerecis.com
Primary Contact Person: Skuli Magnusson,
Contact Title: VP of Regulatory Affairs, QA & Clinical Affairs
Email Direct: sm@kerecis.com
Date Summary Prepared: July 15, 2025
## Device Information
Trade Name (Proprietary): Kerecis SurgiBind (50241)
FDA Device Code: OXH
Categorization: Mesh, surgical, collagen, plastic and reconstructive surgery
Device Class: Class II, 21 CFR § 878.3300 – Surgical Mesh
## Predicate Device
Company Name: Kerecis
Device Name (Proprietary): Kerecis Reconstruct
Device 510(k): K202430
## Reference Device
Company Name: Cook Biotech Inc.
Device Names (Proprietary): Biodesign Anal Fistula Plug, Biodesign Fistula Plug, Biodesign Plastic Surgery Matrix, Biodesign Tissue Graft
Device 510(k): K191696
## Device Description
Kerecis SurgiBind (50241) is a part of a family of devices manufactured by Kerecis® Limited. The device is lyophilized, terminally sterilized, fish skin medical device comprised of biocompatible, resorbable fish skin
KERECIS Limited
2025
Page 1
{5}
(Wild North Atlantic Cod). The device is intended for single use only. Kerecis Reconstruct is now commercially marketed as Kerecis SurgiBind due to a name change. For clarity, this submission will refer to the device as Kerecis SurgiBind, except when specifically referencing the predicate device (K202430). This information is shown in Table 1.
Table 1: Subject device of this submission
| Product Name | 510(k) Proprietary Name/Device Name | 510(k) number |
| --- | --- | --- |
| Kerecis SurgiBind | Kerecis Reconstruct | K202430 |
The subject device remains physically identical to its predicate device, both in design and packaging, as well as for indications for use. The only difference between the subject device and its predicate device is in the device labeling, with the subject device having additional rehydration fluid options included in their instructions for use (IFUs).
## Intended Use / Indications for Use
The indications for use of K202430 is presented in Table 2. This submission does not introduce any changes in the indications for use.
Table 2: Indications for Use
| Product name | Indications for Use |
| --- | --- |
| Kerecis Reconstruct (K202430) | For implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair, or reinforcement in plastic or reconstructive surgery |
## Description of Modification
The only modification from the predicate device is in the device labeling. The modification that triggered this submission is the inclusion of additional rehydration options in the instructions for use (IFU). In addition to saline and Ringer’s solution (which are the rehydration fluids in the predicate device), autologous body fluids (such as blood) are being added as potential rehydration fluid options for the subject device of this submission.
Kerecis is implementing this modification in response to interest within the field regarding the use of these fluids for device rehydration.
## Technological characteristics and substantial Equivalence
The technological characteristics of the subject device of this submission and the proposed predicate device (K202430), are identical in that they are animal-based products suitable for tissue reinforcement. They also share identical device design and principle of operation. The following items are provided to demonstrate substantial equivalence to the predicate devices:
- Risk analysis summary
KERECIS Limited
2025
Page 2
{6}
- Sterilization justification
- Shelf-life justification
- Biocompatibility justification
- Bench testing
KERECIS Limited
2025
Page 3
{7}
Table 3: Substantial Equivalence comparison table
| Device name | Kerecis SurgiBind (Subject Devices) | Kerecis Reconstruct (Predicate Device) | Biodesign Anal Fistula Plug, Biodesign Fistula Plug, Biodesign Plastic Surgery Matrix, Biodesign Tissue Graft (Reference Devices) | Comparison |
| --- | --- | --- | --- | --- |
| Manufacturer | Kerecis Limited | Kerecis Limited | Cook Biotech Incorporated | N/A |
| 510(k) | K251844 | K202430 | K191696 (Bundled) | N/A |
| Prescription, single use only | Unchanged from predicate device | Yes | Yes | Same |
| Product Code | | OXH | OXN, OXH, FTM (Bundled Submission) | Same |
| Indications for Use | | For implantation to reinforce soft tissue where weakness exists, in patients requiring soft tissue repair, or reinforcement in plastic or reconstructive surgery. | Biodesign Anal Fistula Plug: For implantation to reinforce soft tissue where a rolled configuration is required, for repair of anal, rectal and enterocutaneous fistulas. Biodesign Fistula Plug: For implantation to reinforce soft tissue for repair of recto-vaginal or anorectal fistulas. Biodesign Plastic Surgery Matrix: For implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic or reconstructive surgery. | Same as predicate |
KERECIS Limited
{8}
KERECIS Limited
2025
Page 5
| Device name | Kerecis SurgiBind
(Subject Devices) | Kerecis Reconstruct
(Predicate Device) | Biodesign Anal Fistula Plug,
Biodesign Fistula Plug, Biodesign
Plastic Surgery Matrix, Biodesign
Tissue Graft
(Reference Devices) | Comparison |
| --- | --- | --- | --- | --- |
| | | | Biodesign Tissue Graft: For
implantation to reinforce soft
tissue. | |
| Materials | | Atlantic Cod Fish Skin. | Porcine small intestinal submucosa:
(constituents of the extracelluar
matrix) | Same as predicate |
| Supplied sterile? | | Yes | Yes | Same |
| Sterilization Method | | Ethylene Oxide | Ethylene Oxide | Same |
| Intended for single use? | | Yes | Yes | Same |
| Rehydration fluid(s) | Sterile Saline, Ringer’s solution
and autologous body fluids
(such as blood) | Sterile Saline and Ringer’s solution | All Devices in the bundle:
Sterile saline and autologous body
fluids | Different to
predicate,
comparable to
reference devices |
{9}
# Performance Testing
Performance testing for this 510(k) submission was largely leveraged from the Kerecis predicate device, as the subject device shares the same intended use, materials, and manufacturing process. As a result, performance testing is not required, except for rehydration and suture retention tests using autologous body fluid. These specific tests were conducted to address the only modification for the subject device, i.e., the inclusion of autologous body fluids (such as blood) as rehydration agents. The results confirmed that device performance remains consistent and comparable to the predicate device, supporting a determination of substantial equivalence.
# Conclusion
For the purposes of determining substantial equivalence under section 513(i) of the FD&C Act (21 U.S.C. § 360c(i)), Kerecis® SurgiBind (the subject device) shares the same intended use, functions, mode of action, and fundamental scientific technology as the predicate device. Additionally, the subject device is composed of the same materials and is manufactured using identical processes as the predicate device. The sole modification is the inclusion of autologous body fluids (such as blood) as optional rehydration fluids in the device's Instructions for Use (IFU). Because there are no changes to the fundamental scientific technology or intended use, and verification and validation activities have been completed, there is sufficient evidence to demonstrate that the subject device performs comparably to the predicate device currently marketed for the same intended use.
It is concluded that the subject device is substantially equivalent to the predicate device.
KERECIS Limited
2025
Page 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.