K202430 · Kerecis Limited · OXH · May 2, 2021 · General, Plastic Surgery
Device Facts
Record ID
K202430
Device Name
Kerecis Reconstruct
Applicant
Kerecis Limited
Product Code
OXH · General, Plastic Surgery
Decision Date
May 2, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For implantation to reinforce soft tissue where weakness exists, in patients requiring soft tissue repair, or reinforcement in plastic or reconstructive surgery.
Device Story
Kerecis Reconstruct is a fish skin-derived medical device used for physical reinforcement of soft tissue defects or weaknesses. Obtained from Atlantic cod skin via a controlled manufacturing process, the device is biocompatible, non-crosslinked, and resorbable. It is supplied in sterile, rectangular solid forms (4x7 cm, 7x10 cm, 7x20 cm). Used in plastic or reconstructive surgery, the device is implanted by a surgeon to provide structural support; it supports fixation via sutures, tacks, or staples. The device requires re-hydration before use. By reinforcing weakened tissue, it aids in surgical repair and patient recovery.
Clinical Evidence
Evidence includes bench testing and animal studies. Bench testing evaluated tensile strength, suturability, hydration time, thickness, weight, stiffness, and microscopic structure. Animal testing in a GLP laboratory assessed safety (veterinarian health assessment, pathologist tissue response) and efficacy (histology, in-vitro tensile strength of explants). Results showed the device is equivalent in safety and efficacy to the control device for reinforcing soft tissue defects.
Technological Characteristics
Material: Atlantic cod fish skin. Form: Rectangular solid. Principle: Resorbable, non-crosslinked scaffold for soft tissue reinforcement. Fixation: Sutures, tacks, or staples. Sterilization: Ethylene Oxide. Shelf-life: 3 years. Requires re-hydration prior to use.
Indications for Use
Indicated for patients requiring soft tissue repair or reinforcement in plastic or reconstructive surgery where soft tissue weakness exists.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
May 2, 2021
Kerecis Limited Gudmundur Sigurjonsson CEO Eyrargata 2 – PO Box 151 Isafjordur, 400 Iceland
Re: K202430
Trade/Device Name: Kerecis Reconstruct Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OXH Dated: March 25, 2021 Received: April 1, 2021
Dear Gudmundur Sigurjonsson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202430
Device Name
Kerecis Reconstruct
Indications for Use (Describe)
The subject device is indicated for:
For implantation to reinforce soft tissue where weakness exists, in patients requiring soft tissue repair, or reinforcement in plastic or reconstructive surgery.
| Type of Use (Select one or both, as applicable) | <table border="0"><tr><td><div style="display:inline-block;"> <div style="float:left;">☑</div> <div style="float:left;">Reproductive Use (Part 21 CFR 321.1 Subject to Phase 1)</div> </div></td></tr><tr><td><div style="display:inline-block;"> <div style="float:left;">☐</div> <div style="float:left;">Non-Therapeutic Use (21 CFR 321.1 Subject to Phase 1)</div> </div></td></tr></table> | <div style="display:inline-block;"> <div style="float:left;">☑</div> <div style="float:left;">Reproductive Use (Part 21 CFR 321.1 Subject to Phase 1)</div> </div> | <div style="display:inline-block;"> <div style="float:left;">☐</div> <div style="float:left;">Non-Therapeutic Use (21 CFR 321.1 Subject to Phase 1)</div> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:inline-block;"> <div style="float:left;">☑</div> <div style="float:left;">Reproductive Use (Part 21 CFR 321.1 Subject to Phase 1)</div> </div> | | | |
| <div style="display:inline-block;"> <div style="float:left;">☐</div> <div style="float:left;">Non-Therapeutic Use (21 CFR 321.1 Subject to Phase 1)</div> </div> | | | |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Kerecis® Reconstruct Volume 05: 510(k) Summary
Image /page/3/Picture/1 description: The image shows the logo for "Kerecis" in a dark blue font, with the "c" in the middle of the word in orange. Above the logo, the text "K202430 Page 1 of 5" is displayed. The logo is simple and modern, with a focus on the company name.
# Volume 05: 510(k) Summary
Traditional Premarket Notification Submission (510(k)) Summary) Prepared in accordance with 21 CFR § 807.92
# 5.1 Submitter Information
| Sponsor Name: | Kerecis Limited |
|-----------------------------|---------------------------------------------------|
| Sponsor Address: | Eyrargata 2 - PO Box 151, 400 Isafjordur, Iceland |
| Sponsor Telephone: | +354-419-8000 |
| Establishment Registration: | 301060025 |
| Contact Person: | Gudmundur Fertram Sigurjonsson |
| Contact Title: | Chairman |
| Email Direct: | gfs@kerecis.com |
| Date Summary Prepared: | March 19, 2021 |
#### 5.2 Device Information
| Trade Name (proprietary): | Kerecis Reconstruct |
|---------------------------|-----------------------------|
| Common (usual): | Surgical Graft |
| Device Classification: | Class II, 21 CFR § 878.3300 |
| FDA Device Code: | OXH |
## 5.3 Predicate Device and Reference Devices
| Predicate Device | |
|----------------------------|----------------------------------|
| Company Name: | Cook Biotech Inc. |
| Device Name (proprietary): | Biodesign Plastic Surgery Matrix |
| Device 510(k): | K191696 (K161221, K034039) |
| Reference Device #1 | |
| Company Name: | Kerecis |
| Device Name (proprietary): | Kerecis SecureMesh |
| Device 510(k): | K153364 |
{4}------------------------------------------------
Kerecis Reconstruct Volume 05: 510(k) Summary
Image /page/4/Picture/1 description: The image shows the logo for "Kerecis" in a dark blue font, with the "c" in the middle of the word in orange. To the left of the logo is the text "K202430 Page 2 of 5". The logo is simple and modern, and the use of color makes it visually appealing. The logo is likely used to represent the company and its brand.
Reference Device #2 Company Name: Device Name (proprietary): Device 510(k):
Cook Biotech Inc. Biodesign Hernia Graft K133306
# 5.4 Device Description
The subject device is a fish skin medical device indicated for physical reinforcement of a soft tissue defect or weakness.
The subject device is obtained from cod fish skin by a standardized controlled manufacturing process and supplied in a peel-pouch terminally sterile packaging in the following rectangular solid sizes:
- . 4x7 cm
- 7x10 cm
- 7x20 cm .
The subject device is biocompatible, non-crosslinked, and therefore resorbable, strong, flexible, and supports fixation.
#### 5.5 Intended Use
The subject device is indicated for:
- a For implantation to reinforce soft tissue where weakness exists, in patients requiring soft tissue repair, or reinforcement in plastic or reconstructive surgery.
# 5.6 Technological Characteristics
Comparison of the subject device with predicate device (K191696), and the reference devices (K153364 and K133306), demonstrate that it is substantially equivalent with regards to: intended use, materials, design, and operational principle.
# 5.7 Summary of Supporting Evidence for Substantial Equivalence
- . Analysis of 510(k) Substantial Equivalence Decision-Making Process as outlined in FDA's Guidance Document, the 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)].
- . Bench testing was performed, following FDA guidance, "Preparation of a Premarket Notification Application for a Surgical Mesh" (1999), including but not limited to: tensile strength, suturability, hydration time test, thickness measurement, weight measurement, stiffness test (bend test), and microscopic structure analysis. All tests were performed on hydrated, sterile, ready-for-market devices, to confirm that the devices perform as expected under clinical conditions, and there were no negative effects on the mechanical properties.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for "Kerecis" in a bold, sans-serif font. The letters are primarily in a dark blue color, except for the "C" in the middle, which is in a bright orange color. Above the logo, the text "K202430 Page 3 of 5" is displayed in a smaller, regular font.
- . Animal testing was performed in a GLP laboratory, using the subject/test device and a control device. Based on the results of the study: the veterinarian's assessment of animal health (safety), the pathologist's assessment of the tissue response to the device (safety and efficacy), histology, and the in-vitro performance data tensile strength testing of the test and control explants side-by-side (efficacy), the performance of the subject device showed equivalence in safety and efficacy to that of the control device, when used to reinforce soft tissue defects.
## 5.8 Summary Table of Substantial Equivalence
A comparison of the subject device, predicate device, and two (2) reference devices are presented in TABLE 5-1.
| | Subject Device | Predicate<br>Device | Reference<br>Device | Reference<br>Device | Discussion |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Company | Kerecis | Cook Biotech | Kerecis | Cook Biotech | No discussion<br>required |
| Name | Kerecis<br>Reconstruct | Biodesign<br>Plastic Surgery<br>Matrix | Kerecis<br>SecureMesh | Biodesign<br>Hernia Graft | No discussion<br>required |
| 510(k) | Subject Device | K191696 | K153364 | K133306 | Predicate &<br>Reference devices<br>are all legally<br>marketed in the<br>USA |
| Device<br>Classification | II | II | II | II | Identical |
| Regulation | 21 CFR 878.3300 | 21 CFR 878.3300 | 21 CFR 878.3300 | 21 CFR 878.3300 | Identical |
| Product<br>Code | OXH | FTM, OXH | OXE | FTM, OXK | Product Code of<br>subject is a subset<br>of the predicate |
| Intended<br>Use | to reinforce soft<br>tissue where<br>weakness exists<br>in patients<br>requiring soft<br>tissue repair or<br>reinforcement in<br>plastic or<br>reconstructive<br>surgery | to reinforce soft<br>tissue where<br>weakness exists<br>in patients<br>requiring soft<br>tissue repair or<br>reinforcement in<br>plastic or<br>reconstructive<br>surgery | to be used as a<br>staple line<br>buttress | to reinforce soft<br>tissue where<br>weakness exists | Intended Use is<br>identical for the<br>subject &<br>predicate |
| Indications<br>for Use | Indications for<br>use include soft<br>tissue repair or | Indications for<br>use include soft<br>tissue repair or | Reinforcement<br>of staple lines<br>during: | Indications for<br>use include the<br>repair of hernia | Indications for Use<br>are identical for<br>the subject & |
| | reinforcement in<br>plastic or<br>reconstructive<br>surgery. The<br>graft is supplied<br>sterile and is<br>intended for<br>one-time use. | reinforcement in<br>plastic or<br>reconstructive<br>surgery. The<br>graft is supplied<br>sterile and is<br>intended for<br>one-time use. | 1.<br>Bariatric<br>surgical<br>procedures<br>2.<br>Gastric,<br>small<br>bowels and<br>mesentery<br>procedures<br>3.<br>Colorectal<br>and colon<br>procedures<br>4.<br>Lung and<br>bronchus<br>resections | or body wall<br>defects. The<br>graft is supplied<br>sterile and is<br>intended for<br>one-time use. | predicate (Note:<br>the subject device<br>adds example<br>procedures for<br>clarification) |
| Material | Atlantic Cod Fish | Porcine | Atlantic Cod Fish | Porcine | Identical material<br>to reference<br>device |
| Shapes | Rectangular | Rectangular | Custom design<br>for surgical<br>staplers | Rectangular | Subject device is<br>rectangular or<br>shredded. Either<br>configuration is<br>provided sterile in<br>a Tyvek pouch. |
| Size | 4x7 cm<br>7x10 cm<br>7x20cm | 4x7 cm<br>7x10 cm<br>7x20 cm | Multiple sizes<br>for FDA cleared<br>surgical staplers | 5x8cm to<br>30x30cm | Subject device<br>sizes are identical<br>to the predicate<br>device. |
| Surface Area | Maximum<br>140cm² | Maximum<br>140cm2 | Maximum<br>600cm2 | Maximum<br>900cm2 | Subject surface<br>area is within<br>range of SIS<br>devices |
| Operating<br>Principle | Supports<br>fixation by<br>sutures, tacks,<br>or staples | Supports<br>fixation by<br>sutures, tacks,<br>or staples | Supports<br>fixation by<br>sutures, tacks,<br>or staples | Supports<br>fixation by<br>sutures, tacks,<br>or staples | Identical |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
| Shelf-Life | 3 years | 1.5 years | 3 years | 3 years | Identical to<br>reference devices |
| Requires Re-<br>hydration? | Yes | Yes | Yes | Yes | Identical |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the Kerecis logo in blue and orange. The logo is a stylized version of the company name, with the "C" in "Kerecis" being orange. Above the logo, the text "K202430 Page 4 of 5" is visible, indicating this is a page from a document.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the Kerecis logo in blue and yellow. The logo is a stylized wordmark with the letters "k," "e," "r," "e," "c," "i," and "s." The "c" in "Kerecis" is colored yellow, while the rest of the letters are blue. Above the logo, the text "K202430 Page 5 of 5" is visible.
## 5.9 Conclusion
The data provided within this submission support substantial equivalence of the subject device to the predicate device with regards to intended use, technological characteristics including principles of operation, performance characteristics, and device safety.
For use in reinforcement of plastic surgery soft tissue defects, Kerecis Reconstruct is safe, effective, and substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.