K251805 · Siemens Medical Solutions USA, Inc. · JAK · Oct 15, 2025 · Radiology
Device Facts
Record ID
K251805
Device Name
syngo.CT Dual Energy
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
JAK · Radiology
Decision Date
Oct 15, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
syngo.CT Dual Energy is designed to operate with CT images based on two different X-ray spectra. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials. syngo.CT Dual Energy combines images acquired with low and high energy spectra to visualize this information. Depending on the region of interest, contrast agents may be used. The general functionality of the syngo.CT Dual Energy application is as follows: - Bone Marrow 2) - Bone Removal 1) - Brain Hemorrhage 1) - Gout Evaluation 1) - Hard Plaques 1) - Heart PBV - Kidney Stones 1) 2) 3) - Liver VNC 1) - Lung Mono 1) - Lung Perfusion 1) - Lung Vessels 1) - Monoenergetic 1) 2) - Monoenergetic Plus 1) 2) - Optimum Contrast 1) 2) - Rho/Z 1) 2) - SPP (Spectral Post-Processing Format) 1) 2) - SPR (Stopping Power Ratio) 1) 2) - Virtual Non-Calcium (VNCa) 1) 2) - Virtual Unenhanced 1) The availability of each feature depends on the Dual Energy scan mode. 1) This functionality supports data from Siemens Healthineers Photon-Counting CT scanners acquired in QuantumPlus modes. 2) This functionality supports data from Siemens Healthineers Photon-Counting CT scanners acquired in QuantumPeak modes. 3) Kidney Stones is designed to support the visualization of the chemical composition of kidney stones and especially the differentiation between uric acid and non-uric acid stones. For full identification of the kidney stone, additional clinical information should be considered such as patient history and urine testing. Only a well-trained radiologist can make the final diagnosis upon consideration of all available information. The accuracy of identification is decreased in obese patients.
Device Story
Post-processing software application for CT scanners; processes dual-energy or photon-counting CT (PCCT) data. Inputs: CT images acquired with high/low energy spectra. Operation: Performs material decomposition based on energy-dependent attenuation coefficients; generates color overlays, virtual non-contrast (VNC) images, and quantitative maps (e.g., electron density, effective atomic number, stopping power ratio). Used in clinical settings by radiologists to improve visualization of chemical composition (e.g., iodine concentration, bone/soft tissue separation, kidney stone analysis). Output: Visualized images and quantitative data displayed on workstation. Clinical decision-making: Assists in diagnosis of hemorrhage, perfusion, and tissue characterization; provides baseline density measurements. Benefits: Enables material differentiation without additional non-contrast scans; supports clinical assessment of fibrosis/cirrhosis and plaque characterization.
Clinical Evidence
Bench testing and clinical reader studies. Liver VNC (Fat Map) validated on clinical PCCT scans (mean deviation -3.8%, LoAs within +/- 10%). ECV maps validated on phantom data (error <2%) and compared to US-SWE in 25 cases (significant rank correlation). Bone Removal and Hard Plaques evaluated via three-reader studies; bone removal rated sufficient in >90% of cases; hard plaque labeling accurate in >85% of ratings. Quantitative performance of Brain Hemorrhage and Lung PBV compared between DSDE and PCCT using phantoms; results showed comparable accuracy. No clinical prospective trials required; bench/phantom/reader studies support equivalence.
Technological Characteristics
Software-based post-processing application. Operates on CT image data. Algorithms perform material decomposition, image reconstruction, and quantitative analysis (e.g., Rho/Z, SPR). Connectivity: DICOM compliant. Software lifecycle: ANSI AAMI IEC 62304 compliant. Deployment: Workstation-based. No hardware components; purely software.
Indications for Use
Indicated for humans aged 21 years and older. Used for qualitative and quantitative post-processing of CT images acquired with two different X-ray spectra to visualize chemical composition of body materials. Specific applications include bone marrow, bone removal, brain hemorrhage, gout, hard plaques, heart PBV, kidney stones, liver VNC, lung analysis, monoenergetic imaging, optimum contrast, Rho/Z, SPP, SPR, and virtual unenhanced imaging. Contraindications: Accuracy of kidney stone identification is decreased in obese patients.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 15, 2025
Siemens Medical Solutions USA, Inc.
% Kenny M Bello
Regulatory Affairs Professional
810 Innovation Drive
KNOXVILLE, TN 37932
Re: K251805
Trade/Device Name: syngo.CT Dual Energy
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: JAK
Dated: August 8, 2025
Received: September 12, 2025
Dear Kenny M. Bello:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251805 - Kenny M. Bello
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251805 - Kenny M. Bello
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K251805
Device Name
syngo.CT Dual Energy
Indications for Use (Describe)
syngo.CT Dual Energy is designed to operate with CT images based on two different X-ray spectra.
The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials. syngo.CT Dual Energy combines images acquired with low and high energy spectra to visualize this information. Depending on the region of interest, contrast agents may be used.
The general functionality of the syngo.CT Dual Energy application is as follows:
- Bone Marrow 2)
- Bone Removal 1)
- Brain Hemorrhage 1)
- Gout Evaluation 1)
- Hard Plaques 1)
- Heart PBV
- Kidney Stones 1) 2) 3)
- Liver VNC 1)
- Lung Mono 1)
- Lung Perfusion 1)
- Lung Vessels 1)
- Monoenergetic 1) 2)
- Monoenergetic Plus 1) 2)
- Optimum Contrast 1) 2)
- Rho/Z 1) 2)
- SPP (Spectral Post-Processing Format) 1) 2)
- SPR (Stopping Power Ratio) 1) 2)
- Virtual Non-Calcium (VNCa) 1) 2)
- Virtual Unenhanced 1)
The availability of each feature depends on the Dual Energy scan mode.
1) This functionality supports data from Siemens Healthineers Photon-Counting CT scanners acquired in QuantumPlus modes.
2) This functionality supports data from Siemens Healthineers Photon-Counting CT scanners acquired in QuantumPeak modes.
3) Kidney Stones is designed to support the visualization of the chemical composition of kidney stones and especially the differentiation between uric acid and non-uric acid stones. For full identification of the kidney stone, additional clinical information should be considered such as patient history and urine testing. Only a well-trained radiologist can make the final diagnosis upon consideration of all available information. The accuracy of identification is decreased in obese patients.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740
{4}
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 2 of 2
{5}
SIEMENS Healthineers
# K251805 - 510(k) Summary
syngo.CT Dual Energy (VC10)
# 1. Identification of the Submitter
| Submitter / Primary Contact Person | Kenny M Bello
Regulatory Affairs
monsuru.bello@siemens-healthineers.com
+1(202) 856-6099 |
| --- | --- |
| Secondary Contact Person | Clayton Ginn
Regulatory Affairs
clayton.ginn@siemens-healthineers.com
+1 (865) 898-2692 |
| Submitter Address | Siemens Medical Solutions, Inc. USA
Molecular Imaging
810 Innovation Drive
Knoxville, TN 37932
Establishment Registration Number: 1034973 |
| Legal Manufacturer | Siemens Healthcare GmbH
Siemensstr 1
D-91301 Forchheim, Germany
Establishment Registration Number: 3004977335 |
| Importer/Distributor | Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19355
Establishment Registration Number: 2240869 |
# 2. Device Name and Classification
Product Name: syngo.CT Dual Energy
Propriety Trade Name: syngo.CT Dual Energy
Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
CFR Section: 21 CFR §892.1750
Device Class: Class II
Product Code: JAK
# 3. Predicate Devices
## Predicate Device:
Trade Name: syngo.CT Dual Energy
Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
CFR Section: 21 CFR §892.1750
Device Class: Class II
Product Code: JAK
K-Number: K241757
Siemens Medical Solutions USA, Inc.
{6}
SIEMENS Healthineers
# Reference Device:
Trade Name: syngo.via RT Image Suite
Classification Name: Medical charged-particle radiation therapy system
Classification Panel: Radiology
CFR Section: 21 CFR §892.5050
Device Class: Class II
Product Code: MUJ
K-Number: K232799
# 4. Device Description
Dual energy offers functions for qualitative and quantitative post-processing evaluations. syngo.CT Dual Energy is a post-processing application consisting of several post-processing application classes that can be used to improve the visualization of the chemical composition of various energy dependent materials in the human body when compared to single energy CT. Depending on the organ of interest, the user can select and modify different application classes or parameters and algorithms.
Different body regions require specific tools that allow the correct evaluation of data sets. syngo.CT Dual Energy provides a range of application classes that meet the requirements of each evaluation type. The different application classes for the subject device can be combined into one workflow.
The product is intended to be used for at least 21-year-old humans.
# 5. Indications for Use
syngo.CT Dual Energy is designed to operate with CT images based on two different X-ray spectra. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials. syngo.CT Dual Energy combines images acquired with low and high energy spectra to visualize this information. Depending on the region of interest, contrast agents may be used.
The general functionality of the syngo.CT Dual Energy application is as follows:
- Bone Marrow²)
- Bone Removal¹)
- Brain Hemorrhage¹)
- Gout Evaluation¹)
- Hard Plaques¹)
- Heart PBV
- Kidney Stones¹)²)³)
- Liver VNC¹)
- Lung Mono¹)
- Lung Perfusion¹)
- Lung Vessels¹)
- Monoenergetic¹)²)
- Monoenergetic Plus¹)²)
- Optimum Contrast¹)²)
- Rho/Z¹)²)
- SPP (Spectral Post-Processing Format)¹)²)
Siemens Medical Solutions USA, Inc.
{7}
SIEMENS Healthineers
- SPR (Stopping Power Ratio)¹) ²)
- Virtual Non-Calcium (VNCa)¹) ²)
- Virtual Unenhanced¹)
The availability of each feature depends on the Dual Energy scan mode.
1) This functionality supports data from Siemens Healthineers Photon-Counting CT scanners acquired in QuantumPlus modes.
2) This functionality supports data from Siemens Healthineers Photon-Counting CT scanners acquired in QuantumPeak modes.
3) Kidney Stones is designed to support the visualization of the chemical composition of kidney stones and especially the differentiation between uric acid and non-uric acid stones. For full identification of the kidney stone, additional clinical information should be considered such as patient history and urine testing. Only a well-trained radiologist can make the final diagnosis upon consideration of all available information. The accuracy of identification is decreased in obese patients.
# 6. Indications for Use Comparison to the Predicate Device
| Subject Device
syngo.CT Dual Energy
(SOMARIS/8 VC10) | Predicate Device
syngo.CT Dual Energy
(SOMARIS/8 VB80) |
| --- | --- |
| syngo.CT Dual Energy is designed to operate with CT images based on two different X-ray spectra.
The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials.
syngo.CT Dual Energy combines images acquired with low and high energy spectra to visualize this information.
Depending on the region of interest, contrast agents may be used.
The general functionality of the syngo.CT Dual Energy application is as follows:
• Bone Marrow²)
• Bone Removal¹)
• Brain Hemorrhage¹)
• Gout Evaluation¹)
• Hard Plaques¹)
• Heart PBV
• Kidney Stones¹) ²) ³)
• Liver VNC¹)
• Lung Mono¹)
• Lung Perfusion¹)
• Lung Vessels¹)
• Monoenergetic¹) ²)
• Monoenergetic Plus¹) ²)
• Optimum Contrast¹) ²)
• Rho/Z¹) ²) | syngo.CT Dual Energy is designed to operate with CT images based on two different X-ray spectra.
The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials.
syngo.CT Dual Energy combines images acquired with low and high energy spectra to visualize this information.
Depending on the region of interest, contrast agents may be used.
The general functionality of the syngo.CT Dual Energy application is as follows:
• Monoenergetic ¹)
• Brain Hemorrhage ¹)
• Gout Evaluation ¹)
• Lung Vessels ¹)
• Heart PBV
• Bone Removal
• Lung Perfusion ¹)
• Lung Mono ¹)
• Liver VNC ¹)
• Monoenergetic Plus ¹)
• Virtual Unenhanced ¹)
• Bone Marrow
• Hard Plaques
• Rho/Z |
Siemens Medical Solutions USA, Inc.
{8}
SIEMENS Healthineers
| • SPP (Spectral Post-Processing Format)^{1) 2)}
• SPR (Stopping Power Ratio)^{1) 2)}
• Virtual Non-Calcium (VNCa)^{1) 2)}
• Virtual Unenhanced^{1)}
The availability of each feature depends on the Dual Energy scan mode.
1) This functionality supports data from Siemens Healthineers Photon-Counting CT scanners acquired in QuantumPlus modes.
2) This functionality supports data from Siemens Healthineers Photon-Counting CT scanners acquired in QuantumPeak modes.
3) Kidney Stones is designed to support the visualization of the chemical composition of kidney stones and especially the differentiation between uric acid and non-uric acid stones. For full identification of the kidney stone, additional clinical information should be considered such as patient history and urine testing. Only a well-trained radiologist can make the final diagnosis upon consideration of all available information. The accuracy of identification is decreased in obese patients. | • Kidney Stones^{1) 2)}
• SPR (Stopping Power Ratio)
• SPP (Spectral Post-Processing Format)^{1)}
• Optimum Contrast^{1)}
The availability of each feature depends on the Dual Energy scan mode.
1) This functionality supports data from Siemens Healthineers Photon-Counting CT scanners acquired in QuantumPlus modes.
2) Kidney Stones is designed to support the visualization of the chemical composition of kidney stones and especially the differentiation between uric acid and non-uric acid stones. For full identification of the kidney stone, additional clinical information should be considered such as patient history and urine testing. Only a well-trained radiologist can make the final diagnosis upon consideration of all available information. The accuracy of identification is decreased in obese patients. |
| --- | --- |
The changes of the IFU-Statement as compared to the predicate device are depicted in bold font.
- For the changes conducted in SOMARIS/8 VC10, five application classes support Photon-Counting CT scanners: Bone Removal, Hard Plaques, SPR, Rho/Z and Virtual Non-Calcium (VCNa)
- The following classes support in SOMARIS/8 VC10 the QuantumPeak scan-mode: SPR, Rho/Z and Virtual Non-Calcium (VCNa)
- Comparing both IFU-Statements, an additional footnote has been incorporated in order to document which application classes support the QuantumPeak scan-mode. The footnote is as follows: "2) This functionality supports data from Siemens Healthineers Photon-Counting CT scanners acquired in QuantumPeak modes."
# 7. Comparison of Technological Characteristics with the Predicate Device
The differences between the above referenced predicate device are listed at a high-level in the following table:
Siemens Medical Solutions USA, Inc.
{9}
SIEMENS
Healthineers
| Extensions of the subject device | Subject Device | Predicate Device |
| --- | --- | --- |
| SPR (Stopping Power Ratio) | syngo.CT Dual Energy (SOMARIS/8 VC10) | syngo.CT Dual Energy (SOMARIS/8 VB80) |
| | The Rapid Result Technology output type SPR allows for the calculation of stopping power images, which show the expected energy loss of protons in the material at each voxel position.
**Modification:**
Quantum Spectral Imaging (PCCT) and QuantumPeak scan-mode support. | The Rapid Result Technology output type SPR allows for the calculation of stopping power images, which show the expected energy loss of protons in the material at each voxel position. |
| Brain Hemorrhage | The application class visualizes the contrast agent concentration in the case of brain hemorrhage without use of an additional non-contrast scan. Basis for this approach is a material decomposition into contrast agent and soft tissue. The results of this calculation are shown as color overlay to anatomical, grayscale information.
**Modification:**
Warning about non-quantitative use of Quantum Spectral Imaging Scans (PCCT) has been removed. | The application class visualizes the contrast agent concentration in the case of brain hemorrhage without use of an additional non-contrast scan. Basis for this approach is a material decomposition into contrast agent and soft tissue. The results of this calculation are shown as color overlay to anatomical, grayscale information. |
| Lung Analysis
Lung Perfusion | The application class visualizes the contrast agent concentration in the lung without use of an additional non-contrast scan. The basis for this approach is a material decomposition into contrast agent and soft tissue. The results of this calculation are shown as color overlay to anatomical, grayscale information. An optional lung isolation is provided.
**Modification:**
Warning about non-quantitative use of Quantum Spectral Imaging Scans (PCCT) has been removed. | The application class visualizes the contrast agent concentration in the lung without use of an additional non-contrast scan. The basis for this approach is a material decomposition into contrast agent and soft tissue. The results of this calculation are shown as color overlay to anatomical, grayscale information. An optional lung isolation is provided. |
| Bone Removal | This application class removes bone or dense plastic from CT angiography (CTA) data sets. The prerequisite for a bone removal is the decomposition of the material into its component parts: blood, contrast agent, and bone.
**Modification:**
Quantum Spectral Imaging (PCCT) scan-mode support. | This application class removes bone or dense plastic from CT angiography (CTA) data sets. The prerequisite for a bone removal is the decomposition of the material into its component parts: blood, contrast agent, and bone. |
Siemens Medical Solutions USA, Inc.
{10}
SIEMENS Healthineers
| Extensions of the subject device | Subject Device | Predicate Device |
| --- | --- | --- |
| Liver VNC | syngo.CT Dual Energy (SOMARIS/8 VC10) | syngo.CT Dual Energy (SOMARIS/8 VB80) |
| | The Liver VNC application class allows to visualize the contrast agent concentration in the liver without an additional non-contrast scan, even if there are irregular fatty infiltrations or necrotic areas. The application class generates virtual non-contrast (VNC) images by subtracting iodine from the Dual Energy data sets. The VNC images can be used for baseline density measurements. In addition, this application class provides fat map evaluation for contrast enhanced CT-scans as well as for non-contrast scans.
**Modification:**
• Fat Map for Quantum Spectral Imaging Scans (PCCT)
• Support of normalized iodine maps called Extracellular Volume (ECV) | The Liver VNC application class allows to visualize the contrast agent concentration in the liver without an additional non-contrast scan, even if there are irregular fatty infiltrations or necrotic areas. The application class generates virtual non-contrast (VNC) images by subtracting iodine from the Dual Energy data sets. The VNC images can be used for baseline density measurements. In addition, this application class provides fat map evaluation for contrast enhanced CT-scans as well as for non-contrast scans. |
| Hard Plaques | This application class highlights calcified plaques within large vessels even if they have CT values that are comparable to the neighboring contrast medium.
**Modification:**
Quantum Spectral Imaging (PCCT) scan-mode support. | This application class highlights calcified plaques within large vessels even if they have CT values that are comparable to the neighboring contrast medium. |
| Rho/Z | This application class measures electron density as well as effective atomic number and performs a basic characterization of the material of interest.
**Modification:**
Quantum Spectral Imaging (PCCT) and QuantumPeak scan-mode support. | This application class measures electron density as well as effective atomic number and performs a basic characterization of the material of interest. |
The remaining applications classes/functions in syngo.CT Dual Energy remain unchanged compared to the predicate version:
- SPP (Spectral Post-Processing Format)
- Monoenergetic
- Gout Evaluation
- Lung Analysis (Lung Vessels, Lung Mono)
- Heart PBV
- Monoenergetic Plus
- Virtual Unenhanced
- Kidney Stones
- Lung Lobe Segmentation
Siemens Medical Solutions USA, Inc.
{11}
SIEMENS Healthineers
# 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
## Software Verification and Validation
Software Documentation for Enhanced documentation Level per FDA’s Guidance Document “Guidance for the Content of Premarket Submissions for Device Software Functions” issued on June 14, 2023, is also included as part of this submission. The Risk Analysis was completed, and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claim of substantial equivalence.
## Summary of Performance Testing
### Liver VNC – Fat Map for Photon-Counting CT
The feature Liver VNC – Fat Map was tested on clinical PCCT scans with contrast enhancement, comparing the results against fat values calculated from non-contrast CT scans. The limits of agreement (LoAs) between both methods were [-8.6%, -1.0%] with a mean deviation of -3.8%. Thus, the feature Fat Map overestimates the reference value on average, but the LoAs remain within the prescribed tolerance of +/- 10%.
Together with the supporting literature it is concluded that the feature is safe and effective.
### Liver VNC – Extracellular Volume Maps for Photon-Counting CT
The software was validated on phantom data. The test results show that the implementation of ECV maps in syngo.CT Dual Energy is accurate up to an absolute error of 2% ECV, even in the presence of fat in the liver parenchyma.
The ECV was compared against ultrasound derived shear-wave elastography (US-SWE) in 25 clinical cases. A statistical relevant rank correlation between hepatic ECV values and US-SWE was found.
A literature review showed that CT-derived ECV is well investigated and might be useful for the assessment of fibrosis and cirrhosis in the liver.
On conclusion, the device is safe to compute ECV maps from PCCT scans in the equilibrium phase and the ECV value is clinically useful.
### Bone Removal for Photon-Counting CT
The clinical performance of the application class DE Body Bone Removal applied to PCCT data was evaluated with a three-reader study. The readers assessed the quality of the bone removal by means of residual bone and its impact on the image interpretability. The bone removal was rated to be of sufficient quality with no residual bone interfering with image interpretation in over 90% of the ratings. Additionally, the performance regarding the visualization of vessels by means of artifacts induced by the application was assessed. Each reader evaluated over 90% of the cases to be free of major artifacts in vessels.
The clinical performance of the application class DE Head Bone Removal on PCCT data was evaluated with a three-reader study. The readers assessed the quality of the bone removal by means of residual bone and its impact on the image interpretability. The bone removal was concordantly rated to be sufficient with residual bone that interferes with image interpretation in less than 2% of the ratings. Additionally, the performance regarding the visualization of vessels by means of artifacts induced by the application
Siemens Medical Solutions USA, Inc.
{12}
SIEMENS Healthineers
was assessed. Major artifacts were found predominantly for the vertebral arteries and parts of the internal carotid arteries. Thus, applying the DE Head Bone Removal application class to PCCT data may provide limited result quality for vertebral arteries, and segments of the internal carotid arteries close to the carotid canal. This limitation is pointed out in the Instructions for Use.
## Hard Plaques for Photon-Counting CT
The clinical performance of the application class DE Hard Plaques applied to PCCT data was evaluated with a three-reader study. The readers assessed the labelling of hard plaques in the abdominal aorta and iliac arteries by means of true positive and false positive labelling in combination with its clinical implicated type of stenosis. The DE Hard Plaques application class labels most of the relevant plaques in above 85% of the ratings. No case of global failure was present. Furthermore, false positive labelling was found in less than 20% of the ratings. The most severe stenosis implied was moderate in less than 10% of the cases in which false positives labelling occurred. Otherwise, no or mild stenosis was implied.
## VNCa for Photon-Counting CT
A phantom with inserts mimicking red and yellow bone marrow combined with several concentrations of hydroxyapatite (HA) was scanned with photon-counting CT. The test demonstrated that the calcium signal is removed from the material inserts containing HA. Although over-subtraction was noted, the relative contrast between yellow and red marrow at the same HA concentration was retained in the VNCa maps.
## Rho/Z/SPR for Photon-Counting CT
The tests are based on two phantoms with sizes comparable to a human head and a human abdomen. The phantoms were equipped with inserts of known material composition. Measurements were performed in clinically relevant routine protocols at typical dose levels.
The material parameters relative electron density (RED, Rho) and effective atomic number (EAN, Z) were automatically derived with the application syngo.CT DE Rho/Z and provided a basic characterization of the materials of interest. This was demonstrated with tissue-equivalent materials covering a wide range of human tissues (lung, adipose, soft tissue, bone). Furthermore, the test demonstrated that the derived RED values were within ± 3% (with respect to water) and EAN values were within ± 1.
The material parameter stopping power ratio (SPR) was automatically derived with the application syngo.CT DE SPR and provided the stopping-power ratio of a material of interest at a proton energy of 100 MeV. This was demonstrated with tissue-equivalent materials covering a wide range of human tissues (lung, adipose, soft tissue, bone). Furthermore, the test demonstrated that the derived SPR values were within ± 3% with respect to water.
## Brain Hemorrhage and Lung PBV – Quantitative Use with Photon-Counting CT
The quantitative performance of virtual-non contrast (VNC) and iodine material density (IMD) maps computed by DE Brain Hemorrhage was compared between dual-source dual energy (DSDE) and photon-counting CT (PCCT) data. The clinically relevant situation was simulated by a phantom of 20 cm diameter and six material inserts with varying iodine concentration and blood density.
The IMD measurements deviated by -0.2 mg l / ml to 0.3 mg l / ml from the nominal values of the material inserts with no systematic difference between DSDE and PCCT scans. The VNC values showed small
Siemens Medical Solutions USA, Inc.
{13}
SIEMENS Healthineers
variations across the scan modes in the order of 10 HU. Increased blood density resulted in substantially increased VNC values, while iodine enhancement was clearly removed.
In summary, the virtual non-contrast and iodine material density maps created from photon-counting CT scans show a similar accuracy as the maps created from Dual Source Dual Energy scans.
The quantitative performance of Lung PBV computed by DE Lung Analysis was compared between dual-source dual energy (DSDE) and photon-counting CT (PCCT) data for different scan modes. The clinically relevant situation was simulated by a phantom of $20\mathrm{cm}$ diameter and different material inserts mimicking lung tissue with iodine. Additionally, iodine in water at different iodine concentrations were tested.
The maximum absolute deviation from PCCT data is generally lower compared to the one from DSDE for all iodine in water inserts as well as most of the lung tissue and iodine inserts. Exceptions are the lung tissue with $0.0\mathrm{mg / ml}$ and $1.0\mathrm{mg / ml}$ iodine concentration, for the which the deviations from one PCCT mode were slightly higher, but still comparable to the one from DSDE data.
In summary, the iodine enhancement provided by Lung PBV as part of DE Lung Analysis derived from PCCT scans show a comparable accuracy as the one from DSDE scans.
# Risk Analysis
The risk analysis was completed, and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
Siemens hereby certifies that syngo.CT Dual Energy meets the following FDA Recognized Consensus standards listed below:
| Standard | Version | Content | FDA Recognition Number (if applicable) |
| --- | --- | --- | --- |
| ANSI AAMI IEC 62304 | :2006/A1:2016 | Medical device software - Software life cycle processes [Including Amendment 1 (2016)] | 13-79 |
| NEMA PS 3.1 - 3.20 2023e | :2022 | Digital Imaging and Communications in Medicine (DICOM) Set | 12-352 |
| ISO 14971 | :2019 | Application of Risk Management to Medical Devices | 5-125 |
| IEC 62366-1 | Edition 1.1 2020-06 CONSOLIDATED VERSION | Medical devices - Part 1: Application of usability engineering to medical devices | 5-129 |
| ISO 15223-1 | Fourth edition 2021-07 | Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements | 5-134 |
Siemens Medical Solutions USA, Inc.
{14}
SIEMENS Healthineers
| Standard | Version | Content | FDA Recognition Number (if applicable) |
| --- | --- | --- | --- |
| ISO 20417:2021 | First edition 2021-04
Corrected version 2021-12 | Medical devices - Information to be supplied by the manufacturer | 5-135 |
# 9. Conclusion
syngo.CT Dual Energy has the same intended use and similar indication for use as the predicate device. The subject device syngo.CT Dual Energy does not have changes in fundamental scientific technology compared to the predicate devices. The technological characteristics such as image visualization, operating platform, and image measurement are the same as the predicate device. For the subject device, syngo.CT Dual Energy, Siemens used the same testing with the same workflows as used to clear the predicate device. Siemens considers syngo.CT Dual Energy to be as safe, as effective, and with performance substantially equivalent to the commercially available predicate device.
Siemens Medical Solutions USA, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.