K241757 · Siemens Medical Solutions USA, Inc. · JAK · Jan 3, 2025 · Radiology
Device Facts
Record ID
K241757
Device Name
syngo.CT Dual Energy
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
JAK · Radiology
Decision Date
Jan 3, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials. syngo. CT Dual Energy combines images acquired with low and high energy spectra to visualize this information. Depending on the region of interest, contrast agents may be used.
Device Story
syngo.CT Dual Energy is a post-processing software application for CT scanners; processes high and low energy spectral data to visualize material chemical composition. Used by radiologists in clinical settings to generate color overlays, virtual non-contrast images, and quantitative maps (e.g., iodine concentration, stopping power). Supports various clinical workflows including brain hemorrhage assessment, lung perfusion analysis, and kidney stone characterization. Operates on Siemens SOMARIS/8 platform; integrates with Photon-Counting CT (PCCT) and Dual Source CT data. Output assists clinicians in identifying perfusion defects, differentiating stone types, and characterizing tissue without additional non-contrast scans. Benefits include reduced need for repeat scans and enhanced diagnostic visualization of material-specific information.
Clinical Evidence
Clinical performance evaluated via two-reader studies. DE Brain Hemorrhage study (n=28 features) demonstrated complementary visualization of iodine and hemorrhage. Lung Mono study (n=20 cases) showed agreement between clot position and perfusion defects. Lung Analysis study (n=33 cases) confirmed visualization of reduced local iodine content in lung parenchyma/arteries. Software verification and validation testing confirmed all specifications met acceptance criteria.
Technological Characteristics
Post-processing software for CT systems. Operates on SOMARIS/8 platform. Supports DICOM standards. Implements material decomposition algorithms based on energy-dependent attenuation coefficients. Connectivity: Networked/DICOM. Software lifecycle per ANSI AAMI IEC 62304. Risk management per ISO 14971. Usability per IEC 62366-1.
Indications for Use
Indicated for visualization of chemical composition of anatomical regions using dual-energy CT data. Features include Monoenergetic, Brain Hemorrhage, Gout Evaluation, Lung Vessels, Heart PBV, Bone Removal, Lung Perfusion, Lung Mono, Liver VNC, Monoenergetic Plus, Virtual Unenhanced, Bone Marrow, Hard Plaques, Rho/Z, Kidney Stones, SPR, SPP, and Optimum Contrast. Kidney Stones feature aids differentiation of uric acid vs non-uric acid stones; requires clinical correlation. Accuracy decreased in obese patients.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 3, 2025
Siemens Medical Solutions USA, Inc. % Kenny M Bello Regulatory Affairs Professional 810 Innovation Drive KNOXVILLE, TN 37932
Re: K241757
Trade/Device Name: syngo.CT Dual Energy Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: December 5, 2024 Received: December 5, 2024
Dear Kenny M Bello:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241757
Device Name sygno.CT Dual Energy
#### Indications for Use (Describe)
The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials. syngo. CT Dual Energy combines images acquired with low and high energy spectra to visualize this information. Depending on the region of interest, contrast agents may be used.
The general functionality of the syngo.CT Dual Energy application is as follows:
- · Monoenergetic 1)
- · Brain Hemorrhage 1)
- · Gout Evaluation 1)
- · Lung Vessels 1)
- · Heart PBV
- · Bone Removal
- · Lung Perfusion 1)
- · Lung Mono 1 )
- · Liver VNC 1)
- · Monoenergetic Plus 1)
- · Virtual Unenhanced 1)
- Bone Marrow
- · Hard Plaques
- Rho/Z
- · Kidney Stones 1) 2)
- · SPR (Stopping Power Ratio)
- · SPP (Spectral Post-Processing Format) 1)
- · Optimum Contrast 1)
The availability of each feature depends on the Dual Energy scan mode.
1) This functionality supports data from Siemens Healthineers Photon-Counting CT scanners acquired in QuantumPlus modes.
2) Kidney Stones is designed to support the visualization of the chemical composition of kidney stones and especially the differentiation between uric acid stones. For full identification of the kidney stone, additional clinical information should be considered such as patient history and urine testing. Only a well-trained radiologist can make the final diagnosis upon consideration of all available information. The accuracy of identification is decreased in obese patients.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
# 510(k) Summary for K241757
syngo.CT Dual Energy (VB80)
### 1. Identification of the Submitter
| Submitter / Primary Contact Person | Kenny M Bello | |
|------------------------------------|-----------------------------------------------------------------------|--|
| | Regulatory Affairs | |
| | monsuru.bello@siemens-healthineers.com | |
| | +1(202) 856-6099 | |
| Secondary Contact Person | Clayton Ginn | |
| | Regulatory Affairs | |
| | clayton.ginn@siemens-healthineers.com | |
| | +1 (865) 898-2692 | |
| Submitter Address | Siemens Medical Solutions, Inc. USA | |
| | Molecular Imaging | |
| | 810 Innovation Drive | |
| | Knoxville, TN 37932 | |
| | Establishment Registration Number: 1034973 | |
| Legal Manufacturer | Siemens Healthcare GmbH<br>Siemensstr 1<br>D-91301 Forchheim, Germany | |
| | | |
| | | |
| | Establishment Registration Number: 3004977335 | |
| Importer/Distributor | Siemens Medical Solutions USA, Inc. | |
| | 40 Liberty Boulevard | |
| | Malvern, PA 19355 | |
| | Establishment Registration Number: 2240869 | |
### 2. Device Name and Classification
| Product Name: | syngo.CT Dual Energy |
|-----------------------|----------------------------------|
| Propriety Trade Name: | syngo.CT Dual Energy |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
# 3. Predicate Devices
# Predicate Device:
| Trade Name: | syngo.CT Dual Energy |
|-----------------------|----------------------------------|
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
| K-Number: | K232155 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots.
# 4. Device Description
Dual energy offers functions for qualitative and quantitative post-processing evaluations. syngo.CT Dual Energy is a post-processing application consisting of several post-processing application classes that can be used to improve the visualization of the chemical composition of various energy dependent materials in the human body when compared to single energy CT. Depending on the organ of interest, the user can select and modify different application classes or parameters and algorithms.
Different body regions require specific tools that allow the correct evaluation of data sets. syngo.CT Dual Energy provides a range of application classes that meet the requirements of each evaluation type. The different application classes for the subject device can be combined into one workflow.
The modifications for the software version SOMARIS/8 VB80 are listed in section 7.
### 5. Indications for Use
The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials. syngo.CT Dual Energy combines images acquired with low and high energy spectra to visualize this information. Depending on the region of interest, contrast agents may be used.
The general functionality of the syngo.CT Dual Energy application is as follows:
- Monoenergetic 1)
- · Brain Hemorrhage 1)
- · Gout Evaluation 1)
- Lung Vessels 1)
- Heart PBV
- Bone Removal
- · Lung Perfusion 1)
- Lung Mono 1)
- Liver VNC 1)
- · Monoenergetic Plus 1)
- Virtual Unenhanced 1)
- Bone Marrow
- Hard Plaques
- Rho/Z
- Kidney Stones 1) 2)
- SPR (Stopping Power Ratio)
- · SPP (Spectral Post-Processing Format) 1)
- · Optimum Contrast 1)
The availability of each feature depends on the Dual Energy scan mode.
- 1) This functionality supports data from Siemens Healthineers Photon-Counting CT scanners acquired in QuantumPlus modes.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a cluster of orange dots arranged in a pattern.
- 2) Kidney Stones is designed to support the visualization of the chemical com and especially the differentiation between uric acid and non-uric acid stones. For full identification of the kidney stone, additional clinical information should be considered such as patient history and urine testing. Only a well-trained radiologist can make the final diagnosis upon consideration of all available information. The accuracy of identification is decreased in obese patients.
| Subject Device | Predicate Device |
|--------------------------------------------------------------|--------------------------------------------------------------|
| syngo.CT Dual Energy | syngo.CT Dual Energy |
| (SOMARIS/8 VB80) | (SOMARIS/8 VB71) |
| The various materials of an anatomical region of interest | The various materials of an anatomical region of interest |
| have different attenuation coefficients, which depend on the | have different attenuation coefficients, which depend on the |
| used energy. These differences provide information on the | used energy. These differences provide information on the |
| chemical composition of the scanned body materials. | chemical composition of the scanned body materials. |
| syngo.CT Dual Energy combines images acquired with low | syngo.CT Dual Energy combines images acquired with low |
| and high energy spectra to visualize this information. | and high energy spectra to visualize this information. |
| Depending on the region of interest, contrast agents may be | Depending on the region of interest, contrast agents may be |
| used. | used. |
| The general functionality of the syngo.CT Dual Energy | The general functionality of the syngo.CT Dual Energy |
| application is as follows: | application is as follows: |
| • Monoenergetic 1) | • Monoenergetic 1) |
| • Brain Hemorrhage 1) | • Brain Hemorrhage |
| • Gout Evaluation 1) | • Gout Evaluation 1) |
| • Lung Vessels 1) | • Lung Vessels |
| • Heart PBV | • Heart PBV |
| • Bone Removal | • Bone Removal |
| • Lung Perfusion 1) | • Lung Perfusion |
| • Lung Mono 1) | • Liver VNC 1) |
| • Liver VNC 1) | • Monoenergetic Plus 1) |
| • Monoenergetic Plus 1) | • Virtual Unenhanced 1) |
| • Virtual Unenhanced 1) | • Bone Marrow |
| • Bone Marrow | • Hard Plaques |
| • Hard Plaques | • Rho/Z |
| • Rho/Z | • Kidney Stones 1) 2) |
| • Kidney Stones 1) 2) | • SPR (Stopping Power Ratio) |
| • SPR (Stopping Power Ratio) | • SPP (Spectral Post-Processing Format) 1) |
| • SPP (Spectral Post-Processing Format) 1) | • Optimum Contrast 1) |
| • Optimum Contrast 1) | The availability of each feature depends on the Dual Energy |
| The availability of each feature depends on the Dual Energy | scan mode. |
| scan mode. | 1) This functionality supports data from Photon-Counting |
| 1) This functionality supports data from Siemens | CT scanners. |
| Healthineers Photon-Counting CT scanners acquired in | 2) Kidney Stones is designed to support the visualization of |
| QuantumPlus modes. | the chemical composition of kidney stones and |
| 2) Kidney Stones is designed to support the visualization of | especially the differentiation between uric acid and non- |
| the chemical composition of kidney stones and | uric acid stones. For full identification of the kidney |
| especially the differentiation between uric acid and non- | stone, additional clinical information should be |
| uric acid stones. For full identification of the kidney | considered such as patient history and urine testing. |
| stone, additional clinical information should be | Only a well-trained radiologist can make the final |
| considered such as patient history and urine testing. | diagnosis upon consideration of all available |
# 6. Indications for Use Comparison to the Predicate Device
Only a well-trained radiologist can make the final
{8}------------------------------------------------
diagnosis upon consideration of all available information. The accuracy of identification is decreased in obese patients.
information. The accuracy of identific in obese patients.
The changes of the IFU-Statement as compared to the predicate device are depicted in bold font. For the changes conducted in SOMARIS/7 VB80, two application classes support Photon-Counting CT scanners: Brain Hemorrhage and Lung Analysis (represented by Lung Vessels and Lung Perfusion). The additional image type called Lung Mono has been included into the IFU statement as well, to reflect the entire functionality of syngo.CT Dual Energy. Furthermore, footnote 1 has been specified to reflect the possibility that some application classes are able to support data from Siemens Healthineers Photon-Counting CT scanners acquired in QuantumPlus modes. The intention of this change is to reflect the distinction between the photon-counter scan-mode "QuantumPlus" and the additional scan-mode called "QuantumPeak".
# 7. Comparison of Technological Characteristics with the Predicate Device
| Feature | Subject Device | Predicate Device |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | syngo.CT Dual Energy<br>(SOMARIS/8 VB80) | syngo.CT Dual Energy<br>(SOMARIS/8 VB71) |
| Brain Hemorrhage | The application class visualizes the contrast agent<br>concentration in the case of brain hemorrhage<br>without use of an additional non-contrast scan.<br>Basis for this approach is a material<br>decomposition into contrast agent and soft tissue.<br>The results of this calculation are shown as color<br>overlay to anatomical, grayscale information. This<br>output may be used for complimentary<br>visualization of iodine and overlay maps.<br>Modification:<br>Quantum Spectral Imaging (PCCT) scan-mode<br>support. | The application class visualizes the contrast agent<br>concentration in the case of brain hemorrhage<br>without use of an additional non-contrast scan.<br>Basis for this approach is a material decomposition<br>into contrast agent and soft tissue. The results of<br>this calculation are shown as color overlay to<br>anatomical, grayscale information. |
| Lung Analysis<br>- Lung Vessels | This application class visualizes the contrast<br>agent concentration in the lung without use of<br>an additional non-contrast scan. The basis for<br>this approach is a material decomposition into<br>contrast agent and soft tissue. The results of this<br>calculation are shown as color overlay to<br>anatomical, grayscale information.<br>Modification:<br>Quantum Spectral Imaging (PCCT) scan-mode<br>support. | This application class visualizes the contrast agent<br>concentration in the lung without use of an<br>additional non-contrast scan. The basis for this<br>approach is a material decomposition into contrast<br>agent and soft tissue. The results of this calculation<br>are shown as color overlay to anatomical, grayscale<br>information. |
| Feature | Subject Device | Predicate Device |
| | syngo.CT Dual Energy<br>(SOMARIS/8 VB80) | syngo.CT Dual Energy<br>(SOMARIS/8 VB71) |
| - Lung Perfusion | The application class visualizes the contrast<br>agent concentration in the lung without use of<br>an additional non-contrast scan. The basis for<br>this approach is a material decomposition into<br>contrast agent and soft tissue. The results of this<br>calculation are shown as color overlay to<br>anatomical, grayscale information. An optional<br>lung isolation is provided. Lung Mono series<br>should not be interpreted in isolation. | The application class visualizes the contrast agent<br>concentration in the lung without use of an<br>additional non-contrast scan. The basis for this<br>approach is a material decomposition into contrast<br>agent and soft tissue. The results of this calculation<br>are shown as color overlay to anatomical, grayscale<br>information. An optional lung isolation is provided. |
| | Modification:<br>Quantum Spectral Imaging (PCCT) scan-mode<br>support for Lung PBV.<br>Availability of an additional image type (Lung<br>Mono) for dual source dual energy and<br>Quantum Spectral Imaging (PCCT) scan modes. | |
| Monoenergetic Plus | This application class simulates images that are<br>equivalent to images scanned with a single<br>photon energy beam, depending on the energy<br>(keV). | This application class simulates images that are<br>equivalent to images scanned with a single photon<br>energy beam, depending on the energy (keV). |
| | Modification:<br>Dual Source Dual Energy (DSDE) data from PCCT<br>(QuantumPeak) can be loaded. | |
| Bone Marrow | This application class allows to visualize the<br>bone marrow composition based on<br>nonenhanced CT data. It generates virtual non-calcium (VNCa)<br>images by subtracting calcium from the Dual Energy data sets. | This application class allows to visualize the bone<br>marrow composition based on nonenhanced CT<br>data. It generates virtual non-calcium (VNCa)<br>images by subtracting calcium from the Dual Energy<br>data sets. |
| | Modification:<br>Dual Source Dual Energy (DSDE) data from PCCT<br>(QuantumPeak) can be loaded. | |
| Kidney Stones | This application class visualizes the chemical<br>differences between kidney stones by<br>decomposing the kidney stones into its<br>component parts: tissue, uric acid, and oxalate<br>(calcium stone). | This application class visualizes the chemical<br>differences between kidney stones by decomposing<br>the kidney stones into its component parts: tissue,<br>uric acid, and oxalate (calcium stone). |
| | Modification:<br>Dual Source Dual Energy (DSDE) data from PCCT<br>(QuantumPeak) can be loaded. | |
| Virtual Unenhanced | This application class visualizes the contrast<br>agent concentration in soft body tissue without<br>the need of an additional non-contrast scan. | This application class visualizes the contrast agent<br>concentration in soft body tissue without the need<br>of an additional non-contrast scan. |
| | Modification:<br>Dual Spiral/Twin Spiral scan-mode support for<br>Virtual Unenhanced. | |
The differences between the above referenced predicate device are listed at a high-level in the following table:
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.
The remaining applications classes/functions in syngo.CT Dual Energy rema the predicate version:
- SPR (Stopping Power Ratio)
- SPP (Spectral Post-Processing Format)
- Monoenergetic
- Gout Evaluation
- Bone Removal
- Hard Plaques
- Rho/Z
- Lung Lobe Segmentation
### 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Software Verification and Validation
Software Documentation for Enhanced documentation Level per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Device Software Functions" issued on June 14, 2023, is also included as part of this submission. The Risk Analysis was completed, and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claim of substantial equivalence.
### Summary of Performance Testing
### DE Brain Hemorrhage for Photon-Counting CT
We performed a two-reader study to assess the clinical performance of the application class DE Brain Hemorrhage on photon-counting CT (PCCT) data, comparing the reading results on VNC and Iodine images with later follow-up data. The study demonstrated that DE Brain Hemorrhage can provide complementary visualization of iodine and overlay maps as demonstrated in a two reader study where 24 of 28 features were identified by both readers. Of all 28 features, 19 were assessed to be pure hemorrhage, 6 to be pure iodine staining, and 3 to be a mix of both, hemorrhage and iodine.
### Lung Mono for Dual Source Dual Energy
A two-reader study on twenty cases of suspected pulmonary embolisms was performed. The test demonstrated adequate agreement between the position of obstructing clots and the location of perfusion defects in the Lung Mono maps. Thus, the Lung Mono maps can provide complementary visualization for the purposes of identifying and locating perfusion defects caused by obstructing clots. Lung Mono images should not be interpreted alone.
#### Lung Analysis (Lung PBV, Lung Mono, and Lung Vessels) for Photon-Counting CT
A two-reader study on 33 cases of suspected pulmonary embolisms was performed to assess the performance of Lung PBV, Lung Mono, and Lung Vessel maps.
The test demonstrated that the three image types visualize reduced local iodine content, either in the lung parenchyma or in the lung arteries. The test was based on obstructing clots, which are the cause of pulmonary embolism and can be identified in conventional CT images by trained radiologists.
#### Risk Analysis
The risk analysis was completed, and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a cluster of orange dots.
verification and validation of the device was found acceptable to support equivalence.
Siemens hereby certifies that syngo.CT Dual Energy meets the following FDA Recognized Consensus standards listed below:
| Standard | Version | Content | FDA<br>Recognition<br>Number<br>(if applicable) |
|-----------------------------|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| ANSI AAMI IEC<br>62304 | :2006/A1:2016 | Medical device software - Software life cycle<br>processes [Including Amendment 1 (2016)] | 13-79 |
| NEMA PS 3.1 -<br>3.20 2022d | :2022 | Digital Imaging and Communications in<br>Medicine (DICOM) Set | 12-349 |
| ISO 14971 | :2019 | Application of Risk Management to Medical<br>Devices | 5-125 |
| IEC 62366-1 | Edition 1.1<br>2020-06<br>CONSOLIDATED<br>VERSION | Medical devices - Part 1: Application of<br>usability engineering to medical devices | 5-129 |
| ISO 15223-1 | Fourth edition<br>2021-07 | Medical devices - Symbols to be used with<br>information to be supplied by the<br>manufacturer - Part 1: General requirements | 5-134 |
| ISO 20417:2021 | First edition<br>2021-04<br>Corrected<br>version 2021-<br>12 | Medical devices - Information to be supplied<br>by the manufacturer | 5-135 |
# 9. Conclusion
syngo.CT Dual Energy has the same intended use and similar indication for use as the predicate device. The subject device syngo.CT Dual Energy does not have changes in fundamental scientific technology compared to the predicate devices. The technological characteristics such as image visualization, operating platform, and image measurement are the same as the predicate device. For the subject device, syngo.CT Dual Energy, Siemens used the same testing with the same workflows as used to clear the predicate device. Siemens considers syngo.CT Dual Energy to be as safe, as effective, and with performance substantially equivalent to the commercially available predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.