NIM Essence™ EMG Endotracheal Tube (NIMEID060); NIM Essence™ EMG Endotracheal Tube (NIMEID065); NIM Essence™ EMG Endotracheal Tube (NIMEID070); NIM Essence™ EMG Endotracheal Tube (NIMEID075); NIM Essence™ EMG Endotracheal Tube (NIMEID080)
K251672 · Medtronic Xomed, Inc. · ETN · Feb 10, 2026 · Ear, Nose, Throat
Device Facts
Record ID
K251672
Device Name
NIM Essence™ EMG Endotracheal Tube (NIMEID060); NIM Essence™ EMG Endotracheal Tube (NIMEID065); NIM Essence™ EMG Endotracheal Tube (NIMEID070); NIM Essence™ EMG Endotracheal Tube (NIMEID075); NIM Essence™ EMG Endotracheal Tube (NIMEID080)
Applicant
Medtronic Xomed, Inc.
Product Code
ETN · Ear, Nose, Throat
Decision Date
Feb 10, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Indications for Use
The EMG Endotracheal Tube is intended for use as a means of providing both an open airway for patient ventilation and for intraoperative monitoring of EMG activity of the intrinsic laryngeal musculature when connected to an appropriate EMG monitor.
Device Story
Flexible, reinforced PVC endotracheal tube with inflatable cuff; features integrated Flexible Printed Circuit (FPC) electrodes wrapped around shaft. Electrodes contact laryngeal muscles near vocal cords to capture EMG signals during surgery. Device connects to external EMG monitoring consoles (e.g., Medtronic NIM 3.0 or NIM Vital) to facilitate intraoperative nerve monitoring. Used in surgical settings; operated by clinicians. Output displayed on external monitor to assist in nerve location and monitoring; helps prevent nerve damage during procedures. Stainless steel reinforcement provides structural rigidity to prevent kinking. Sterile, single-use device.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including shelf-life validation, EO sterilization validation (ISO 11135), biocompatibility (ISO 10993-1, ISO 18562-1), electrical safety (ANSI/AAMI ES 60601-1/IEC 60601-1), EMC (IEC 60601-1-2), and mechanical/functional performance testing per ISO 5361.
Technological Characteristics
Reinforced PVC endotracheal tube with inflatable cuff; stainless steel reinforcement. FPC electrodes with Flexible Printed Carbon Overprint. Type BF applied part. Sterilized via Ethylene Oxide (EtO). Connectivity: Wired connection to external EMG monitoring console. Complies with ISO 10993-1, ISO 18562-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62366-1, and ISO 5361.
Indications for Use
Indicated for continuous monitoring of nerves supplying laryngeal musculature during surgical procedures. Not intended for postoperative use.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 10, 2026
Medtronic Xomed, Inc.
Emily Davis
Principal Regulatory Affairs Specialist
6743 Southpoint Drive North
Jacksonville, Florida 32216
Re: K251672
Trade/Device Name: NIM Essence™ EMG Endotracheal Tube
Regulation Number: 21 CFR 874.1820
Regulation Name: Surgical Nerve Stimulator/Locator
Regulatory Class: Class II
Product Code: ETN
Dated: May 30, 2025
Received: May 30, 2025
Dear Emily Davis:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251672 - Emily Davis
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251672 - Emily Davis
Page 3
Sincerely,
Joyce C. Lin -S
for Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251672 | |
| Device Name NIM Essence™ EMG Endotracheal Tube | |
| Indications for Use (Describe) The EMG Tube is indicated for use where continuous monitoring of the nerves supplying the laryngeal musculature is required during surgical procedures. The EMG tube is not intended for postoperative use. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Medtronic
Medtronic Xomed, Inc.
6743 Southpoint Drive N.
Jacksonville, FL 32216
United States of America
www.medtronic.com
Tel. 904-296-9600
# 510(k) Summary
Company (submitter):
Medtronic Xomed, Inc.
6743 Southpoint Drive North
Jacksonville, Florida 32216 USA
Telephone Number: (904) 296-9600
Date Prepared:
February 9, 2026
510(k) Number
K251672
Official Correspondent:
Emily Davis
Principal Regulatory Affairs Specialist
Email Address: emily.r.davis@medtronic
Alternative Contact:
Leo Chen
Regulatory Affairs Manager
Email Address: leo.chen@medtronic.com
I. DEVICE
| Proprietary (Trade) Name: | NIM Essence™ EMG Endotracheal Tube |
| --- | --- |
| Common Name: | Stimulator, Nerve |
| Regulation Name: | Surgical nerve stimulator/locator |
| Regulation Number: | 21 CFR 874.1820 |
| Product Code: | ETN |
| Classification: | 2 |
| Panel: | Ear, Nose, & Throat (ENT) |
II. PREDICATE DEVICE
The NIM Essence™ Endotracheal Tube are substantially equivalent in intended use and technological characteristics to the following predicate device:
| 510(k) Number | 510(k) Device Name | 510(k) Clearance Date |
| --- | --- | --- |
| K112686 | NIM Trivantage™ EMG Endotracheal Tube | June 27, 2012 |
NOTE: Product under K112686 was submitted and cleared under the trade name "Next Gen EMG Endotracheal Tube (final name TBD)" where the final name was determined to be "NIM Trivantage™" shortly after 510(k) clearance.
NIM Essence™ EMG Endotracheal Tube
Traditional 510(k)
{5}
Medtronic
Medtronic Xomed, Inc.
6743 Southpoint Drive N.
Jacksonville, FL 32216
United States of America
www.medtronic.com
Tel. 904-296-9600
# III. DEVICE DESCRIPTION
The Medtronic NIM Essence™ EMG Endotracheal Tube (ETT) is a flexible, reinforced PVC endotracheal tube with an inflatable cuff. FPC (Flexible Printed Circuit) is attached and wrapped around the main shaft of the endotracheal tube, and silicone cover tube overlays on the FPC. The FPC electrodes are exposed only for a short distance, approximately 30 mm, slightly superior to the cuff. The electrodes are designed to make contact with the laryngeal muscles around the patient's vocal cords to facilitate electromyographic (EMG) monitoring of the laryngeal musculature during surgery when connected to an EMG monitoring device. Both the tube and the cuff are manufactured from material that allows the tube to readily conform to the shape of the patient's trachea with minimal trauma to tissues. The NIM Essence™ EMG Endotracheal Tube is packaged as a sterile single-use device.
The Medtronic NIM™ 3.0 (cleared via K083124 on February 27, 2009) and NIM Vital™ (cleared via K200759 on October 28, 2020) nerve monitors are the recommended EMG monitors for use with the EMG Endotracheal Tube. All references to connections and technical specifications for EMG monitor contained in the instructions are made with references to the NIM™ family of devices. Medtronic recommends that the user consult the NIM™ User's Guide for determining proper settings of the NIM™ console and instructions for intraoperative EMG monitoring and motor nerve location and stimulation. The NIM Essence™ EMG Endotracheal Tube is Type BF applied part.
# IV. INTENDED USE
The EMG Endotracheal Tube is intended for use as a means of providing both an open airway for patient ventilation and for intraoperative monitoring of EMG activity of the intrinsic laryngeal musculature when connected to an appropriate EMG monitor.
# V. INDICATIONS FOR USE
The EMG Tube is indicated for use where continuous monitoring of the nerves supplying the laryngeal musculature is required during surgical procedures. The EMG tube is not intended for postoperative use.
# VI. SUBSTANTIAL EQUIVALENCE
The EMG Endotracheal tube is substantially equivalent to the predicate device based on comparison of indications for use and technological characteristics where additions to the subject devices instructions for use were made being the only difference. Technological Characteristics of our device compared to each predicate is shown in the following table and in the Substantial Equivalence section of this submission.
NIM Essence™ EMG Endotracheal Tube
Traditional 510(k)
Page 2 of 6
{6}
Medtronic
Medtronic Xomed, Inc.
6743 Southpoint Drive N.
Jacksonville, FL 32216
United States of America
www.medtronic.com
Tel. 904-296-9600
Substantial Equivalence/Device Comparison
| Feature/Attribute | NIM Essence™ EMG Endotracheal Tube (Subject Device) | NIM Trivantage™ EMG Endotracheal Tube / K112686 (Predicate Device) | Comparison |
| --- | --- | --- | --- |
| Product Code | ETN | ETN | Same |
| Regulation Number | 21 CFR 874.1820 | 21 CFR 874.1820 | Same |
| Regulation Description | Surgical nerve stimulator/locator | Surgical nerve stimulator/locator | Same |
| Classification | Class II | Class II | Same |
| Common Name | Stimulator, Nerve | Stimulator, Nerve | Same |
| Indications for Use | The EMG Tube is indicated for use where continuous monitoring of the nerves supplying the laryngeal musculature is required during surgical procedures. The EMG tube is not intended for postoperative use. | The EMG Tube is indicated for use where continuous monitoring of the nerves supplying the laryngeal musculature is required during surgical procedures. The EMG tube is not intended for postoperative use. | Same |
| Single Use | Yes | Yes | Same |
| Shelf Life | 3 Years | 4 Years | Difference, see below discussion 1 |
| Principles of Operation | Electrical stimulation | Electrical stimulation | Same |
| Tube Inner Diameter (ID) | 6mm-8mm | 5mm-9mm | Similar, the subject device provides the tube inner diameter size within the range of the predicate device. |
NIM Essence™ EMG Endotracheal Tube
Traditional 510(k)
Page 3 of 6
{7}
Medtronic
Medtronic Xomed, Inc.
6743 Southpoint Drive N.
Jacksonville, FL 32216
United States of America
www.medtronic.com
Tel. 904-296-9600
| Feature/Attribute | NIM Essence™ EMG Endotracheal Tube (Subject Device) | NIM Trivantage™ EMG Endotracheal Tube / K112686 (Predicate Device) | Comparison |
| --- | --- | --- | --- |
| Number of Laryngeal Surface Electrodes | 4 | 4 | Same |
| Electrode Surface Material | Flexible Printed Carbon Overprint | Conductive Silver Ink | Difference, see below discussion 2 |
| Tube Material | PVC | PVC | Same |
| Cuff Material | PVC | PVC | Same |
| Reinforcing Material | Stainless Steel | None | Difference, see below discussion 3 |
| Sterilization Method | EtO | EtO | Same |
| Duration of Use | Limited (less than 24 hours) | Limited (less than 24 hours) | Same |
| Biocompatibility | Yes, according to ISO 10993-1 and ISO 18562-1 | Yes, according to ISO 10993-1 and ISO 18562-1 | Same |
## Difference Discussion 1:
The subject device NIM Essence™ EMG Endotracheal Tube declares 3 years as the shelf life. Accelerating aging test report and real time aging test report have proved 3 years shelf life of the subject device. Package accelerating aging test report and package real time aging test report prove that the subject device package can maintain a sterile barrier for the entirety of the proposed shelf-life. There is no impact of the shelf life difference to the device safety or performance.
## Difference Discussion 2:
Firstly, the flexible printed circuit (FPC) as a part of NIM Essence EMG endotracheal tube is biocompatible and it can meet the requirements for biological safety performance. The NIM Essence tube complies with ISO 10993-1 and FDA biocompatibility guidance 2023. Cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity tests are performed on the subject device to find it is biocompatible.
NIM Essence™ EMG Endotracheal Tube
Traditional 510(k)
Page 4 of 6
{8}
Medtronic
Medtronic Xomed, Inc.
6743 Southpoint Drive N.
Jacksonville, FL 32216
United States of America
www.medtronic.com
Tel. 904-296-9600
Secondly, NIM Essence EMG endotracheal tubes with FPC design are assessed for electrical safety and electromagnetic compatibility and can meet ANSI AAMI ES 60601-1/IEC 60601-1, IEC 60601-1-2 and IEC TS 60601-4-2.
Thirdly, NIM Essence EMG endotracheal tubes with FPC design can comply with IEC 60601-1-6 and IEC 62366-1. NIM Essence EMG Tube Usability Engineering Plan and Summary Report proves that NIM Essence EMG tube has been found to be adequately safe and effective for the intended users, uses and use environments.
## Difference Discussion 3:
The difference in the design of the subject device and predicate device is in the reinforcement. The subject device uses the stainless steel as the reinforcing material to make tube more rigid, thus, reduces softness. The technological characteristic differences do not affect safety and effectiveness as demonstrated through performance testing. The NIM Essence™ EMG endotracheal tube shall not kink after bending the tube 90° around a pre-defined 30mm radius of curvature. For the collapse test, the specified steel ball shall pass freely through the NIM Essence tube after conditionings. The reinforcing material is to make the tube more rigid. The stainless steel used as the reinforcing material in the Medtronic endotracheal tubes for more than 30 years. Moreover, PVC tube with stainless steel reinforcing material from Well Lead Medical was cleared under K073383 on July 10, 2008. Well Lead Medical manufactures a similar PVC tube with stainless steel reinforcing material. The bench testing performed verifies that the performance and materials of the subject device are substantially equivalent in terms of critical performance characteristics to the predicate device.
## VII. PERFORMANCE DATA
The following non-clinical testing was performed to demonstrate substantial equivalence to the predicate device.
1. Stability Test for single use device – NIM Essence ETT was validated to achieve 3 years shelf life for sterility and performance.
2. Sterilization Test – NIM Essence ETT is sterilized using a validated EO sterilization process which complies with ISO 11135 Sterilization of health-care products - Ethylene oxide - Requirements for the development validation and routine control of a sterilization process for medical devices - Amendment 1: Revision of Annex E Single batch release.
3. Biocompatibility Test – NIM Essence ETT passed all required testing under ISO 10993-1, FDA guidance “Use of international Standard ISO 10993-1” issued on September 8, 2023 and ISO 18562-1.
4. Electrical Safety Test per ANSI/AAMI ES 60601-1/IEC 60601-1 and Electromagnetic Compatibility (EMC) Test per IEC 60601-1-2 and IEC TS 60601-4-2.
NIM Essence™ EMG Endotracheal Tube
Traditional 510(k)
{9}
Medtronic
Medtronic Xomed, Inc.
6743 Southpoint Drive N.
Jacksonville, FL 32216
United States of America
www.medtronic.com
Tel. 904-296-9600
5. Performance Test (Bench) – NIM Essence ETT passed all pre-defined testing for mechanical and electrical functionality, compatibility with NIM Vital and NIM 3.0, and per ISO 5361.
6. Performance Test (Usability) – NIM Essence ETT testing has been found to be adequately safe and effective for the intended users, uses and use environments while passing compliance to IEC 62366-1 and IEC 60601-1-6.
## Clinical Testing
Not applicable - No clinical study is deemed necessary since the substantial equivalence has been sufficiently demonstrated by non-clinical studies.
## VIII. CONCLUSION
Based upon the supporting data summarized above, we concluded that the subject device NIM Essence™ EMG Endotracheal Tube is substantially equivalent to the legally-marketed predicate device K112686 and does not raise new or different concerns or questions of safety and effectiveness related to the predicate device. Therefore, it is the opinion of the submitter that the NIM Essence™ EMG ETT are as safe, as effective and performs as well as the legally marketing predicate device.
NIM Essence™ EMG Endotracheal Tube
Traditional 510(k)
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.