K112686 · Medtronic Xomed, Inc. · ETN · Jun 27, 2012 · Ear, Nose, Throat
Device Facts
Record ID
K112686
Device Name
NEXT GEN EMG ENDOTRACHEAL TUBE
Applicant
Medtronic Xomed, Inc.
Product Code
ETN · Ear, Nose, Throat
Decision Date
Jun 27, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Indications for Use
The Next Generation EMG Endotracheal Tube is intended for use as a means of providing both an open airway for patient ventilation and for intraoperative monitoring of EMG activity of the laryngeal musculature when connected to an appropriate EMG monitor. The EMG Tube is indicated for use when continuous monitoring of the integrity of the nerves supplying the laryngeal musculature is required during surgical procedures.
Device Story
Device is a PVC endotracheal tube with inflatable cuffs and four integrated conductive silver electrodes; used for airway management and intraoperative EMG monitoring of laryngeal musculature. Electrodes contact vocal cords to capture EMG signals; signals transmitted to external multi-channel EMG neuromonitoring device. Used in surgical settings by clinicians; output allows real-time monitoring of nerve integrity to guide surgical decision-making and prevent nerve damage. Benefits include simultaneous ventilation and nerve monitoring.
Clinical Evidence
Bench testing included biocompatibility (cytotoxicity, intracutaneous reactivity, sensitization) and functional testing (compliance, electrode impedance, wire harness, product integration, ink adhesion). Animal study confirmed performance equivalence to predicates for monitoring nerve integrity during surgery.
Technological Characteristics
PVC (DEHP-free) endotracheal tube; four conductive silver ink electrodes; inflatable high-volume, low-pressure cuffs. Connectivity via external multi-channel EMG neuromonitoring device. Sterile, single-use. Biocompatibility per ISO 10993-1; compliance testing per ISO 5361:1999.
Indications for Use
Indicated for patients requiring continuous intraoperative monitoring of laryngeal nerve integrity during surgical procedures.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
k112686
JUN 2 7 2012
Medtronic Xomed, Inc. Section 5.0 510(k) Summary
Traditional 510(k) Submission, Next Gen EMG Endotracheal Tube
# Section 5.0 510(k) Summary
A. 510(k) Owner
Medtronic Xomed, Inc 6743 Southpoint Drive North Jacksonville, Florida 32216-0980 USA 904-279-7586 904-296-2386 (FAX)
B. Contact Information
Marek Pawlowski Senior Regulatory Affairs Specialist Medtronic Xomed, Inc marek.pawlowski@medtronic.com
Next Gen EMG Endotracheal Tube
C. Date Summary Prepared
D. Proprietary Name
E. Device Name
Trade name:
Common/Usual Name: Classification Name:
Next Gen EMG Endotracheal Tube (final name TBD) EMG ET Tube, EMG Endotracheal Tube Surgical nerve stimulator/locator (21 CFR 874.1820, Product Code ETN, Class II)
#### F. Predicate Devices:
| Trade Name: | NuVasive® NeuroVision® EMG Endotracheal Tube |
|-------------------------|------------------------------------------------------------------------------------|
| Common/Usual Name: | EMG ET Tube |
| Classification Name: | Surgical nerve stimulator/locator<br>(21 CFR 874.1820, Product Code ETN, Class II) |
| Premarket Notification: | K094054 |
November 10, 2011
Xomed EMG Endotracheal Tube Trade Name: EMG ET Tube Common/Usual Name: Classification Name: Surgical nerve stimulator/locator (21 CFR 874.1820, Product Code ETN, Class II) K925640 Premarket Notification:
## G. Purpose of Submission:
:
Design change in the material of the tube from silicone in the predicate device Xomed EMG Tube to PVC in the proposed device and design change of electrodes from stainless steel wire in the predicate device to conductive silver ink in the proposed device.
{1}------------------------------------------------
K112686
Traditional 510(k) Submission, Next Gen EMG Endotracheal Tube Medtronic Xomed, Inc. Section 5.0 510(k) Summary
## H. Device Description
The Next Gen EMG Endotracheal Tube is a standard-size PVC (DEHP-free polyvinyl chloride) endotracheal tube with inflatable high volume, low pressure cuffs. Each tube is imprinted with four (two pairs) conductive silver electrodes. The conductive silver is printed on the PVC of the main shaft of the endotracheal tube and exposed only for a short distance, approximately 40 mm, slightly superior to the cuff, for contacting the vocal cords. The electrodes are designed to make contact with the patient's vocal cords to facilitate electromyographic (EMG) monitoring of the laryngeal musculature during surgery when connected to a multi-channel EMG neuromonitoring device.
#### I. Intended Use/Indications for Use:
#### Indications for Use:
The EMG Tube is indicated for use when continuous monitoring of the integrity of the nerves supplying the laryngeal musculature is required during surgical procedures.
#### Intended Use:
The Next Generation EMG Endotracheal Tube is intended for use as a means of providing both an open airway for patient ventilation and for intraoperative monitoring of EMG activity of the laryngeal musculature when connected to an appropriate EMG monitor.
## J. Substantial Equivalence
The subject device: Next Gen EMG Endotracheal Tube is substantially equivalent to predicate devices cleared by FDA for commercial distribution in the United States. A comparison of technological characteristics in areas including design, material composition, function, packaging and sterilization is presented in the table below:
| | Subject Device<br>Next Gen EMG<br>Endotracheal Tube | Predicate Device(s) | |
|--------------------------------|-----------------------------------------------------|-----------------------------------------------------------------|---------------------------------------------|
| | | NuVasive®<br>NeuroVision® EMG<br>Endotracheal Tube<br>(K094054) | Xomed EMG Endotracheal<br>Tube<br>(K925640) |
| Laryngeal Surface<br>Electrode | Yes | Yes | Yes |
| Endolaryngeal<br>Location | Yes | Yes | Yes |
| Number of<br>Electrodes | 4 | 4 | 4 |
| Electrode Surface<br>Material | Conductive Silver Ink | Conductive Silver Ink | Stainless Steel Wire |
| Tube & Cuff<br>Material | PVC | PVC | Silicone |
| Reinforcing Material | None | None | Stainless Steel Wire |
| Tube Dimensions | Various Dimensions | Various Dimensions | Various Dimensions |
| Sterilization &<br>Packaging | Sterile, single use only | Sterile, single use only | Sterile, single use only |
{2}------------------------------------------------
-k 112 686
Traditional 510(k) Submission, Next Gen EMG Endotracheal Tube Medtronic Xomed, Inc. Section 5.0 510(k) Summary
#### K. Conclusion
In house laboratories conducted studies verifying the design of the proposed device. Studies followed scientific protocols. The data were determined to be scientifically valid under 21 CFR 860.7.
Nonclinical testing was performed to demonstrate that the proposed Next Gen EMG Tube is substantially equivalent to the predicate devices.
The following testing was performed:
- Biocompatibility testing per ISO 10993-1 requirements, including: .
- cytotoxicity testing (per ISO 10993-5:2009) O
- intracutaneous reactivity testing (per ISO 10993-10:2010) o
- sensitization testing (per ISO 10993-10:2010) o
- Functional testing, including:
- o compliance testing per ISO 5361:1999
- electrode impedance test O
- wire harness test o
- product integration testing o
- o marking ink adhesion test.
Functional testing was performed using an animal model. The animal study provided evidence that the proposed device performed substantially equivalent to the predicate devices during monitoring of nerve integrity during surgical procedures.
The data presented in this Premarket Notification support that the subject device is safe and effective and performs in the same manner as the predicate devices when used in accordance with the labeled directions for use and for the specified indication(s).
The risks of the subject device, as well as the benefits to the patient, are the same as those . attributed to the use of the predicate devices.
A direct comparison of key technological characteristics demonstrates that the proposed Next Gen EMG Endotracheal Tube is substantially equivalent to the predicate devices in areas including design, material composition, function, packaging and sterilization.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like bird with three stripes forming its body and wings. The bird is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged around the circumference of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Medtronic Xomed c/o Mr. Marek Pawlowski Senior Regulatory Affairs Specialist 6743 Southpoint Dr. North Jacksonville, FL 32216
JUN 2 7 2012
Re: K112686
Trade/Device Name: Next Generation EMG Endotracheal Tube Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical nerve stimulator/locator Regulatory Class: Class II Product Code: ETN Dated: June 15, 2012 Received: June 18, 2012
Dear Mr. Pawlowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Marek Pawlowski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fdagov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
E.L. Rummell MD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K112 686
Traditional 510(k) Submission, Next Gen EMG Endotracheal Tube Medtronic Xomed, Inc. Section 4.0 Indications for Use Statement
# Section 4.0 Indications for Use Statement
## Indications for Use
510(k) Number (if known): KIIQ 660
Next Generation EMG Endotracheal Tube (Final name to be Device Name: _ determined).
Intended Use/Indications for Use:
The Next Generation EMG Endotracheal Tube is intended for use as a means of providing both an open airway for patient ventilation and for intraoperative monitoring of EMG activity of the laryngeal musculature when connected to an appropriate EMG monitor.
The EMG Tube is indicated for use when continuous monitoring of the integrity of the nerves supplying the laryngeal musculature is required during surgical procedures.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
John Doucett
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear,
Division of Ophthalmic, Neurologices (Division Sign-Off) Division of Opinal
K112686
510(k) Number.
Image /page/5/Picture/18 description: The image shows the text "Prescription Use (Per 21 CFR 801.109)" in a bold, sans-serif font. The text is oriented diagonally from the bottom left to the upper right of the image. A dark line crosses out the text.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.