This product is suitable for making invisible orthodontic appliances by additive manufacturing (light curing 3D printing) process. The orthodontic appliance is designed for orthodontic treatment. It uses continuous gentle force to adjust tooth position, correct malocclusion, and maintain the results of completed orthodontic treatment. It can also be used for printing splints and night guards.
Device Story
Additively Manufactured Aligner Resin is a light-curing acrylic resin (acrylate oligomers/monomers) used to 3D print custom orthodontic appliances; including clear aligners, splints, and night guards. The resin is polymerized via UV light (385 nm or 405 nm). The workflow involves 3D printing, de-supporting, cleaning, and post-curing. Used in professional dental settings; prescribed by dental practitioners. Aligners are worn by patients for 20-22 hours/day, typically changed every 7-10 days to apply continuous gentle force for tooth movement. The device enables personalized orthodontic treatment and retention, benefiting patients by providing a removable, aesthetic alternative to traditional braces.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing per ISO 20795-2:2013 (Homogeneity, Surface Properties, Color, Porosity, Flexural Strength, Flexural Modulus, Water Solubility/Sorption) and biocompatibility testing per ISO 10993-1:2018 and ISO 7405:2018 (Cytotoxicity, Sensitization, Irritation, Acute/Subchronic Systemic Toxicity, Genotoxicity).
Technological Characteristics
Light-cured acrylate-based resin; polymerized at 385 nm or 405 nm. Mechanical properties: Flexural Modulus ~877.49 MPa, Ultimate Flexural Strength ~39.72 MPa. Complies with ISO 20795-2:2013. Non-sterile; 24-month shelf life. Compatible with validated 3D printers.
Indications for Use
Indicated for patients requiring orthodontic treatment to adjust tooth position, correct malocclusion, maintain orthodontic results, or for the fabrication of dental splints and night guards.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 27, 2025
Aidite (Qinhuangdao) Technology Co., Ltd.
% Boyle Wang
General Manager
Shanghai Truthful Information Technology Co., Ltd.
Room 1801, No. 161 East Lu Jiazui Rd., Pudong
Shanghai, 200120
CHINA
Re: K251415
Trade/Device Name: Additively Manufactured Aligner Resin
Regulation Number: 21 CFR 872.5470
Regulation Name: Orthodontic Plastic Bracket
Regulatory Class: Class II
Product Code: NXC, MQC
Dated: July 29, 2025
Received: July 30, 2025
Dear Boyle Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251415 - Boyle Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251415 - Boyle Wang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251415 | |
| Device Name Additively Manufactured Aligner Resin | |
| Indications for Use (Describe) This product is suitable for making invisible orthodontic appliances by additive manufacturing (light curing 3D printing) process. The orthodontic appliance is designed for orthodontic treatment. It uses continuous gentle force to adjust tooth position, correct malocclusion, and maintain the results of completed orthodontic treatment. It can also be used for printing splints and night guards. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
# 510(k) Summary
# K251415
This summary is submitted in accordance with 21 CFR 807.92.
## 1.0 Submission Sponsor
Name: Aidite (Qinhuangdao) Technology Co., Ltd.
Address 1: No.9 Dushan Road, Economic and Technological Development Zone, Qinhuangdao City, 066004 Hebei P.R.China
Address 2: No. 56 Tianchi Road, Economic and Technological Development Zone, Qinhuangdao City, 066004 Hebei P.R. China
Contact: Mr. Ye Zhanfu
Tel: +86 335 8587898
Fax: +86 335 8587198
## Designated Submission Correspondent
Contact: Mr. Boyle Wang
Address: Shanghai Truthful Information Technology Co., Ltd.
Room 1801, No. 161 East Lu Jiazui Rd., Pudong Shanghai, 200120 China
Tel: +86-21-50313932
Email: info@truthful.com.cn
Date of Preparation: Jul.29th, 2025
## 2.0 Device Information
Trade name: Additively Manufactured Aligner Resin
Common name: Additive Manufacturing (Light Curing) Orthodontie Resin
Model: DPAS-C
Classification name: Material, Tooth Shade, Resin
Production code: NXC,MQC
Regulation number: 21 CFR 872.5470
Classification: Class II
Panel: Dental
## 3.0 Identification of Predicate Device and Reference Device
Predicate Device 1:
{5}
510(k) Summary
510(k) Number: K212680
Trade/Product Name: LuxCreo Clear Aligner System
Manufacturer: LUXCREO INC.
Predicate Device 2:
510(k) Number: K221022
Trade/Product Name: NextDent Ortho Flex
Manufacturer: Vertex-Dental BV
## 4.0 Device Description
Additively Manufactured Aligner Resin is custom plastic aligner system which are a series of doctor prescribed clear removable aligners that are used as alternative treatment for the alignment of maloccluded or misaligned teeth. This series of aligners gently move the patient's teeth in small increments from their original state to a treated state.
Additively Manufactured Aligner Resin is a light-curing resin used to print orthodontic appliances. The resin is a light yellow transparent liquid. The appliance printed by a 3D printer is de-supported, cleaned and post-cured to obtain an orthodontic appliance that can correct the patient's malocclusion. This product is a light-curing 3D printing resin composed of acrylate resin oligomers and acrylate monomers as well as initiators and additives.
Additive manufacturing (light cured) orthodontic resin is a light cured acrylic resin commonly used in the manufacture of orthodontic appliances. The resin is stored in a black HDPE bottle according to the weight of the resin liquid. The color of the resin is colorless or yellowish liquid, polymerized by UV light at 385 nm or 405 nm.
## 5.0 Indication for Use Statement
This product is suitable for making invisible orthodontic appliance by additive manufacturing process (light curing 3D printing process). The orthodontic appliance is designed for orthodontic treatment. It uses continuous gentle force to adjust tooth position, correct malocclusion, and maintain the results of completed orthodontic treatment.
It can also be used for printing splint and night guard.
## 6.0 Non-clinical Test Conclusion
**Bench Testing:**
- The most applicable standard for mechanical characteristics determination of the subject device is FDA-recognized version ISO 20795-2:2013 Dentistry – Base polymers – Part 2: Orthodontic base polymers.
Performance testing including Homogeneity, Surface Properties, Forming
2 of page 6
{6}
510(k) Summary
Performance, Color, No Porosity, Flexural Strength, Flexural Modulus, Ultimate Flexural Strength, Water Solubility and Water Sorption.
All test results of above items were similar to that of the predicate device that meet the requirements of ISO 20795-2:2013. The performance characteristics of the subject device are comparable to those of the predicate device for this particular indication and raise no new or different questions of safety and effectiveness. Therefore, the subject device and the predicate device are substantially equivalent in physical properties.
- Printers Validation: Different printers are validated for processing (3D printing) using the candidate device. Additional printers post-510(k) clearance will be added to the labelling by means of the Quality Systems and within the validation plan presented in this 510(k).
## Biocompatibility Testing:
The nature of body contact of materials used in the design of the Additively Manufactured Aligner Resin were classified as being Surface Devices in contact with mucosal membrane with a permanent contact duration of >30 days, repeated use. The biocompatibility testing was performed according to FDA currently-recognized versions of biocompatibility consensus standards ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process and ISO 7405:2018 Dentistry – Evaluation of biocompatibility of medical devices used in dentistry.
The following biological safety aspects have been addressed:
- Cytotoxicity – ISO 10993-5
- Sensitization – ISO 10993-10
- Irritation – ISO 10993-23
- Acute Systemic Toxicity – ISO 10993-11
- Subchronic Systemic Toxicity – ISO 10993-11
- Genotoxicity – ISO 10993-3
Sterility and Shelf-Life Testing:
The device is provided non-sterile.
Based on shelf-life testing, the Additively Manufactured Aligner Resin has an unopened shelf life of 2 years. Once opened, the resin should be used within 60 days and must be protected from light exposure.
## 7.0 Technological Characteristics and Substantial Equivalence
The following table shows similarities and differences of use, design, and material between our device and the predicate devices.
Table 1- Comparison of Technology Characteristics
3 of page 6
{7}
510(k) Summary
| Item | Subject Device | | Predicate Device 1 | Predicate Device 2 | | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| 510(k) No. | K251415 | | K212680 | K221022 | | -- |
| Product Name | Additively Manufactured Aligner Resin | | LuxCreo Clear Aligner System | NextDent Ortho Flex | | -- |
| Product Code | NXC | MQC | NXC | MQC | KMY | Same |
| Regulation No. | 21 CFR 872.5470 | NA | 21 CFR 872.5470 | NA | 21 CFR 872.5525 | Same |
| Class | Class II | | Class II | Unclassified (pre amendment) | Class 1 | Same |
| Intended Use/Indication for Use | This product is suitable for making invisible orthodontic appliance by additive manufacturing process (light curing 3D printing process). The orthodontic appliance is designed for orthodontic treatment. It uses continuous gentle force to adjust tooth position, correct malocclusion, and maintain the results of completed orthodontic treatment. It can also be used for printing splint and night guard. | | LuxCreo Clear Aligner System is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e. all second molars). The LuxCreo Clear Aligner System repositions teeth by way of continuous gentle force. | NextDent Ortho Flex is a 3D print resin intended for the manufacturing of 3D printed dental splints and retainers. To retain the regulated dentition. NextDent Ortho Flex is intended exclusively for professional dental work. | | Same* (Analysis 1) |
| Mode of Action | Alignment of teeth by application of continuous gentle force, by sequential use of preformed plastic trays | | Alignment of teeth by application of continuous gentle force, by sequential use of preformed plastic trays | Not publiclt available | | Same with Predicate Device 1 |
| Method of Use | Each preformed plastic tray is worn by the patient as prescribed by the dental practitioner, usually for 7-10 days, usually no more than 15 days prior to using the next sequential aligner tray | | Each preformed plastic tray is worn by the patient as prescribed by the dental practitioner, usually a few weeks prior to using the next sequential aligner tray | Not publiclt available | | Same with Predicate Device 1 |
| Prescription | Rx | | Rx | Rx | | Same |
4 of page 6
{8}
510(k) Summary
| Use | | | | |
| --- | --- | --- | --- | --- |
| Duration of Use | Each set of aligners can be worn for approximately 2 weeks of 20-22 hours of wear per day or according to doctor’s prescription. | 20-22 hours/day | Not publiclt available | Same with Predicate Device 1 |
| Method of Manufacturing | Light-cured 3D printing | Light-cured 3D printing | Light-cured 3D printing | Same |
| Software Used for Ordering Workflow | Yes | Yes | Yes | Same |
| Application | Removable | Removable | Removable | Same |
| Materials | Light-cured polyurethane resin | Light-cured polyurethane resin | acrylate-based UV light cure resin | Similar |
| Shelf-Life | 24 months | Six-month | 18 months | Different (Analysis 2) |
| Performance Testing | Per ISO 20795-2:2013 | ISO 20795-2:2013 | ISO 20795-2:2013 | Same |
| Flexural Modulus | Average 877.49 MPa | 804 ± 64 MPa | Not publiclt available | Different (Analysis 3) |
| Ultimate Flexural Strength | Average 39.72 MPa | 23.6 ± 1.9 MPa | Not publiclt available | |
| Water Solubility | Average 3.05μg/mm³ | 3.668±1.0748 μg/mm³ | Not publiclt available | |
| Water Sorption | Average 29.94μg/mm³ | 19.952±6.6719 μg/mm³ | Not publiclt available | |
| Sterile | Non-sterile | Non-sterile | Non-sterile | Same |
| Biocompatibility | ISO 10993-5 In vitro Cytotoxicity
ISO 10993-10 Oral Mucosa Irritation
ISO 10993-23 Skin Irritation
ISO 10993-11 Acute Systemic Toxicity
ISO 10993-11 Subacute Systemic Toxicity
ISO 10993-3 Genotoxicity | ISO 10993-5 In vitro Cytotoxicity
ISO 10993-10 Oral Mucosa Irritation
Pyrogen ISO 10993-10 Skin Irritation in Rabbits
ISO 10993-10 Skin Sensitization in Guinea Pigs (Maximization Test)
ISO 10993-11 Acute Systemic Toxicity | • Cytotoxicity
• Sensitization
• Irritation or intra cutaneous reactivity
• Subacute/sub chronic systemic toxicity
• Genotoxicity
• Chemical characterization | Same |
5 of page 6
{9}
510(k) Summary
## Analysis*:
1) The Indication for Use of the subject device, Additively Manufactured Aligner Resin, is consistent with the combined indications of the identified predicate devices. The subject device incorporates the orthodontic therapeutic purpose of Predicate Device 1 and the dental splint manufacturing capability of Predicate Device 2. There are no new indications introduced, and the overall intended use remains consistent with legally marketed predicate devices, supporting substantial equivalence.
2) The difference in shelf life does not impact their substantial equivalence. The shelf-life validation report of the subject device demonstrates the stability of the products' performance over time, ensuring their reliability and safety during their respective shelf lives.
3) There are measurable differences among the Flexural Modulus and Ultimate Flexural Strength. The Flexural Modulus and Ultimate Flexural Strength of the subject device is better than the predicate device because the average values of the subject device were statistically significantly higher than that of the predicate device.
Water Solubility and Water Sorption values of the subject device and the predicate devices were slight different, all tested materials conform to the requirements of ISO 20795-2:2013, the differences observed are not expected to adversely affect the safety or effectiveness of the subject device.
Therefore, the subject device is considered substantially equivalent to the predicate devices in terms of performance characteristics.
## 8.0 Summary of Clinical Test
Clinical testing was not required for this submission.
## 9.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the subject device is substantially equivalent to the legally marketed predicated device.
6 of page 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.