K212680 · LuxCreo, Inc. · NXC · May 31, 2022 · Dental
Device Facts
Record ID
K212680
Device Name
LuxCreo Clear Aligner System
Applicant
LuxCreo, Inc.
Product Code
NXC · Dental
Decision Date
May 31, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LuxCreo Clear Aligner System is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e. all second molars). The LuxCreo Clear Aligner System repositions teeth by way of continuous gentle force.
Device Story
The LuxCreo Clear Aligner System consists of a series of custom, removable, clear plastic aligners designed to reposition teeth through continuous gentle force. The system is prescribed by dental health professionals. The manufacturing process utilizes 3D stereolithographic printing with specialty liquid polyurethane resins; printed aligners undergo UV-light curing to achieve final mechanical properties. Aligners are packaged in a specific sequence for patient use, typically worn 20-22 hours per day. The device is used in a clinical/home context where the physician manages the treatment plan and the patient wears the aligners to achieve incremental tooth movement. The system relies on software for the ordering and manufacturing workflow to translate the prescribed tooth displacement into the physical aligner design.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including biocompatibility (ISO 10993-3, -5, -6, -10, -11, USP Pyrogen), physical/mechanical property testing (flexural strength, modulus, hardness, stress relaxation, water solubility/sorption), and design verification/validation.
Technological Characteristics
Light-cured polyurethane resin; additive manufacturing via DLP/SLA 3D printing; UV-cured; removable aligner form factor; prescription use; software-assisted ordering/manufacturing workflow. Biocompatibility per ISO 10993 standards. Mechanical properties: Flexural Strength 23.6 ± 1.9 MPa (ISO 20795-2:2013); Flexural Modulus 1106.47±13.23 MPa (ASTM D790).
Indications for Use
Indicated for treatment of tooth malocclusion in patients with permanent dentition (all second molars).
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Luxcreo Inc % Ming-Yie Jan Principle Consultant RusCert Technology Co., Ltd 8F., No. 187, Lequn 2nd Rd. Zhongshan Dist. Taipei City, 10462 Taiwan
Re: K212680
Trade/Device Name: LuxCreo Clear Aligner System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: February 24, 2022 Received: March 2, 2022
Dear Ming-Yie Jan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K212680
Device Name LuxCreo Clear Aligner System
Indications for Use (Describe)
The LuxCreo Clear Aligner System is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e. all second molars). The LuxCreo Clear Aligner System repositions teeth by way of continuous gentle force.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div style="display:flex; align-items:center;"> <div style="margin-right:5px;"> <span style="font-size: 20px;">☑</span> </div> <span>Prescription Use (Part 21 CFR 201 Subpart D)</span> </div> | |
| <div style="display:flex; align-items:center;"> <div style="margin-right:5px;"> <span style="font-size: 20px;">☐</span> </div> <span>Over-The-Counter Use (21 CFR 201.66)</span> </div> | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for LuxCreo. The logo is in a sans-serif font, with the letters in black except for the 'X', which is orange. The logo is simple and modern, and the colors are eye-catching.
### SECTION 5-510(k) SUMMARY
This 510 (K) summary is being submitted in accorndance with requirements of Title 21,CFR Section 807.92.
- K212680 A. 510(k) NUMBER
| B. | DATE PREPARED May 24th, 2022 |
|----|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| C. | SUBMITTER LUXCREO INC.<br>940 Old County Road, Belmont, CA 94002<br>Registration Number: 3016237718<br>FEI Number: 3016237718<br>Tel:+1 650-3360227 |
| D. | CONTACT PERSON Primary Contact Person<br>C.O.O.<br>Mike Yang<br>+1 650-336-0888<br>Mike.yang@luxcreo.com<br><br>Second Contact<br>Project Manager<br>Jethro Wu<br>0970532485<br>Jethro.Wu@LuxCreo.com |
| E. | DEVICE Proprietary Name: LuxCreo Clear Aligner System<br>Common Name: Aligner, Sequential<br>Product Code: NXC<br>Regulation Number: 21 CFR 872.5470<br>Regulation Name: Orthodontic Plastic Bracket<br>Device Class: Class II<br>Review Panel: Dental |
| F. | INDICATION(S) FOR USE LuxCreo Clear Aligner System is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e. all second molars). The LuxCreo Clear Aligner System repositions teeth by way of continuous gentle force |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for LuxCreo. The logo is black, except for the X, which is orange. The logo is simple and modern.
| G. | PRIMARY<br>PREDICATE DEVICE | Proprietary Name:<br>Common Name:<br>Product Code:<br>Regulation Number:<br>Regulation Name:<br>510(k) Number:<br>510(k) Submitter:<br>Device Class:<br>Review Panel: | ULab Systems Dental Aligner Kit<br>Sequential Aligner<br>NXC<br>21 CFR 872.5470<br>Orthodontic Plastic Bracket<br>K192596<br>uLab Systems, Inc.<br>Class II<br>Dental |
|----|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| H. | SECONDARY<br>PREDICATE DEVICE | Proprietary Name:<br>Common Name:<br>Product Code:<br>Regulation Number:<br>Regulation Name:<br>510(k) Number:<br>510(k) Submitter:<br>Device Class:<br>Review Panel: | ClearCorrect System<br>Sequential Aligner<br>NXC<br>21 CFR 872.5470<br>Orthodontic Plastic Bracket<br>K113618<br>ClearCorrect LLC<br>Class II<br>Dental |
| I. | PRIMARY<br>REFERNCE DEVICE | Proprietary Name:<br>Common Name:<br>Regulation Number:<br>Classification<br>Product Code:<br>Subsequent Product<br>Code:<br>510(k) Number:<br>510(k) Submitter:<br>Device Class:<br>Review Panel: | KeyPrint KeySplint Soft<br>Mouthguard, Prescription/ Positioner,<br>Tooth, Preformed<br>No associated regulation/ 21 CFR<br>872.5525<br>MQC<br>KMY<br>K183598<br>Keystone Industries<br>Unclassified, Class I<br>Dental |
| J. | SECONDARY<br>REFERENCE<br>DEVICE | Proprietary Name:<br>Common Name:<br>Regulation Number: | LuxaPrint Ortho Plus<br>Prescription Mouthguard/ Resin, Denture,<br>Relining, Repairing, Rebasing<br>No associated regulation/ 21 CFR<br>872.3760 |
SECTION 5-510(k) SUMMARY Page 2 of 14 pages
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for LuxCreo. The logo is black, except for the X, which is orange. The text is sans-serif and bolded.
| Classification | MQC |
|--------------------|----------------------------|
| Product Code: | |
| Subsequent Product | EBI |
| Code: | |
| 510(k) Number: | K210940 |
| 510(k) Submitter: | DMG Digital Enterprises SE |
| Device Class: | Unclassified, Class II |
| Review Panel: | Dental |
K. DEVICE Dental health professionals achieve orthodontic tooth movement DECRIPTION through prescription of aligners which apply force to the patient's teeth so that each tooth follows a prescribed, predetermined displacement. LuxCreo Clear Aligner System is custom plastic aligner system which are a series of doctor prescribed clear removable aligners that are used as alternative treatment for the alignment of maloccluded or misaligned teeth. This series of aligners gently move the patient's teeth in small increments from their original state to a treated state
> A 3D printer, based on a 3D stereolithographic drawing, prints the predetermined shape of each aligner. Final polymerization is achieved by placing the printed aligner in a UV-light curing box. 3D printing uses specialty liquid resins, which help the aligners achieve mechanical properties similar to thermoplastics, and increased processability.
The aligners are packaged and labeled according to the sequence in which they are intended for use, determined by the prescribing dental health professional. The finished set of aligners is shipped with six-month-shelf-life to the prescribing physician, who is accountable for ensuring the patient uses the device properly and safely.
The LuxCreo Clear Aligner System mechanism of operation, and software usage, are identical to the predicate devices, and support a determination of substantial equality. Both the LuxCreo Clear Aligner System and the predicate devices are manufactured from a biocompatible, non-sterile polyurethane materials that supports a determination of substantial equality.
- L. COMPARISON OF TECHNOLOGICAL
LuxCreo Clear Aligner System and predicate devices (K192596 and K113618) are:
SECTION 5-510(k) SUMMARY Page 3 of 14 pages
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for LUXCreo. The word "Lux" is in black font, with the "X" in orange. The word "Creo" is in black font.
CHARACTERISTICS WITH THE PREDICATE DEVICES
- Same intended use/ indication(s) for use 1.
- Same intended population 2.
- Same mode of action 3.
- Same method of use 4.
- న్. Same duration of use
- Required using software for ordering workflow 6.
- 7. Removable devices
- 8. Different manufacturing process
- 9. Different materials
LuxCreo Clear Aligner System and reference devices (K183598 and K210940) are:
- Same light-cured 3 D printing 1.
- Similiar DLP/ SLA 3D printer UV cured 3D printing resin 2. for medical device manufacturing
- 3. Similar biocompatibility evaluation and testing.
- Having software involved during the manufacturing 4. ordering and manufacturing process.
The manufacturing process of LuxCreo Clear Aligner System is a light-cured, 3D printing process with light-cured polyurethane resin which is different from the thermoforming process of predicate devices. The biocompatibility risks have been considered. The biocompatibility tests indicated that the lightcured polyurethane resin is safe to use. The light-cured 3D printing process has also been used in the manufacturing of devices in dentistry. The difference in manufacturing process and base materials did not raise other safety and efficacy concerns.
LuxCreo Clear Aligner System and predicate devices (K192596 and K113618) are all the same intended use/ indication(s) for use, intended population, mode of action, method of use, removable devices which requires using a software for ordering workflow. The difference in manufacturing process and materials have been tested and validated. LuxCreo Clear Aligner System performed biocompatibility testing, physical and chemical properties testing of light-cured polyurethane resin.
Both KeyPrint KeySplint Soft, K183598, and LuxaPrint Ortho Plus, K210940, are using DLP/ SLA 3D printer with UV cured 3D printing resin for the intended use of orthodontic and dental appliances such as mouthguards, nightguards, splints and
SECTION 5-510(k) SUMMARY Page 4 of 14 pages
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for LuxCreo. The logo is in a sans-serif font, with the letters in black except for the 'X', which is orange. The logo is simple and modern.
> repositioners. The similar additive manufacturing and based materials between the LuxCreo Clear Aligner System and reference devices are substantially equivalent without extra concerns on their difference of properties.
As presented in the 510(k) and summarized herein, LUXCREO INC. concluded that LuxCreo Clear Aligner System is substantially equivalent to the predicate devices ULab Systems Dental Aligner Kit and the ClearCorrect System (K192596 and K113618) along with the reference devices KeyPrint KeySplint Soft (K183598) and LuxaPrint Ortho Plus (K210940).
#### SECTION 5-510(k) SUMMARY Page 5 of 14 pages
{8}------------------------------------------------
(650) 336-0888 940 Old County Road, Belmont, CA 94002 www.luxcreo.com info@luxcreo.com
### M. A COMPARISON TABLE WITH PREDICATE DEVICES
| Items | Proposed Device<br>LuxCreo Clear Aligner System | Primary Predicate Device<br>ULab Systems Dental Aligner Kit | Secondary Predicate Device<br>ClearCorrect System | Substantial<br>Equivalence<br>Comparison<br>Assessment |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| 510(k) Number<br>Product Code<br>Device<br>Classification | TBD<br>NXC<br>Class II | K192596<br>NXC<br>Class II | K113618<br>NXC<br>Class II | N/A<br>Same<br>Same |
| Intended Use/<br>Indication(s) for<br>Use | LuxCreo Clear Aligner System<br>is indicated for the treatment of<br>tooth malocclusion in patients<br>with permanent dentition (i.e.<br>all second molars). The<br>LuxCreo Clear Aligner System<br>repositions teeth by way of<br>continuous gentle force. | The uLab Systems Dental Aligner is<br>indicated for the alignment of<br>permanent teeth during orthodontic<br>treatment of malocclusions by way of<br>continuous gentle forces. | The ClearCorrect System is<br>indicated for the treatment of<br>tooth malocclusion in patients<br>with permanent.dentition (i.e.<br>all second molars). The<br>ClearCorrect System positions<br>teeth by way of<br>continuous gentle force. | Same |
| Mode of<br>Action | Alignment of teeth by<br>application of continuous<br>gentle force, by sequential use<br>of preformed plastic trays | Each preformed plastic tray is worn in<br>sequence by the patient as prescribed<br>by the dental practitioner. Orthodontic<br>movement occurs through<br>continuous gentle forces applied to the<br>dentition as each tooth follows the<br>programmed displacement based on a<br>doctor's prescription. | Alignment of teeth by<br>application of continuous<br>gentle<br>force, by sequential use of<br>preformed plastic trays | Same |
| Method of<br>Use | Each preformed plastic tray is<br>worn by the patient as<br>prescribed by the dental<br>practitioner, usually a few<br>weeks prior to using the next | During the orthodontic treatment, each<br>preformed plastic aligner is worn in<br>sequence by the patient as prescribed<br>by the dental practitioner,<br>moving the patient's teeth gradually to | Each preformed plastic tray is<br>worn by the patient as<br>prescribed by the dental<br>practitioner, usually a few<br>weeks prior to using the next | Same |
| Items | Proposed Device<br>LuxCreo Clear Aligner System | Primary Predicate Device<br>ULab Systems Dental Aligner Kit | Secondary Predicate Device<br>ClearCorrect System | Substantial<br>Equivalence<br>Comparison<br>Assessment |
| | sequential aligner tray | the ideal position. | sequential aligner tray | |
| OTC/Rx | Rx | Rx | Rx | Same |
| Duration of Use | 20-22 hours/day | Each set of aligners can be worn for<br>approximately 2 weeks of 20-22 hours<br>of wear per day or according to doctor's<br>prescription. | 20-22 hours/day | Same |
| Method of<br>Manufacturing | Light-cured 3 D printing | Thermoforming | Thermoforming | Different |
| Software Used<br>for Ordering<br>Workflow | Yes | Yes | Yes | Same |
| Application | Removable | Removable | Removable | Same |
| Biocompatibility | ISO 10993-5 In vitro<br>Cytotoxicity<br>ISO 10993-10 Oral Mucosa<br>Irritation<br>Pyrogen<br>ISO 10993-10 Skin Irritation in<br>Rabbits<br>ISO 10993-10 Skin<br>Sensitization in Guinea Pigs<br>(Maximization Test)<br>ISO 10993-11 Acute Systemic<br>Toxicity | Raw materials:<br>Cytotoxicity Elution -MEM<br>Intracutaneous/Intradermal Reactivity<br>Maximization for Delayed-Type<br>Hypersensitivity<br>Oral Mucosa Irritation Test<br>Final product:<br>Cytotoxicity Elution -MEM | ISO 10993-5 Cytotoxicity<br>ISO 10993-10 Intracutaneous<br>reactivity, oral mucosa<br>irritation test, maximization<br>test for delayed type<br>hypersensitivity | Similar |
| Material | Light-cured polyurethane resin | Zendura A (thin thermoformed<br>polyurethane) or Zendura FLX<br>(copolyester and polyurethane<br>composite) | Thermoplastic polyurethane<br>resin | Different |
| Items | Proposed Device<br>LuxCreo Clear Aligner System | Primary Predicate Device<br>ULab Systems Dental Aligner Kit | Secondary Predicate Device<br>ClearCorrect System | Substantial<br>Equivalence<br>Comparison<br>Assessment |
| Design | Image: LuxCreo Clear Aligner System labeled G2 | Image: ULab Systems Dental Aligner Kit | Image: ClearCorrect System | Similar<br>design |
SECTION 5-510(k) SUMMARY
Page 6 of 14 pages
{9}------------------------------------------------
### (650) 336-0888 940 Old County Road, Belmont, CA 94002 www.luxcreo.com
info@luxcreo.com
### SECTION 5-510(k) SUMMARY
Page 7 of 14 pages
{10}------------------------------------------------
### LUX Creo
### (650) 336-0888 940 Old County Road, Belmont, CA 94002 www.luxcreo.com info@luxcreo.com
#### Substantial Equivalence Comparison Table-Predicate Devices-- Mechanical Properties N.
| Items | Proposed Device<br>LuxCreo Clear Aligner System | Primary Predicate Device<br>ULab Systems Dental Aligner Kit<br>K192596 | Secondary Predicate Device<br>ClearCorrect System<br>K113618 | Substantial<br>Equivalence<br>Comparison<br>Assessment |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Ultimate<br>Flexural<br>Strength | 23.6 ± 1.9 MPa | N/A | 36.8±1.1 MPa (ISO 20795-2:2013) | Different |
| Flexural<br>Modulus | 1106.47±13.23 MPa (ASTM<br>D790)<br>804 ± 64 MPa (ISO 20795-<br>2:2013) | N/A | 1122±24 MPa (ISO 20795-2:2013) | Different |
| Shore D<br>Hardness | 21.63±0.38 HD | N/A | 21.33±0.06 HD | Similar |
| Stress<br>Relaxation | 37.3 ± 0.3%<br>No cracking was found. | N/A | 25.5 ± 0.4%<br>No cracking was found. | Different |
| Items | Proposed Device<br>LuxCreo Clear Aligner System | Primary Reference Device<br>KeyPrint KeySplint Soft | Secondary Reference Device<br>LuxaPrint Ortho Plus | Substantial<br>Equivalence<br>Comparison<br>Assessment |
| 510(k) Number | TBD | K183598 | K210940 | N/A |
| Product Code | NXC | MQC/ KMY | MQC/ EBI | Different |
| Device<br>Classification | Class II | Unclassified/ Class II | Unclassified/ Class II | Different |
| Intended Use/<br>Indication(s) for<br>Use | LuxCreo Clear Aligner System<br>is indicated for the treatment of<br>tooth malocclusion in patients<br>with permanent dentition (i.e.<br>all second molars). The<br>LuxCreo Clear Aligner System<br>repositions teeth by way of<br>continuous gentle force. | The KeyPrint® KeySplint Soft™ device is<br>indicated for the fabrication of<br>orthodontic and dental appliances such<br>as mouthguards, nightguards, splints<br>and repositioners. | For the fabrication of<br>orthodontic and dental<br>appliances such as<br>mouthguards, nightguards,<br>splints and positioners. The<br>product is for use with<br>DLP/SLA printers that work at<br>wavelengths of 385 nm or 405<br>nm. | Different |
| Mode of<br>Action | Alignment of teeth by<br>application of continuous<br>gentle force, by sequential use<br>of preformed plastic trays | It depends on the indications of final<br>device. | The device consists of raw<br>material that is used to<br>fabricate removable custom<br>dental appliances, such as<br>orthodontic splints and/or<br>mouthguards. The finished<br>devices can be used to<br>support tooth stabilization<br>following active<br>orthodontic treatment,<br>and/or for the relief of<br>bruxism or snoring. The<br>device is used in<br>conjunction with a | Different |
| Items | Proposed Device<br>LuxCreo Clear Aligner System | Primary Reference Device<br>KeyPrint KeySplint Soft | Secondary Reference Device<br>LuxaPrint Ortho Plus | Substantial<br>Equivalence<br>Comparison<br>Assessment |
| | | | compatible scanner, 3D<br>printer, and curing unit. | |
| Method of<br>Use | Each preformed plastic tray is<br>worn by the patient as<br>prescribed by the dental<br>practitioner, usually a few<br>weeks prior to using the next<br>sequential aligner tray | Dental office/clinic, dental laboratory<br>for appliance design and fabrication;<br>home use for the resulting dental<br>appliance. | Dental office/clinic, dental<br>laboratory for appliance<br>design and fabrication; home<br>use for the resulting dental<br>appliance. | Different |
| OTC/Rx | Rx | Rx | Rx | Same |
| Software Used<br>for Ordering<br>Workflow | Yes, a software is involved<br>during the manufacturing<br>ordering and manufacturing<br>process.<br>This device does not contain<br>Software. | Yes, a software is involved during the<br>manufacturing ordering and<br>manufacturing process.<br>This device does not contain<br>Software. | Yes, a software is involved<br>during the manufacturing<br>ordering and manufacturing<br>process.<br>This device does not contain<br>Software. | Same |
| Biocompatibility | ISO 10993-3:2014<br>Genotoxicity test<br>ISO 10993-5 In vitro<br>Cytotoxicity<br>ISO 10993-6 Subchronic<br>systemic toxicity<br>ISO 10993-10 Oral Mucosa<br>Irritation<br>Pyrogen<br>ISO 10993-10 Skin Irritation in<br>Rabbits<br>ISO 10993-10 Skin<br>Sensitization in Guinea Pigs | ISO 10993-5 Cytotoxicity<br>ISO 10993-10 Sensitization<br>ISO 10993-10 Irritation | ISO 10993-1<br>Biocompatibility<br>Assessment<br>ISO 10993-5 Cytotoxicity<br>ISO 10993-10 Irritation<br>ISO 10993-10<br>Sensitization<br>ISO 10993-3 Genotoxicity | Equivalent |
| Items | Proposed Device<br>LuxCreo Clear Aligner System | Primary Reference Device<br>KeyPrint KeySplint Soft | Secondary Reference Device<br>LuxaPrint Ortho Plus | Substantial<br>Equivalence<br>Comparison<br>Assessment |
| | (Maximization Test)<br>ISO 10993-11 Acute Systemic<br>Toxicity<br>ISO 10993-11 Subchronic<br>systemic toxicity | | | |
| Material | Light-cured polyurethane resin | Light-cured polyurethane resin | Photocurable (meth)-<br>acrylate-based polymer<br>resin | Equivalent |
| Design | G2<br>Image: Clear aligner | Image: Clear aligner | Image: Clear aligner | Similar |
#### Substantial Equivalence Comparison Table-Reference Devices-Basic Information 0.
SECTION 5-510(k) SUMMARY Page 8 of 14 pages
{11}------------------------------------------------
### (650) 336-0888 940 Old County Road, Belmont, CA 94002 www.luxcreo.com
info@luxcreo.com
### SECTION 5-510(k) SUMMARY
Page 9 of 14 pages
{12}------------------------------------------------
### (650) 336-0888 940 Old County Road, Belmont, CA 94002 www.luxcreo.com
info@luxcreo.com
### SECTION 5-510(k) SUMMARY
Page 10 of 14 pages
{13}------------------------------------------------
### (650) 336-0888 940 Old County Road, Belmont, CA 94002 www.luxcreo.com
info@luxcreo.com
#### Substantial Equivalence Comparison Table- Reference Devices-Manufacturing Information: P.
| Items | Proposed Device<br>LuxCreo Clear Aligner<br>System | Primary Reference Device<br>KeyPrint KeySplint Soft<br>K183598 | Secondary Reference Device<br>LuxaPrint Ortho Plus<br>K210940 | Substantial<br>Equivalence<br>Comparison<br>Assessment |
|-----------------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|--------------------------------------------------------|
| Method of<br>Manufacturing | Light-cured 3 D printing | Light-cured 3 D printing | Light-cured 3 D printing | Same |
| Technical<br>Specifications of<br>3D printing | DLP/ SLA 3D printer UV<br>cured 3D printing resin for<br>medical device manufacturing. | DLP/ SLA 3D printer UV cured 3D<br>printing resin for medical device<br>manufacturing. | DLP/ SLA 3D printer UV<br>cured 3D printing resin for<br>medical device manufacturing. | Same |
### SECTION 5-510(k) SUMMARY
Page 11 of 14 pages
{14}------------------------------------------------
(650) 336-0888 940 Old County Road, Belmont, CA 94002 www.luxcreo.com info@luxcreo.com
| Items | Proposed Device<br>LuxCreo Clear Aligner<br>System | Primary Reference Device<br>KeyPrint KeySplint Soft<br>K183598 | Secondary Reference Device<br>LuxaPrint Ortho Plus<br>K210940 | Substantial<br>Equivalence<br>Comparison<br>Assessment |
|--------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------|---------------------------------------------------------------|--------------------------------------------------------|
| Manufacturing<br>process<br>workflow | The raw 3D printing material in<br>combination with LuxCreo's<br>manufacturing system. | Raw 3D printing material | Raw 3D printing material | Different |
#### Q. Substantial Equivalence Comparison Table-Physical and Mechanical Properties
| Items | Proposed Device<br>LuxCreo Clear Aligner System | Primary Reference Device<br>KeyPrint KeySplint Soft<br>K183598 | Secondary Reference Device<br>LuxaPrint Ortho Plus<br>K210940 | Substantial<br>Equivalence<br>Comparison<br>Assessment |
|-------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------|------------------------------------------------------------------------------------|--------------------------------------------------------|
| Flexural Strength | 40.02±0.08 MPa (ASTM D790)<br><br>23.6 ± 1.9 MPa (ISO 20795-2:2013) | 44-47 MPa (ASTM D790) | 80 – 93 MPa (385 nm)<br>77 - 90 MPa (405 nm)<br>(ISO 4049 & ISO 20795-2) | Different |
| Flexural<br>Modulus | 1106.47±13.23 MPa (ASTM D790)<br><br>804 ± 64 MPa (ISO 20795-2:2013) | 1100-1400 MPa (ASTM D790)<br>135-200 MPa (ISO 20795-2) | 1.9 - 2.4 GPa (385 nm)<br>2.1 - 2.5 GPa (405 nm)<br>(ISO 4049 & ISO 20795-2) | Different |
| Shore D<br>Hardness | 21.63±0.38 HD | 80-85 MPa (ASTM D2240) | ≥ 60 (Cured material) | Different |
| Ultimate Flexural<br>Strength | Ultimate flexural strength is<br>23.6 ± 1.9 MPa. | 2.6—4.4 MPa (ISO 20795-2) | N/A | Different |
| Water Solubility | 3.668±1.0748 µg/mm³ | 4.8 µg/mm³ | 1.5 - 3.6 µg/mm³ (385 nm)<br>0.7 - 2.1 µg/mm³ (405 nm)<br>(ISO 4049 & ISO 20795-2) | Different |
| Water Sorption | 19.952±6.6719 µg/mm³ | 18 µg/mm³ | 18.6 – 20.3 (385 nm)<br>17.1 - 17.7 µg/mm³ (405nm)<br>(ISO 4049 & ISO 20795-2) | Similar |
SECTION 5-510(k) SUMMARY
Page 12 of 14 pages
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for LuxCreo. The logo is in a sans-serif font, with the "X" in "Lux" being orange and the rest of the letters being black. The logo is simple and modern.
- R. SUMMARY OF PERFORMANCE DATA
The following performance data were provided to demonstrate the safetv and efficacy:
- A. ISO 10993-3:2014 Genotoxicity test
- ISO 10993-5 In vitro C vtotoxicitv B.
- ISO 10993-6:2016 Subcutaneous Implantation Test C.
- D. ISO 10993-10 Oral Mucosa Irritation
- USP Pyrogen Study E.
- ISO 10993-10 Skin Irritation F.
- G. ISO 10993-10 Skin Sensitization (Maximization Test)
- ISO 10993-11 Acute Systemic Toxicity H.
- I. ISO 10993-11:2017 Sub-chronic systemic toxicity
- J. Transportation and accelerating ageing tests were validated and completed.
- Physical, chemical and mechanical properties were tested. K.
- Design verification, validation and manufacturing validation L. were completed. All the results meet the product specification requirements.
- છ SUBSTANTIAL EQUIVALENCE CONCLUSION
LuxCreo Clear Aligner System, ULab Systems Dental Aligner Kit, and the ClearCorrect System are intended to provide force to the user's teeth gently, and in small increments, from their original misalignment to their final treated position, for improved dental alignment. The difference in manufacturing process and base materials have not raised extra safety and performance concerns, based on the relevant tests and evaluations provided in this submission.
Both KeyPrint KeySplint Soft, K183598, and LuxaPrint Ortho Plus, K210940, are using DLP/ SLA 3D printer with UV cured 3D printing resin for the intended use of orthodontic and dental appliances such as mouthguards, nightguards, splints and repositioners. The additive manufacturing and based materials used for the LuxCreo Clear Aligner System and reference devices are substantially equivalent without extra concerns on their difference of properties.
Based upon the information presented in this section, LUXCREO INC. concludes that the LuxCreo Clear Aligner System is
SECTION 5-510(k) SUMMARY Page 13 of 14 pages
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for LuxCreo. The logo is in a sans-serif font, with the letters in black except for the 'X', which is orange. The logo is simple and modern.
substantially equivalent to predicate devices in regard to indications for use, design, and technology.
#### SECTION 5-510(k) SUMMARY Page 14 of 14 pages
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.