The PediFlex™ Flexible Nail System is intended for fixation of diaphyseal fractures of long bones where the medullary canal is narrow, or flexibility of the implant is required. This includes upper extremity fractures in all patients and lower extremity fractures in pediatric or small stature patients. In pediatric patients, the flexibility of the nail allows it to be inserted at a point that does not disturb or disrupt the growth plate.
Device Story
PediFlex™ Flexible Nail System consists of stainless steel (316L) and titanium alloy (Ti-6Al-4V) flexible nails, interlocking clamps, clamp screws, and end caps; used for diaphyseal long bone fracture fixation. Implants are MR conditional. System provides mechanical stabilization of bone fractures; clamps and end caps are optional accessories. Used in clinical/surgical settings by orthopedic surgeons. Nails inserted into medullary canal; flexibility allows insertion points avoiding pediatric growth plates. Output is physical stabilization of bone fragments to facilitate healing. Benefits include reduced soft tissue irritation via end caps and growth plate preservation in pediatric patients.
Clinical Evidence
Bench testing only. Performance supported by sterilization/cleaning validation (AAMI ST72/ST98, ISO 11137, ISO 11737), package design verification (ASTM D4332/D4169/F2096), shelf life validation (ASTM F1980), usability validation (ISO 11607, IEC 62366-1), and MR conditional testing (ASTM F2052/F2213/F2182/F2119). Biocompatibility was assessed via existing data from predicate devices.
Technological Characteristics
Materials: 316L stainless steel and Ti-6Al-4V titanium alloy. Components: flexible nails (1.5-4.5mm diameter), interlocking clamps, clamp screws, end caps. MR conditional. Sterilization: validated for sterile and non-sterile options per ISO 11137. Mechanical fixation principle.
Indications for Use
Indicated for fixation of diaphyseal long bone fractures in patients requiring flexible implants or with narrow medullary canals; includes upper extremity fractures in all patients and lower extremity fractures in pediatric or small stature patients.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 24, 2025
OrthoPediatrics Corp.
Jessica Powers
Regulatory Affairs Specialist
2850 Frontier Dr.
Warsaw, Indiana 46582
Re: K251362
Trade/Device Name: PediFlex™ Flexible Nail System
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener
Regulatory Class: Class II
Product Code: HTY
Dated: April 30, 2025
Received: May 1, 2025
Dear Jessica Powers:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251362 - Jessica Powers
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251362 - Jessica Powers
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CHRISTOPHER FERREIRA -S
Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251362 | |
| Device Name PediFlex™ Flexible Nail System | |
| Indications for Use (Describe) The PediFlex™ Flexible Nail System is intended for fixation of diaphyseal fractures of long bones where the medullary canal is narrow, or flexibility of the implant is required. This includes upper extremity fractures in all patients and lower extremity fractures in pediatric or small stature patients. In pediatric patients, the flexibility of the nail allows it to be inserted at a point that does not disturb or disrupt the growth plate. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
OrthoPediatrics, Corp.
PediFlex™ Flexible Nail System
Traditional 510(k)
K251362
OrthoPediatrics
# 510(k) Summary
I. Submitter
| Submission: | Traditional 510(k) Premarket Notification |
| --- | --- |
| Applicant: | OrthoPediatrics Corp. |
| Applicant Address: | 2850 Frontier Drive, Warsaw, IN 46582 |
| Establishment Registration Number: | 3006460162 |
| Contact: | Jessica Powers |
| Contact Phone: | (574) 268-6379 |
| Date Prepared: | June 23, 2025 |
II. Device
| Device Trade Name: | PediFlex™ Flexible Nail System |
| --- | --- |
| Common Name: | Intramedullary Nails and Accessories |
| Regulation Number: | 21 CFR 888.3040 |
| Device Classification Name: | Smooth or threaded metallic bone fixation fastener |
| Product Code: | HTY: pin, fixation, smooth |
| Device Classification: | II |
| Classification Panel: | Orthopedic |
III. Predicate Device and Reference Device
Substantial equivalence is claimed to the following predicate devices:
Primary Predicate Device:
- PediFlex™ Flexible Nail System (K201838, OrthoPediatrics Corp.)
IV. Device Description
The PediFlex™ Flexible Nail System includes stainless steel and titanium flexible nails, interlocking clamps, clamp screws and end caps for the application of aiding bone fracture repair and healing. All the implants are MR conditional. The system is implanted using Class II and Class I exempt surgical instruments.
The PediFlex™ flexible nails are provided in 316L stainless steel and titanium alloy (Ti-6Al-4V). The 316L stainless steel flexible nails are available in diameters ranging from 1.5 mm to 4.0 mm in increments of 0.5 mm. The titanium alloy (Ti-6Al-4V) flexible nails are available in diameters ranging from 1.5 mm to 4.5 mm in increments of 0.5 mm.
The PediFlex™ interlocking clamps are designed to hold flexible nails in place after insertion by clamping the nail and being fixed to the bone. The use of the interlocking clamps is optional. The interlocking
Page 1 of 3
{5}
OrthoPediatrics, Corp.
PediFlex™ Flexible Nail System
Traditional 510(k)
K251362
OrthoPediatric
clamps are manufactured from both 316L stainless steel and titanium alloy (Ti-6Al-4V) materials. The 316L stainless steel interlocking clamps are available in right and left configurations, in diameters of 3.0 mm, 3.5 mm and 4.0 mm which are to be used with 316L stainless steel flexible nails in diameters of 3.0 mm, 3.5 mm and 4.0 mm. The 316L stainless steel clamp screws are available in 15 mm, 25 mm, and 35 mm in length. The titanium alloy (Ti-6Al-4V) interlocking clamps are available in right and left configurations, in diameters of 3.0 mm, 3.5 mm, 4.0 mm and 4.5 mm which are to be used with titanium alloy (Ti-6Al-4V) flexible nails in diameters of 3.0 mm, 3.5 mm, 4.0 mm and 4.5 mm. The titanium alloy (Ti-6Al-4V) clamp screws are available in 15 mm, 25 mm, and 35 mm in length. The material of the clamp must match the material of the nail and screw in the surgery.
The PediFlex™ end caps are designed to affix to the exposed nail tip to help prevent soft tissue irritation. The use of the end cap is optional. They are not a structural element and impart no additional strength to the construct. The subject end caps are manufactured in titanium alloy (Ti-6Al-4V) and are available in diameters ranging from 1.5mm to 4.5mm with 0.5mm increments which are to be used with titanium alloy (Ti-6Al-4V) flexible nails in diameters ranging from 1.5mm to 4.5mm with 0.5mm increments. The material of the end cap must match the material of the nail used in the surgery.
The flexible nails, interlocking clamps, clamp screws and end caps in PediFlex™ Flexible Nail System are offered both sterile and non-sterile. All other Class II implants and instruments of the PediFlex™ Flexible Nail System are offered non-sterile to be sterilized by the end-user.
# V. Indications for Use
The PediFlex™ Flexible Nail System is intended for fixation of diaphyseal fractures of long bones where the medullary canal is narrow, or flexibility of the implant is required. This includes upper extremity fractures in all patients and lower extremity fractures in pediatric or small stature patients. In pediatric patients, the flexibility of the nail allows it to be inserted at a point that does not disturb or disrupt the growth plate.
# VI. Comparison of Technological Characteristics
The subject PediFlex™ Flexible Nail System and predicate device share identical intended use, patient population, principles of operation, and all fundamental technological characteristics. There are some differences between the predicate and subject devices in terms of sterilization, packaging, shelf life, and MR conditional labeling. However, those differences are supported by successful testing provided in this submission. Therefore, such differences do not raise new questions of safety and effectiveness.
# VII. Performance Data
The sterilization, packaging and shelf life of the subject devices were supported by the following verification and validation activities:
- Sterilization and Cleaning Validation following AAMI ST72, AAMI ST98, ASTM F3127, ISO 11137-1, ISO 11137-2, ISO 11737-1, ISO 11737-2, ISO 11737-3 and ISO 19227.
- Package Design Verification following ASTM D4332, ASTM D4169, ASTM F2096, ASTM F2203, ASTM F88, ASTM F1886, ISO 11607-1, ISO 11607-2 and ISO 15415/15416.
- Shelf Life Validation following ASTM F1980, ASTM F1886, ASTM F2096, ASTM F2203, ASTM F88, ISO 11607-1, ISO 11607-2 and ISO 15415/15416.
Page 2 of 3
{6}
OrthoPediatrics, Corp.
PediFlex™ Flexible Nail System
Traditional 510(k)
K251362
OrthoPediatrics
- Usability Validation following ISO 11607-1 and IEC-62366-1
An engineering analysis has been conducted that supports that there is no impact of sterilization or aging on the products' performance and functionality.
The biocompatibility of the non-sterile predicate devices was assessed and cleared under previous 510(k)s. A biocompatibility assessment has been conducted and upon review of all available information regarding the devices and proposed new packaging materials, including how the device and packaging materials may be impacted by the updated cleaning, packaging, and sterilization flow, no further biocompatibility testing is deemed necessary.
The implants of the PediFlex™ Flexible Nail System have been evaluated for use in an MR environment in following ASTM F2052, ASTM F2213, ASTM F2182, ASTM F2119, and FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment - Guidance for Industry and Food and Drug Administration Staff" issued on October 10, 2023. The implants were determined to be MR Conditional and will be labeled as such.
Results of the performance testing demonstrate substantially equivalent performance of the subject device as compared to the predicate.
## VIII. Conclusion
The information and data provided within the submission support that the PediFlex™ Flexible Nail System is substantially equivalent to the predicate device.
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.