The PediFlex™ Flexible Nail System is intended for fixation of diaphyseal fractures of long bones where the medullary canal is narrow or flexibility of the implant is required. This includes upper extremity fractures and lower extremity fractures in pediatric or small stature patients, the flexibility of the nail allows it to be inserted at a point that does not disturb or disrupt the growth plate.
Device Story
PediFlex Flexible Nail System comprises stainless steel (316L) and titanium alloy (Ti-6Al-4V) flexible nails, optional interlocking clamps, and clamp screws; used for internal fixation of diaphyseal long bone fractures. Nails are inserted into the medullary canal; interlocking clamps secure nails to bone. System is used in clinical settings by surgeons; implants are MR Conditional. Benefits include fracture stabilization while allowing insertion points that avoid growth plate disruption in pediatric patients. System includes reusable and single-use instruments; all components provided non-sterile and require user sterilization.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation testing included mechanical construct testing (ASTM F543), usability validation, biocompatibility (ISO 10993-1, -5, -10, -17), and MR compatibility (ASTM F2052, F2119, F2182, F2213).
Indicated for fixation of diaphyseal long bone fractures in patients requiring flexible implants or having narrow medullary canals; includes upper extremity fractures in all patients and lower extremity fractures in pediatric or small stature patients.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Orthopediatrics PediFlex™ Flexible Nail System (K082375)
Orthopediatrics PediFlex™ Flexible Nail System (K081097)
Submission Summary (Full Text)
{0}------------------------------------------------
29 January 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
OrthoPediatrics Corp. Yan Li Regulatory Affairs Associate Manager 2850 Frontier Drive Warsaw, Indiana 46582
Re: K201838
Trade/Device Name: PediFlex Flexible Nail System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HTY Dated: December 21, 2020 Received: December 23, 2020
Dear Yan Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name PediFlex™ Flexible Nail System
### Indications for Use (Describe)
The PediFlex™ Flexible Nail System is intended for fixation of diaphyseal fractures of long bones where the medullary canal is narrow or flexibility of the implant is required. This includes upper extremity fractures and lower extremity fractures in pediatric or small stature patients, the flexibility of the nail allows it to be inserted at a point that does not disturb or disrupt the growth plate.
Type of Use (Select one or both, as applicable)
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------|
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for OrthoPediatrics. The logo features a cartoon drawing of a child with a big smile. The child is holding a banana in their left hand. The text "OrthoPediatrics" is written in a playful font to the right of the child.
# 510(k) Summary
### Submitter I.
| Submission: | Traditional 510(k) Premarket Notification |
|------------------------------------|-------------------------------------------|
| Applicant: | OrthoPediatrics Corp. |
| Applicant Address: | 2850 Frontier Drive, Warsaw, IN 46582 |
| Establishment Registration Number: | 3006460162 |
| Contact: | Yan Li |
| Contact Phone: | (574) 267-0864 |
| Date Prepared: | January 28, 2021 |
#### II. Device
| Device Trade Name: | PediFlex™ Flexible Nail System |
|------------------------|--------------------------------|
| Regulation Number: | 21 CFR 888.3040 |
| Product Code: | HTY |
| Device Classification: | II |
| Common Name: | Pin, Fixation, Smooth |
| Classification Panel: | Orthopedic |
#### III. Predicate Device
Substantial equivalence is claimed to the following predicate devices:
## Primary predicate:
- Orthopediatrics PediFlex™ Flexible Nail System, K082375, OrthoPediatrics Corp. ●
# Secondary predicate:
- Orthopediatrics PediFlex™ Flexible Nail System, K081097, OrthoPediatrics Corp. ●
#### IV. Device Description
The PediFlex™Flexible Nail System includes stainless steel and titanium flexible nails, interlocking clamps, and clamp screws for the application of aiding bone fracture repair and healing. The PediFlex™Flex™Flexible Nail System was evaluated for use in an MR Environment and were determined to be MR Conditional. The system is implanted using Class I exempt instruments.
The PediFlex™ flexible nails are provided in 316L stainless steel and titanium alloy (Ti-6A1-4V). The 316L stainless steel flexible nails are available in diameters ranging from 1.5 mm to 4.0 mm
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for OrthoPediatrics. The logo features a cartoon drawing of a smiling face with a banana-shaped body and a small hand. The text "OrthoPediatrics" is written in a playful, handwritten font, with the "Ortho" in blue and the "Pediatrics" in red.
in increments of 0.5 mm. The titanium allov (Ti-6Al-4V) flexible nails are available in diameters ranging from 1.5 mm to 4.5 mm in increments of 0.5 mm. The tip of the flexible nails is available in two designs, the round tip and the advanced round tip. Both tip designs are provided in 316L stainless steel and titanium alloy (Ti-6Al-4V) in various diameters and lengths. The PediFlex™ interlocking clamps are designed to hold flexible nails in place after insertion by clamping the nail and being fixed to the bone. Using the interlocking Clamps is optional in the surgery. The interlocking clamps are manufactured from both 316L stainless steel and titanium allov (Ti-6Al-4V) materials. The 316L stainless steel interlocking Clamps are available in right and left configurations, in diameters of 3.0 mm, 3.5 mm and 4.0 mm which are to be used with 316L stainless steel flexible nails in diameters of 3.0 mm. 3.5 mm and 4.0 mm. The 316L stainless steel clamp screws are available in 15 mm, and 35 mm in length. The titanium alloy (Ti-6A1-4V) interlocking Clamps are available in right and left configurations, in diameters of 3.0 mm, 3.5 mm, 4.0 mm and 4.5 mm which are to be used with titanium alloy (Ti-6AI-4V) flexible nails in diameters of 3.0 mm, 3.5 mm, 4.0 mm and 4.5 mm. The titanium alloy (Ti-6Al-4V) clamp screws are available in 15 mm, 25 mm, and 35 mm in length. The material of the clamp must match the material of the nail and screw in the surgery.
The system is implanted using Class I exempt instruments. The Class II instruments in the system include hex drivers which are for installing the clamp screws and the hex easy out which is to remove the clamp screws if the hex strips. The instruments in the system are Class I exempt.
The implants in the PediFlex™ Flexible Nail System are for single use only and will be provided non-sterile. The instruments in the PediFlex™ Flexible Nail System are reusable, except for the Hex Easy Out instrument which is single use. Instruments are also provided non-sterile. The devices must be sterilized by the end user before use.
### V. Indications for Use
The PediFlex™ Flexible Nail System is intended for fixation of diaphyseal fractures of long bones where the medullary canal is narrow or flexibility of the implant is required. This includes upper extremity fractures in all patients and lower extremity fractures in pediatric or small stature patients. In pediatric patients, the flexibility of the nail allows it to be inserted at a point that does not disturb or disrupt the growth plate.
### VI. Comparison of Technological Characteristics
The PediFlex™ Flexible Nail System and predicate devices share identical indications for use, which is intended for fixation of diaphyseal fractures of long bones where the medullary canal is narrow or flexibility of the implant is required. This includes upper extremity fractures in all patients and lower extremity fractures in pediatric or small stature patients. In pediatric patients, the flexibility of the nail allows it to be inserted at a point that does not disturb or disrupt the growth plate.
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for OrthoPediatrics. The logo features a smiling face with a blue outline and a green body. The text "OrthoPediatrics" is written in a playful, handwritten font, with "Ortho" in blue and "Pediatrics" in red.
The PediFlex™ Flexible Nail System and the predicate devices also share the same principles of operation in the same anatomical sites, materials of construction, body contact and duration of contact, sterilization, packaging, and fundamental technological characteristics; and they have similar dimensions and design features for implants and components types.
The modifications to the nail tip design for the stainless steel and titanium flexible nails in the PediFlex™ Flexible Nail System are line extensions of the flexible nails in the PediFlex™ Flexible Nail System cleared in K081097 and K082375. They have same materials of construct, dimensions, principal of operation and the only difference is the tip design. which does not create a new worst case or affect the strength of the implant. Therefore, these minor differences in tip design do not raise new questions of safety or effectiveness.
Components introduced in this submission include the interlocking clamps and clamp screws. The clamp and clamp screws are designed to hold flexible nails in place after insertion by clamping the nail and being fixed to the bone. Using the interlocking clamps is optional in the surgery. Successful performance testing has been conducted on the clamps and clamp screws to demonstrate their substantial equivalence.
The predicate devices have not been evaluated in an MR environment. The subject device PediFlex™ Flexible Nail System was evaluated for use in an MR Environment and was determined to be MR Conditional.
### VII. Performance Data
Performance testing was conducted to demonstrate that the PediFlex™ Flexible Nail System meets the design input requirements identified based on the intended use of the device, including the needs of the user and patient, and where appropriate, applicable standards. The following performance data is provided in support of the substantial equivalence determination:
Verification Testing:
- VerRep-1001-0003 PediFlex ILC Construct Test ●
- VerRep-1001-0009 ILC ASTM F543
Validation Testing:
- ValRep-1001-0001 PediFlex Advanced Validation ●
- ValRep-1001-0004 PediFlex ILC Validation ●
- ValRep-1001-0008 PediFlex ILC Usability ●
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for OrthoPediatrics. The logo features a cartoon drawing of a smiling face with a banana body and a small hand. The text "OrthoPediatrics" is written in a playful, handwritten font, with "Ortho" in blue and "Pediatrics" in red.
The PediFlex™ Flexible Nail System implants were evaluated for use in an MR Environment and were determined to be MR Conditional. The following testing data were provided in support of MR compatibility.
- . MRER-1001-001 PediFlex Advanced MRER
- VerRep-9999-0007 MR Compatibility Force and Torque ●
- VerRep-9999-0008 MR Compatibility Artifact Testing ●
- VerRep-9999-0014 MR Heating PediFlex .
Biocompatibility testing for the PediFlex™ Flexible Nail System was performed in conformance with ISO 10993-1. The following test reports are provided to support the biocompatibility of the subject devices.
- VerRep-0999-0006 Solid IM Nails Biocompatibility .
- VerRep-0999-0009 Instrument Biocompatibility ●
| A list of FDA recognized standards used for the evaluation of the performance of the devices | |
|----------------------------------------------------------------------------------------------|--|
| have been listed in the table. | |
| FDA<br>Recognition<br>Number | Standard Reference | Standard Title |
|------------------------------|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| 11-327 | ASTM F543-17 | Standard Specification and Test Methods for Metallic<br>Medical Bone Screws |
| 8-422 | ASTM F2052-15 | Standard Test Method for Measurement of Magnetically<br>Induced Displacement Force on Medical Devices in the<br>Magnetic Resonance Environment |
| 8-153 | ASTM F2119-07/(R)2013 | Standard Test Method for Evaluation of MR Image Artifacts<br>from Passive Implants |
| 8-227 | ASTM F2182-11a | Standard Test Method for Measurement of Radio Frequency<br>Induced Heating on or Near Passive Implants During<br>Magnetic Resonance Imaging |
| 8-466 | ASTM F2213-17 | Standard Test Method for Measurement of Magnetically<br>Induced Torque on Medical Devices in the Magnetic<br>Resonance Environment |
| 5-114 | ANSI AAMI IEC<br>62366-1:2015 | Medical Devices - Application of usability engineering to<br>medical devices |
| 2-220 | ANSI AAMI ISO<br>10993-1:2018 | Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management process<br>(ISO 10993-1:2009) |
| 2-245 | ANSI AAMI ISO<br>10993-5:2009/(R)2014 | Biological evaluation of medical devices - Part 5: Tests for<br>in vitro cytotoxicity |
{7}------------------------------------------------
Image /page/7/Picture/2 description: The image shows the logo for OrthoPediatrics. The logo features a cartoon drawing of a child with a smiling face and a hand holding a banana. The text "OrthoPediatrics" is written in a playful, handwritten font, with the "Ortho" in blue and the "Pediatrics" in red.
| FDA<br>Recognition<br>Number | Standard Reference | Standard Title |
|------------------------------|-----------------------------------------|---------------------------------------------------------------------------------------------------------|
| 2-174 | ANSI AAMI ISO<br>10993-10: 2010/(R)2014 | Biological evaluation of medical devices – Part 10: Tests for<br>irritation and skin sensitization |
| 2-237 | ANSI AAMI ISO<br>10993-17: 2002/(R)2012 | Biological evaluation of medical devices. Establishment of<br>allowable limits for leachable substances |
# VIII. Conclusion
The information provided above supports that the PediFlex™Flexible Nail System is substantially equivalent to the predicate devices. Information and data provided within the submission support the differences between the subject and predicate devices. Therefore, it is concluded that the PediFlex™ Flexible Nail System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.