K251354 · UV Smart Technologies B.V. · SCS · Jan 29, 2026 · General Hospital
Device Facts
Record ID
K251354
Device Name
UV Smart D60
Applicant
UV Smart Technologies B.V.
Product Code
SCS · General Hospital
Decision Date
Jan 29, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6511
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K251354 · Jan 29, 2026
UV Smart D60
UV Smart Technologies B.V.
Clinical-use TEE probes from hospital settings
In-use testing was performed on TEE probes previously used in a clinical setting to demonstrate the achievement of high-level disinfection (HLD) after processing in the UV Smart D60.
In-Use Testing; Clinical setting; TEE probes; High-level disinfection
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
In-Use Testing; Retrospective/Observational assessment of clinically used instruments
TEE probes used in a clinical setting
Not applicable for this study
Achievement of HLD on clinically used TEE probes
Indications for Use
The UV Smart D60 is indicated for use in a healthcare environment to achieve a high level disinfection (HLD) of external surfaces of Transesophageal Echocardiogram (TEE) probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths.
Device Story
UV Smart D60 is a chemical-free, UV-C disinfection chamber for TEE probes. System uses eight UV-C lamps and reflective walls to deliver a fixed 120-second disinfection cycle. Input: pre-cleaned TEE probes. Operation: probe suspended in chamber; system recognizes registered instruments via traceability system; cycle initiates only if instrument is approved. Monitoring: dual-sensor system (photodiode for UV-C dose, current sensor for power) verifies efficacy per cycle. Output: high-level disinfected probe ready for patient use. Used in healthcare facilities (hospitals, clinics) by trained personnel. Benefits: provides validated HLD without chemicals, ensuring patient safety through consistent, monitored disinfection.
Clinical Evidence
Bench testing only. No clinical data. Evidence includes: biocompatibility (ISO 10993-1), electrical safety (IEC 61010-1, IEC 61010-2-040), EMC (IEC 61326-1, IEC 60601-1-2), UV-C hierarchy testing (most resistant: Aspergillus brasiliensis), potency testing (sporicidal, bactericidal, fungicidal, mycobactericidal, virucidal efficacy against 10 pathogens), and simulated use testing demonstrating >6 log reduction of Bacillus subtilis on worst-case TEE probe geometries. Optical verification confirmed MED delivery at worst-case locations.
Technological Characteristics
UV-C disinfection chamber; 8 UV-C lamps; reflective chamber walls; UV-C-transmitting instrument holders. Fixed 120s cycle. Dual-sensor monitoring (photodiode for UV-C dose, current sensor for power). No active ventilation. Electrical safety per IEC 61010-1/2-040. EMC per IEC 61326-1/60601-1-2. Software-controlled cycle with instrument traceability and interlocks.
Indications for Use
Indicated for high level disinfection of external surfaces of TEE probes without lumens or deep indentations/channels in healthcare environments. Intended for use by trained personnel.
Regulatory Classification
Identification
An ultraviolet radiation disinfection chamber device is intended to disinfect patient contacting medical devices using UV radiation after the device has been cleaned. Disinfection of the medical device is achieved within an enclosed chamber through the exposure to UV radiation. The Germitec Chronos® is an automated medical device consisting of a sealed disinfection chamber, incorporating multiple UV-C lamps and reflective walls, indicated for use in a healthcare environment to achieve a high-level disinfection of surfaces of external, transvaginal, and transrectal ultrasound probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths.
Special Controls
In combination with the general controls of the FD&C Act, the ultraviolet radiation disinfection chamber device is subject to the following special controls:
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 29, 2026
UV Smart Technologies B.V.
Thijs Kea
Director
Patrijsweg 74
Rijswijk, ZH 2289EX
Netherlands
Re: K251354
Trade/Device Name: UV Smart D60
Regulation Number: 21 CFR 880.6511
Regulation Name: Ultraviolet Radiation Disinfection Chamber Device
Regulatory Class: Class II
Product Code: SCS
Dated: December 29, 2025
Received: December 29, 2025
Dear Thijs Kea:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251354 - Thijs Kea
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251354 - Thijs Kea
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
KATHARINE SEGARS -S
Katharine Segars, PhD.
Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251354 | |
| Device Name UV Smart D60 | |
| Indications for Use (Describe) The UV Smart D60 is indicated for use in a healthcare environment to achieve a high level disinfection (HLD) of external surfaces of Transesophageal Echocardiogram (TEE) probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
UVsmart
# 510(K) SUMMARY K251354
## 1. ADMINISTRATIVE INFORMATION
| 510(k) Owner | UV Smart Technologies B.V. |
| --- | --- |
| Address | Patrijsweg 74, Rijswijk, Zuid-Holland, The Netherlands |
| Phone Number | +31 85 060 9800 |
| Fax Number | N/A |
| Company Representative | Thijs Kea – Director |
| Additional Company Representatives | Daan Hoek – Director |
| Email | info@uv-smart.com |
| Summary Preparation Date | 29 January 2026 |
## 2. SUBJECT DEVICE INFORMATION
| Type of 510(k) Submission | Traditional |
| --- | --- |
| UV Smart Trade Name | UV Smart D60 |
| Subject Device K-Number | K251354 |
| Common Name | UV Smart D60 |
| Classification Name | Ultraviolet Radiation Disinfection Chamber Device |
| Product Code | SCS |
| Regulation Number | 21 CFR 880.6511 |
| Regulatory Class | Class II |
| Review Panel | General Hospital |
## 3. PREDICATE DEVICE INFORMATION
| Predicate Device Name | Germitec Chronos |
| --- | --- |
| Predicate Device Clearance Number | DEN230067 |
| Common Name | Chronos |
| Classification Name | Ultraviolet Radiation Disinfection Chamber Device |
| Product Code | SCS |
| Regulation Number | 21 CFR 880.6511 |
| Regulatory Class | Class II |
| Review Panel | General Hospital |
UV Smart K251354
{5}
UVsmart
# 4. DEVICE DESCRIPTION
The UV Smart D60 is indicated for use in a healthcare environment to achieve a high level disinfection (HLD) of external surfaces of Transesophageal Echocardiogram (TEE) probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths (hereafter collectively referred to as the Intended Load).
The UV Smart D60 provides chemical-free high level disinfection using Ultraviolet-C (UV-C) light within a fully enclosed disinfection chamber. The system features eight UV-C lamps, UV-C-reflective chamber walls, and UV-C-transmitting instrument holders designed to suspend the intended load in an optimal position. The fixed disinfection cycle time of 120 seconds is the only critical determinant of disinfection efficacy. The disinfection efficacy is verified during each cycle by two independent threshold sensors—one for UV-C dose and one for power consumption.
The UV Smart D60 is intended to be used in healthcare facilities by trained personnel. Prior to disinfection, each soiled instrument must be cleaned at the bedside and manually cleaned in accordance with the original equipment manufacturer (OEM) instructions for use. Under normal operating conditions, the UV Smart D60 has an expected service life of ten years, with preventive maintenance and lamp replacement required annually, as specified in the UV Smart D60's Instructions for Use.
To ensure that the intended load is both physically compatible with the disinfection chamber and capable of receiving the required HLD UV-C dose (Minimum Effective Dose or MED) on its surfaces, the intended load is approved for use with the UV Smart D60. Once approved, the intended load is released for use in the UV Smart D60's traceability system. The UV Smart D60 will not initiate a disinfection cycle unless a registered instrument is recognized, ensuring that only approved instruments can be processed.
Each disinfection cycle accommodates a single instrument, which is suspended entirely within the D60's disinfection chamber—ensuring that all surfaces, including handles and attached cables, receive the MED for validated high level disinfection.
# 5. INDICATIONS FOR USE
The UV Smart D60 is indicated for use in a healthcare environment to achieve a high level disinfection (HLD) of external surfaces of Transesophageal Echocardiogram (TEE) probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths.
# 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
For ease of reference and comparison between the subject device and predicate device, an overview is provided in Table 1 - Technological Characteristics Comparison.
UV Smart K251354
{6}
UVsmart
Table 1 - Technological Characteristics Comparison
| Characteristic | UV Smart D60 | Predicate Device | Comparison |
| --- | --- | --- | --- |
| Regulation Number | 880.6511 | 880.6511 | Same |
| Regulatory Class | Class II | Class II | Same |
| Product Code | SCS | SCS | Same |
| Regulation Name | Ultraviolet radiation chamber disinfection device | Ultraviolet radiation chamber disinfection device | Same |
| Device Property | Disinfection Device | Disinfection Device | Same |
| Intended Use / Indications for Use | The UV Smart D60 is indicated for use in a healthcare environment to achieve a high level disinfection (HLD) of external surfaces of Transesophageal Echocardiogram (TEE) probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths. | Chronos is indicated for use in a healthcare environment to achieve a high level disinfection of surfaces of external, transvaginal, and transrectal ultrasound probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths. | Different
The UV Smart D60 is intended for TEE probes. |
| Input | TEE probes that have been pre-cleaned according to OEM instructions and hospital protocol. | External, transvaginal, and transrectal ultrasound probes that have been pre-cleaned according to OEM instructions and hospital protocol. | Different
The UV Smart D60 is intended for TEE probes. |
| Output | High level disinfected instruments, ready for use on patients. | High level disinfected instruments, ready for use on patients. | Same |
| Output Measurements | UV-C Dose Verification via photodiode and Power Consumption monitoring via current sensor. | UV-C Dose Verification via multiple photodiodes. | Similar |
| Independent Monitoring System | Dual-sensor: photodiode for UV-C dose and current sensor for power consumption verification. | Photodiode-based UV-C dose monitoring (multiple sensors). | Different
The UV Smart D60 utilizes a variety of sensors for independent monitoring. |
UV Smart K251354
Page 3 of 9
{7}
UVsmart
| Characteristic | UV Smart D60 | Predicate Device | Comparison |
| --- | --- | --- | --- |
| Report Structure | Detailed report on touchscreen interface including pass/fail status, dose, power, time, date, and traceability. | Pass/fail message on device interface; full report available via optional connected interface. | Different
The UV Smart D60 provides detailed traceability on the interface. |
| Intended Users | Physicians or other licensed practitioners in healthcare institutions, such as clinics, hospitals, healthcare facilities, residential care facilities and long-term care services. | Physicians or other licensed practitioners in healthcare institutions, such as clinics, hospitals, healthcare facilities, residential care facilities and long-term care services. | Same |
| Ventilation | No ventilation required. Temperature monitoring and control. | Integrated ventilation. | Different
The UV Smart D60 does not require ventilation. |
| Intended Use Environment | Healthcare institutions (Clinics, hospitals, healthcare facilities, residential care facilities and long-term care services) | Healthcare institutions (Clinics, hospitals, healthcare facilities, residential care facilities and long-term care services) | Same |
| Software device that operates on off-the-shelf hardware | No | No | Same |
Both the subject device (UV Smart D60) and the predicate device share similar indications for use, with the primary difference being the intended load for disinfection. The subject device is indicated for high level disinfection of TEE probes, whereas the predicate device is intended for the disinfection of external, transvaginal, and transrectal ultrasound probes.
Another difference lies in the disinfection cycle operation: the predicate device adjusts its cycle duration based on the load, while the subject device uses a fixed disinfection cycle time of 120 seconds. Additionally, the subject device accommodates the entire instrument—including both the insertion portion and the connecting cables—within the disinfection chamber.
UV Smart K251354
Page 4 of 9
{8}
UVsmart
. All differences have been thoroughly assessed and supported through appropriate verification and validation testing.
# 7. NONCLINICAL TESTING
For ease of reference and comparison between performance testing and data of the subject device and predicate device, an overview is provided in Table 2 - Performance Testing Overview.
Table 2 - Performance Testing Overview
| Test | Acceptance Criteria | Result |
| --- | --- | --- |
| Biocompatibility | An evaluation of the surface materials of representative TEE probes following exposure to the UV-C disinfection cycle was provided. Biocompatibility risk evaluation and testing has been conducted in accordance with the FDA 2020 Biocompatibility Guidance document, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," | Pass |
| Software & Cybersecurity | Software testing was conducted as per FDA guidance document, "Content a/premarket submissions for Device Software Functions." Cybersecurity testing was conducted as per FDA guidance documents, "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" & "Post-market Management of Cybersecurity in Medical Devices." | Pass |
| Electromagnetic Compatibility & Electrical Safety | Electromagnetic Compatibility (EMC) of the UV Smart D60 device has been evaluated in accordance with IEC 61326-1, Electrical equipment for measurement, control and laboratory use -EMC requirements -Part I: General requirements, and IEC 60601-1-2, Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances Requirements and tests. Electrical safety testing of the UV Smart D60 device was completed in accordance with IEC 61010-1, Safety requirement for electrical equipment for measurement, control and laboratory use. Part 1: | Pass |
UV Smart K251354
{9}
UVsmart
| Test | Acceptance Criteria | Result |
| --- | --- | --- |
| | General requirements and IEC 61010-2-040, Safety requirements for electrical equipment for measurement, control and laboratory use. Part 2-040: Particular requirement for sterilizers and washer-disinfectors used to treat medical materials. | |
| UV-C Hierarchy Testing | Performance testing has been conducted to support a side-by-side evaluation of microorganism UV-C resistance. Each microorganism was tested at a series of varying UV-C doses in the UV Smart D60 device.
The panel of microorganisms included in UV-C resistance hierarchy testing included bacterial spores, bacteria, mycobacteria, yeast/molds (spores) and Trichomaceae family (spore forming molds).
A hierarchy of UV-C resistance was determined for the UV Smart D60 device. Based on the results obtained, the most resistant microorganism tested was Aspergillus brasiliensis, of the Trichomaceae family (spore forming molds). | Pass |
| Potency Testing | The UV Smart D60 device was tested for potency using the standard array of microbicidal testing to support device efficacy.
Modified testing was developed to demonstrate microbicidal potency for the device as there exists no current standard method in place for UV-C disinfection processes. Potency testing was conducted to demonstrate sporicidal, bactericidal, fungicidal, mycobactericidal and virucidal efficacy of the device.
Specifically:
Mycobacterium terrae
Tricophyton mentagrophytes
Salmonella choleraesuis
Staphylococcus aureus
Pseudomonas aeruginosa
Adenovirus Type 5
Murine Norovirus
Poliovirus Type 1
Herpes Simplex Type 2
Simian Virus 40 | Pass |
UV Smart K251354
Page 6 of 9
{10}
UVsmart
| Test | Acceptance Criteria | Result |
| --- | --- | --- |
| Simulated Use Testing | Simulated use testing was conducted with worst-case TEE probes to demonstrate that the UV Smart D60 can reliably and consistently achieve a > 6 log reduction of the most resistant clinically relevant microorganism (Bacillus subtilis) following exposure to the disinfection cycle.
Based on the completed hierarchy of resistance testing and the results of potency testing, Bacillus subtilis was identified as the most resistant bacterial spore. Based on resistance and clinical relevance, it was chosen for use in Simulated Use testing.
Worst-case representatives of a TEE probe (mucous membrane contacting) were selected based on their geometric complexity and difficulty to consistently achieve UV-C HLD.
Worst-case locations were identified based on difficulty for UV-C exposure to penetrate and therefore posed the greatest challenge to microbial inactivation. For each of these worst-case locations, a > 6 log microbial reduction was demonstrated for the worst-case instruments. | Pass |
| Optical Verification on Worst-Case probes | Worst-case TEE probe associated "cold spot" locations on each instrument, underwent optical verification studies to demonstrate that each location received the minimum effective dose (MED) UV-C exposure. All optical measurements at these geometrically worst-case locations were shown to receive the MED following exposure to the UV Smart D60 device’s 120 second disinfection cycle. | Pass |
| In-Use Testing | TEE probes used in a clinical setting were subjected to the UV Smart D60’s high level disinfection cycle.
The probes were cleaned using cleaning solutions routinely used by the hospitals prior to exposure in the UV Smart D60 system. In-Use testing demonstrated the achievement of HLD on these clinical used TEE probes. | Pass |
| UV-C Leak test | Leakage testing verified that the chamber door prevents UV-C escape. | Pass |
UV Smart K251354
Page 7 of 9
{11}
UVsmart
| Test | Acceptance Criteria | Result |
| --- | --- | --- |
| Door Interlock Test | Door interlock performance testing was conducted in accordance with IEC 61010-1 requirements. | Pass |
| Verification of ventilation contamination risk control | The UV Smart D60 does not include any form of active ventilation or airflow exchange within its disinfection chamber.
Internal temperature is continuously monitored throughout the disinfection cycle, and a built-in safeguard disables the device if elevated temperatures are detected. Disinfection cannot resume until the internal temperature returns to the safe operating range.
As a result, no additional risk of microbial contamination is introduced by ventilation, and no filtration or exhaust validation is required. | Pass |
| Equipment Surface Temperature | Performance testing was conducted to demonstrate the ability of the UV Smart D60 device to control temperatures within the chamber to ensure the inserted instruments do not pose burn risk to the operator or patient. Testing was conducted in accordance with the requirements of IEC 60601-2-37 ultrasound probes) and IEC 60601-2-18 (endoscopic equipment). | Pass |
| Independent Cycle Monitoring System Performance Testing | Performance testing was conducted to demonstrate the ability of the UV Smart D60 device's independent cycle monitoring system to accurately determine if a completed disinfection cycle is accurately identified as a success or failure. Testing was conducted to simulate conditions both where the target dose was reached and conditions where the target dose was not actually reached. In all test runs the independent monitoring system was able to identify successful and unsuccessful cycles. | Pass |
| Instrument 3D Scanning Validation | Validation testing was completed to demonstrate that the optical scanning process used is able to accurately capture the geometric complexity of TEE probes. Testing included a range of instrument geometries to challenge the system. Documents provided demonstrate that the method used to acquire 3D scanned files was sufficiently robust to support the optical simulation verification process. | Pass |
UV Smart K251354
Page 8 of 9
{12}
UVsmart
| Test | Acceptance Criteria | Result |
| --- | --- | --- |
| Optical Simulation Verification | Optical Simulation verification was conducted to evaluate the interaction between the geometric variations of the TEE probes, and the UV-C dose distribution when exposed to the UV Smart D60 disinfection chamber. Validation of the simulation included worst-case TEE probes identified for each dose group. This validation testing confirmed that the MED was systematically received at the three worst-case-location of the instrument when exposed to the UV-C disinfection cycle. | Pass |
# 8. CLINICAL TESTING
Not applicable for the proposed device.
# 9. CONCLUSION
The conclusions drawn from the nonclinical and clinical tests demonstrates that the UV Smart D60 is as safe, as effective and performs as well as or better than the legally marketed predicate device Germitec Chronos (DEN230067).
UV Smart K251354
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.