DEN230067 · Germitec · SCS · Aug 28, 2024 · General Hospital
Device Facts
Record ID
DEN230067
Device Name
Chronos®
Applicant
Germitec
Product Code
SCS · General Hospital
Decision Date
Aug 28, 2024
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 880.6511
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
DEN230067 · Aug 28, 2024
Chronos®
Germitec
Clinical-use ultrasound probes (transvaginal, transrectal, and external)
In-use testing was conducted to demonstrate the achievement of high-level disinfection (HLD) on ultrasound probes that had been used in a clinical setting, reflecting real-world clinical practice.
In-Use Testing; Observational/Performance evaluation in clinical setting
Clinically used transvaginal, transrectal, and external ultrasound probes
Not applicable for this study
Achievement of high-level disinfection (HLD)
Indications for Use
Chronos® is indicated for use in a healthcare environment to achieve a high-level disinfection of surfaces of external, transvaginal, and transrectal ultrasound probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths.
Device Story
Chronos is an ultraviolet (UV) radiation disinfection chamber for ultrasound probes. Device operates by placing cleaned probes into an enclosed chamber; UV radiation exposure achieves high-level disinfection. Used in healthcare environments; operated by clinical staff. System includes safety features to prevent UV exposure. Output is a disinfected probe surface. Benefits include rapid, automated disinfection of patient-contacting medical devices, reducing cross-contamination risk. Device requires prior cleaning of probes before UV cycle.
Clinical Evidence
No clinical data provided. Evidence basis relies on non-clinical performance testing, including resistance testing, potency testing against various microorganisms (spores, bacteria, mycobacteria, viruses), simulated use testing under worst-case conditions, in-use testing, photobiological safety validation, and material compatibility testing.
Technological Characteristics
Ultraviolet radiation disinfection chamber. Employs UV radiation for microbial reduction. Includes safety features to prevent UV exposure. Requires software for cycle control and safety monitoring. Subject to electromagnetic compatibility, electrical, and thermal safety standards. Materials must be compatible with repeated UV exposure.
Indications for Use
Indicated for high-level disinfection of external, transvaginal, and transrectal ultrasound probes without lumens or deep channels/indentations in healthcare environments.
Regulatory Classification
Identification
An ultraviolet radiation disinfection chamber device is intended to disinfect patient contacting medical devices using UV radiation after the device has been cleaned. Disinfection of the medical device is achieved within an enclosed chamber through the exposure to UV radiation. The Germitec Chronos® is an automated medical device consisting of a sealed disinfection chamber, incorporating multiple UV-C lamps and reflective walls, indicated for use in a healthcare environment to achieve a high-level disinfection of surfaces of external, transvaginal, and transrectal ultrasound probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths.
Special Controls
In combination with the general controls of the FD&C Act, the ultraviolet radiation disinfection chamber device is subject to the following special controls:
Submission Summary (Full Text)
{0}------------------------------------------------
## DE NOVO CLASSIFICATION REQUEST FOR GERMITEC CHRONOS
### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Ultraviolet radiation disinfection chamber device. An ultraviolet radiation disinfection chamber device is intended to disinfect patient contacting medical devices using UV radiation after the device has been cleaned. Disinfection of the medical device is achieved within an enclosed chamber through the exposure to UV radiation.
NEW REGULATION NUMBER: 21 CFR 880.6511
CLASSIFICATION: Class II
PRODUCT CODE: SCS
### BACKGROUND
DEVICE NAME: Germitec Chronos®
SUBMISSION NUMBER: DEN230067
DATE DE NOVO RECEIVED: October 3, 2023
### SPONSOR INFORMATION:
Germitec 19 Rue Vauban 33000 Bordeaux, France
### INDICATIONS FOR USE
Chronos® is indicated for use in a healthcare environment to achieve a high-level disinfection of surfaces of external, transvaginal, and transrectal ultrasound probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths.
## LIMITATIONS
The Germitec Chronos® is intended to be marketed for Over-The-Counter (OTC) use.
PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS
{1}------------------------------------------------
## DEVICE DESCRIPTION
The Germitec Chronos® device is intended to provide Ultraviolet C (UV-C) chemical-free highlevel disinfection (HLD) of ultrasound probes within an enclosed disinfection chamber. As defined in FDA Guidance - Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, found at https://www.fda.gov/media/80265/download, HLD is a lethal process utilizing a sterilant under less than sterilizing conditions. The process kills all forms of microbial life except for large numbers of bacterial spores. The subject device is intended for use in health care facilities by trained personnel. The Chronos® is an automated medical device consisting of a sealed disinfection chamber, incorporating multiple UV-C lamps and reflective walls. UV-C dose is identified as the only critical determinant of the effectiveness of the disinfection cycle.
The software is used to determine achievement of the disinfection total dose and acceptance of the automated HLD cycle. The total dose delivered to the surface of each ultrasound probe undergoing. the disinfection process is continuously monitored by two optical sensors, UV-C sensitive photodiodes. The control system continuously takes readings from the sensors, tracks progress of the ongoing disinfection cycle, and terminates the disinfection cycle when the cycle-controlling photodiodes indicate that the pre-defined UV-C dose has been achieved. Additionally, the device utilizes an independent cycle monitoring step through the use of a third UV-C photodiode. This independent probe is used to verify that the pre-defined UV-C HLD dose has been delivered.
Image /page/1/Picture/3 description: The image shows a medical device with a user interface panel on the left side. On the right side, there is a probe holder and barcode reader opening, a disinfection chamber, and a door. The device appears to be designed for disinfecting medical probes and reading barcodes.
# Pictures of Chronos® - Door Closed (left) and Door Open (right):
Onto each ultrasound probe for which compatibility has been validated, a ring is placed onto the ultrasound cable. This ring is used to set the probe into the proper position within the chamber, as well as to include an identifier barcode. This barcode is read by the Chronos® software, and
{2}------------------------------------------------
provides identification of both the appropriate disinfection dose as well as traceability for the ultrasound probe undergoing disinfection. The encoding of the barcode to each specific probe model and installation of the ring on each probe can only be performed by an authorized service representative from Germitec.
Image /page/2/Picture/1 description: The image shows a barcode reader and its components. The top left image shows the front view of the barcode reader, with labels pointing to the window, barcode sticker, ring, and probe cable. The top right image shows a side view of the barcode reader, with a label pointing to the barcode reader opening. The bottom image shows a close-up view of the ring receptacle and the ring, with labels pointing to each component.
Barcode ring location:
# SUMMARY OF BENCH STUDIES
## REPROCESSING, STERILITY AND SHELF-LIFE
The Germitec Chronos® is provided non-sterile and does not require additional reprocessing or sterilization. The device requires regular maintenance in order to continue to function as intended.
Annual preventive maintenance activities are defined for general system checks, device cleaning (internal and external components), calibration of optical sensors (UV-C photodiodes) and replacement of the ventilation filter. Additionally, the UV-C lamps must be replaced on a two-year interval, or every 100 hours of cumulative use, whichever comes first. The software provides an indication to the operator of UV-C lamp usable life remaining.
## BIOCOMPATIBILITY
The Germitec Chronos® device is designed to provide liquid-free, high-level disinfection of reusable ultrasound probes using UV-C as the germicidal agent. Since UV-C does not remain on the ultrasound probe following completion of an HLD cycle, there is no indirect contact with the patient or personnel conducting HLD reprocessing. The sponsor has
{3}------------------------------------------------
provided an evaluation of the surface materials of representative ultrasound probes following exposure to the UV-C disinfection cycle. Biocompatibility risk evaluation and testing has been conducted in accordance with the FDA 2020 Biocompatibility Guidance document, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," found at https://www.fda.gov/media/142959/download
## SOFTWARE & CYBERSECURITY
The Germitec Chronos® device incorporates a software-based control system used to control and monitor the delivery of UV-C dose required for high level disinfection of reusable ultrasound probes. The control system includes an ultrasound probe identifying system that captures key information about the probe, using a barcode reader, to link it to each disinfection. In addition to controlling the HLD cycle, the Chronos® device software controls system operation, ensures that the device is operating as expected, and communicates potential cycle errors to the user, as well as providing warnings about system status, condition, and reminders regarding preventive maintenance. The software testing was conducted as per FDA guidance document, "Content of premarket submissions for Device Software Functions, " found at https://www.fda.gov/media/153781/download.
Cybersecurity testing was conducted as per FDA guidance documents. "Cybersecurity in Medical Devices: Ouality System Considerations and Content of Premarket Submissions" found at https://www.fda.gov/media/119933/download, and "Postmarket Management of Cybersecurity in Medical Devices, " at https://www.fda.gov/media/95862/download.
# ELECTROMAGNETIC COMPATIBILITY & ELECTRICAL SAFETY
Electromagnetic Compatibility (EMC) of the Chronos® device has been evaluated in accordance with IEC 61326-1, Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements, and IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests.
Electrical safety testing of the Chronos® device was completed in accordance with IEC 61010-1. Safety requirement for electrical equipment for measurement, control and laboratory use. Part 1: General requirements and IEC 61010-2-040, Safety requirements for electrical equipment for measurement, control and laboratory use. Part 2-040: Particular requirement for sterilizers and washer-disinfectors used to treat medical materials. Testing has been successfully completed per the referenced standards. including electrical, mechanical resistance to shock/impact, and mechanical safety testing.
## PERFORMANCE TESTING - BENCH
The sponsor conducted the following performance tests to support that the device can achieve its intended use:
{4}------------------------------------------------
#### UV-C Hierarchy of Resistance .
Performance testing has been conducted to support a side-by-side evaluation of microorganism UV-C resistance. Each microorganism was tested at a series of varying UV-C doses in the Chronos® device. For the purposes of hierarchy testing, a panel of microorganisms was selected which represented a combination of organisms currently used as surrogates for other HLD processes or organisms reported to be particularly resistant to UV-C in literature. The panel of microorganisms included in UV-C resistance hierarchy testing included bacterial spores, bacteria, mycobacteria, yeast/molds (spores) and Trichomaceae family (spore forming molds). A hierarchy of UV-C resistance was determined for the Chronos® device. Based on the results obtained, the most resistant microorganism tested was Aspergillus brasiliensis, of the Trichomaceae family (spore forming molds).
#### Potency Testing .
The Chronos® device was tested for potency using the standard array of microbicidal testing to support device efficacy. Modified testing was developed to demonstrate microbicidal potency for the device as there exists no current standard method in place for UV-C disinfection processes. Potency testing was conducted to demonstrate sporicidal, bactericidal, fungicidal, mycobactericidal and virucidal efficacy of the device. Microbicidal potency testing included the following microorganisms.
| Microbicidal Activity | Organisms Challenged |
|-----------------------|-----------------------------------------|
| Sporicidal | Bacillus subtilis |
| | Bacillus pumilus |
| | Clostridium sporogenes |
| Bactericidal | Deinococcus radiodurans |
| | Salmonella enterica |
| | Staphylococcus aureus |
| | Pseudomonas aeruginosa |
| | Paraburkholderia fungorum |
| Mycobactericidal | Mycobacterium terrae |
| | Mycobacterium bovis |
| | Mycobacterium avium |
| Fungicidal | Candida auris |
| | Candida albicans |
| | Trichophyton mentagrophyte (spore form) |
| Virucidal | Poliovirus Type 1 |
| | Adenovirus Type 5 |
| | Murine Norovirus |
| | Vaginal Herpes |
| | Polyomavirus |
| | HPV16 / HPV18 |
{5}------------------------------------------------
| Sporicidal Activity | Organisms Challenged |
|-------------------------------------------|----------------------------------------------------------------|
| Trichomaceae family (spore forming molds) | Aspergillus brasiliensis<br>Penicillium polonicum (spore form) |
Based on the testing documentation provided, the results support the determination of microbicidal potency for the device.
#### ● Simulated Use Testing
Simulated use testing was conducted over a range of worst-case ultrasound probes to demonstrate that the Chronos® can reliably and consistently achieve a > 6 log reduction of the surrogate microorganism (Bacillus subtilis) following exposure to the disinfection cycle. Based on the completed hierarchy of resistance testing and the results of potency testing, Bacillus subtilis was identified as the most resistant bacterial spore and determined to be the most appropriate for use in Simulated Use testing. Representative transvaginal and transrectal (mucous membrane contacting), and external (skin contacting) ultrasound probes were selected based on their geometric complexity and difficulty to consistently achieve UV HLD.
For each probe, the three worst-case locations were identified based on difficulty for UV-C exposure to penetrate and therefore posed the greatest challenge to microbial inactivation. For each of these worst-case locations, a > 6 log microbial reduction was demonstrated for the worst-case ultrasound probes over a range of challenging geometries.
#### Optical Verification on Worst-Case Probes .
Worst-case ultrasound probes and associated "cold spot" locations on each probe, underwent optical verification studies to demonstrate that each location received the minimum effective dose (MED) of (b)(4) |mJ/cm2 UV-C exposure. All optical measurements at these geometrically worst-case locations were shown to receive the MED following exposure to the Chronos® device automated disinfection cycles.
#### In-Use Testing .
Ultrasound probes used in a clinical setting were subjected to the Chronos® automated high level disinfection cycles. Transvaginal and transrectal probes were used with single use probe protective sheath and gel. Transvaginal and transrectal probes were cleaned using cleaning solutions routinely used by hospital. External probes were used without a sheath and only underwent a dry wipe prior to exposure in the Chronos® system. In-Use testing demonstrated the achievement of HLD on these clinical used ultrasound probes.
{6}------------------------------------------------
#### UV-C Leak Test .
The sponsor has conducted performance testing to support to demonstrate that the chamber door prevents UV-C leakage from the disinfection chamber during a routine disinfection cycle. Documentation was provided to show that the amount of UV-C leakage remained < [b)( nW/cm2.
#### . Door Interlock Test
Door interlock performance testing was conducted in accordance with IEC 61010-1 requirements.
#### . Verification of ventilation contamination risk control
Performance of the Chronos® device ventilation was conducted to support control of excessive heat build up while mitigating potential for probe contamination from outside air. Testing provided documented evidence verifying the ability of the ventilation system to limit microbial contamination entering through the ventilation system.
#### . Probe Surface Temperature
Performance testing was conducted to demonstrate the ability of the Chronos® device to control temperatures within the chamber to ensure ultrasound probes do not pose burn risk to the operator or patient. Testing was conducted in accordance with IEC 60601-2-37.
#### . Independent Cycle Monitoring System Performance Testing
Performance testing was conducted to demonstrate the ability of the Chronos® device's independent cycle monitoring system to accurately to determine if a completed disinfection cycle is accurately identified as a success or failure. Testing was conducted to simulate conditions both where the target dose was reached and conditions where the target dose was not actually reached. In all test runs the independent monitoring system was able to identify successful and unsuccessful cycles.
#### Probe 3D Scanning Validation .
Validation testing was completed to demonstrate that the optical scanning process used is able to accurately capture the geometric complexity of ultrasound probes. Testing included a range of ultrasound probe geometries to challenge the system. Document provided demonstrated that the method used to acquire probe 3D scanned files was sufficiently robust to support the optical simulation verification process.
{7}------------------------------------------------
#### Optical Simulation Verification .
Optical Simulation verification was conducted to evaluate the interaction between the geometric variations of the ultrasound probes and the UV-C dose distribution when exposed to the Chronos® disinfection chamber. Validation of the simulation included the three worst-case probes identified for each dose group. This validation testing confirmed that the MED was systematically received at the three worst-case-location of the probe when exposed to the automated pre-defined UV-C disinfection cycle.
# LABELING
The labeling consists of a product label and User Manual.
Labeling for this device is in accordance with the special controls listed below.
## RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of an ultraviolet radiation disinfection chamber device and the measures necessary to mitigate these risks.
| Risks to Health | Mitigation Measures |
|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Patient cross-contamination due to<br>device failure or operator error<br>leading to inadequate microbial<br>reduction | Non-clinical performance testing<br>Software verification, validation, and hazard analysis<br>Labeling |
| Exposure to UV radiation, leading<br>to skin and eye damage | Non-clinical performance testing<br>Biocompatibility evaluation<br>Software verification, validation, and hazard analysis<br>Labeling |
| Incompatibility of the materials<br>intended to be disinfected leading<br>to reduced performance or<br>premature failure | Non-clinical performance testing<br>Biocompatibility analysis<br>Software verification, validation, and hazard analysis<br>Labeling |
| Electrical shock or thermal<br>hazards | Non-clinical performance testing<br>Electrical safety testing<br>Thermal safety testing |
| Interference with other devices | Electromagnetic compatibility testing<br>Electrical safety testing |
# SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the ultraviolet radiation disinfection chamber device is subject to the following special controls:
{8}------------------------------------------------
- (1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
- (i) Resistance testing must demonstrate the hierarchy of resistance to UV radiation using a panel of microorganisms that are clinically relevant and those known to be resistant to UV radiation;
- (ii) Potency testing against bacterial spores, vegetative bacteria, mycobacteria, yeast/molds (spores), spore forming molds and virus, must demonstrate that the UV disinfection device is able to achieve microbial log reduction of a panel of clinically relevant and UV resistant microorganisms, as appropriate for the intended level of disinfection;
- (iii) Simulated use testing must evaluate device performance under worst-case use and device conditions, including selection of appropriate challenge microorganism(s), microbial inoculum with residual soil, device complexity, exposure dose, and time, as appropriate for the intended level of disinfection;
- (iv) In-use testing must evaluate device performance under real-world clinical use conditions:
- Performance testing must demonstrate the photobiological safety of any lamps or (v) lamp systems;
- Performance testing must validate safety features intended to prevent exposure; (vi) and
- (vii) Performance testing must characterize the long-term material compatibility of UV radiation on clinically relevant surfaces and/or devices following repeated UV disinfection cycle exposure.
- (2) Biocompatibility evaluation must demonstrate safe residual levels of chemicals on medical device surfaces and/or gaseous byproducts in air.
- Software verification, validation, and hazard analysis must be performed for any (3) software components.
- (4) Performance data must demonstrate the electromagnetic compatibility (EMC), thermal and electrical safety of the device.
- (5) The labeling must include:
- (i) Warnings and instructions to ensure the device is used for UV disinfection purposes, and is not to be used as a sterilant or as part of a sterilization process for medical devices:
- (ii) User setup, operating and maintenance procedures;
- Description of the required preparation of equipment for disinfection in the UV (iii) chamber device;
- (iv) UV hazard warning labels:
- Explanation of device safety features; and (v)
- Information regarding material compatibility. (vi)
{9}------------------------------------------------
# BENEFIT-RISK DETERMINATION
# Benefit:
The sponsor demonstrated that the probable benefits include the following:
- . Chemical-free, automated. High Level Disinfection (HLD) device for the surfaces of external, transvaginal, and transrectal probes. HLD reprocessing is performed following routine cleaning activities of the reprocessed ultrasound.
- Reduced disinfection duration. .
- . Traceability of the disinfection process.
# Risks:
The risks of the device are based on nonclinical laboratory testing described above.
The main probable risk for the subject device is that a failure in the automated reprocessing cycle does not achieve the required microbial reduction, and high level disinfection is not achieved. Probes which have not been adequately disinfected may lead to patient crosscontamination, and an increased health risk associated with infections. Further, additional risks associated with UV exposure, electrical shock and interference with other devices, as well as thermal burns associated with probes following the disinfection cycle have also been identified. Additionally, cybersecurity risks have been identified for the device. These risks have all been mitigated through completion of performance testing, warnings, precautions, and special controls. Human factors testing was deemed not required based on an evaluation of potential high-risk use scenarios and the anticipated level of user interface needed to support routine operation of the device. In addition, In-Use testing was provided to demonstrate real world use conditions.
## Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
## Benefit/Risk Conclusion
In conclusion, given the testing provided and available information summarized above, for the following indication statement:
Chronos® is indicated for use in a healthcare environment to achieve a high-level disinfection of surfaces of external, transvaginal, and transrectal ultrasound probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths.
The probable benefits outweigh the risks for the Germitec Chronos®. The device provides benefits, and the risks can be mitigated by the use of general controls and the identified special controls.
{10}------------------------------------------------
# CONCLUSION
The De Novo request for the Germitec Chronos® is granted and the device is classified as follows:
Product Code: SCS Device Type: Ultraviolet radiation disinfection chamber device Regulation Number: 21 CFR 880.6511 Class: II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.