Shenzhen Fansizhe Science and Technology Co., Ltd.
Product Code
OHT · General, Plastic Surgery
Decision Date
Jul 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Intense Pulsed Light (IPL) System is an over-the-counter device intended for the removal of unwanted body hair.
Device Story
Over-the-counter IPL device for permanent hair reduction; utilizes Xenon arc flashlamp to emit light pulses; operates below skin surface without cutting/pulling. Input: user-initiated activation via switch or auto-emission; touch chip detects skin contact to enable pulses; skin color sensor (select models) prevents operation on unsuitable skin tones. Output: light pulses for hair follicle treatment. Used in home environment by adult patients. Cooling care mode (select models) reduces heat; dual pulse technology (select models) provides single/dual pulse options. Healthcare provider involvement is absent; device output affects hair growth reduction; benefits include non-invasive hair removal with minimal pain.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence includes biocompatibility testing (ISO 10993-5, -10, -23), electrical safety and EMC testing (IEC 60601-1, -1-2, -1-11, -2-83), and photobiological eye safety testing (IEC 62471). Software verification and validation confirmed mitigation of hazards to acceptable levels.
Technological Characteristics
Materials: ABS, PC. Energy source: Xenon arc flashlamp. Wavelength: 560 (±20nm) to 1200nm. Output: 1-5 intensity levels. Features: Touch sensor for skin contact, skin color sensor (select models), cooling care (select models), dual pulse technology (select models). Connectivity: Standalone. Sterilization: Not required. Software: Microprocessor-controlled.
Indications for Use
Indicated for removal of unwanted body hair in adults aged 18 and older with Fitzpatrick skin types I-IV.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Hand-held IPL device (IPL Home Use Hair Removal Device), Model: FZ-608, FZ-608G, FZ-100, FZ-200 (K230739)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 14, 2025
Shenzhen Fansizhe Science and Technology Co., Ltd.
% Youshan Gong
RA Specialist
Feiying Drug & Medical Consulting Technical Service Group
Rm 2401 Zhenye International Business Center
No. 3101-90, Qianhai Road
Shenzhen, Guangdong 518052
China
Re: K251173
Trade/Device Name: Intense Pulsed Light (IPL) System (T033KQ, T033KD, T033KF, T033MQ, T033MD, T033MF, T055KQ, T055KD, T055KH, T002AQ, T002AD, T002AF, T002BQ, T002BD, T002BF, T050KQ, T050KD, T050KF)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHT
Dated: March 28, 2025
Received: April 16, 2025
Dear Youshan Gong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251173 - Youshan Gong
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K251173 - Youshan Gong
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.07.14
14:48:12 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251173 | |
| Device Name Intense Pulsed Light (IPL) System (Model(s): T033KQ, T033KD, T033KF, T033MQ, T033MD, T033MF, T055KQ, T055KD, T055KH, T002AQ, T002AD, T002AF, T002BQ, T002BD, T002BF, T050KQ, T050KD, T050KF) | |
| Indications for Use (Describe) The Intense Pulsed Light (IPL) System is an over-the-counter device intended for the removal of unwanted body hair. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Page 1 of 13
# K251173 510(k) Summary
“510(k) Summary” as required by 21 CFR Part 807.92.
Date Prepared: 2025-7-10
## I. Submitter
Shenzhen Fansizhe Science And Technology Co., Ltd
2nd, 3rd And 4th floors, No.11, Henggangxia Industrial Zone, Xiangshan Community, Xinqiao Street, Bao'an District, Shenzhen Guangdong, CN 518125
Post code: 518125
Endang Cheng
Title: General Manager
Tel: +86 18617061709
E-mail: ChengEndang1001@163.com
## II. Device
Name of Device: Intense Pulsed Light (IPL) System
Model(s): T033KQ, T033KD, T033KF, T033MQ, T033MD, T033MF, T055KQ, T055KD, T055KH, T002AQ, T002AD, T002AF, T002BQ, T002BD, T002BF, T050KQ, T050KD, T050KF
Common or Usual Name: Light Based Over-The-Counter Hair Removal
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: II
Product Code: OHT
Regulation Number: 21 CFR 878.4810
## III. Predicate Device and Reference Device
{5}
Page 2 of 13
> Predicate Device
| Manufacturer | Predicate Device | 510(k) Number |
| --- | --- | --- |
| Shenzhen Fansizhe Science And Technology Co., Ltd | Intense Pulsed Light (IPL) System, Model: T013C, T015C, T015K | K221569 |
> Reference Device 1
| Manufacturer | Reference Device | 510(k) Number |
| --- | --- | --- |
| Shenzhen Ulike Smart Electronics Co.,Ltd. | Ice Cooling IPL Hair Removal Device (UI20 DB, UI20 RE, UI20 GP, UI20 PW, UI20S DB, UI20S RE, UI20S PW, UI20S GP, UI20 WH, UI20 GR, UI20 BS, UI20 MP, UI20 BL, UI20 PN, UI20 BR, UI20WG, UI20S WH, UI20S GR, UI20S BS, UI20S MP) | K241998 |
> Reference Device 2
| Manufacturer | Reference Device | 510(k) Number |
| --- | --- | --- |
| Shenzhen IONKA Medical Technology Co., Ltd. | Hand-held IPL device (IPL Home Use Hair Removal Device), Model: FZ-608, FZ-608G, FZ-100, FZ-200 | K230739 |
IV. Device Description
Intense Pulsed Light (IPL) System, models: T033KQ, T033KD, T033KF, T033MQ, T033MD, T033MF, T055KQ, T055KD, T055KH, T002AQ, T002AD, T002AF, T002BQ, T002BD, T002BF, T050KQ, T050KD, T050KF, are a small over-the-counter device for the permanent reduction of hair growth based on Intense Pulsed Light (IPL). It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the power adapter and its IPL emission activation is by a switch or auto light emission.
{6}
Intense Pulsed Light (IPL) System, all models, contains a Xenon arc flashlamp, and a touch chip to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot emit the treatment light pulses.
Intense Pulsed Light (IPL) System, models: T002AQ, T002AD, T002AF, contains a Xenon arc flashlamp, a touch chip to detect appropriate skin contact, and a skin color sensor to detect the skin color. In "Skin Color Recognition mode", the device's skin sensor automatically detects skin tone for your protection. If your skin tone is not in tone table suitable for treatment, the device must not be used. You need to identify your skin tone before treatment according to skin tone table, and confirm whether the product is applicable to you after the skin color sensor detects a skin tone.
Based on the cooling technology, Intense Pulsed Light (IPL) System, models: T033KQ, T033KD, T033KF, T055KQ, T055KD, T055KH, T050KQ, T050KD, T050KF, has cooling care functions. When the cooling care mode is enabled, it can reduce the excessive heat generated on the skin by the photon irradiation and do cooling compresses during hair removal.
Based on the dual pulse technology, Intense Pulsed Light (IPL) System, models: T033KQ, T033KD, T033KF, T033MQ, T033MD, T033MF, T055KQ, T055KD, T055KH, T050KQ, T050KD, T050KF, has single pulse and dual pulse functions.
The Intense Pulsed Light (IPL) System includes main unit, an adaptor and goggles.
The device is intended to be used for adults aged over 18.
# V. Indications for Use
Intense Pulsed Light (IPL) System is an over-the-counter device intended for removal of unwanted body hair.
# VI. Materials
| Model | Contacted Component Name | Materials |
| --- | --- | --- |
| T033KQ, T033KD, T033KF, T033MQ, T033MD, T033MF, T055KQ, T055KD, T055KH, T002AQ, | Host of machine (including air outlet, treatment window, air inlet, buttons) | ABS, PC |
| T033KF, T033MQ, T033MD, T033MF, T055KQ, T055KD, T055KH, T002AQ, | Host of machine (including air outlet, treatment window, air inlet, buttons) | ABS, PC |
{7}
Page 4 of 13
T002AD, T002AF,
T002BQ, T002BD,
T002BF, T050KQ,
T050KD, T050KF
# VII. Comparison of Technological Characteristics With the Predicate Device
The Intense Pulsed Light (IPL) System has the same intended use, mode of action and similar operational characteristics as the predicate device. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the Intense Pulsed Light (IPL) System may be found substantially equivalent to its predicate device.
Intense Pulsed Light (IPL) System is compared with the following Predicate Devices in terms of intended use, design, material, specifications and performance:
| Comparison Elements | Subject Device | | | | | | Predicate Device | Reference Device 1 | Reference Device 2 | Remark |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Model | T033KQ T033KD T033KF | T033MQ T033MD T033MF | T055KQ T055KD T055KH | T002AQ T002AD T002AF | T002BQ T002BD T002BF | T050KQ T050KD T050KF | T013C | UI20 DB | FZ-608 FZ-608G FZ-100 FZ-200 | |
| 510(k) Number | K251173 | | | | | | K221569 | K241998 | K230739 | / |
| Trade name | Intense Pulsed Light (IPL) System (T033KQ, T033KD, T033KF, T033MQ, T033MD, T033MF, T055KQ, T055KD, T055KH, T002AQ, T002AD, T002AF, T002BQ, T002BD, T002BF, T050KQ, T050KD, | | | | | | Intense Pulsed Light (IPL) System (Model: | Ice Cooling IPL Hair Removal Device (UI20 DB, | Hand-held IPL device (IPL Home Use Hair | / |
{8}
| Comparison Elements | Subject Device | | | | | | Predicate Device | Reference Device 1 | Reference Device 2 | Remark |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Model | T033KQ T033KD T033KF | T033MQ T033MD T033MF | T055KQ T055KD T055KH | T002AQ T002AD T002AF | T002BQ T002BD T002BF | T050KQ T050KD T050KF | T013C | UI20 DB | FZ-608 FZ-608G FZ-100 FZ-200 | |
| | T050KF) | | | | | | T013C, T015C, T015K) | UI20 RE, UI20 GP, UI20 PW, UI20S DB, UI20S RE, UI20S PW, UI20S GP, UI20 WH, UI20 GR, UI20 BS, UI20 MP, UI20 BL, UI20 PN, UI20 BR, UI20WG, UI20S WH, UI20S GR, UI20S BS, UI20S MP) | Removal Device), Model: FZ-608, FZ-608G, FZ-100, FZ-200 | |
| Manufacturer | Shenzhen Fansizhe Science And Technology Co., Ltd | | | | | | Shenzhen Fansizhe Science And Technology Co., Ltd | Shenzhen Ulike Smart Electronics Co., Ltd. | Shenzhen IONKA Medical Technology Co., Ltd. | / |
| Regulation number | 21 CFR 878.4810 | | | | | | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Product code | OHT | | | | | | OHT | OHT | OHT | Same |
| Device classification | Class II | | | | | | Class II | Class II | Class II | Same |
| Indication for use/ | The Intense Pulsed Light (IPL) System is an over-the-counter device | | | | | | The Intense | Ice Cooling IPL | IPL Home Use | Same |
{9}
| Comparison Elements | Subject Device | | | | | | Predicate Device | Reference Device 1 | Reference Device 2 | Remark |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Model | T033KQ
T033KD
T033KF | T033MQ
T033MD
T033MF | T055KQ
T055KD
T055KH | T002AQ
T002AD
T002AF | T002BQ
T002BD
T002BF | T050KQ
T050KD
T050KF | T013C | UI20 DB | FZ-608 FZ-608G
FZ-100 FZ-200 | |
| Intended use | intended for the removal of unwanted body hair. | | | | | | Pulsed Light (IPL) System is an over-the-counter device intended for the removal of unwanted body hair. | Hair Removal Device with sapphire treatment window is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. | Hand-held IPL device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. | |
| Prescription or OTC | OTC | | | | | | OTC | OTC | OTC | Same |
| Applicable skin | Fitzpatrick skin types I-IV | | | | | | Fitzpatrick skin | Fitzpatrick Skin | Fitzpatrick Skin | Same |
Page 6 of 13
{10}
| Comparison Elements | Subject Device | | | | | | Predicate Device | Reference Device 1 | Reference Device 2 | Remark |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Model | T033KQ T033KD T033KF | T033MQ T033MD T033MF | T055KQ T055KD T055KH | T002AQ T002AD T002AF | T002BQ T002BD T002BF | T050KQ T050KD T050KF | T013C | UI20 DB | FZ-608 FZ-608G FZ-100 FZ-200 | |
| | | | | | | | types I-IV | Types I-V | Types I-V | |
| Sterilization | Not required | | | | | | Not required | Not required | Supplied by external adapter | Same |
| Light source | Intense Pulsed Light | | | | | | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Same |
| Energy medium | Xenon Arc lamp | | | | | | Xenon Arc lamp | Xenon Arc lamp | Xenon Arc lamp | Same |
| Wavelength range | 560 (±20nm) ~1200nm | 560 (±20nm) ~1200nm | 560 (±20nm) ~1200nm | 560 (±20nm) ~1200nm | 560 (±20nm)~1200nm | 560 (±20nm)~1200nm | 510~1200nm | 550nm~1200nm | 510nm~1200nm | Similar |
| Max. Output energy (J) | 16.7J | 17.7J | 15.8J | 16.6J | 16.6J | 17.2J | 16.1J | 26J | 10J | Similar |
| Max. Energy density (J/cm2) | 5.1J/cm2 | 5.4J/cm2 | 5.3J/cm2 | 4.2J/cm2 | 4.2J/cm2 | 5.2J/cm2 | 4.03J/cm2 | 6.67J/cm2 | 5.43J/cm2 | Similar |
| Spot size | 3.3cm2 | 3.3cm2 | 3.0cm2 | 4.0cm2 | 4.0cm2 | 3.3cm2 | 4cm2 | 3.9cm2 | 3.0cm2 | Similar |
| Pulse duration | 0.4~12ms | 0.4~12ms | 0.4~12ms | 0.4~12ms | 0.4~12ms | 0.4~12ms | 4~12ms | 0.88~3.20ms | 0.5~0.8 ms | Similar |
| Output intensity level | 1~5 levels | 1~5 levels | 1~5 levels | 1~5 levels | 1~5 levels | 1~3 levels | 1~5 levels | 1~10 levels | 1~9 levels | Same |
| Pulsing control | Finger switch | | | | | | Finger switch | Finger switch | Finger switch | Same |
| Software/Firmware/Microprocessor | Yes | | | | | | Yes | Yes | Yes | Same |
{11}
| Comparison Elements | Subject Device | | | | | | Predicate Device | Reference Device 1 | Reference Device 2 | Remark |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Model | T033KQ
T033KD
T033KF | T033MQ
T033MD
T033MF | T055KQ
T055KD
T055KH | T002AQ
T002AD
T002AF | T002BQ
T002BD
T002BF | T050KQ
T050KD
T050KF | T013C | UI20 DB | FZ-608 FZ-608G
FZ-100 FZ-200 | |
| Control? | | | | | | | | | | |
| Operating environment | Temperature: 5-26°C
Relative humidity: 20-90%,
without condensation
Atmospheric Pressure: 70
kPa to 106 kPa | | | | | | Temperature: 5-30°C
Relative humidity: 20-90%,
without condensation
Atmospheric Pressure: 70
kPa to 106 kPa | Temperature: 5-30°C
Relative humidity: 5-90%,
Atmospheric Pressure: 80 kPa to 106 kPa | Not public | Same |
| Storage and transportation environment | Temperature: -20 -55 °C
Relative humidity: 5-95%,
without condensation
Atmospheric Pressure:70
kPa to 106 kPa | | | | | | Temperature: -20 -55 °C
Relative humidity: 5-95%,
without condensation
Atmospheric | Temperature: -10 -55 °C
Relative humidity: 5-90%,
Atmospheric Pressure:50 kPa to 106 kPa | Not public | Same |
Page 8 of 13
{12}
| Comparison Elements | Subject Device | | | | | | Predicate Device | Reference Device 1 | Reference Device 2 | Remark |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Model | T033KQ
T033KD
T033KF | T033MQ
T033MD
T033MF | T055KQ
T055KD
T055KH | T002AQ
T002AD
T002AF | T002BQ
T002BD
T002BF | T050KQ
T050KD
T050KF | T013C | UI20 DB | FZ-608 FZ-608G
FZ-100 FZ-200 | |
| | | | | | | | Pressure:70
kPa to 106 kPa | | | |
| Electrical safety | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-83 | | | | | | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-83 | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-83 | ANSI AAMI ES60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-83 | Same |
| Eye safety | IEC 62471 | | | | | | IEC 62471 | IEC 62471 | IEC 62471 | Same |
| Biocompatibility | ISO 10993-5
ISO 10993-10
ISO 10993-23 | | | | | | ISO 10993-5
ISO 10993-10 | ISO 10993-5
ISO 10993-10
ISO 10993-23 | ISO 10993-5
ISO 10993-10 | Same |
VIII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
1) Biocompatibility Testing
The biocompatibility evaluation for the body-contacting components of the subject device was conducted in accordance with the “Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process'”, as recognized by FDA. The following testing was performed to, and passed, including:
- ISO 10993-5:2009, Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
Page 9 of 13
{13}
> ISO 10993-10:2021, Biological evaluation of medical devices –Part 10: Tests for skin sensitization
> ISO 10993-23:2021, Biological evaluation of medical devices –Part 23: Tests for skin irritation
## 2) Electrical Safety and EMC
Electrical safety and EMC testing was performed to, and passed, the following standards:
- IEC 60601-1 Medical electrical equipment –Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-11 Medical Electrical Equipment –Part 1: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- IEC 60601-2-83 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
## 3) Eye Safety
- IEC 62471 Photobiological safety of lamps and lamp systems
## 4) Software Verification and Validation
Software documentation consistent with **moderate level** of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
## IX. Conclusions
Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device Intense Pulsed Light (IPL) System is as safe, as effective, and performs as well as the legally marketed predicate device and reference device.
Page 10 of 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.